# Copran Zri (K232952)

_Whitepeaks Dental Solutions GmbH · EIH · Jun 24, 2024 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232952

## Device Facts

- **Applicant:** Whitepeaks Dental Solutions GmbH
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Jun 24, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Copran Zri consists of machinable zirconia discs for the preparation of full ceramic crowns, onlays and 3-and 4-unit bridges and inlay bridges (anterior and molar.)

## Device Story

Copran Zri consists of pre-sintered zirconia blanks (3Y-TZP) used for the fabrication of dental restorations. The device is supplied as discs in 95mm and 98mm diameters with thicknesses ranging from 10-25mm, available in white or pre-shaded variants. In a dental laboratory setting, a technician places the blank into a CAD/CAM milling machine. The machine mills the zirconia into the desired restoration shape based on digital design data. Following milling, the restoration undergoes sintering at temperatures exceeding 1450°C to achieve final material properties. The finished restoration is then placed by a dentist. The device provides a durable, biocompatible ceramic material for restorative dentistry, benefiting patients by restoring tooth function and aesthetics.

## Clinical Evidence

Bench testing only. Physical properties (bending strength, solubility, radioactivity, thermal expansion) tested per ISO 6872:2015; all parameters met standard requirements. Biocompatibility assessment conducted in accordance with ISO 10993-1:2018.

## Technological Characteristics

Material: 3Y-TZP zirconia (ZrO2); ISO 6872:2015 Type II Class 5. Form factor: Pre-sintered discs (95/98mm diameter, 10-25mm thickness). Processing: CAD/CAM milling followed by sintering (>1450°C). Non-sterile. Tetragonal crystal morphology. Flexural strength: 1400 MPa. Fracture toughness: 5 MPa·m1/2. Solubility: 6.7 µg/cm². CTE: 10.2 10-6/K.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- Ivoclar IPS e.max ZirCad MO ([K051705](/device/K051705.md))

## Reference Devices

- Whitepeaks Copran Zr ([K092496](/device/K092496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 24, 2024

Whitepeaks Dental Solutions GmbH % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172

Re: K232952

Trade/Device Name: Copran Zri Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: April 1, 2024 Received: April 2, 2024

Dear Angela Blackwell:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232952

Device Name Copran Zri

Indications for Use (Describe)

Copran Zri consists of machinable zirconia discs for the preparation of full ceramic crowns, onlays and 3-and 4-unit bridges and inlay bridges (anterior and molar.)

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# K232952 Whitepeaks Dental Solutions Copran Zri discs 510K Summary April 1, 2024

Name and Address: Whitepeaks Dental Solutions GmbH Alfredstr. 81 45130 Essen Germany Contact Person: Oliver Puckert Email: o.puckert@whitepeaks-dental.de Telephone: +49(281) 206458-11

Name of device: Whitepeaks Dental Solutions Copran Zri discs Common name: zirconia blanks Classification Name: Porcelain powder for clinical use CFR: 21 CFR 872.6660 Primary Product Code: EIH

### Submission Contact:

Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com

#### Device Description:

Pre-sintered zirconia blanks for the fabrication of individual dental restorations.

Copran Zri disks come in both white and pre-shaded light, medium or intense in 95 and 98 mm diameter in thicknesses from 10-25mm.

#### Indications for Use:

Copran Zri consists of machinable zirconia discs and blocks for the preparation of full ceramic crowns, onlays and 3-and 4-unit bridges and inlay bridges (anterior and molar.)

Testing Summary: The physical properties of Copran Zri were tested according to ISO 6872:2015 and all parameters meet the standard. Testing included bending strength, solubility, radioactivity and thermal expansion coefficient.

A biocompatibility assessment of Copran Zri was done in accordance with ISO 10993-1:2018.

Mechanism of Action: Used in a milling machine for fabrication of a dental restoration

{5}------------------------------------------------

## Predicate Device: Ivoclar IPS e.max ZirCad MO K051705

## Reference Devices: Whitepeaks Copran Zr K092496

#### Substantial Equivalence:

The zirconia blanks have similar ingredients to the predicate and reference devices, the same indications for use, and similar physical parameter testing.

| Name                | Copran Zri                                                                                                                                                           | Copran Zri Light, Medium and Intense                                                                                                                                 | Copran Zr Reference Device                                                                                                                                                                                                                                                                                                                                                                    | Ivoclar ZirCad MO Predicate Device                                                                                                                                          |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number         | K232952                                                                                                                                                              | K232952                                                                                                                                                              | K092496                                                                                                                                                                                                                                                                                                                                                                                       | K051705                                                                                                                                                                     |
| Common Name         | Zirconia blanks                                                                                                                                                      | Zirconia blanks                                                                                                                                                      | Zirconia blanks                                                                                                                                                                                                                                                                                                                                                                               | Zirconia blanks                                                                                                                                                             |
| Classification Name | Porcelain powder for clinical use                                                                                                                                    | Porcelain powder for clinical use                                                                                                                                    | Porcelain powder for clinical use                                                                                                                                                                                                                                                                                                                                                             | Porcelain powder for clinical use                                                                                                                                           |
| Class               | II                                                                                                                                                                   | II                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                                                                                                                            | II                                                                                                                                                                          |
| Product Code        | EIH                                                                                                                                                                  | EIH                                                                                                                                                                  | EIH                                                                                                                                                                                                                                                                                                                                                                                           | EIH                                                                                                                                                                         |
| CFR                 | 872.6660                                                                                                                                                             | 872.6660                                                                                                                                                             | 872.6660                                                                                                                                                                                                                                                                                                                                                                                      | 872.6660                                                                                                                                                                    |
| Indications for Use | Copran Zri consists of machinable zirconia discs for the preparation of full ceramic crowns, onlays and 3-and 4-unit bridges and inlay bridges (anterior and molar.) | Copran Zri consists of machinable zirconia discs for the preparation of full ceramic crowns, onlays and 3-and 4-unit bridges and inlay bridges (anterior and molar.) | Copran Zr Zirconia blanks are presintered blanks for CAD CAM or manual milling, made from biocompatible, tetragonal and polycrystalline zirconiumdioxyde. Milling blanks designed for: Crown frameworks in the anterior and posterior areas Bridge frameworks in the anterior and posterior areas Primary conical crowns and telescopic crowns Cantilevered bridges with a max, of one pontic | IPS e.max ZirCAD consists of machinable zirconia blocks for the preparation of full ceramic crowns, onlays and 3-and 4-unit bridges and inlay bridges (anterior and molar.) |

Zirconia discs from Whitepeaks Dental Solutions

{6}------------------------------------------------

|                                   |                                                                                                              |                                                                                                              | having a premolar<br>width<br>Inlays, Onlays,<br>Veneers                                                     |                                                                                                    |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Device Description                | Pre-sintered<br>zirconia blanks for<br>the fabrication of<br>individual dental<br>restorations.              | Pre-sintered<br>zirconia blanks for<br>the fabrication of<br>individual dental<br>restorations.              | Pre-sintered<br>zirconia blanks for<br>the fabrication of<br>individual dental<br>restorations.              | Pre-sintered<br>zirconia blanks<br>for the<br>fabrication of<br>individual dental<br>restorations. |
| Basic Design                      | Discs                                                                                                        | Discs                                                                                                        | Discs and blocks                                                                                             | Discs and blocks                                                                                   |
| Mechanism of<br>Action            | Used in a milling<br>machine for<br>fabrication of a<br>dental restoration                                   | Used in a milling<br>machine for<br>fabrication of a<br>dental restoration                                   | Used in a milling<br>machine for<br>fabrication of a<br>dental restoration                                   | Used in a milling<br>machine for<br>fabrication of a<br>dental<br>restoration                      |
| Materials                         | ISO 6872:2015<br>Type II Class 5                                                                             | ISO 6872:2015<br>Type II Class 5                                                                             | ISO 6872:2015<br>Type II Class 5                                                                             | ISO 6872:2015<br>Type II Class 5                                                                   |
| Formula                           | ZrO2 with others                                                                                             | ZrO2 with others                                                                                             | ZrO2 with others                                                                                             | ZrO2 with others                                                                                   |
| Crystal<br>Morphology             | Tetragonal                                                                                                   | Tetragonal                                                                                                   | Tetragonal                                                                                                   | Tetragonal                                                                                         |
| Material                          | 3Y-TZP                                                                                                       | 3Y-TZP                                                                                                       | 3Y-TZP                                                                                                       | 3Y-TZP                                                                                             |
| Processing                        | Sintering at<br>temperature ><br>1450°C                                                                      | Sintering at<br>temperature ><br>1450°C                                                                      | Sintering at<br>temperature ><br>1450°C                                                                      | Sintering at<br>temperature ><br>1450°C                                                            |
| Disc Dimensions                   | Discs 95mm<br>diameter in<br>thicknesses 10-<br>25mm<br>Discs 98mm<br>diameter in<br>thicknesses 10-<br>25mm | Discs 95mm<br>diameter in<br>thicknesses 10-<br>25mm<br>Discs 98mm<br>diameter in<br>thicknesses 10-<br>25mm | Discs 95mm<br>diameter in<br>thicknesses 10-<br>25mm<br>Discs 98mm<br>diameter in<br>thicknesses 10-<br>25mm | Discs 98mm<br>diameter in<br>thicknesses 10-<br>25mm                                               |
| Single Use                        | Yes                                                                                                          | Yes                                                                                                          | Yes                                                                                                          | Yes                                                                                                |
| Color                             | None                                                                                                         | Pre-shaded                                                                                                   | None                                                                                                         | Pre-shaded                                                                                         |
| Sterility                         | Non-sterile                                                                                                  | Non-sterile                                                                                                  | Non-sterile                                                                                                  | Non-sterile                                                                                        |
| Flexural Strength<br>MPa          | 1400                                                                                                         | 1400                                                                                                         | 1400                                                                                                         | 1150                                                                                               |
| Fracture<br>Toughness<br>MPa·m1/2 | 5                                                                                                            | 5                                                                                                            | 5                                                                                                            | >5.1                                                                                               |
| Solubility µg/cm²                 | 6.7                                                                                                          | 6.7                                                                                                          | 6.7                                                                                                          | <100                                                                                               |
| CTE (25-500°C)<br>10-6/K          | 10.2                                                                                                         | 10.2                                                                                                         | 10.2                                                                                                         | 10.5 ± 0.5                                                                                         |

{7}------------------------------------------------

| Radioactivity | < 0.0117 Bq·g-1 | < 0.0117 Bq·g-1 | < 0.0117 Bq·g-1 | Not present in<br>labeling |
|---------------|-----------------|-----------------|-----------------|----------------------------|
|               | 238U            | 238U            | 238U            |                            |

Conclusion: Copran Zri are substantially equivalent to the predicate device. They have the same indications for use, mechanism of action and similar physical properties and formula. Both the subject devices, the predicate devices and the reference devices physical properties meet ISO 6872:2015. Differences in shapes and sizes between the subject devices and the predicate device are covered by the reference device. Slight differences in diameter do not change their substantial equivalence to the predicate devices as these differences do not change the materials properties in any way, it is just to allow for machine fit.

---

**Source:** [https://fda-staging.innolitics.com/device/K232952](https://fda-staging.innolitics.com/device/K232952)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).