The Trio-CT® Triple Lumen Catheter is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis and apheresis. The third internal lumen is intended for infusion, power injection of contrast media and central venous pressure monitoring. The catheter is intended to be inserted in the jugular, femoral or subclavian vein as required. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media.
Device Story
Short-term (less than 30 days) triple-lumen dialysis catheter; thermosensitive polyurethane construction; symmetric/tapered tip design. Three lumens: two for hemodialysis/apheresis (venous/arterial); one independent lumen for infusion, power injection of contrast media, central venous pressure monitoring, blood draws, and medication infusion. Inserted via jugular, femoral, or subclavian vein using Seldinger method. Operated by physicians or qualified healthcare professionals in clinical settings. Attachable suture wing for securement. Facilitates immediate central venous access; enables high-flow hemodialysis and contrast media delivery (max 5ml/sec).
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), mechanical hemolysis testing, and a 30-day in vivo thromboresistance study in sheep.
Technological Characteristics
Thermosensitive polyurethane; 13.5F diameter; 12-24cm lengths; three lumens. Symmetric/tapered tip. EO sterilization. Seldinger insertion method. No software or electronic components.
Indications for Use
Indicated for adult patients requiring short-term (less than 30 days) vascular access for hemodialysis and apheresis, and for infusion, power injection of contrast media, and central venous pressure monitoring via a third lumen. Not intended for pediatric patients.
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 16, 2023
Medical Components Inc. (Dba MedComp) Danielle McKinney Regulatory Affairs Associate II 1499 Delo Drive Harleysville, Pennsylvania 19438
Re: K232945
Trade/Device Name: Trio-CT® Triple Lumen Catheter w/ Curved Extensions Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: Class II Product Code: NIE Dated: September 20, 2023 Received: September 20, 2023
Dear Danielle McKinney:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Gema Gonzalez -S
Maura Rooney, MS Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232945
Device Name
Trio-CT® Triple Lumen Catheter w/ Curved Extensions
Indications for Use (Describe)
· The Trio-CT® Triple Lumen Catheter is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis and apheresis. The third internal lumen is intended for infusion, power injection of contrast media and central venous pressure monitoring.
• The catheter is intended to be inserted in the jugular, femoral or subclavian vein as required. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MedComp. The logo consists of a red abstract shape on the left, followed by the word "med" in red, and "COMP" in black. The logo is simple and modern, and the colors are eye-catching.
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Control Center (DCC) - W066-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
# 510(k) Summary Traditional 510(k): Original
# 1. Submitter Information:
510(k) Submitted By:
- Medical Components Inc. (Dba MedComp) 1499 Delp Drive Harleysville, PA 19438 United States of America
- Primary Correspondent: Danielle McKinney, Regulatory Affairs Associate II regulatory@medcompnet.com 843-467-5661
- Date Prepared: September 20, 2023
#### 2. Subject Device Information:
| Trade Name: | Trio-CT® Triple Lumen Catheter w/ Curved Extensions |
|-----------------------|---------------------------------------------------------|
| Common Name: | Triple Lumen Catheter |
| Regulation Name: | Blood Access Device and Accessories,<br>21 CFR 876.5540 |
| Device Class: | Class II |
| Classification Panel: | Gastroenterology/Urology |
| Product Code: | NIE |
#### 3. Predicate Device Information:
| Trade Name (predicate): | Trio-CT™ Triple Lumen Catheter |
|----------------------------|-------------------------------------------------------------------------|
| Common Name (predicate): | Triple Lumen Catheter |
| Classification Regulation: | Catheter, Hemodialysis, Triple Lumen, Non-Implanted,<br>21 CFR 876.5540 |
| Device Class: | Class II |
| Classification Panel: | Gastroenterology/Urology |
| Product Code: | NIE |
| Predicate 510(k) Number: | 510(k) K183219 |
| Predicate 510(k) Owner: | Medical Components Inc. (Dba MedComp) |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "K232945" and "Page 2 of 5" at the top right corner. Below this text is the logo for MedComp. The logo is in red and black. The image appears to be a page from a document.
#### 4. Device Description:
The Trio-CT® Triple Lumen Catheter is a short-term (less than 30 days) dialysis catheter made of thermosensitive polyurethane. The catheter has three separate lumens allowing continuous blood flow. The venous (blue) and arterial (red) lumens may be used for hemodialysis and apheresis treatments. The middle (purple) lumen is independent from the two dialysis lumens, and may be used for intravenous therapy, power injection of contrast media, central venous pressure monitoring, blood draws and infusion of medications. The catheter is available with straight or curved extensions in a variety of lengths to accommodate physician preference and clinical needs. The attachable suture wing can be used to provide additional catheter securement and to minimize movement at the exit site.
#### 5. Indications for Use
- The Trio-CT® Triple Lumen Catheter is indicated for use in attaining short-term (less . than 30 days) vascular access for hemodialysis and apheresis. The third internal lumen is intended for infusion, power injection of contrast media and central venous pressure monitoring.
- . The catheter is intended to be inserted in the jugular, femoral or subclavian vein as required. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media.
# 6. Summary of Technological Characteristics
The subject device has similar technological characteristics as the predicate device. The Trio-CT® Triple Lumen Catheters facilitate hemodialysis, intravenous administration of fluids or medications, and power injection of contrast media in patients in whom immediate central venous vascular access for short-term hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. The catheter is not intended for use in pediatric patients. The Trio-CT® Triple Lumen Catheters utilize a symmetric/tapered tip design which facilitates insertion without a peelable introducer. The Trio-CT® Triple Lumen Catheters are made from medical-grade chemically-stable materials and are provided sterile and nonpyrogenic. The catheters can be inserted via the internal jugular vein, femoral vein or subclavian vein. While there are differences in the anatomic reference points utilized for properly identifying the insertion site and risks associated with each site, once the insertion site is identified, the implantation process utilizes the Seldinger insertion method to place the catheter. Regardless of insertion point selected, Trio-CT® Triple Lumen Catheters are to be inserted, manipulated and removed by a qualified, licensed physician or other qualified health care professional under direction of a physician utilizing strict aseptic technique. Based on the similarity of these technological characteristics, there are no changes that affect the safety or effectiveness of the Trio-CT® Triple Lumen Catheter w/ Curved Extensions.
# 7. Substantial Equivalence Discussion
The Substantial Equivalence Comparison Table (Table 2) below provides evidence to facilitate the substantial equivalence determination between the Trio-CT® Triple Lumen Catheter w/ Curved Extensions to the predicate device Trio-CT™ Triple Lumen Catheter.
{6}------------------------------------------------
K232945 Page 3 of 5
Image /page/6/Picture/1 description: The image shows the logo for MedComp. The logo consists of a red icon on the left, followed by the word "med" in red and the word "COMP." in black. The icon is a stylized shape that resembles a house or building.
#### Table 2: Substantial Equivalence Comparison
| Description | Subject Device | Predicate Device | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Device Trade<br>Name | Trio-CT® Triple Lumen Catheter w/ Curved<br>Extensions | Trio-CT™ Triple Lumen Catheter | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| 510(k)<br>submitter/holder | Medical Components Inc. (Dba MedComp) | Medical Components Inc. (Dba MedComp) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| 510(k) Number | K###### | K183219 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Indications for<br>Use | The Trio-CT® Triple Lumen Catheter is indicated<br>for use in attaining short-term (less than 30 days)<br>vascular access for hemodialysis and apheresis.<br>The third internal lumen is intended for infusion,<br>power injection of contrast media and central<br>venous pressure monitoring.<br>The catheter is intended to be inserted in the<br>jugular, femoral or subclavian vein as required.<br>The maximum recommended infusion rate is<br>5ml/sec for power injection of contrast media. | The Trio-CT® Triple Lumen Catheter is indicated<br>for use in attaining short-term (less than 30 days)<br>vascular access for hemodialysis and apheresis.<br>The third internal lumen is intended for infusion,<br>power injection of contrast media and central<br>venous pressure monitoring.<br>The catheter is intended to be inserted in the<br>jugular, femoral or subclavian vein as required. The<br>maximum recommended infusion rate is 5ml/sec<br>for power injection of contrast media. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Prescription/Over<br>-the-Counter Use | Prescription Use | Prescription Use | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Target Population | Adult | Adult | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Duration of Use | Short Term | Short Term | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Sterilization<br>Method | 1x and/or 2x EO Sterilized | 1x and/or 2x EO Sterilized | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Technical Specifications | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Catheter French<br>Size | 13.5F | 13.5F | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Inner Diameters | Venous - 0.088" nominal<br>Arterial – 0.088" nominal<br>Infusion - 0.042" nominal | Venous - 0.088" nominal<br>Arterial - 0.088" nominal<br>Infusion - 0.042" nominal | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Number of<br>Lumens | Three (3) | Three (3) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Catheter Lengths | 12cm, 15cm, 20cm and 24cm | 12cm, 15cm, 20cm, 24cm and 30cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Priming Volume<br>(cc) | Length Center Arterial Venous 12cm 0.4 cc 1.4 cc 1.4 cc 15cm 0.4 cc 1.5 cc 1.5 cc 20cm 0.5 cc 1.7 cc 1.7 cc 24cm 0.5 cc 1.9 cc 1.9 cc | | | | | | | | | | | | | | | | | | | | | Length Center Arterial Venous 12cm 0.4 cc 1.2 cc 1.2 cc 15cm 0.4 cc 1.3 cc 1.3 cc 20cm 0.5 cc 1.5 cc 1.5 cc 24cm 0.5 cc 1.6 cc 1.6 cc 30cm 0.6 cc 1.9 cc 1.9 cc | | | | | | | | | | | | | | | | | | | | | | | | |
| Power Injection<br>Pressure (psi) | Max<br>Indicated<br>Power<br>Injection<br>Flow-Rate Average<br>Max<br>Catheter<br>Pressure<br>During Max<br>Indicated<br>Power<br>Injection<br>Flow Rate Average<br>Max Burst<br>Pressure Range of<br>Max Burst<br>Pressures 5 ml/sec 92 psi 314 psi 277-425<br>psi | | | | | | | | | Max<br>Indicated<br>Power<br>Injection<br>Flow-Rate Average<br>Max<br>Catheter<br>Pressure<br>During Max<br>Indicated<br>Power<br>Injection<br>Flow Rate Average<br>Max Burst<br>Pressure Range of<br>Max Burst<br>Pressures 5 ml/sec 84 psi 420 psi 332-446<br>psi | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
{7}------------------------------------------------
K232945 Page 4 of 5
Image /page/7/Picture/1 description: The image shows the logo for MedComp. The logo consists of a red abstract shape on the left, followed by the word "med" in red and the word "COMP" in black. There is a period after the word "COMP".
| Description | Subject Device | | | | Predicate Device | | | |
|-------------------------------------|-------------------------------------|-----|-----|-----|-------------------------------------|-----|-----|-----|
| | Flow Rate (ml/min) | | | | Flow Rate (ml/min) | | | |
| | | 200 | 300 | 400 | | 200 | 300 | 400 |
| Flow (ml/min) vs<br>Pressure (mmHg) | 12CM Venous | 20 | 41 | 61 | 12CM Venous | 20 | 30 | 51 |
| | Arterial | -29 | -44 | -69 | Arterial | -21 | -40 | -60 |
| | 15CM Venous | 21 | 40- | 59 | 15CM Venous | 20 | 40 | 63 |
| | Arterial | -30 | -45 | -70 | Arterial | -26 | -46 | -70 |
| | 20CM Venous | 30 | 45 | 72 | 20CM Venous | 21 | 40 | 62 |
| | Arterial | -30 | -49 | -74 | Arterial | -30 | -50 | -70 |
| | 24CM Venous | 30 | 46 | 71 | 24CM Venous | 30 | 50 | 76 |
| | Arterial | -30 | -50 | -80 | Arterial | -33 | -50 | -80 |
| | | | | | 30CM Venous | 30 | 51 | 84 |
| | | | | | Arterial | -33 | -59 | -90 |
| Principles of Operation | | | | | | | | |
| Insertion Site | Jugular, femoral or subclavian vein | | | | Jugular, femoral or subclavian vein | | | |
# 8. Summary of Non-Clinical Testing and Standards
#### 8.1 Biocompatibility
A biocompatibility evaluation was performed in accordance with the FDA Guidance 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Based on this evaluation, the following tests were conducted:
- ISO Muscle Implantation Study in Rabbits 4 Weeks ●
- USP Rabbit Pyrogen Study, Material Mediated ●
- ASTM Hemolysis Study ●
- SC5b-9 Complement Activation Assay ●
- ISO Guinea Pig Maximization Sensitization Test ●
- ISO Intracutaneous Study in Rabbits ●
- Cytotoxicity Study Using the ISO Elution Method
- ISO Acute Systemic Toxicity Study in Mice ●
- ISO Systemic Toxicity Study in the Rat, Repeated Parenteral Administration of Two ● Extracts, 28 Days
- Bacterial Reverse Mutation Study ●
- Genotoxicity: Mouse Lymphoma Assay ●
- Infrared Spectroscopy ●
- Mechanical Hemolysis Testing of Triple Lumen Catheters ●
- In Vivo GLP 30 Day Thromboresistance Study in Sheep- Jugular Vein
#### 8.2 Performance Testing
Design validation performance testing was leveraged from the predicate device manufactured by Medical Components, Inc. Design verification performance testing was completed to confirm performance criteria of the subject device. Clinical testing was not required nor performed to support the substantial equivalence of these devices.
{8}------------------------------------------------
#### 9. Conclusion
In conclusion, based on the non-clinical testing performed, the subject device Trio-CT® Triple Lumen Catheter w/ Curved Extensions is substantially equivalent to the predicate device Trio-CT™ Triple Lumen Catheter [510(k) K183219] and does not raise any concern for the safety and effectiveness of the subject device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.