K232933 · Healkee Medical Pte, Ltd. · FRA · May 16, 2024 · General Hospital
Device Facts
Record ID
K232933
Device Name
AirKEE P900
Applicant
Healkee Medical Pte, Ltd.
Product Code
FRA · General Hospital
Decision Date
May 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AirKEE P900 is a medical ultraviolet air purifier intended for medical purposes to capture and destroy bacteria, mold, and viruses in the air through the multi-stage air filtration system and exposure to ultraviolet radiation.
Device Story
AirKEE P900 is a mobile medical ultraviolet air purifier for use in clinical environments (hospitals, clinics, dental facilities). It draws air through a multi-stage filtration system: primary G4 filter (large particles), nano-material filter, UV-C lamps (microorganism inactivation via TiO2 photocatalyst), activated carbon filter, and HEPA H13 filter (0.1-micron particles). A centrifugal fan moves air through the system. The device is operated via a touchscreen control panel allowing fan speed adjustment, timer settings, and UV/lighting control. The device provides continuous air purification, reducing airborne pathogens to benefit patients and staff by improving air quality. Safety features include non-removable output grills and interlock switches to prevent direct UV exposure or fan contact.
Clinical Evidence
Bench testing only. Performance validated via microorganism log reduction studies in a 20m³ chamber, achieving 4-log (99.99%) reduction for S. aureus, E. coli, K. pneumoniae, P. aeruginosa, S. marcescens, A. Niger, and Influenza A (H1N1/H3N3) at specified fan speeds. HEPA fractional efficiency confirmed at 99.99% for 0.1-0.2µm particles per IEST-RP-CC001.6. Electrical safety and EMC testing passed per IEC 60601-1-2 and UL 507. Ozone emissions verified <0.05 ppm per UL 867.
Technological Characteristics
Mobile air purifier; multi-stage filtration (G4, nano-material, activated carbon, HEPA H13); UV-C LED light source (254.7nm, 552µW/cm²); TiO2 photocatalyst; centrifugal fan; touchscreen interface. Standards: UL 507 (fans), IEC 60601-1-2 (EMC), IEST-RP-CC001.6 (HEPA), UL 867 (ozone). Power: 110V. Connectivity: Standalone.
Indications for Use
Indicated for use in hospitals, medical facilities, medical clinics, nursing facilities, and dental facilities to capture and destroy bacteria, mold, and viruses in the air. Suitable for use by healthcare professionals and lay users.
Regulatory Classification
Identification
A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 16, 2024
HealKEE Medical Pte Ltd % Shanshan Liu Charles & International LLC 45 Ashford St Allston, Massachusetts 02134
Re: K232933
Trade/Device Name: AirKEE P900 Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: April 20, 2024 Received: April 22, 2024
Dear Shanshan Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Stephen A. Digitally signed by Stephen A. Anisko -S Date: 2024.05.16 Anisko -S 15:31:19 -04'00'
for: Christopher Dugard Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232933
Device Name AirKEE P900
#### Indications for Use (Describe)
AirKEE P900 is a medical ultraviolet air purifier intended for medical purposes to capture and destroy bacteria, mold, and viruses in the air through the multi-stage air filtration system and exposure to ultraviolet radiation.
AirKEE P900 has been demonstrated to eliminate the following test microorganisms entrained on the subject device under the following exposure conditions:
| Test Item | Avg. max Log Reduction/Entrainment<br>Time (min) (in a 20m3 chamber) |
|----------------------------------|------------------------------------------------------------------------------------------|
| Bacteria Staphylococcus aureus | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Escherichia coli | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Klebsiella pneumoniae | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Pseudomonas aeruginosa | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Serratia marcescens | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Mold<br>Aspergillus Niger | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Virus<br>Influenza A virus, H1N1 | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Virus<br>Influenza A virus, H3N3 | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K232933 - HealKee Medical Pte Ltd
This 510(k) Summary is in conformance with 21CFR 807.92
| Submitter: | HealKee Medical Pte Ltd<br>11 North Buona Vista Drive, Level 08-09<br>The Metropolis Tower 2, Singapore 138589 |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Shanshan Liu, Consultant<br>Charles & International LLC<br>Email: sliu@charlieintl.com<br>Phone: 617-893-3456 |
| Company Contact: | William Tao<br>Title: Founder & CEO<br>Email: william.tao@healkee.com<br>Phone: +65 6808 7770 |
| Date Prepared: | May 1st, 2024 |
| Trade Name: | AirKEE P900 |
| Regulation Number: | 21 CFR 880.6500 |
| Regulation Name: | Medical Ultraviolet Air Purifier |
| Regulatory Class: | Class II |
| Common Name: | Air Purifier |
| Classification Name: | Purifier, Air, Ultraviolet, Medical |
| Review Panel: | General Hospital |
| Product Code: | FRA |
# Predicate Device(s):
| | Predicate |
|---------------------------|-------------------|
| Trade Name | Aura Storm |
| 510(k) Submitter / Holder | Invictus Lighting |
| 510(k) Number | K212644 |
| Regulation Number | 21 CFR 880.6500 |
| Classification | Class II |
| Classification Panel | General Hospital |
| Product Code | FRA |
{5}------------------------------------------------
### Device Description:
AirKEE P900 is a mobile multi-function medical ultraviolet air purifier. It is used to destroy bacteria in the air by exposure to ultraviolet radiation. It may be used for medical purposes which can include hospitals, medical facilities, medical clinics, nursing facilities, and dental facilities. AirKEE P900 is controlled via a touchscreen control panel on the machine.
AirKEE P900's main components consist of:
Multi-stage air treatment
- a filtration system with primary filter, HEPA filter, activated carbon filter, and nano- O material filter
- UV lamps that generate UV-C irradiation to inactivate and eliminate microorganisms o
- a motor/impeller to move air through the filtration system o
System control
- o a touch panel interface to indicate the working status of the device and consumables
AirKEE P900's multi-stage air treatment: Air enters through the bottommost chamber and fans direct the air in an upwards manner. Firstly, air passes through primary G4 filter, where large particulate matter will be removed. G4 filtered air then passes through patented nano-material filter for further removal. Microorganisms will be inactivated by UV, and photocatalysts. Air then passes through activated carbon filter. Lastly, HEPA H13 filter eliminates 99.99% of the remaining particles (0.1 microns in diameter). The treated air then exits through the top in an upwards manner.
The control panel allows reading, troubleshooting functions, and settings of various function. Its enhanced user interface with touchscreen function allows easier navigation of the system.
## Indications for Use:
AirKEE P900 is a medical ultraviolet air purifier intended for medical purposes to capture and destroy bacteria, mold, and viruses in the air through the multi-stage air filtration system and exposure to ultraviolet radiation.
AirKEE P900 has been demonstrated to eliminate the following test microorganisms entrained on the filter of the subject device under the following exposure conditions:
| Test Item | Avg. max Log Reduction/Entrainment Time (min) (in a 20m³ chamber) |
|---------------------------------|------------------------------------------------------------------------------------------|
| Bacteria Staphylococcus aureus | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Escherichia coli | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Klebsiella pneumoniae | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Pseudomonas aeruginosa | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Bacteria Serratia marcescens | Log 4 (99.99%)/60 minutes @ high fan speed<br>Log 4 (99.99%)/120 minutes @ low fan speed |
| Mold Aspergillus Niger | Log 4 (99.99%)/60 minutes @ high fan speed |
{6}------------------------------------------------
| | | Log 4 (99.99%)/120 minutes @ low fan speed |
|-------|-------------------------|--------------------------------------------|
| Virus | Influenza A virus, H1N1 | Log 4 (99.99%)/60 minutes @ high fan speed |
| Virus | Influenza A virus, H3N3 | Log 4 (99.99%)/120 minutes @ low fan speed |
{7}------------------------------------------------
| Item | AirKEE P900<br>Subject Device | Aura Storm<br>(K212644)<br>Predicate | Comparison |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Type | Medical Ultraviolet Air Purifier | Medical Ultraviolet Air Purifier | Identical |
| Product Code | FRA | FRA | Identical |
| Classification Regulation | 21 CFR 880.6500 | 21 CFR 880.6500 | Identical |
| Class | II | II | Identical |
| Rx/OTC | OTC | OTC | Identical |
| Indications for Use | AirKEE P900 is a medical ultraviolet air<br>purifier intended for medical purposes to<br>capture and destroy bacteria, mold, and<br>viruses in the air through the multi-stage air<br>filtration system and exposure to ultraviolet<br>radiation.<br><br>AirKEE P900 has been demonstrated to<br>eliminate the following test microorganisms<br>entrained on the filter of the subject device<br>under the following exposure conditions:<br><br>Staphylococcus aureus<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Escherichia coli<br>Log 4 (99.99%)/60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Klebsiella pneumoniae<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed; | The Aura Storm air purifier is a device<br>intended for medical<br>purposes that is used to capture and<br>destroy bacteria and viruses in the air<br>through the multi-stage filtration system<br>and exposure to ultraviolet radiation.<br><br>The Aura Storm air purifier has been<br>demonstrated to destroy<br>the following bacteria: Staphylococcus<br>albicans.<br><br>Staphylococcus aureus,<br>and Escherichia Coli; and virus: A/PR8/34<br>H1N1 virus entrained on the filter of the<br>subject device under the following<br>exposure conditions:<br>Average Maximum log<br>reduction / entrainment time (minutes) at<br>Fan Speed 4. Room Temperature test: Log<br>4<br>(99.99%) / 60 minutes. | Similar<br>The Indications for<br>use of the devices<br>are identical. The<br>devices have<br>different<br>performance metrics<br>for the chosen<br>microorganisms. |
| | Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Serratia marcescens<br>Log 4 (99.99%)/60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Aspergillus Niger<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Influenza A virus, H1N1<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed<br>Influenza A virus, H3N3<br>Log 4 (99.99%)/ 120 minutes at low fan<br>speed | reduction / entrainment time (minutes) at<br>Fan Speed 1. Room Temperature test: Log<br>4<br>(99.99%) / 120 minutes. | |
| User | Healthcare Professional<br>Lay User | Healthcare Professional<br>Lay User | Identical |
| Environment for Use | Hospitals, medical<br>facilities, medical clinics,<br>nursing facilities, and<br>dental facilities. | Hospitals, medical<br>facilities, medical clinics,<br>nursing facilities, and<br>dental facilities. | Identical |
| Placement | AirKEE P900 can be placed anywhere in a<br>room for general air treatment.<br><br>AirKEE P900 is designed for rooms<br>up to 3000 ft3. | Aura Storm is designed for rooms up to<br>2773 square feet. | Similar |
| User Control | Touch panel with selection of operating<br>mode (Manual, Silent), 3 fan speeds, UV<br>lamp on/off, medical lighting on/off. | Touch panel with 4 manual fan settings,<br>automode, UV Lamp on/off, Anion<br>on/off, lock and<br>timer. | Similar |
| Software | Basic Firmware; used to turn the unit on, off, change fan speed, and other administrative functions (timer, UV) | Basic Firmware, used to turn the unit on, off, and change fan speed. | Similar |
| Mechanism of Action | UV light of sufficient energy (UV-C) activates a TiO2 lined photocatalyst that destroys microorganisms entrained on the filter through a photochemical reaction, plus the addition of a pre-filter, activated carbon and patented nano-material filter, and HEPA filter. | UV light of sufficient energy (UV-C) activates a TiO2 lined photocatalyst that destroys microorganisms entrained on the filter through a photochemical reaction, plus the addition of a pre-filter and HEPA filter. | Identical |
| -Installation | Free standing | Free standing | Identical |
| Filter | Primary Filter<br>G4 filter to trap larger particles HEPA Filter HEPA (H13) filter eliminates 99.99% of particles 0.1 microns in diameter Activated Carbon and Patented Nanomaterial Filter | Pre-Filter Synthetic screen mesh type added prior to HEPA Designed to trap larger particles and keep them out of the HEPA Dimensions: 14 in x 15 in x 0.125 in HEPA Filter MERV 13 Dimensions: 14 in x 15 in x 0.6875 in Catalytic Filter Patent Pending Hybrid Oxydizer with proprietary Dualaction Catalyst Carbon/Cold Catalyst Oxidier Filter Dimensions: 14' x 15" x 0.625" | Similar |
| Photocatalyst | Titanium Dioxide | Proprietary Catalyst | Similar |
| Light Source | UV Type: UV-C UV Light Source: LED | UV-C Light Source: LED Wavelength: 253.7nm | Identical |
| | • Wavelength: 254.7nm<br>• Total of 36 6W ultraviolet tubes with<br>single light intensity of 92µW/cm²<br>• Total light intensity: 552µW/cm² | • Total of 2 UV- C tube lamps (1 per<br>side)<br>• Total UV Power: 8.0 W | |
| Air Source | Centrifugal Fan | Centrifugal Fan | Identical |
| Flow Control | Three speeds (low, medium, high) | 4 speeds (low, medium, high, boost) and<br>auto mode provide up to 370 CFM | Similar |
| Device Air Changes Per<br>Hour (ACH) | 13 ACH on high fan speed in a 3000 ft3<br>room | 5.5 ACH on high fan speed (speed 4), in a<br>4000 ft3 room | Similar |
| Fan Exposure<br>Safety Features | Non-removable grill at air output to prevent<br>the user from accessing spinning fan<br>without tools. Safety feature confirmed by<br>UL 507. | Non-removable grill at air output and the<br>switch safety feature at the inlet prevent<br>the user from accessing spinning fan<br>without tools. Safety feature confirmed by<br>ETL tested to UL 507. | Similar |
| UV Light Exposure<br>Safety Features | Touchscreen monitor allows user to switch<br>off the UV lights prior to opening filter<br>doors. Besides, UV lights are also located at<br>the inner side of the machine which<br>prevents direct UV exposure. | There are two sets of Safety switches on<br>the Aura Storm. The first is on both outer<br>doors where the magnetic switch will<br>disengage and the unit will not turn on.<br>A secondary switch in the Aura Storm<br>filter and if the filter is either improperly<br>installed or<br>the filter is missing, the unit will not<br>operate. The unit will not operate with a<br>generic filter and an Invictus filter must be<br>used for the system to work. These<br>switches have been designed to protect<br>the user from any possibility of exposure<br>to direct contact with UV light. Safety<br>feature confirmed by ETL to UL 507<br>safety standard. | Similar |
| Input Voltage | 110V ±10% | 120 V | Identical |
| Current | 3.6 Amps | 0.55 Amps | Similar |
| Power Consumption | 783W ±5W | Up to 65 W | Similar |
| Dimensions | Unit dimension:<br>16.8"W x 26.4"L x 80.7"H<br>HEPA Filter (cylindrincal):<br>9.8" diameter x 10.0"H<br>Primary Filter:<br>23.0"x 6.6" x 1.2"<br>Activated Carbon / Nanomaterial Filter:<br>23.4" x 10.7" x 1.1" | Outer frame dimensions:<br>23in(H) x 18.2in(W) x 10.6in(L)<br>Filter dimensions:<br>Pre-Filter:<br>14 in x 15 in x 0.125 in<br>HEPA Filter:<br>14 in x 15 in x 0.6875 in<br>Catalytic Filter:<br>14 in x 15 in x 0.1875 in<br>Carbon/Cold Catalyst<br>Oxidizer Filter:<br>14 in x 15in x 0.625 in | Similar |
| Conformance with<br>Standards | UL 507 Standard for<br>Electrical Fans<br>IEC 60601-1 Basic Safety<br>and Essential Performance<br>IEC 60601-1-2 EMC. EMC for<br>Medical Devices<br>IEST-RP-CC001.6 HEPA and<br>ULPA Filters | UL 507 Standard for<br>Electrical Fans<br>IEC 60601-1-2 EMC<br>EMC for Medical<br>Devices | Similar |
| Device Performance<br>microorganism | AirKEE P900 has been demonstrated to<br>eliminate the following test microorganisms<br>entrained on the filter of the subject device<br>under the following exposure conditions:<br>Staphylococcus aureus<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed; | The Aura Storm air purifier has been<br>demonstrated to destroy<br>the following bacteria: Staphylococcus<br>albicans.<br>Staphylococcus aureus,<br>and Escherichia Coli; and virus: A/PR8/34<br>H1N1 virus entrained on the filter of the | Similar |
| | | | |
| | Escherichia coli<br>Log 4 (99.99%)/60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Klebsiella pneumoniae<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Pseudomonas aeruginosa<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Serratia marcescens<br>Log 4 (99.99%)/60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Aspergillus Niger<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed / 120 minutes at low fan speed;<br>Influenza A virus, H1N1<br>Log 4 (99.99%)/ 60 minutes at high fan<br>speed<br>Influenza A virus, H3N3<br>Log 4 (99.99%)/ 120 minutes at low fan<br>speed | subject device under the following<br>exposure conditions:<br>Average Maximum log<br>reduction / entrainment time (minutes) at<br>Fan Speed 4. Room Temperature test: Log 4<br>(99.99%) / 60 minutes.<br>Average Maximum log<br>reduction / entrainment time (minutes) at<br>Fan Speed 1. Room Temperature test: Log 4<br>(99.99%) / 120 minutes. | |
| Device Performance<br>Electrical and EMC | UL 507 Standard for<br>Electrical Fans<br><br>IEC 60601-1-2 EMC. EMC for<br>Medical Devices<br><br>IEST-RP-CC001.6 HEPA and<br>ULPA Filters | UL 507 Standard for<br>Electrical Fans<br><br>IEC 60601-1-2 EMC.<br>EMC for Medical Devices.<br><br>ISO 29463 H13 ISO 35H<br>Filtration testing | Similar |
| Device Performance<br>Ozone release level | UL 867 - Electrostatic Air Cleaners | Unknown, not mentioned | Different<br>AirKEE P900 has<br>done additional |
| | | | ozone testing to<br>demonstrate the<br>AirKEE P900 unit is<br>able to operate at<br>less than 0.05 ppm a<br>its highest fan speed. |
## Comparison of Technological Characteristics with Predicate Device:
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
## Summary of Non-Clinical Tests
| Test Name | Applicable Standards | Purpose | Acceptance Criteria | Results |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|----------------------------------------------------------------------------------------------------------------|
| Microorganism<br>Performance | Internal Standards | To understand the log reduction<br>rate for Staphylococcus aureus,<br>Escherichia coli, Klebsiella<br>pneumoniae, Pseudomonas<br>aeruginosa, Serratia marcescens,<br>Aspergillus Niger, and<br>Influenza A virus. | 4 Log reduction (99.99%) | 4 Log reduction in 60<br>minutes at high fan speed;<br>and 4 Log reduction in 120<br>minutes at low fan speed. |
| Fractional<br>Efficiency | IEST-RP-CC001.6-<br>HEPA and ULPA<br>Filters | Fractional efficiency testing was<br>performed on the AirKEE filter<br>per IEST-RP-CC001.6 Type H<br>test to determine the fractional<br>efficiency percentage of varying<br>size ranges. | Per Standard particles | Filter fractional efficiency<br>percentage of 99.99% at 0.1-<br>0.2µm. |
| Electrical safety<br>and<br>electromagnetic<br>compatibility<br>testing | UL 507 Standard for<br>Electrical Fans<br>IEC 60601-1-2 EMC.<br>EMC for Medical<br>Devices | General requirements for basic<br>safety and essential<br>performance including UV light<br>leakage and intensity | Per Standard | Pass |
| Ozone | UL 867 - Electrostatic<br>Air Cleaners | Ozone testing was performed<br>per UL 867 by monitoring the<br>ozone concentration in a test | Per Standard | Testing demonstrates the<br>AirKEE P900 unit is able to |
{14}------------------------------------------------
| | chamber at the highest fan | operate at less than 0.05 ppm |
|--|----------------------------|-------------------------------|
| | speed. | at its highest fan speed. |
{15}------------------------------------------------
### Conclusion
The conclusions drawn from the non-clinical testing demonstrate that the subject device, AirKEE P900 is as safe, as effective, and performs as well as or better than the legally marketed predicate, K212644 Class II (21 CFR 880.6500), product code FRA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.