Software as a Medical Device, Therapeutic, Pediatric
Indications for Use
VERVE is indicated for the treatment for convergence insufficiency in patients aged 11-35 under the supervision of a medical professional by providing therapeutic visual procedures in a virtual reality setting.
Device Story
VERVE is a software-as-a-medical-device (SaMD) providing vision therapy for convergence insufficiency (CI) via virtual reality. The system utilizes an off-the-shelf tablet and a third-party VR headset with integrated eye tracking. The device delivers therapeutic visual procedures that modify vergence demand, mirroring office-based vergence and accommodative therapy (OBVAT). During sessions, the headset tracks eye movement to measure treatment outcomes. Data is processed on a cloud-based server, with results displayed on the tablet for the supervising medical professional. The output allows clinicians to monitor patient progress and symptom reduction, facilitating clinical decision-making regarding the continuation or adjustment of therapy. The device benefits patients by providing a structured, VR-based environment to improve near point of convergence and fusional vergence, thereby reducing CI-related symptoms.
Clinical Evidence
Prospective clinical study of 25 patients (ages 10-33) with CI. Treatment: 1 hour, twice weekly for 6 weeks using VERVE v3.0.1. Primary endpoints: Near point of convergence (NPC), amplitude of accommodation, positive/negative fusional vergence, and CISS score. Results: NPC improved from 10.8±3.8 cm to 4.9±1.7 cm; positive fusional vergence improved from 11.4±5.7 Δ to 19.2±6.6 Δ; CISS scores improved from 37.2±7.8 to 18.7±6.8. 72% achieved success (normal ranges), 28% improved, 0% non-responders.
Technological Characteristics
SaMD utilizing virtual reality. Hardware: off-the-shelf tablet and third-party VR headset with integrated eye tracking. Connectivity: cloud-based server for data analysis. Software development follows IEC 62304. Treatment principle: modification of disparity vergence demand.
Indications for Use
Indicated for treatment of convergence insufficiency in patients aged 11-35 under medical professional supervision.
Regulatory Classification
Identification
A fixation device is an AC-powered device intended for use as a fixation target for the patient during ophthalmological examination. The patient directs his or her gaze so that the visual image of the object falls on the fovea centralis (the center of the macular retina of the eye.)
{0}
![LOGO]
OculoMotor Technologies
% Allison Komiyama, Ph.D., RAC
VP, MedTech Innovation
RQM+
2251 San Diego Ave., Suite B-257
San Diego, California 92110
Re: K232930
Trade/Device Name: VERVE
Regulation Number: 21 CFR 886.1290
Regulation Name: Fixation Device
Regulatory Class: Class I
Product Code: SBN
Dated: May 7, 2024
Received: May 7, 2024
Dear Allison Komiyama, Ph.D., RAC:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K232930 - Allison Komiyama, Ph.D., RAC
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{2}
K232930 - Allison Komiyama, Ph.D., RAC
Page 3
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K232930
Device Name
VERVE
Indications for Use (Describe)
VERVE is indicated for the treatment for convergence insufficiency in patients aged 11-35 under the supervision of a medical professional by providing therapeutic visual procedures in a virtual reality setting.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
OculoMotor Technologies
## 510(k) Summary
K232930
DATE PREPARED
June 11, 2024
MANUFACTURER AND 510(k) OWNER
OculoMotor Technologies
155 Village Boulevard, Suite 130
Princeton, NJ 08540
Telephone: +1 (973) 558-5456
Official Contact: Chang Yaramothu, PhD, CTO
REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, PhD, RAC
RQM+
Telephone: +1 (412) 816-8253
Email: akomiyama@rqmplus.com
DEVICE INFORMATION
Proprietary Name/Trade Name: VERVE
Common Name: Fixation Device
Regulation Number: 21 CFR 886.1290
Class: Class I
Product Code: SBN
Premarket Review: Ophthalmic Devices (DHT1A)
Review Panel: Ophthalmic
PREDICATE DEVICE IDENTIFICATION
VERVE is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
| --- | --- | --- |
| K051684 | EYEPORT Vision Training System/ Exercise Your Eyes Inc. | ✓ |
The predicate device has not been subject to a design related recall.
Page 1
{5}
OculoMotor Technologies
## **DEVICE DESCRIPTION**
Virtual Eye Rotation Vision Exercises (VERVE) is a software as a medical device (SaMD) intended to provide vision therapy using virtual reality. The device is indicated for prescription use in patients 11 years or older who have convergence insufficiency (CI). VERVE is provided pre-installed on an off-the-shelf tablet and a third-party headset with integrated eye tracking. Data analysis is performed on VERVE's cloud-based server and results are displayed on the tablet.
VERVE uses virtual reality to incorporate effective elements from office-based vergence and accommodative therapy (OBVAT) to treat CI. Traditional OBVAT includes modifying the amount of eye rotation required to complete each exercise, also known as disparity vergence demand. VERVE can modify the vergence demand within a session while the user engages with the software device.
## **INDICATIONS FOR USE**
VERVE is indicated for the treatment for convergence insufficiency in patients aged 11-35 under the supervision of a medical professional by providing therapeutic visual procedures in a virtual reality setting.
## **COMPARISON OF TECHNOLOGICAL CHARACTERISTICS**
OculoMotor Technologies (OMT) believes VERVE is substantially equivalent to the predicate device based on the information summarized here:
The subject and predicate devices have the same intended use and employ the same fundamental treatment methodology. Both are fixation devices intended to treat convergence insufficiency. The main differences between the subject and predicate devices are the intended use population and their physical designs. The subject device is indicated specifically for patients 11 years of age and older whereas the predicate does not specify pediatric use. Additionally, the subject device relies on a virtual reality headset and tablet to administer treatment while the predicate device uses alternating red and blue lights on a long plastic rod. The subject device uses the headset to perform eye tracking during treatment to measure treatment outcomes whereas the predicate device does not have that feature.
## **SUMMARY OF NON-CLINICAL TESTING**
No FDA performance standards have been established for VERVE. The following tests were performed to demonstrate safety based on current industry standards:
### *Software Verification*
The software development and testing were executed with consideration to IEC 62304 *Medical device software — Software life cycle processes*.
Page 2
{6}
OculoMotor^{}[] Technologies
## SUMMARY OF CLINICAL TESTING
OMT conducted a clinical study to evaluate VERVE's performance in the treatment of convergence insufficiency. Patients diagnosed with convergence insufficiency were enrolled into the VERVE clinical trial. All 25 patients (ages 10 to 33) completed the study and no dropouts occurred. Treatment (using VERVE software version 3.0.1 on the PICO Neo3 Pro Eye headset) was administered twice a week for 6 weeks, and each treatment session lasted 1 hour. Near point of convergence, amplitude of accommodation, positive fusional vergence, negative fusional vergence, and the convergence insufficiency symptom score (CISS) were recorded before and after treatment.
Results demonstrated VERVE significantly improved near point of convergence (baseline was $10.8 \pm 3.8$ cm, outcome was $4.9 \pm 1.7$ cm, change was $5.9 \pm 4.0$ cm; normal is considered 6 cm or less measured from the bridge of the nose along the midline) and positive fusional vergence (baseline was $11.4 \pm 5.7$ $\Delta$, outcome was $19.2 \pm 6.6$ $\Delta$, change was $7.8 \pm 10.5$ $\Delta$; normal is considered 15 $\Delta$ or more) and improvements were clinically relevant. VERVE also significantly improved visual symptoms assessed through the CISS (baseline was $37.2 \pm 7.8$ points, outcome was $18.7 \pm 6.8$ points; a score of 21 points or greater is considered clinically symptomatic for young adults (16 points for pediatric) and the higher the score the more symptomatic the patient).
Treatment success in the study was defined as reaching normal ranges for near point of convergence and positive fusional vergence. Overall, 72% (18 out of 25 patients) were categorized as having achieved success, 28% (7 out of 25 patients) were categorized as having improved, and 0% (0 out of 25 patients) were categorized as non-responders.
## CONCLUSION
Based on the testing performed, including bench testing, a clinical evaluation and software verification and validation, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use and performance characteristics for the proposed VERVE are assessed to be substantially equivalent to the predicate device.
Page 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.