# enspire 300 Series Automated Endoscope Reprocessor (K232918)

_STERIS Corporation · NZA · Oct 3, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232918

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [NZA](/productcode/NZA.md)
- **Decision Date:** Oct 3, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The enspire™ 300 Series Automated Endoscope Reprocessor System is intended to effectively provide a pressure and channel monitor, clean, high-level disinfect, rinse and air purge validated immersible, reusable, semi-critical, heat sensitive medical devices including, but not limited to, flexible endoscopes and non-channeled naso-endoscopes. The validated cleaning process replaces manual cleaning for endoscopes other than duodenoscopes. Manual Cleaning of endoscopes is required prior to placement in high level disinfection only cycle. The enspire 300 Series Automated Endoscope Reprocessor System uses Revital-Ox PAA High Level Disinfectant to provide high level disinfection of validated immersible, reusable, semi-critical, heat sensitive medical devices. It automatically mixes the Part B solutions, high level disinfects the load during a controlled cycle and rinses the load. The wash phase of the enspire 300 Series Automated Endoscope Reprocessor System cycle uses only Revital-Ox 2X Concentrate Enzymatic Detergent to perform cleaning. The Revital-Ox PA High Level Disinfectant (HLD) is a two-part solution, which when mixed, is intended to provide high level disinfection of validated immersible, semi-critical, heat sensitive medical devices including, but not limited to, flexible endoscopes and non-channeled nasoendoscopes, when used in the enspire 300 AER.

## Device Story

Automated endoscope reprocessor (AER) for cleaning and high-level disinfection of flexible endoscopes/accessories. User inputs scope profile, operator/case/physician IDs; system performs automated pressure monitoring and channel flow coefficient evaluation against stored reference values. System uses Revital-Ox 2X Concentrate Enzymatic Detergent for wash phase and Revital-Ox PAA High-Level Disinfectant for disinfection. Automated injection of chemicals monitored by flow meter; rinse with 0.2-micron filtered water; final air purge with oil-free filtered compressed air. Microprocessor-controlled; provides cycle documentation. Used in clinical settings to ensure device safety for reuse; reduces manual labor and potential for human error in reprocessing; maintains device integrity via continuous pressure monitoring.

## Clinical Evidence

Bench testing only. Performance testing confirmed that the replacement pre-filter material does not negatively impact device performance compared to the predicate.

## Technological Characteristics

Microprocessor-controlled AER. Materials: stainless steel, silicone, polypropylene, PVDF. Uses Revital-Ox 2X Enzymatic Detergent and Revital-Ox PAA HLD. Features: automated pressure monitoring, channel flow monitoring, 0.2-micron water filtration, oil-free compressed air purge. Connectivity: barcode scanner, touchscreen interface, optional external printer. Cycle parameters (contact time, temp, volume) are fixed and non-alterable by operator.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- enspire 300 Series Automated Endoscope Reprocessor ([K230560](/device/K230560.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 3, 2023

Steris Corporation Nalepka Jennifer Manager, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060

Re: K232918

Trade/Device Name: enspire 300 Series Automated Endoscope Reprocessor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NZA Dated: September 19, 2023 Received: September 19, 2023

Dear Nalepka Jennifer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K232918

Device Name

enspire 300 Series Automated Endoscope Reprocessor

Indications for Use (Describe)

The enspire™ 300 Series Automated Endoscope Reprocessor System is intended to effectively provide a pressure and channel monitor, clean, high-level disinfect, rinse and air purge validated immersible, reusable, semi-critical, heat sensitive medical devices including, but not limited to, flexible endoscopes and non-channeled naso-endoscopes. The validated cleaning process replaces manual cleaning for endoscopes other than duodenoscopes. Manual Cleaning of endoscopes is required prior to placement in high level disinfection only cycle.

The enspire 300 Series Automated Endoscope Reprocessor System uses Revital-Ox PAA High Level Disinfectant to provide high level disinfection of validated immersible, reusable, semi-critical, heat sensitive medical devices. It automatically mixes the Part B solutions, high level disinfects the load during a controlled cycle and rinses the load. The wash phase of the enspire 300 Series Automated Endoscope Reprocessor System cycle uses only Revital-Ox 2X Concentrate Enzymatic Detergent to perform cleaning.

The Revital-Ox PA High Level Disinfectant (HLD) is a two-part solution, which when mixed, is intended to provide high level disinfection of validated immersible, semi-critical, heat sensitive medical devices including, but not limited to, flexible endoscopes and non-channeled nasoendoscopes, when used in the enspire 300 AER.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for STERIS. The word "STERIS" is written in a bold, sans-serif font. Below the word is an image of several horizontal, wavy lines stacked on top of each other. The lines are blue. 

## 510(k) Summary K232918 enspire 300 Series Automated Endoscope Reprocessor

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600

Contact: Jennifer Nalepka Manager, Regulatory Affairs Tel: 440-392-7458

Summary Date: October 2, 2023

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{5}------------------------------------------------

#### Device Name 1.

Trade Name:

Device Classification: Common/usual Name: Classification Name: Classification Number: Product Code:

enspire 300 Series Automated Endoscope Reprocessor Class II Endoscope Cleaner and Reprocessor Accessories, Germicide, Cleaning, For Endoscopes 21 CFR 876.1500 NZA

#### 2. Predicate Device

enspire 300 Series Automated Endoscope Reprocessor, K230560

#### Description of Device 3.

The enspire 300™ Series Automated Endoscope Reprocessor (AER) is a medical device processing system used for cleaning and high level disinfection of immersible, reusable, semi-critical, heat-sensitive devices such as flexible endoscopes and their accessories. The system consists of the enspire 300 AER. Revital-Ox 2X Concentrate Enzymatic Detergent (R2X), and Revital-Ox Peracetic Acid High Level Disinfectant.

The enspire 300 series AER is an automated, self-contained device for the effective cleaning and high level disinfection of semi-critical medical devices and their accessories. Prior to placement in the processor, users will be instructed to perform bedside (point of use) cleaning and manual leak testing. The devices will not require manual pre-cleaning prior to processing, with the exception of duodenoscopes which will still require manual cleaning per the manufacturer's written instructions. Channel Connectors, as identified in STERIS labeling, are used to facilitate the delivery of the R2X detergent, HLD solution and rinse water to internal channels of devices that have them.

On the first process of an endoscope, the user must create a scope profile for the endoscope in the AER. The creation of the scope profile saves key endoscope attributes to the AER which are used to monitor the cycle during processing of the device. Once the device is positioned in the enspire 300 AER and optional operator ID, case ID, procedure ID and physician information is inputted, the operator initiates the processing cycle during which the Processor will create and maintain the conditions necessary for effective cleaning and high level disinfection. At the beginning of the processing cycle, an automated pressure monitor is performed to confirm the integrity of the flexible endoscope. In parallel with the pressure monitor and prior to initiation of the cleaning phase, the processing system will evaluate the flow coefficient of each individual channel and compare to their respective reference value stored in the device profile created prior to first processing. The enspire 300 AER maintains inflation of the processed device throughout the process to prevent ingress of fluid in the event of any loss of integrity. At the end of the processing cycle, the cleaned and high level disinfected devices are rinsed with 0.2 micron filtered potable water followed by a dried, oil free, filtered compressed air purge to evacuate rinse water from the endoscope channels. The AER, which is micro-processor controlled and continually monitored, provides documentation of each cycle.

{6}------------------------------------------------

The enspire 300 AER utilizes Revital-Ox 2X Concentrate Enzymatic Detergent for cleaning and Revital-Ox Peracetic Acid High Level Disinfectant for high level disinfection. Revital-Ox PAA HLD is a two part solution consisting of Part A, 15% PAA, and Part B, conditioner.

#### 4. Indications for Use

The enspire 300 Automated Endoscope Reprocessor is intended to effectively provide a pressure and channel monitor, clean, high-level disinfect, rinse and air purge validated immersible, reusable, semi-critical, heat sensitive medical devices including, but not limited to, flexible endoscopes and non-channeled nasoendoscopes. The validated cleaning process replaces manual cleaning for endoscopes other than duodenoscopes. Manual Cleaning of endoscopes is required prior to placement in high level disinfection only cycle.

The enspire 300 Automated Endoscope Reprocessor uses Revital-Ox PA High Level Disinfectant to provide high level disinfection of validated immersible, reusable, semi-critical, heat sensitive medical devices. It automatically mixes the Part A and Part B solutions; high level disinfects the load during a controlled cycle and rinses the load. The wash phase of the enspire 300 Series Automated Endoscope Reprocessor cycle uses only Revital-Ox 2X Concentrate Enzymatic Detergent to perform cleaning.

The Revital-Ox PA High Level Disinfectant (HLD) is a two-part solution, which when mixed, is intended to provide high level disinfection of validated immersible, reusable, semi-critical, heat sensitive medical devices including, but not limited to, flexible endoscopes and non-channeled naso-endoscopes, when used in the enspire 300 AER.

#### Technological Characteristic Comparison Tables ട്.

| Feature                                 | Proposed<br>enspire 300 Series AER                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate<br>enspire 300 Series AER<br>(K230560)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison                                                                      |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Indications for<br>Use                  | The enspire 300 Automated<br>Endoscope Reprocessor is<br>intended to effectively provide a<br>pressure and channel monitor,<br>clean, high-level disinfect, rinse<br>and air purge validated<br>immersible, reusable, semi-<br>critical, heat sensitive medical<br>devices including, but not<br>limited to, flexible endoscopes<br>and non-channeled naso-<br>endoscopes. The validated<br>cleaning process replaces<br>manual cleaning for endoscopes<br>other than duodenoscopes.                                                                                                                                                                              | The enspire 300 Automated<br>Endoscope Reprocessor is<br>intended to effectively provide a<br>pressure and channel monitor,<br>clean, high-level disinfect, rinse<br>and air purge validated<br>immersible, reusable, semi-<br>critical, heat sensitive medical<br>devices including, but not<br>limited to, flexible endoscopes<br>and non-channeled naso-<br>endoscopes. The validated<br>cleaning process replaces<br>manual cleaning for endoscopes<br>other than duodenoscopes.                                                                                                                                                                                                | Identical                                                                       |
| Feature                                 | Proposed<br>enspire 300 Series AER                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate<br>enspire 300 Series AER<br>(K230560)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison                                                                      |
|                                         | Manual Cleaning of endoscopes<br>is required prior to placement in<br>high level disinfection only<br>cycle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Manual Cleaning of endoscopes<br>is required prior to placement in<br>high level disinfection only<br>cycle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                 |
|                                         | The enspire 300 Automated<br>Endoscope Reprocessor uses<br>Revital-Ox PA High Level<br>Disinfectant to provide high<br>level disinfection of validated<br>immersible, reusable, semi-<br>critical, heat sensitive medical<br>devices. It automatically mixes<br>the Part A and Part B solutions;<br>high level disinfects the load<br>during a controlled cycle and<br>rinses the load. The wash phase<br>of the enspire 300 Series<br>Automated Endoscope<br>Reprocessor cycle uses only<br>Revital-Ox 2X Concentrate<br>Enzymatic Detergent to perform<br>cleaning.                                                                                             | The enspire 300 Automated<br>Endoscope Reprocessor uses<br>Revital-Ox PA High Level<br>Disinfectant to provide high<br>level disinfection of validated<br>immersible, reusable, semi-<br>critical, heat sensitive medical<br>devices. It automatically mixes<br>the Part A and Part B solutions;<br>high level disinfects the load<br>during a controlled cycle and<br>rinses the load. The wash phase<br>of the enspire 300 Series<br>Automated Endoscope<br>Reprocessor cycle uses only<br>Revital-Ox 2X Concentrate<br>Enzymatic Detergent to perform<br>cleaning.                                                                                                               |                                                                                 |
| Operating<br>Principles /<br>Technology | The enspire 300 AER provide<br>delivery of solutions and fluids<br>to endoscopes and their<br>accessories. Revital-Ox 2X<br>Concentrate Enzymatic<br>detergent and Revital-Ox PAA<br>HLD are introduced into the<br>enspire 300 AER to provide<br>cleaning and high level<br>disinfection. Decontamination<br>cycle is used to help with<br>routine maintenance and to help<br>prevent contamination of the<br>Reprocessor during period of<br>inactivity. Descaling Cycle is<br>used to help prevent and remove<br>scale build-up in the<br>Reprocessor.                                                                                                         | The enspire 300 AER provide<br>delivery of solutions and fluids<br>to endoscopes and their<br>accessories. Revital-Ox 2X<br>Concentrate Enzymatic<br>detergent and Revital-Ox PAA<br>HLD are introduced into the<br>enspire 300 AER to provide<br>cleaning and high level<br>disinfection. Decontamination<br>cycle is used to help with<br>routine maintenance and to help<br>prevent contamination of the<br>Reprocessor during period of<br>inactivity. Descaling Cycle is<br>used to help prevent and remove<br>scale build-up in the<br>Reprocessor.                                                                                                                           | Identical                                                                       |
| Process<br>Parameters                   | The cycle parameters cannot be<br>altered by operator. The critical<br>process parameters are:<br>● Contact Time<br>● Use Dilution Temp<br>● Water Volume<br>● Chemical Volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The cycle parameters cannot be<br>altered by operator. The critical<br>process parameters are:<br>● Contact Time<br>● Use Dilution Temp<br>● Water Volume<br>● Chemical Volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                       |
| Feature                                 | Proposed<br>enspire 300 Series AER                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate<br>enspire 300 Series AER<br>(K230560)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison                                                                      |
|                                         | • Channel irrigation<br>flow/pressure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | • Channel irrigation<br>flow/pressure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                 |
| Process<br>Monitors                     | Channel irrigation flow alarms<br>when endoscope channel is<br>obstructed or disconnected, if<br>pressure too low or too high, or<br>if sensor defect is detected.<br>Chemical volume alarms if<br>chemical injected volume is too<br>high, too low, chemical pump is<br>defective or chemical container<br>has inadequate volume to<br>complete the cycle. Water<br>volume alarms when water level<br>is too high for HLD.<br>Temperature alarm is triggered<br>when temperature is too high,<br>sump takes too long to heat,<br>RTD variance exceeded or a<br>sensor is defective. Alarm is<br>triggered if Real Time Clock<br>error is detected.               | Channel irrigation flow alarms<br>when endoscope channel is<br>obstructed or disconnected, if<br>pressure too low or too high, or<br>if sensor defect is detected.<br>Chemical volume alarms if<br>chemical injected volume is too<br>high, too low, chemical pump is<br>defective or chemical container<br>has inadequate volume to<br>complete the cycle. Water<br>volume alarms when water level<br>is too high for HLD.<br>Temperature alarm is triggered<br>when temperature is too high,<br>sump takes too long to heat,<br>RTD variance exceeded or a<br>sensor is defective. Alarm is<br>triggered if Real Time Clock<br>error is detected.                                 | Identical                                                                       |
| Design<br>Features                      | • Intended for use with<br>Revital-Ox 2X Concentrate<br>Enzymatic Detergent and<br>Revital-Ox PAA HLD<br>• Microprocessor controlled<br>• Internal components<br>constructed of stainless steel,<br>silicone, polypropylene and<br>PVDF.<br>• Processor provides pre-<br>filtered water and the<br>proposed filters have a new<br>support material.<br>• 0.2 micron filtered water<br>used for rinsing<br>• Automated injection of<br>cleaning and HLD solutions,<br>with accuracy monitored by a<br>flow meter<br>• Uses dried, oil free and<br>filtered compressed air for<br>Air Purge<br>• Automated endoscope<br>Pressure monitor<br>• Monitors channel flow | • Intended for use with<br>Revital-Ox 2X Concentrate<br>Enzymatic Detergent and<br>Revital-Ox PAA HLD<br>• Microprocessor controlled<br>• Internal components<br>constructed of stainless steel,<br>silicone, polypropylene and<br>PVDF.<br>• Processor provides pre-<br>filtered water<br>• 0.2 micron filtered water<br>used for rinsing<br>• Automated injection of<br>cleaning and HLD solutions,<br>with accuracy monitored by a<br>flow meter<br>• Uses dried, oil free and<br>filtered compressed air for<br>Air Purge<br>• Automated endoscope<br>Pressure monitor<br>• Monitors channel flow<br>• Includes a bar code scanner;<br>employs touchscreen display<br>interface | Similar - function<br>and<br>biocompatibility of<br>pre-filter is<br>unchanged. |
| Feature                                 | Proposed<br>enspire 300 Series AER                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate<br>enspire 300 Series AER<br>(K230560)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison                                                                      |
|                                         | Includes a bar code scanner; employs touchscreen display interface Separate, optional printer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Separate, optional printer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                 |
| Cycle Parameters                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Comparison                                                                      |
| Pressure<br>Monitor                     | Performed at the beginning of<br>the cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Performed at the beginning of<br>the cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                       |
| Flow check                              | Performed at the beginning and<br>end of the cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Performed at the beginning and<br>end of the cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                       |
| Cleaning<br>Phase<br>Temperature        | 50-55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 50-55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                       |
| Cleaning<br>Phase<br>Exposure time      | 4.5 minutes wash                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 4.5 minutes wash                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                       |
| Rinse phase<br>after cleaning           | 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                       |
| HLD phase<br>Temperature                | 50-55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 50-55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                       |
| HLD phase<br>exposure time              | 3 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 3 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                       |
| Rinse Phase<br>after HLD<br>phase       | 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 30 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                       |
| Number of<br>rinses                     | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                       |
| Air purge                               | Performed between each phase<br>and at the end of the cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Performed between each phase<br>and at the end of the cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                       |
| Accessories                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Comparison                                                                      |
| Detergent                               | Revital-Ox 2X Concentrate<br>Enzymatic Detergent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Revital-Ox 2X Concentrate<br>Enzymatic Detergent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                       |
| HLD                                     | Revital-Ox PAA HLD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Revital-Ox PAA HLD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                       |
| Chemical<br>Indicator                   | enspire 300 series AER Process<br>monitor. Peracetic acid dose<br>indicator for routine monitoring<br>of enspire 300 AER using<br>Revital-Ox PAA HLD.<br>Chemical reaction on indicator<br>pad to produce color change.                                                                                                                                                                                                                                                                                                                                                                                                                                           | enspire 300 series AER Process<br>monitor. Peracetic acid dose<br>indicator for routine monitoring<br>of enspire 300 AER using<br>Revital-Ox PAA HLD.<br>Chemical reaction on indicator<br>pad to produce color change.                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                       |
| Scope<br>Connectors /<br>Flow Units     | Scope connectors are required<br>for all lumens of endoscopes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Scope connectors are required<br>for all lumens of endoscopes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                       |
| Operator<br>Maintenance                 | Periodic replacement of water<br>filters. Periodic replacement of<br>printer tape if using the external<br>printer option. Running the<br>decontamination cycle every 24<br>hours.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Periodic replacement of water<br>filters. Periodic replacement of<br>printer tape if using the external<br>printer option. Running the<br>decontamination cycle every 24<br>hours.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                       |

## Table 1. Predicate Device Comparison Table

{7}------------------------------------------------

## STERIS Traditional 510(k) PREMARKET NOTIFICATION enspire 300 Series Automated Endoscope Reprocessor

{8}------------------------------------------------

## STERIS Traditional 510(k) PREMARKET NOTIFICATION enspire 300 Series Automated Endoscope Reprocessor

{9}------------------------------------------------

## STERIS Traditional 510(k) PREMARKET NOTIFICATION enspire 300 Series Automated Endoscope Reprocessor

{10}------------------------------------------------

#### 6. Summary of Non-Clinical Testing

Shown in Table 2 is the new testing that was performed to evaluate the modified device.

| Verification Activity | Description |
|-----------------------|-------------|
|-----------------------|-------------|

| Test                                                | Acceptance Criteria                                                | Result |
|-----------------------------------------------------|--------------------------------------------------------------------|--------|
| Performance testing with<br>replacement pre-filters | The modification does not affect the<br>performance of the device. | Pass   |

#### 7. Conclusion

Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device (K230560), Class II (21 CFR 876.1500), product code NZA.

---

**Source:** [https://fda-staging.innolitics.com/device/K232918](https://fda-staging.innolitics.com/device/K232918)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).