K232888 · Suzhou Haishen Medical Device Associates Co., Ltd. · ETN · Dec 21, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K232888
Device Name
Disposable Laryngeal Electrodes
Applicant
Suzhou Haishen Medical Device Associates Co., Ltd.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Dec 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The disposable laryngeal electrodes are intended to be used as a disposable, self-adhesive electrodes attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures.
Device Story
Disposable laryngeal electrodes consist of conductive polymer film with medical-grade ink, polypropylene connector, and cable assembly. Electrodes attach to surface of endotracheal tube; neutral adhesive electrode placed on patient shoulder. Device connects to external nerve monitor/EMG unit. Used in OR during thyroid, anterior cervical, carotid endarterectomy, and brain surgeries. Monitors recurrent laryngeal nerve and vagus nerve activity. Provides continuous EMG signal input to nerve monitor; healthcare provider observes EMG output on monitor to assess nerve integrity during surgery. Helps prevent nerve damage by providing real-time feedback to surgeon.
Clinical Evidence
Bench testing only. Evaluated adhesive property, electrode impedance (≤100 Ω), and electrical safety (IEC 60601-1). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity) confirmed non-cytotoxic and non-irritating. Sterilization validated per ISO 11135 (SAL 10^-6).
Technological Characteristics
Medical-grade ink on polyester substrate, conductive polymer film, polypropylene connector. 2 or 4 channel configurations. Non-invasive. ETO sterilization (ISO 11135). Electrical safety per IEC 60601-1. Biocompatibility per ISO 10993.
Indications for Use
Indicated for continuous EMG monitoring of the larynx during surgical procedures, including thyroid, anterior cervical, carotid endarterectomy, and brain surgeries, in patients requiring endotracheal intubation.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
December 21, 2023
Suzhou Haishen Medical Device Associates Co., Ltd % Xiaoqing Xue Registration Engineer Sinow Medical AS Høyteknologisenteret Thormøhlens Gate 55 Bergen, 5008 Norway
Re: K232888
Trade/Device Name: Disposable Laryngeal Electrodes Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: November 24, 2023 Received: November 24, 2023
Dear Xiaoqing Xue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232888
Device Name Disposable Laryngeal Electrode
Indications for Use (Describe)
The disposable laryngeal electrodes are intended to be used as a disposable, self-adhesive electrodes attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures.
Type of Use (Select one or both, as applicable)
| <label><input checked="true" type="checkbox"/><span>For activities under 21 CFR 312.2, Sponsor-Investigator</span></label> | <label><input type="checkbox"/><span>Compassionate Use, including Treatment Use</span></label> |
|----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
# K232888
#### 1. Submitter
Submitter: Suzhou Haishen United Medical Device Associates Co., Ltd Address: 3F, 4F 401, 4F 402, 5F, Building 10, No. 168 Majian Road, Suzhou, Jiangsu Province, China Telephone: +86 18921623557 Email: 785445240@gq.com Contact Person: Leyi Dai Date Prepared: Dec.15, 2023
#### 2. Designated Submission Correspondent
Company: Sinow Medical AS Address: Vestre Fantoftåsen 44, 5072, Bergen, Norway Contact Person: Xiaoqing Xue Telephone: +86 15161196032 Email: xue(@bergemed.com
## 3. Subject Device
Trade name: Disposable Laryngeal Electrodes 510(K) number: K232888 Regulatory Class: II Regulation Number: 21 CFR 874.1820 Product Codes: ETN Classification Panel: Ear Nose & Throat Classification Name: Stimulator, Nerve
## 4. Primary Predicate Device
Manufacturer: Inomed Medizintechnik GmbH Trade name: Inomed Adhesive Laryngeal Electrode 510(K) Number: K091874 Regulation Number: 21 CFR 874.1820 Product Codes: ETN Classification Panel: Ear Nose & Throat Classification Name: Stimulator, Nerve
## 5. Device Description
The Disposable Laryngeal Electrodes are single used electrodes constructed from an medical-grade ink with a conductive polymer film covered by an insulating coating, a connector made of polypropylene and a cable assembly. The electrodes are designed to monitor the recurrent nerve during thyroid, anterior cervical, carotid endartecrectomy surgery and vagus nerve monitoring during brain surgery. The monitoring is performed with a surface electrode that is attached to an endotracheal tube placed on the vocal cord. A neutral adhesive electrode is placed on the patient's
{4}------------------------------------------------
shoulder. Both adhesive electrodes are connected to a reusable recording cable and to the nerve monitor.
# 6. Intended Use/Indications for Use
The Disposable Laryngeal Electrodes are intended to be used as a disposable, self-adhesive electrode attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures.
| Attribute | Disposable Laryngeal<br>Electrode (Subject<br>Device) | Inomed Adhesive<br>Laryngeal Electrode<br>(Predicate device<br>K091874) | Comment |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Regulation<br>Number | 21 CFR 874.1820 | 21 CFR 874.1820 | Same |
| Regulation Name | Surgical nerve<br>stimulator/locator | Surgical nerve<br>stimulator/locator | Same |
| Regulatory Class | Class II | Class II | Same |
| Product Code | ETN | ETN | Same |
| Indications for<br>Use | The Disposable Laryngeal<br>Electrodes are intended to<br>be used as a disposable,<br>self-adhesive electrode<br>attached to an endotracheal<br>tube and positioned for<br>continuous EMG<br>monitoring of the larynx<br>during surgical procedures. | The Inomed Adhesive<br>Laryngeal Electrodes are<br>intended to be used as a<br>disposable, self-adhesive<br>electrode attached to an<br>endotracheal tube and<br>positioned for continuous<br>EMG monitoring of the<br>larynx during surgical<br>procedures. | Same |
| Monitoring site | Trachea/larynx | Trachea/larynx | Same |
| Monitoring type | Continuous EMG<br>monitoring | Continuous EMG<br>monitoring | Same |
| May be used with<br>all commercial<br>EMG units | Yes | Yes | Same |
| Method of<br>electrode<br>attachment | Attached to the surface of<br>the endotracheal tube | Attached to the surface of<br>the endotracheal tube | Same |
| Number of<br>electrodes<br>utilized | 2 | 2 | Same |
| Number of<br>channels | 2, 4 | 2 | Different 1 |
| Device design | -Medical grade inks<br>suspended in a polyester<br>substrate | -Medical grade inks<br>suspended in a polyester<br>substrate | Same |
| | -polypropylene connector<br>-cable assembly | -polypropylene connector<br>-cable assembly | |
| Electrical<br>insulation | Electrical insulation on all<br>surfaces until the head of<br>the electrode | Electrical insulation on all<br>surfaces until the head of<br>the electrode | Same |
| Single use only | Yes | Yes | Same |
| Safety<br>characteristics | Non-invasive | Non-invasive | Same |
| Sterilization | ETO | ETO | Same |
# 7. Comparison of Technological Characteristics
{5}------------------------------------------------
Discussion of difference !: The subject device included two and four channels type.The predicate device only have two channels type. Although differences between numbers of channel, the subject two and four channel device have the same design, materials and intended performance, only increasing two identical channels. Furthermore, the four channel device is selected as representative sample to performed electrical safety testing according to IEC 60601-1 to ensure meet the electrical safety requirement. Therefore, we considered that this difference does not raise new questions of safety or effectiveness.
#### Non-Clinical Testing 8.
## 8.1. Performance Testing-Bench
Suzhou Haishen United Medical Device Associates Co., Ltd has performed the following non-clinical laboratory testing to determine substantial equivalence.
| Test | Acceptable criteria | Result |
|---------------------|------------------------------------------------------------------------------------------------------------------|--------|
| Adhesive property | The electrode should be able to be firmly adhered. | Pass |
| Electrode Impedance | Each path of the electrode should be able to conduct,<br>impedance: ≤100 Ω . | Pass |
| Security | Normal state (μΑ):<br>AC: ≤0.01mA (10μA) and DC: ≤0.1mA(100μA)<br>Application part pressurized state (mA): ≤5mA | Pass |
| Sterility testing | The products were sterile both before and after aged. | Pass |
The essential performance and safety of Disposable Laryngeal Electrode was tested for performance in accordance with internal requirements. The adhesive property, electrode Impedance and security were evaluated for subject and predicate device to ensure that the subject device performed as intended and supported substantial equivalence to the predicate devices.
## 8.2. Biocompatibility testing
The contact classification for the Disposable Laryngeal Electrode of the subject device is a surface contact device that directly contact with the mucosal membrane ( ≤ 24h). The biocompatibility evaluation for the subject device was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and in accordance with FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff. The results of biocompatibility
{6}------------------------------------------------
testing included in the table below demonstrate that the device meets biological safety requirements per ISO 10993-1 for mucosal membrane contact device.
| Test | Standard | Result |
|---------------------------|--------------|-----------------------------|
| In vitro cytotoxicity | ISO 10993-5 | Non-cytotoxic |
| Skin sensitization | ISO 10993-10 | Non-sensitive |
| Intracutaneous reactivity | ISO 10993-23 | Non- irritation |
| Acute systemic toxicity | ISO 10993-11 | Non-acute systemic toxicity |
| Pyrogen | ISO 10993-11 | Non-pyrogenic |
## 8.3. Sterilization and shelf-life testing
The subject device is sterilized using Ethylene Oxide. The sterilization validation has been performed in accordance with the principles of ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. A sterility assurance level (SAL) of 10th has been demonstrated. The device meets EO residuals per ISO 10993-7.
A shelf-life of 3 years has been established based on accelerated aging testing.
#### 8.4. Electrical safety testing
Suzhou Haishen United Medical Device Associates Co., Ltd has performed the electrical safety testing according to IEC 60601-1 Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance. Testing were performed and confirmed that the subject device is meet the requirement.
#### 9. Conclusion
The Disposable Laryngeal Electrode is substantially equivalent to the predicate device when evaluating intended use and technological characteristics. The subject device has the identical intended use as the predicate device and only has a minor technological difference that does not raise different questions of safety or effectiveness. The nonclinical performance data submitted in the 510(k) demonstrates the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.