← Product Code [NXC](/productcode/NXC) · K232887

# Invisalign Palatal Expander System (K232887)

_Align Technology, Inc. · NXC · Dec 15, 2023 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232887

## Device Facts

- **Applicant:** Align Technology, Inc.
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Dec 15, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K232887 · Dec 15, 2023 | Invisalign Palatal Expander System | Align Technology, Inc. | Commercially available product data | Real-world data from the commercially available product was analyzed to evaluate the clinical performance of the IPE System as part of the evidence supporting substantial equivalence. | Real-world data; Clinical performance; Post-market study |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Real-world data post-market study; Retrospective analysis of commercially available product | Patients treated with the Invisalign Palatal Expander System | Not applicable for this study | Clinical performance |

## Indications for Use

The Invisalign Palatal Expander System is intended for use in the treatment of dental and/or skeletal malocclusion in primary, mixed or permanent dentition.

## Device Story

Invisalign Palatal Expander (IPE) System is a series of removable, 3D-printed thermoplastic orthodontic devices; designed to expand narrow maxilla in small increments; followed by a holding phase using a passive retainer. System includes Invisalign Palatal Expanders, Invisalign Palatal Holders, and an Attachment Template. Proprietary 3D software enables CAD designers to generate device shapes and quantities based on doctor-prescribed expansion amounts and holding durations. Used in dental clinics by orthodontists/dentists; patients wear devices to apply continuous outward force to the maxilla. Output is a patient-matched series of expanders and holders. Benefits include non-invasive, sequential expansion of the maxilla to correct malocclusion. Healthcare providers use the system to achieve skeletal/dental expansion without traditional metal expansion screws.

## Clinical Evidence

Early feasibility study and real-world data analysis conducted. Study observed desired active expansion of upper jaw width in all subjects with no unanticipated or serious adverse events reported. Bench testing included durability, expander fatigue, insertion/removal force, bioburden, and packaging validation. Biocompatibility assessed per ISO 10993-1 and ISO 7405.

## Technological Characteristics

Thermoplastic polyamide material; additive manufacturing (3D printing); intraoral form factor. Proprietary 3D software for treatment planning and device design. Biocompatibility per ISO 10993-1 and ISO 7405. Software verification and validation per IEC 62304.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- Invisalign System ([K220287](/device/K220287.md))
- Dentaurum Expansion Screws ([K935154](/device/K935154.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 15, 2023

Align Technology, Inc. Sindhura Gaddamanugu Sr. Regulatory Affairs Specialist II 2820 Orchard Parkway San Jose, California 95134

Re: K232887

Trade/Device Name: Invisalign Palatal Expander System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: September 18, 2023 Received: September 18, 2023

Dear Sindhura Gaddamanugu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232887

#### Device Name Invisalign Palatal Expander System

#### Indications for Use (Describe)

The Invisalign Palatal Expander System is indicated for the orthodontic treatment of malocclusion. The system is used for rapid expansion and subsequent holding of skeletal and/or dental narrow maxilla (upper jaw, dental arch and tooth, palate) with primary, mixed, or permanent dention during orthopedic treatment in children or adolescents. In adults, it is to be used in conjunction with surgery or other interventions when necessary.

| Type of Use (Select one or both, as applicable)                                                                                                                   |                                                                                                                                                                  |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <span style="font-size:20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div style="display:flex; align-items:center;"> <span> <span style="font-size:20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Section 7: 510(K) Summary - K232887

# Align Technology's Invisalign System Palatal Expander (IPE) System

#### General Information

| 510(k) Sponsor          | Align Technology, Inc.                                                                                     |
|-------------------------|------------------------------------------------------------------------------------------------------------|
| Address                 | 2820 Orchard Parkway<br>San Jose, CA 95134                                                                 |
| FDA Registration Number | 2953749                                                                                                    |
| Contact Person          | Sindhura Gaddamanugu<br>Sr. Regulatory Affairs Specialist II, Regulatory Affairs<br>Align Technology, Inc. |
|                         | cc: Shweta Daga<br>Director, Regulatory Affairs<br>Align Technology Inc.                                   |
| Contact Information     | Email: sgaddamanugu@aligntech.com<br>Phone: +1 408-470-1000<br>Fax: +1408-470-1010                         |
| Date Prepared           | September 18, 2023                                                                                         |

#### Name of Modified Device and Name/Address of Sponsor

| Name of Device          | Invisalign Palatal Expander (IPE) System                           |
|-------------------------|--------------------------------------------------------------------|
| Name/Address of Sponsor | Align Technology, Inc.<br>2820 Orchard Parkway, San Jose, CA 95134 |
| Trade/Proprietary Name  | Invisalign Palatal Expander (IPE) System                           |
| Common Name             | Palatal Expander, Sequential                                       |
| Classification Name     | Orthodontic Plastic Bracket                                        |
| Regulation Number       | 21 CFR 872.5470                                                    |
| Product Code            | NXC                                                                |
| Regulatory Class        | II                                                                 |

# Primary Predicate Device

| Name of Device          | Invisalign System                                                  |
|-------------------------|--------------------------------------------------------------------|
| Name/Address of Sponsor | Align Technology, Inc.<br>2820 Orchard Parkway, San Jose, CA 95134 |
| Trade/Proprietary Name  | Invisalign System                                                  |
| Common Name             | Aligner, Sequential                                                |
| Classification Name     | Orthodontic Plastic Bracket                                        |
| Regulation Number       | 21 CFR 872.5470                                                    |
| Product Code            | NXC                                                                |
| Regulatory Class        | II                                                                 |

ALIGN CONFIDENTIAL

{5}------------------------------------------------

| Name of Device          | Dentaurum Expansion Screws             |
|-------------------------|----------------------------------------|
| Name/Address of Sponsor | Dentauraum, Inc.                       |
| Trade/Proprietary Name  | Dentaurum Expansion Screws             |
| Common Name             | Retainer, Screw Expansion, Orthodontic |
| Classification Name     | Orthodontic appliance and accessories  |
| Regulation Number       | 21 CFR 872.5410                        |
| Product Code            | DYJ                                    |
| Regulatory Class        | I                                      |

#### Secondary Predicate Device

# Purpose of the Traditional 510(k) Notice

The purpose of this Traditional 510(k) notice is to request clearance for the Subject device. the Invisalign Palatal Expander (IPE) System. Additionally, this Traditional 510(k) introduces the proprietary, 3D Shape generation software, and labeling documentation associated with the Invisalign Palatal Expander System.

# Intended Use/Indications For Use

The intended use of the Subiect device is as follows:

The Invisalign Palatal Expander System is intended for use in the treatment of dental and/or skeletal malocclusion in primary, mixed or permanent dentition.

The Indications for Use of the subject device is as follows:

The Invisalign Palatal Expander System is indicated for the orthodontic treatment of malocclusion. The system is used for rapid expansion and subsequent holding of skeletal and/or dental narrow maxilla (upper jaw, dental arch and teeth, palate) with primary, mixed, or permanent dentition during orthodontic or orthopedic treatment in children or adolescents. In adults, it is to be used in conjunction with surgery or other interventions when necessary.

The general purpose of the subject device and the primary predicate, i.e. Invisalign System and the secondary predicate device (Dentaurum Expansion Screws) is the treatment of malocclusion and the subject device's intended use falls within the intended use of the predicate devices. The subject device and the primary predicate device are indicated for the treatment of malocclusion. Additionally, the subject device and the secondary predicate device are used for rapid expansion of narrow maxilla. While the Indication for Use is similar, both the subject and primary predicate device have the same intended use that is used in the treatment of dental or skeletal malocclusion in primary, mixed or permanent dentition.

{6}------------------------------------------------

# Device Description

The subject device, Invisalign Palatal Expander (IPE) System (hereafter referred to as the IPE System) is a modification to the legally marketed multiple predicate devices. Invisalign System (Class II; Product Code: NXC; K220287) as Primary and Dentaurum Expansion Screws/Hyrax Expander – Hyrax ® and Hyrax neo® (Class I, Product Code: DYJ, K935154) as Secondary.

The IPE System consists of the Invisalign Palatal Expanders, Invisalign Palatal Holders, the Attachment Template and the proprietary 3D Shape generation software.

Invisalign Palatal Expanders are a staged series of removable orthodontic devices designed to expand the patient's skeletal and/or dental narrow maxilla (upper jaw, dental arch, teeth and palate) in small increments to an optimal position determined by the doctor. Once the desired clinical outcome of the expansion phase has been achieved, patients progress to the holding phase.

The Invisalign Palatal Holder is a copy of the last stage of the expansion phase designed to hold the maxilla post active expansion, to allow the maxilla to stabilize before the patient progresses to the next phase of treatment (retention, phase 2 or other treatment).

The proprietary Align internal personnel-facing 3D software enables Align's computer-aided design (CAD) designers to generate the shape and quantity/stages of the device. Using this software, CAD designers create digital files for the incremental stages of the doctorprescribed expansion amount and the design and quantity of holders per the prescribed holding duration. Each device in the series is fabricated via additive manufacturing (3D printing).

The Attachment Template (also a component of the primary predicate device, the Invisalign System) enables correct placement and bonding of attachments made of dental composite (material provided by doctor) to the tooth surface for IPE engagement and retention.

#### Comparison with Predicate Devices

In accordance with 21 CFR 807.92(a)(6) a summary of the technological characteristics' comparison of the proposed subject device to the predicate devices is provided below.

#### Technological Characteristics Comparison with the Predicate Devices

The IPE System provides an additional method for palate expansion during malocclusion treatment as achieved with the primary and secondary predicate devices. The technological differences are limited to the material, additive manufacturing, and the 3D Software used to deliver a patient-matched device to meet its proposed intended use.

{7}------------------------------------------------

|                                                         | Subject Device<br>Invisalign Palatal<br>Expander System                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary<br>Predicate Device<br>Invisalign System<br>(K220287)                                                                                                                                                                                                                                                                                                                                                                                                           | Secondary<br>Predicate Device<br>Dentaurum<br>Expansion<br>Screws<br>(K935154)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Substantial<br>Equivalency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use, User Population                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended<br>Use                                         | The Invisalign Palatal<br>Expander System is<br>intended for use in the<br>treatment of dental<br>and/or skeletal<br>malocclusion in<br>primary, mixed or<br>permanent dentition.                                                                                                                                                                                                                                                                                                                                                  | The Invisalign<br>System is intended<br>for use in the<br>treatment of dental<br>or skeletal<br>malocclusion in<br>primary, mixed<br>and permanent<br>dentition.                                                                                                                                                                                                                                                                                                        | The Expansion<br>Screw is intended<br>for use in the<br>treatment for<br>palatal expansion<br>in all dentition<br>types.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same as primary<br>predicate device<br>Similar to<br>secondary<br>predicate device<br>The subject<br>device's intended<br>use falls within the<br>intended use of<br>the secondary<br>predicate device.                                                                                                                                                                                                                                                                                                            |
| Indications<br>for Use                                  | The Invisalign Palatal<br>Expander System is<br>indicated for the<br>orthodontic treatment<br>of malocclusion. The<br>system is used for<br>rapid expansion and<br>subsequent holding of<br>skeletal and/or dental<br>narrow maxilla (upper<br>jaw, palate, dental<br>arch and teeth) with<br>primary, mixed, or<br>permanent dentition<br>during orthodontic or<br>orthopedic treatment<br>in children or<br>adolescents. In adults,<br>to be used in<br>conjunction with<br>surgery or other<br>interventions when<br>necessary. | The Invisalign<br>System is<br>indicated for the<br>orthodontic<br>treatment of<br>malocclusion.                                                                                                                                                                                                                                                                                                                                                                        | The device is<br>indicated for case<br>of transverse<br>under-development<br>of the upper jaw.                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar to primary<br>predicate device<br>Same as the<br>secondary<br>predicate device<br>The subject<br>device and the<br>primary predicate<br>device have<br>similar Indication<br>for Use and both<br>are indicated for<br>the treatment of<br>malocclusion.<br>Both the subject<br>device and the<br>secondary<br>predicate device<br>have the same<br>indication for use<br>to expand the<br>narrow<br>maxilla/upper jaw.                                                                                     |
| Patient<br>Population                                   | Children, Adolescents<br>and Adults                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Children,<br>Adolescents and<br>Adults                                                                                                                                                                                                                                                                                                                                                                                                                                  | All patients for<br>whom a fixed<br>appliance with an<br>expansion screw<br>can be placed as<br>part of orthodontic<br>treatment.                                                                                                                                                                                                                                                                                                                                                                                                                                | Same as primary<br>predicate device<br>Similar to<br>secondary<br>predicate device<br>The subject<br>device patient                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                         | Subject Device<br>Invisalign Palatal<br>Expander System                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary<br>Predicate Device<br>Invisalign System<br>(K220287)                                                                                                                                                                                                                                                                                                                                                                                                           | Secondary<br>Predicate Device<br>Dentaurum<br>Expansion<br>Screws<br>(K935154)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Substantial<br>Equivalency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | population is the<br>same as the<br>primary predicate<br>device and the<br>population is a<br>subset of<br>secondary<br>predicate device<br>patient population.                                                                                                                                                                                                                                                                                                                                                    |
| Use<br>Location                                         | Dental intraoral<br>devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Dental intraoral<br>devices.                                                                                                                                                                                                                                                                                                                                                                                                                                            | Dental intraoral<br>devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as primary<br>and secondary<br>predicate devices                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| OTC or<br>Prescription<br>(Rx) Device                   | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same as primary<br>and secondary<br>predicate devices                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Principle of<br>Operation                               | Expanders:<br>Sequential expanders<br>apply continuous<br>outward<br>pressure/force to the<br>maxilla.                                                                                                                                                                                                                                                                                                                                                                                                                             | Aligners:<br>Sequential<br>aligners apply<br>continuous gentle<br>force to the teeth<br>and/or position<br>mandible forward.                                                                                                                                                                                                                                                                                                                                            | Hyrax has an<br>expansion screw<br>suspended above<br>the palatal<br>vault/gingiva. A key<br>is used to turn the<br>expansion screws<br>(quarter turn - 90<br>degrees rotation)<br>on a specified<br>schedule per the<br>doctor's discretion,<br>and this keeps the<br>pressure on both<br>halves of the<br>jawbone causing<br>them to widen and<br>move apart. The<br>number of quarter<br>turns per day is<br>instructed by the<br>doctor to the<br>patient and/or<br>parent/supervising<br>adult per the<br>treatment plan for<br>rapid or slow<br>expansion. | Similar to primary<br>and secondary<br>predicate devices.<br>Subject device<br>operates same as<br>primary predicate<br>device in the<br>context of<br>removability,<br>exerting<br>continuous force,<br>and sequential<br>devices.<br>Subject device<br>and secondary<br>predicate operate<br>similarly regarding<br>separating the<br>mid-palatal suture<br>and movement of<br>maxillary halves.<br>Subject device<br>and secondary<br>predicate device<br>operate similarly<br>by exerting<br>continuous force. |
|                                                         | Subject Device<br>Invisalign Palatal<br>Expander System                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary<br>Predicate Device<br>Invisalign System<br>(K220287)                                                                                                                                                                                                                                                                                                                                                                                                           | Secondary<br>Predicate Device<br>Dentaurum<br>Expansion<br>Screws<br>(K935154)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Substantial<br>Equivalency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                         | Holders: The palatal<br>holder is a copy of last<br>stage of the<br>expansion phase.                                                                                                                                                                                                                                                                                                                                                                                                                                               | Retainers: The<br>retainers are a<br>copy of the last<br>stage of the<br>aligners.<br>Note: Invisalign<br>Retainers are not<br>part of the primary<br>predicate device<br>Invisalign System<br>(K220287) as they<br>are classified as a<br>Class I under<br>21CFR872.5525.                                                                                                                                                                                              | The expansion<br>screw-based<br>device should<br>remain in the<br>mouth after active<br>expansion without<br>turning the screw<br>for approximately 2<br>- 3 months for<br>stabilization<br>purposes and to<br>avoid relapse.                                                                                                                                                                                                                                                                                                                                    | Similar to the<br>primary and<br>secondary<br>predicate devices<br>Subject device<br>has the same<br>principle of<br>operation of<br>maintaining and<br>preventing<br>movement using<br>passive devices<br>as both predicate<br>devices.                                                                                                                                                                                                                                                                           |
|                                                         | Technological Characterization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Software                                                | Produces 3D-model<br>file of the IPE devices<br>based on the digital<br>scan and<br>doctor/dental<br>practitioner's<br>prescription.<br>3D Software of the<br>subject device is an<br>internal Align facing<br>only shape generation<br>software which<br>includes software<br>algorithms that are<br>used to determine the<br>shape and calculate<br>the quantity of devices<br>required per the<br>doctor prescribed<br>expansion distance<br>and holding duration.                                                              | Produces 3D-<br>model file of the<br>PVS impression or<br>the digital scan.<br>Identifies the<br>individual teeth<br>that requires<br>treatment (i.e.,<br>repositioning).<br>Creates a<br>treatment plan<br>(i.e., 3-D models<br>that represent the<br>treatment plan)<br>which is reviewed<br>by the treating<br>dental practitioner<br>using ClinCheck's<br>doctor facing<br>function to reject<br>or request<br>modifications to<br>the set-up prior to<br>approval. | Not Applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar to the<br>primary predicate.<br>The 3D software<br>used in the<br>subject and<br>predicate devices<br>are both used for<br>treatment planning<br>and designing the<br>shape of the<br>device. The<br>subject device<br>utilized a<br>simplified version<br>of 3D software<br>based on the<br>purpose and<br>design of the<br>device.                                                                                                                                                                       |
| Materials                                               | Thermoplastic<br>polyamide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Thermoplastic<br>polymer                                                                                                                                                                                                                                                                                                                                                                                                                                                | The screws are<br>made of stainless<br>steel (DIN 1.4301)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar to the<br>primary predicate<br>device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                         | Subject Device<br>Invisalign Palatal<br>Expander System                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary<br>Predicate Device<br>Invisalign System<br>(K220287)                                                                                                                                                                                                                                                                                                                                                                                                           | Secondary<br>Predicate Device<br>Dentaurum<br>Expansion<br>Screws<br>(K935154)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Substantial<br>Equivalency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Durability                                              | Durability was tested<br>to verify the<br>performance of the<br>IPE device. All IPE<br>devices maintained<br>the engagement on<br>attachments, without<br>deformation, cracks,<br>chips or breaks.                                                                                                                                                                                                                                                                                                                                 | Performance Testing<br>No new durability<br>testing was<br>required with the<br>primary predicate<br>device pre-market<br>notification<br>(K220287).<br>Durability testing<br>was performed on<br>previously cleared<br>Invisalign System<br>and testing results<br>met all acceptance<br>criteria.                                                                                                                                                                     | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same as primary<br>predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Stiffness/<br>Force<br>system                           | The subject device<br>was tested to verify<br>that the device is<br>structurally stiff and<br>can deliver the<br>required force in the<br>lateral direction on the<br>posterior teeth. The<br>test results met the<br>acceptance criteria<br>and performed as<br>intended.                                                                                                                                                                                                                                                         | No new force<br>system testing was<br>required with the<br>primary predicate<br>device pre-market<br>notification<br>(K220287).<br>Force mechanical<br>verification testing<br>demonstrated with<br>previously cleared<br>Invisalign System<br>provides adequate<br>force systems on<br>the dentition for its<br>intended use.                                                                                                                                          | The mechanical<br>behavior<br>(stiffness/force) of<br>Hyrax was used as<br>one of the<br>acceptance criteria<br>for the subject<br>device bench<br>testing.                                                                                                                                                                                                                                                                                                                                                                                                      | Similar to Primary<br>Predicate device<br>Same as secondary<br>predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Packaging                                               | The subject device<br>was tested to validate<br>that the packaging<br>can protect the<br>System from<br>exposure to all<br>relevant shipping and<br>handling scenarios.<br>The testing confirmed<br>that the product<br>integrity was<br>maintained during                                                                                                                                                                                                                                                                         | No new packaging<br>validation testing<br>was required with<br>the primary<br>predicate device<br>pre-market<br>notification<br>(K220287).<br>The Packaging<br>validation testing<br>on previously                                                                                                                                                                                                                                                                      | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar to Primary<br>Predicate device<br>Both the subject<br>and primary<br>predicate devices<br>use the same<br>primary<br>packaging, and<br>packaging testing<br>on the subject<br>device confirmed                                                                                                                                                                                                                                                                                                             |
| Subject Device<br>Invisalign Palatal<br>Expander System | Primary<br>Predicate Device<br>Invisalign System<br>(K220287)                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary<br>Predicate Device<br>Dentaurum<br>Expansion<br>Screws<br>(K935154)                                                                                                                                                                                                                                                                                                                                                                                          | Substantial<br>Equivalency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| shipping and handling<br>scenarios.                     | cleared Invisalign<br>System<br>demonstrated that<br>the device integrity<br>was maintained<br>during the various<br>shipping and<br>handling<br>scenarios.                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | that device<br>integrity is<br>maintained during<br>all relevant<br>shipping and<br>handling<br>scenarios                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Retention                                               | The subject device<br>was tested to verify<br>that the device can<br>retain engagement<br>with attachments on<br>posterior teeth while<br>the device is under<br>compression during<br>an active expansion<br>and holding period.<br>The test results met<br>the acceptance<br>criteria and performed<br>as intended.                                                                                                                                                                                                              | No new testing<br>was required with<br>the primary<br>predicate device<br>pre-market<br>notification<br>(K220287).<br>Retention testing<br>performed on<br>previously cleared<br>Invisalign System<br>demonstrated that<br>the device<br>performed as<br>intended.                                                                                                                                                                                                      | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar to_Primary<br>Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Insertion<br>and<br>Removal<br>Force                    | The Insertion and<br>Removal Force test<br>was conducted to<br>verify that the forces<br>required to insert and<br>remove device are<br>significantly less than<br>the bond force of the<br>attachment on the<br>tooth.                                                                                                                                                                                                                                                                                                            | No force system<br>testing was<br>required with the<br>primary predicate<br>device pre-market<br>notification<br>(K220287).<br>The testing<br>performed on<br>previously cleared<br>Invisalign System<br>demonstrated that<br>the device<br>performed as<br>intended.                                                                                                                                                                                                   | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar to_Primary<br>Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Human<br>Factors &<br>Usability                         | Human factors &<br>Usability testing was<br>conducted on the<br>subject device to<br>validate that it is                                                                                                                                                                                                                                                                                                                                                                                                                           | No human factors<br>& usability<br>validation testing<br>was required for<br>the primary                                                                                                                                                                                                                                                                                                                                                                                | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar to_Primary<br>Predicate device<br>Based on<br>available data                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                         | Subject Device<br>Invisalign Palatal<br>Expander System                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Primary<br>Predicate Device<br>Invisalign System<br>(K220287)                                                                                                                                                                                                                                                                                                                                                                                                           | Secondary<br>Predicate Device<br>Dentaurum<br>Expansion<br>Screws<br>(K935154)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Substantial<br>Equivalency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                         | adequately designed<br>for its intended users,<br>uses and use<br>environments                                                                                                                                                                                                                                                                                                                                                                                                                                                     | predicate device<br>pre-market<br>notification<br>(K220287).<br>Human Factors &<br>Usability validation<br>on previously<br>cleared Invisalign<br>System has been<br>found to be<br>adequately<br>designed for the<br>intended users,<br>uses, and use<br>environments.                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | both the subject<br>device and<br>primary predicate<br>device perform as<br>intended and<br>additional device<br>modifications to<br>the user interface<br>are not needed<br>and would not<br>further reduce<br>risk.                                                                                                                                                                                                                                                                                              |
| Clinical<br>Testing                                     | Not Applicable.<br>An early feasibility<br>study was conducted<br>which concluded that<br>desired active<br>expansion of the<br>upper jaw width was<br>observed in all the<br>subjects. There were<br>no unanticipated or<br>serious adverse<br>events reported from<br>this study.<br>Additionally, real-<br>world data was also<br>analyzed from<br>commercially<br>available product.                                                                                                                                           | Not Applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Not Applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar to primary<br>predicate and<br>secondary<br>predicate devices.<br>The subject<br>device is a Class<br>II medical device<br>and clinical testing<br>was not warranted<br>to support<br>substantial<br>equivalency with<br>the primary and<br>secondary<br>predicate devices.                                                                                                                                                                                                                                |

# Table 7-1: Substantial Equivalency Overview

#### ALIGN CONFIDENTIAL

{8}------------------------------------------------

{9}------------------------------------------------

# ALIGN CONFIDENTIAL

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

# Performance Data

The IPE System underwent a complete set of functional and performance testing, including, durability, expander fatigue/insertion & removal, bioburden, and packaging validation testing. Clinical performance of the IPE System was also evaluated in an early feasibility study and realworld data post-market study. Based on the performance testing outcomes, the subject device, 

{13}------------------------------------------------

the IPE System is substantially equivalent to its predicates and there are no new risks from the differences between the subject device and the predicate devices.

# Biocompatibility

Biocompatibility testing for the proposed IPE System was completed and assessed in accordance with ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process, and ISO 7405, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry. The results demonstrate that the IPE System does not pose any significant biologicals risks and is considered safe for its intended use in the intra-oral cavity.

#### Software Testing

In accordance with IEC 62304. Medical device software - Software life cycle processes; successful software verification and validation (V&V) testing at the unit, integration, and system level was performed to qualify the orthodontic software component of the subject device.

# Substantial Equivalence

The subject device is similar to its predicate devices in that it has:

- intended use, same as primary predicate device and similar to secondary predicate o device;
- . indications for use, similar to primary predicate device and same as secondary predicate device:
- principles of operation, similar to both predicate devices; and
- technological characteristics, similar to primary predicate device ●

The differences in the indications for use, principles of operation and technological characteristics between the subject device and its predicate devices do not raise different questions of safety or efficacy. Therefore, the subject device is found to be substantially equivalent to the legally marketed primary predicate device, Invisalign System (K220287) and secondary predicate device, Dentaurum Expansion Screws.

#### Conclusion

Align Technology's Invisalign Palatal Expander (IPE) System has the same intended use and similar indications for use as the previously cleared primary predicate, Invisalign System (K220287). The IPE System has similar intended use and same indications for use as the secondary predicate device Dentaurum Expansion Screws. The difference in principles of operation and technological characteristics between the subject device and the predicate devices were evaluated using standard test methods and do not raise new issues of safety or effectiveness. Performance data (bench and clinical) and biocompatibility testing demonstrate that the subject device is as safe and effective as both predicate devices. Thus, the

Invisalign Palatal Expander (IPE) System is substantially equivalent to both predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K232887](https://fda-staging.innolitics.com/device/K232887)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com