← Product Code [GCJ](/productcode/GCJ) · K232866

# Disposable Abdominal Trocars  (Bladed Trocars, Bladeless Trocars, Optical Trocars) (K232866)

_Jiangsu Channel Medical Device Co., Ltd. · GCJ · Nov 7, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232866

## Device Facts

- **Applicant:** Jiangsu Channel Medical Device Co., Ltd.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Nov 7, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars) is used to puncture the abdominal wall of human body during laparoscopy and operation to establish the working channel for abdominal surgery.

## Device Story

Disposable surgical trocar system; used to puncture abdominal wall during laparoscopy; establishes working channel for endoscopic instruments. Available in bladed, bladeless, and optical (visible) configurations. Device consists of cannula assembly (sleeve, stopcock valve, gas check valve, lip seal) and obturator (fixed base, tube, tip). Some models include smoke evacuation function. Operated by surgeons in clinical settings. Device inserted into skin incision, punctured into abdominal cavity; obturator removed to create surgical channel. Provides port of entry for instruments; optical version allows direct visualization of abdominal wall layers during insertion. Single-use, EO sterilized.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1, -5, -7, -10, -11, -12, and -23. Sterilization validated per ISO 11135. Performance testing included obturator compatibility, insertion/cannula stability, air leakage, and insertion/removal force.

## Technological Characteristics

Materials: Stainless steel, Polycarbonate (PC), ABS. Energy: Mechanical (manual insertion). Dimensions: 5-15mm diameter, 110-150mm length. Connectivity: None. Sterilization: Ethylene Oxide (EO). Design: Cannula with universal seal, valve, and stopcock; optional smoke evacuation (CNTCII series).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Unimicro Trocar kit ([K141594](/device/K141594.md))
- Trocar ([K172038](/device/K172038.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 7, 2023

Jiangsu Channel Medical Device Co., Ltd % Jarvis Wu Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave Shanghai 200122 China

Re: K232866

Trade/Device Name: Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars)

Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: September 15, 2023 Received: September 15, 2023

Dear Mr. Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Digitally signed by Mark Trumbore -S Trumbore -S Date: 2023.11.07
11.07 -05 08:42:27 -05'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices 

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232866

#### Device Name

Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars)

Indications for Use (Describe)

The Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars) is used to puncture the abdominal wall of human body during laparoscopy and operation to establish the working channel for abdominal surgery.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### Jiangsu Channel Medical Device Co., Ltd. Fourth Floor, G3,999# Qujia Road, Qiandeng Town, Kunshan, Jiangsu, China 215341

#### K232866

#### 510(K) Summary

Document prepared date: 2023/11/04

A. Applicant:

Contact Person: Eric Huang Title: General Manager Jiangsu Channel Medical Device Co., Ltd. Address: Fourth Floor, G3,999# Qujia Road, Qiandeng Town, Kunshan, Jiangsu, China 215341 Tel: +0086-512-36857161 Email: chen.huang@channel-mdi.com

Submission Correspondent: Primary Contact: Mr. Jarvis Wu Title: Regulatory Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: jiawei.wu@sungoglobal.com

Secondary Contact: Mr. Raymond Luo Title: Senior Regulatory Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050

Email: zxfda(@sungoglobal.com

#### B. Device:

Trade Name: Disposable Abdominal Trocars (Bladed Trocars, Optical Trocars, Optical Trocars) Common Name: Disposable Surgical Trocar /Cannula

Regulatory Information

Classification Name: Laparoscope, General & Plastic Surgery Classification: Class II Product code: GCJ Regulation Number: 876.1500

### C. Predicate device:

Primary Predicate Device: K141594 Product name: Unimicro Trocar kit Model: Auto-Locking Trocar, Hasson Trocar, Bladeless Trocar. Unimicro Medical Systems (ShenZhen) Co., Ltd.

Secondary Predicate Device: K172038 Product name: Trocar, Model: Auto-Locking Trocar, Bladeless Trocar, Visible Trocar 

{5}------------------------------------------------

WickiMed (Huizhou) Medical Equipment Manufacturing Co.,Ltd.

## D. Intended use of the device:

The Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars) is used to puncture the abdominal wall of human body during laparoscopy and operation to establish the working channel for abdominal surgery.

### E. Device Description:

The Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars) are available in total of eight (8) sizes.

The Disposable Abdominal Trocars (Bladed Trocars), knows as Auto-Locking Trocar, has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. The Disposable Abdominal Trocars (Bladed Trocars) is available in three (3) sizes: 5mm, 10mm and 12mm. The cannula assembly has a universal seal, a valve, and a stopcock. This device has a bladed tip with an internal shield, which is designed to cover the cutting edges once the body cavity has been entered.

The Disposable Abdominal Trocars (Bladeless Trocars) has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. The Disposable Abdominal Trocars (Bladeless Trocars) is available in one (1) size: 5mm.

The Disposable Abdominal Trocars (Optical Trocars), knows as Visible Trocar, has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. This device has a blunt tip, which is designed for open Laparoscopy. The Disposable Abdominal Trocars (Optical Trocars) is available in four (4) sizes: 5mm, 10mm, 12mm. This device allows direct visualization of the abdominal wall layers when the trocar is traversed, which offers a safe and rapid option of primary trocar. The cannula assembly has a universal seal, a valve and a stopcock.

Based on the filter valve, the Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars) are divided into two (2) product series: CNTCI series without smoke evacuation function and CNTCII series with smoke evacuation. The cannula consists of sleeve, stopcock valve, valve lever, gas check valve and lip seal. The obturator consists of fixed base, obturator tube and obturator tip. There are various specifications available depending on the inner diameter and length of sleeve.

| Device       | Subject Device                              | Primary Predicate Device                         | Secondary Predicate<br>Device                                     |
|--------------|---------------------------------------------|--------------------------------------------------|-------------------------------------------------------------------|
| Manufacturer | Jiangsu Channel Medical<br>Device Co., Ltd. | Unimicro Medical Systems<br>(ShenZhen) Co., Ltd. | WickiMed (Huizhou)<br>Medical Equipment<br>Manufacturing Co.,Ltd. |
| 510K number  | K232866                                     | K141594                                          | K172038                                                           |
| Product Name | Disposable Abdominal<br>Trocars             | Unimicro Trocar kit                              | Trocar                                                            |

## F. Comparison with predicate device

{6}------------------------------------------------

K232866

Jiangsu Channel Medical Device Co., Ltd.

| Fourth Floor, G3,999# Qujia Road, Qiandeng Town, Kunshan, Jiangsu, China 215341 |  |  |
|---------------------------------------------------------------------------------|--|--|
|                                                                                 |  |  |

| Classification                      |                                 | Class II Device (21 CFR 876.1500)                                                                                                                                                                                                                           | Class II Device (21 CFR 876.1500)                                                                                                                                                                                          | Class II Device (21 CFR 876.1500)                                                                                                                                                                           |
|-------------------------------------|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use                        |                                 | The Disposable Abdominal<br>Trocars (Bladed Trocars,<br>Bladeless Trocars, Optical<br>Trocars) is used to puncture<br>the abdominal wall of human<br>body during laparoscopy and<br>operation to establish the<br>working channel for<br>abdominal surgery. | The Unimicro Trocar kit,<br>Model: Auto-Locking<br>Trocar, Hasson Trocar, and<br>Bladeless Trocar has<br>application in a variety of<br>endoscopic procedures to<br>provide a port of entry for<br>endoscopic instruments. | The Trocar Models: Auto-<br>Locking Trocar, Bladeless<br>Trocar and Visible Trocar,<br>has application in a variety<br>of endoscopic procedures<br>to provide a port of entry<br>for endoscopic instruments |
| Principles of<br>operation          |                                 | Trocar inserted into the skin<br>incision, and punctured into<br>the abdominal cavity.<br>Removed the puncture cone<br>and made a surgical channel.                                                                                                         | Trocar inserted into the skin<br>incision, and punctured into<br>the abdominal cavity.<br>Removed the puncture cone<br>and made a surgical<br>channel.                                                                     | Trocar inserted into the<br>skin incision, and<br>punctured into the<br>abdominal cavity.<br>Removed the puncture<br>cone and made a surgical<br>channel.                                                   |
| Model                               |                                 | Auto-Locking Trocar<br>(Bladed Trocar)                                                                                                                                                                                                                      | Auto-Locking Trocar                                                                                                                                                                                                        | /                                                                                                                                                                                                           |
|                                     |                                 | Bladeless Trocar                                                                                                                                                                                                                                            | Bladeless Trocar                                                                                                                                                                                                           | /                                                                                                                                                                                                           |
|                                     |                                 | Visible Trocar<br>(Optical Trocar)                                                                                                                                                                                                                          | /                                                                                                                                                                                                                          | Visible Trocar                                                                                                                                                                                              |
| Mainly<br>Structure                 |                                 | Cannula, (blade), universal<br>seal, valve, stopcock                                                                                                                                                                                                        | Cannula, (blade), universal<br>seal, valve, stopcock                                                                                                                                                                       | Cannula, (blade), universal<br>seal, valve, stopcock                                                                                                                                                        |
| Specification                       |                                 | Diameter :5-15mm<br>Length :110-150mm                                                                                                                                                                                                                       | Diameter :5-12mm<br>Length: 70-120mm                                                                                                                                                                                       | Diameter :5-12mm<br>Length :100mm                                                                                                                                                                           |
| Patient-<br>contacting<br>structure | ATR                             | Blade, Cutting head, Cannula                                                                                                                                                                                                                                | Blade, Cutting head,<br>Cannula                                                                                                                                                                                            | Blade, Cutting head<br>Cannula                                                                                                                                                                              |
|                                     | BTR                             | Cutting head, Cannula                                                                                                                                                                                                                                       | Cutting head, Cannula                                                                                                                                                                                                      | Cutting head, Cannula                                                                                                                                                                                       |
|                                     | VTR                             | Cutting head, Cannula                                                                                                                                                                                                                                       | /                                                                                                                                                                                                                          | Cutting head, Cannula                                                                                                                                                                                       |
| Patient-<br>contacting<br>Materials | ATR                             | Stainless Steel, PC                                                                                                                                                                                                                                         | Stainless Steel, PC, ABS                                                                                                                                                                                                   | Stainless Steel, PC, ABS                                                                                                                                                                                    |
|                                     | BTR                             | PC,                                                                                                                                                                                                                                                         | PC, ABS                                                                                                                                                                                                                    | PC, ABS                                                                                                                                                                                                     |
|                                     | VTR                             | PC                                                                                                                                                                                                                                                          | /                                                                                                                                                                                                                          | PC                                                                                                                                                                                                          |
| Safety<br>standards                 |                                 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-7<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-12<br>ISO 10993-23<br>ISO 11135                                                                                                                                      | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-7<br>ISO 10993-10<br>ISO 10993-12<br>ISO 11135-1                                                                                                                                   | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-7<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-12<br>ISO 11135                                                                                                      |
| Performance<br>testing items        | Obturator compatibility         | Obturator compatibility                                                                                                                                                                                                                                     | Obturator compatibility                                                                                                                                                                                                    |                                                                                                                                                                                                             |
|                                     | Insertion & cannula stability   | Insertion & cannula stability                                                                                                                                                                                                                               | Insertion & cannula stability                                                                                                                                                                                              |                                                                                                                                                                                                             |
|                                     | Air leakage                     | Air leakage                                                                                                                                                                                                                                                 | Air leakage                                                                                                                                                                                                                |                                                                                                                                                                                                             |
|                                     | Trocar Insertion/ Removal force | Trocar Insertion/ Removal force                                                                                                                                                                                                                             | Trocar Insertion/ Removal force                                                                                                                                                                                            |                                                                                                                                                                                                             |
| Sterilization                       | EO Sterilized                   | EO Sterilized                                                                                                                                                                                                                                               | EO Sterilized                                                                                                                                                                                                              |                                                                                                                                                                                                             |
| Disposable                          | Yes                             | Yes                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                        |                                                                                                                                                                                                             |

{7}------------------------------------------------

Jiangsu Channel Medical Device Co., Ltd. Fourth Floor, G3,999# Qujia Road, Qiandeng Town, Kunshan, Jiangsu, China 215341

## Different analysis:

The subject and predicate devices have the similar intended use. The subject and predicate devices designs are nearly identical. Both are single-use devices, and structure and technology characteristic are identical. The differences in specification and patient-contacting materials between the subject and predicate devices do not raise conc questions of safety and effectiveness.

## G. Non-Clinical Testing

A series of tests were performed to assess the safety and effectiveness of the Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars). Biocompatibility tests were conducted in accordance with ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-7:2008, ISO 10993-10:2021, ISO 10993-11:2017, ISO 10993-12:2021 and ISO 10993-23:2021. The biocompatibility tests included Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute systemic toxicity and Material-Mediated Pyrogenicity. Sterilization was performed per ISO 11135:2014.

The performance testing conducted on subject device and predicate devices are listed below:

- Obturator Compatibility A
- A Insertion & Cannula Stability
- A Air Leakage
- Trocar Insertion/ Removal force A

All the test results demonstrate Disposable Abdominal Trocars, Bladeless Trocars, Optical Trocars) meet the requirements of its pre-defined acceptance criteria and intended uses.

## H. Clinical Test Conclusion

No clinical study is included in this submission.

# I. Conclusion

Based on the comparison and analysis above, the subject device Disposable Abdominal Trocars (Bladed Trocars, Bladeless Trocars, Optical Trocars), is as safe, as effective, and performs as well as the legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K232866](https://fda-staging.innolitics.com/device/K232866)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
