K232816 · Shenzhen LE Medical Technology Co., Ltd. · DPS · Jun 7, 2024 · Cardiovascular
Device Facts
Record ID
K232816
Device Name
Electrocardiograph, model: ECG301
Applicant
Shenzhen LE Medical Technology Co., Ltd.
Product Code
DPS · Cardiovascular
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of ECG301 is to acquire ECG signals from adult and pediatric patients(above 3 years old) through body surface ECG electrodes. The electrocardiograph is intended to be used only in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by ECG301 can help users to diagnose heart disease. ECG301 does not include any automated measurement or analysis of the ECG signal, and does not produce interpretative statements. The ECG recorded is not intended to be used for measurement or analysis of the ECG signal, or for the automated generation of interpretative statements.
Device Story
Electrocardiograph (ECG301) acquires 12-lead ECG signals simultaneously via body surface electrodes; device features LCD touchscreen, functional buttons, and built-in thermal dot-matrix recorder for report printing; powered by AC or internal rechargeable battery; used in hospitals/healthcare facilities by physicians and trained professionals; device provides raw ECG signal recordings to assist clinicians in heart disease diagnosis; device explicitly excludes automated measurement, analysis, or interpretative statements; clinical benefit derived from providing high-quality ECG data for professional interpretation.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ANSI AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-25 confirmed safety and performance specifications.
Indicated for acquisition of ECG signals from adult and pediatric patients (above 3 years old) via body surface electrodes in hospitals or healthcare facilities by trained professionals to assist in heart disease diagnosis.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
Shenzhen Le Medical Technology Co., Ltd. % You Yijie Manager Oimmig Medical Consulting Service Co., Ltd. RM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu Guangzhou, Guangdong 510663 China
Re: K232816
Trade/Device Name: Electrocardiograph, model: ECG301 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: May 6, 2024 Received: May 6, 2024
Dear You Yijie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Kimberly Crowley
For: Jennifer Shih Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K232816
Device Name Electrocardiograph, model: ECG301
The intended use of ECG301 is to acquire ECG signals from adult and pediatric patients(above 3 years old) through body surface ECG electrodes. The electrocardiograph is intended to be used only in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by ECG301 can help users to diagnose heart disease. ECG301 does not include any automated measurement or analysis of the ECG signal, and does not produce interpretative statements. The ECG recorded is not intended to be used for measurement or analysis of the ECG signal, or for the automated generation of interpretative statements.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 5: 510(k) Summary
# 1. Submitters Information
#### Establishment Registration Information
| Name: | Shenzhen Le medical Technology Co., Ltd. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 5F, Block A, XinYongFeng Industrial Park,<br>No.2,Tianbao Road, YinRenShi Community, ShiYan<br>Sub-District, Baoan District, ShenZhen, China. |
| Contact Person of applicant: | |
| Name: | wuyaxiong |
| Address: | 5F, Block A, XinYongFeng Industrial Park,<br>No.2,Tianbao Road, YinRenShi Community, ShiYan<br>Sub-District, Baoan District, ShenZhen, China. |
| TEL: | +86 13316869982 |
| Email: | Wuyaxiongda@163.com |
| Contact Person of the Submission: | |
| Name: | You Yijie |
| Address: | RM.406, Building C, Run Science Park, No.18<br>Shenzhou Road, Huangpu, Guangzhou, Guangdong<br>510663 P.R.China |
| TEL: | (+86)020-82245821 |
| FAX: | (+86)020-82245821 |
| Email: | jet.you@gimmiq-med.com |
#### Date prepared: Sep. 12, 2023
### 2. Device Information
| Trade Name: | Electrocardiograph |
|-----------------------|------------------------|
| Model: | ECG301 |
| Classification name: | Electrocardiograph |
| Common or Usual Name: | Cardiovascular Devices |
| Review panel: | Cardiovascular |
| Product code: | DPS |
| Regulation Class: | II |
| Regulation Number: | 870.2340 |
# 3. Predicate Device Information
| 510(k)<br>submitter/holder: | Edan Instruments, Inc. |
|-----------------------------|-----------------------------------|
| 510(K) Number: | K171943 |
| Trade Name: | Electrocardiographs |
| Model: | SE-301, SE-300A, SE-300B, SE-300G |
{5}------------------------------------------------
Classification name: Electrocardiograph Review panel: Cardiovascular Product code: DPS Regulation Class: Regulation Number: 870.2340
## 4. Device description
Electrocardiograph, models: ECG301, is to acquire multi-channel ECG signals from adult and pediatric patients(above 3 years old) through body surface ECG electrodes. The device can gather and record ECG signals of 12 leads simultaneously. The cardiogram recorded by ECG301 can help users to diagnose heart disease. ECG301 does not include any automated measurement or analysis of the ECG signal, and does not produce interpretative statements. The ECG recorded is not intended to be used for measurement or analysis of the ECG signal, or for the automated generation of interpretative statements.
Electrocardiograph, models: ECG301, has LCD display with touchscreen and functional buttons. The device can be powered from AC power or through an embedded re-chargeable battery. The device contains built-in thermal dot-matrix recorder for ECG report printing. The Electrocardiograph includes main unit and an AC power cord.
### 5. Indications for Use
The intended use of ECG301 is to acquire ECG signals from adult and pediatric patients(above 3 years old) through body surface ECG electrodes. The electrocardiograph is intended to be used only in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by ECG301 can help users to diagnose heart disease. ECG301 does not include any automated measurement or analysis of the ECG signal, and does not produce interpretative statements. The ECG recorded is not intended to be used for measurement or analysis of the ECG signal, or for the automated generation of interpretative statements.
# 6. Summary of technological characteristics of device compared to the
| SE<br>Comparisons | Subject device<br>(Electrocardiograph,<br>model: ECG301) | Predicate device<br>(Electrocardiographs, Model: SE-<br>301, SE-300A, SE-300B, SE-3) | Discussion of<br>difference | |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510K Number | K232816 | K171943 | / | |
| Classification | 21CFR 870.2340 | 21CFR 870.2340 | Same | |
| Product Code | DPS | DPS | Same | |
| Model | ECG301 | SE-301, SE-300A, SE-300B, SE-3 | / | |
| Indication for use | The intended use of ECG301 is to<br>acquire ECG signals from adult and<br>pediatric patients(above 3 years<br>old) through body surface ECG<br>electrodes. The electrocardiograph<br>is intended to be used only in<br>hospitals or healthcare facilities by<br>doctors and trained healthcare<br>professionals. The cardiogram<br>recorded by ECG301 can help<br>users to diagnose heart disease. | The intended use of the 3-Channel<br>Electrocardiograph is to acquire ECG<br>signals from adult and pediatric<br>patients (beginning at birth through 21<br>years of age) through body surface<br>ECG electrodes. The<br>electrocardiograph is intended to be<br>used only in hospitals or healthcare<br>facilities by doctors and trained<br>healthcare professionals. The<br>cardiogram recorded by the 3-Channel | Similar, the subject<br>device has same<br>indication about<br>ECG signals with<br>12-lead<br>synchronous<br>acquisition and<br>record through<br>body surface ECG<br>electrodes with<br>predicate device | |
| | ECG301 does not include any<br>automated measurement or<br>analysis of the ECG signal, and<br>does not produce interpretative<br>statements. The ECG recorded is<br>not intended to be used for<br>measurement or analysis of the<br>ECG signal, or for the automated<br>generation of interpretative<br>statements. | Electrocardiograph can help users to<br>analyze and diagnose heart disease.<br>However, the ECG with measurements<br>and interpretive statements is offered<br>to clinicians on an advisory basis only. | not have ECG with<br>measurements and<br>interpretive<br>statements.<br>Therefore,<br>indication for use of<br>subject<br>device(ECG301) is<br>covered by<br>predicate device,<br>the difference does<br>not raise new<br>questions of safety<br>and effectiveness. | |
| Safety Standards | ANSI AAMI ES 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-25 | IEC 60601-1<br>IEC 60601-1-2<br>IEC/EN60601-2-25 | Same | |
| Anti-electric-shock<br>type | Class I with external power supply<br>and internal power supply | Class I with internal power supply | Different<br>(Discussion is<br>indicated in D1) | |
| Anti-electric-shock<br>degree | Type CF | Type CF | Same | |
| Degree of<br>protection against<br>harmful ingress of<br>water | Ordinary equipment (Sealed<br>equipment without liquid proof) | Ordinary equipment (Sealed<br>equipment without liquid proof) | Same | |
| Degree of safety of<br>application in the<br>presence of<br>flammable gas | Equipment not suitable for use in<br>the presence of flammable gas | Equipment not suitable for use in the<br>presence of flammable gas | Same | |
| Working mode: | Continuous operation | Continuous operation | Same | |
| EMC: | CISPR 11 Group 1, Class A | CISPR 11 Group 1, Class A | Same | |
| Environment Specifications | | | | |
| Temperature | | | | |
| Transport | -20℃~55℃ | -20℃~55℃ | Same | |
| Storage | -20℃~55℃ | -20°C ~ +55°C | Same | |
| Working | 5°C ~40°C | 5°C ~40°C | Same | |
| Relative Humidity: | | | | |
| Transport &<br>Storage | 25% RH ~ 93% RH Non-<br>Condensing | 25% RH ~ 95% RH Non-Condensing | Similar | |
| Working | 25% RH ~ 85% RH Non-<br>Condensing | 25% RH ~ 80% RH<br>Non-Condensing | Similar | |
| Atmospheric Pressure: | | | | |
| Transport &<br>Storage | 700 hPa ~ 1060 hPa | 700 hPa ~ 1060 hPa | Same | |
| Working | 800 hPa ~ 1060 hPa | 800 hPa ~ 1060 hPa | Same | |
| Power Supply Specifications | | | | |
| Mains Supply: | Operating voltage =100 V - 240 V | Operating voltage =100 V-115 V~/220<br>V-240 V~ | Different | |
| | Operating frequency = 50 Hz / 60<br>Hz | Operating frequency = 50 Hz / 60 Hz | (Discussion is<br>indicated in D2) | |
| | input power = 90 VA | input power = 35 VA | | |
| Built-in Lithium<br>Battery Pack: | Rated voltage = 7.4 V | Rated voltage = 14.4 V | | |
| | Rated capacity =4400 mAh<br>When the battery is fully charged,<br>ECG301 can work normally 2<br>hours. | Rated capacity =1600 mAh<br>When the battery is fully charged, SE-3<br>can work normally 6 hours. | Different<br>(Discussion is<br>indicated in D3) | |
| | Cycle life ≥ 300 times | Cycle life ≥ 300 times | | |
| Performance Specifications | | | | |
| Recording | | | | |
| Recorder: | Thermal dot-matrix recorder | Thermal dot-matrix recorder | Same | |
| Recorder Paper: | Rolled thermal paper | Folded thermal paper Rolled thermal<br>paper | Same | |
| Effective Width: | 80mm | 72 mm | Different<br>(Discussion is<br>indicated in D4) | |
| Paper Speed: | 12.5 mm/s, 25 mm/s, 50mm/s<br>(±3% ) | 5 mm/s, 6.25 mm/s, 10 mm/s, 12.5<br>mm/s, 25 mm/s, 50 mm/s(±3%) | Covered in<br>predicate device | |
| Accuracy of data: | ±5% (x-axis), ±5% (y-axis) | ±5% (x-axis), ±5% (y-axis) | Same | |
| HR Recognition | | | | |
| Technique: | Peak-Peak Detection | Peak-Peak Detection | Same | |
| HR Range: | 30 BPM ~ 300 BPM | 30 BPM ~ 300 BPM | Same | |
| Accuracy: | ±1 BPM | ±1 BPM | Same | |
| ECG Unit | | | | |
| Leads: | Standard 12 leads | Standard 12 leads | Same | |
| Acquisition Mode: | Simultaneously 12 leads | Simultaneously 12 leads | Same | |
| A/D Resolution: | 12 bits | 12 bits | Same | |
| Time Constant: | ≥ 3.2 s | ≥ 3.2 s | Same | |
| Frequency<br>Response: | 0.05 ~ 150Hz(-3db) | 0.05 Hz ~ 150 Hz (-3 dB) | Same | |
| Input Impedance: | ≥50MΩ ( 10Hz ) | ≥50 MΩ (10 Hz) | Same | |
| Input Circuit<br>Current: | ≤0.05 µA | ≤0.05 µA | Same | |
| Input Voltage<br>Range | ≤±10 mVpp | ≤±5 mVpp | Different<br>(Discussion is<br>indicated in D5) | |
| Calibration Voltage: | 1mV±2% | 1 mV±3% | Different<br>(Discussion is<br>indicated in D6) | |
| DC Offset Voltage: | ±500 mV | ±500 mV | Same | |
| Noise: | < 15μVp-p | ≤12.5µVp-p | Different<br>(Discussion is<br>indicated in D7) | |
| Multi-channel<br>Crosstalk | ≤0.5 mm | ≤0.5 mm | Same | |
| Filter | AC Filter: Off /50Hz/60Hz | AC Filter: On / Off | Different | |
| | DFT Filter: Off/0.05Hz / 0.5Hz…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.