Hygenix

K232797 · Shenzhen Hygenix Industrial Co., Ltd. · FXX · Dec 11, 2023 · General, Plastic Surgery

Device Facts

Record IDK232797
Device NameHygenix
ApplicantShenzhen Hygenix Industrial Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateDec 11, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Hygenix is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This device is disposable, non-sterile and for single use only.

Device Story

Hygenix is a 3-ply, flat-pleated, disposable surgical mask; intended for use by healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulates. Device consists of inner/outer spun-bond polypropylene layers and a melt-blown polypropylene filter layer; includes a PC-coated aluminum nose wire for fit; secured by polyester/spandex ear loops. Used in clinical settings for infection control. Provides physical barrier against fluids and particulates; benefits patient and staff by reducing cross-contamination risk. Single-use, non-sterile.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F2100-19, ASTM F2101-19 (BFE >98%), ASTM F2299 (PFE >98%), ASTM F1862 (Fluid resistance 160 mmHg), EN 14683 (Delta P <6.0 mm H2O/cm2), and 16 CFR 1610 (Flammability Class 1). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

Technological Characteristics

3-ply flat-pleated mask. Materials: Spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter), PC-coated aluminum nose wire, polyester/spandex ear loops. Dimensions: 175x95 mm. Performance: ASTM F2100 Level 3. Biocompatibility: ISO 10993-1, -5, -10. Non-sterile, single-use.

Indications for Use

Indicated for patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. December 11, 2023 Shenzhen Hygenix Industrial Co., Ltd % Jerry Doane US Agent FDAbasics 5815 SW 11th Court Rd Ocala, Florida 34473 Re: K232797 Trade/Device Name: Hygenix Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 21, 2023 Received: September 12, 2023 Dear Jerry Doane: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Allan Guan -S For Bifeng Qian, M.D., Ph.D. Assistant Director {2}------------------------------------------------ DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232797 Device Name Hygenix ### Indications for Use (Describe) The Hygenix is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms. body fluids and particulate material. These face intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This device is disposable, non-sterile and for single use only. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | ☐ Registration (Part 41 CFR 229 Subpart B) | ☑ One-Time Correction (49 CFR 229 Subpart C) | |--------------------------------------------|----------------------------------------------| |--------------------------------------------|----------------------------------------------| | | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary (K232797) [AS REQUIRED BY 21CFR807.92] #### I. APPLICANT INFORMATION | Submitter's Name | Shenzhen Hygenix Industrial Co., Ltd | |---------------------------------|--------------------------------------------------------------------------------------------------| | Submitter's Address | 3rd Floor, Building A3, Fang Xing Technology Park,<br>Longgang District, Shenzhen, China, 518100 | | Name of Contact Person | Jia Du | | Designation | General Manager | | Division Name of Contact Person | Quality Management | | Contact Number | +86-755-89788050 | | Contact E-mail | Eva.du@adurochina.com | | Date of Summary Prepared | 07/Nov/2023 | #### II. DEVICE DETAILS | Device Trade Name | Hygenix | |----------------------------|-----------------------| | Device Common Name | 3 Ply Disposable Mask | | Model(s) | Ear Loop | | Device Classification name | Mask, Surgical | | Regulation Number | 21 CFR 878.4040 | | Device Class | Class II | | Product Code | FXX | # III. PREDICATE DEVICE DETAILS | Device Trade Name | Disposable Medical Face Mask (ear loop) | |--------------------------|-------------------------------------------------------| | Device Manufacturer Name | Xiantao Topmed Nonwoven Protective Products Co., Ltd. | {5}------------------------------------------------ | Model(s) | Ear Loop | |-------------------|-----------------| | 510(k) Number | K221196 | | Regulation Number | 21 CFR 878.4040 | | Device Class | Class II | | Product Code | FXX | #### DEVICE DESCRIPTION IV. The Hygenix is a 3 Ply flat- pleated and Disposable surgical mask. The inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The nose wire contained in the proposed device is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of PC coated Aluminium wire. The model ear loop is held in place over the user's mouth and nose by two elastic ear loops welded to the facemask. The ear loops are made of Polyester and Spandex. The dimensions of the Mask are: length- 175 ± 5 mm and width- 95 ± 5 mm. These masks are single-use, disposable devices, provided non-sterile. #### V. INDICATIONS FOR USE The Hygenix is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This device is disposable, non-sterile and for single use only. ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE | SI. No | Features compared | Proposed Device | Predicate Device | Result | |----------------------|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | General Information | | | | | | 1. | 510(k) Number | K232797 | K221196 | - | | 2. | Manufacturer | Shenzhen Hygenix Industrial<br>Co., Ltd | Xiantao Topmed Nonwoven<br>Protective Products Co., Ltd. | - | | 3. | Common Name | 3 Ply Disposable Mask | Disposable Medical Face Mask<br>(ear loop) | Same | | SI.<br>No | Features<br>compared | Proposed Device | Predicate Device | Result | | 4. | Classification Name | Mask, Surgical | Mask, Surgical | Same | | 5. | Classification and<br>Regulation number | Class II, 21 CFR 878.4040 | Class II, 21 CFR 878.4040 | Same | | 6. | Product Code | FXX | FXX | Same | | 7. | Indications For Use | The Hygenix is intended to be<br>worn to protect both the patient<br>and health care personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This device is<br>disposable, non-sterile and for<br>single use only. | The Disposable Medical Face<br>Masks are intended to be worn<br>to protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This is a single use,<br>disposable device(s), provided<br>non-sterile. | Same | | 8. | Model specifications | 3 ply flat-pleated masks with<br>ear loops | Ear loop, Flat pleated,3 layers | Same | | | Materials | | | | | 9. | Outer layer | Spun bond Polypropylene<br>(SBPP) | Spun bond Polypropylene | Same | | 10. | Filter layer | Melt Blown Polypropylene<br>(MBPP) | Melt Blown Polypropylene | Same | | 11. | Inner layer | Spun bond Polypropylene<br>(SBPP) | Spun bond Polypropylene | Same | | 12. | Nose wire | PC coated Aluminium wire | Galvanized iron wire coated by<br>PE | Similar | | 13. | Ear loop | Polyester and Spandex | Polyester + Spandex | Same | | 14. | Mask color | Blue | Blue | Same | | 15. | Dimensions | Length- 175 ± 5 mm<br>Width- 95 ± 5 mm | Length- 175 ± 5 mm<br>Width- 95 ± 5 mm | Same | | SI.<br>No | Features<br>compared | Proposed Device | Predicate Device | Result | | 16. | OTC Use | Yes | Yes | Same | | 17. | Sterility | Non-sterile | Non-sterile | Same | | 18 | Reusability | Single use | Single use | Same | | 19 | ASTM F2100 Level | Level 3 | Level 3 | Same | | Non-Clinical Testing | | | | | | 20. | Fluid resistance | Pass at 160 mmHg | Pass at 160 mmHg | Same | | 21. | Flammability | Class 1 | Class 1 | Same | | 22. | Particulate Filtration<br>Efficiency (PFE) | ≥ 98% | ≥ 98% | Same | | 23. | Bacterial Filtration<br>Efficiency (BFE) | ≥ 98% | ≥ 98% | Same | | 24. | Differential pressure<br>( $Δ$ Ρ) | < 6.0 mm H2O/cm² | < 6.0 mm H2O/cm² | Same | | 25. | Biocompatibility<br>Testing | Non-cytotoxic<br>Non-irritating<br>Non-sensitizing | Non-cytotoxic<br>Non-irritating<br>Non-sensitizing | Same | ### Table 1: General Comparison {6}------------------------------------------------ {7}------------------------------------------------ ## VII. SUMMARY OF NON-CLINICAL TESTING ### Performance Tests Hygenix is subjected to the following performance tests according to the requirements provided in the guidance Surgical Masks - Premarket Notification [510(k)] Submissions. - . Fluid resistance - . Bacterial filtration efficiency - . Particulate filtration efficiency - . Differential pressure - . Flammability ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks {8}------------------------------------------------ ASTM F2101-19 Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus. ASTM F2299 / F2299M - 03(2017) Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres EN 14683 (Annex C): 2019 Medical Face Masks - Requirements And Test Methods ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity) 16 CFR 1610 Standard for the Flammability of clothing textiles ### Biocompatibility The materials used in the Hygenix is biocompatible based on the biocompatibility tests mentioned in the guidance Surgical Masks - Premarket Notification [510(k)] Submissions: - . In-vitro Cytotoxicity - . Skin irritation - . Skin Sensitization These tests are performed according to ISO 10993-1:2018, Biological Evaluation of Medical Devices - Part 1, Evaluation and Testing within a Risk Management Process. ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. ### Table 2: Test Summary | SI.<br>No | Purpose | Reference Standard | Acceptance<br>criteria | Result | |-----------|--------------------------------------------|-----------------------------------|------------------------|--------------------------| | 1. | Fluid resistance | ASTM F1862/ F1862M-17 | Pass at 160 mmHg | Pass at 160 mmHg | | 2. | Particulate Filtration<br>Efficiency (PFE) | ASTM F2299 / F2299M -<br>03(2017) | ≥ 98% | >98%<br>Pass | | 3. | Bacterial Filtration<br>Efficiency (BFE) | ASTM F2101-19 | ≥ 98% | > 98%<br>Pass | | 4. | Differential pressure (ΔP) | EN 14683 (Annex C):<br>2019 | < 6.0 mm H₂O/cm² | < 6.0 mm H₂O/cm²<br>Pass | {9}------------------------------------------------ | SI.<br>No | Purpose | Reference Standard | Acceptance<br>criteria | Result | |-----------|-----------------------|--------------------|------------------------|-----------------| | 5. | Flammability | 16 CFR 1610 | Class 1 | Class 1<br>Pass | | 6. | In-vitro Cytotoxicity | ISO 10993-5:2009 | Non-cytotoxic | Pass | | 7. | Skin Irritation | ISO 10993-10:2010 | Non-irritating | Pass | | 8. | Skin Sensitization | ISO 10993-10:2010 | Non-sensitizing | Pass | ## VIII. Clinical Test Data No clinical study is included in this submission. #### CONCLUSION IX. The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...