The Wearable Breast Pump, Model S18 is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
Device Story
Wearable breast pump for single-user home use; powered by internal rechargeable Li-ion battery. Device consists of pump motor, milk collection parts, adjustment buckle, and charging cable. Operates via diaphragm pump creating vacuum cycles (inhale/exhaust); milk flows into bottle via pressure differential. User interface includes power, mode selection, and suction level adjustment buttons with LED status indicators. Features three modes (Massage, Stimulation, Expression) with nine suction levels each; includes 30-minute automatic shutdown. User operates device to express milk; output is collected breast milk. Benefits include portable, hands-free milk expression for lactating women.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, 60601-1-11, 60601-1-2), battery safety (IEC 62133-2), and software verification/validation. Performance testing confirmed vacuum levels, cycle frequencies, backflow protection, battery capacity, button functionality, and automatic shutdown over a 3-year simulated use-life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 6, 2024
Ningbo Dearevery Electronic Technology Co., LTD Kelong Zhang General Manager No3-1, Gongji South Road, Hudi Village, Linshan Town Yuyao, Zhejiang 315460 CHINA
Re: K232774 Trade/Device Name: Wearable Breast Pump, Model S18 Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: May 5, 2024 Received: May 6, 2024
Dear Kelong Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, PhD, Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232774
Device Name Wearable Breast Pump, Model S18
Indications for Use (Describe)
The Wearable Breast Pump, Model S18 is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – K232774
| 1. Submitter Information: | Ningbo Dearevery Electronic Technology Co., Ltd.<br>No3-1, Gongji South Road, Hudi Village, Linshan Town, Yuyao,<br>Zhejiang, China.<br>Tel :+8613858311121 Fax :+86-574-63233915<br>Email:1020666000@qq.com |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Contact Person: | Kelong Zhang(General Manager) |
| 3. Date prepared: | June 5, 2024 |
| 4. Device Information: | Device Name: Wearable Breast Pump, Model S18<br>Common Name: Powered breast pump<br>Regulation Number: 21 CFR 884.5160<br>Regulation Name: Powered Breast Pump<br>Product Code: HGX (Pump, Breast, Powered)<br>Regulatory Class: Class II |
| 5. Predicate Device<br>Information: | Device Name: Electric breast pump, model RH338<br>Manufacturer: Cixi Ruihong Electric Appliance Co., Ltd.<br>510(k) Number: K201903.<br>The predicate device has not been subject to a design-related recall. |
| 6. Device Description: | The Wearable Breast Pump, Model S18 is an electrically powered breast pump to be used in home environment by a single user. The device is provided non-sterile and can be re-used by a single user. The device is electrically powered by an internal rechargeable Li-ion battery. It can be charged using 5V DC adaptor and the device is designed to be not used during charging.<br>The device consists of the pump motor, milk collection parts, adjustment buckle and charging cable. The device functionality is controlled using an embedded firmware. The subject device has three modes of operation - Expression mode, Stimulation mode and Massage mode, and each mode of operation allows for switching between nine suction levels.<br>The device has three buttons, a power button to turn the device on/off or switch between the modes, a '+ plus' button to increase the suction level and a '- minus' to decrease the suction level within a mode. The device includes an LED status display for the device operating mode and battery status indicator on the pump body. The operating mode LED changes to specific color based on the mode of operation. The device is designed to automatically shut down after 30 minutes of continuous operation.<br>The device operates on a suction(vacuum) cycle that consists of an inhale (letting in air to release vacuum) and an exhaust (removal of air to create vacuum) process. When the breast pump is powered on, the vacuum pump exhausts the air inside the cavity. Once the suction strength reaches the set value, the vacuum pump stops and the valve for the milk opens and expressed milk flows into the bottle due to a pressure difference. |
{5}------------------------------------------------
All milk contacting components of the device are compliant with 21 CFR 177.
The Wearable Breast Pump, Model S18 is intended to express milk from 7. Indications for Use: lactating women in order to collect milk from their breasts. The device is intended for a single user.
#### 8. Predicate Device Comparison
The following table compares the intended use and technological characteristics of the subject and predicate device.
| | Subject Device | Predicate Device | Comment |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | Wearable Breast Pump,<br>Model S18 | Electric Breast Pump,<br>Model RH-338 | |
| | K232774 | K201903 | |
| Product Code | HGX (pump, breast, powered) | HGX (pump, breast, powered) | Same |
| Regulation Number | 21CFR 884.5160, Powered<br>breast pump | 21CFR 884.5160, Powered<br>breast pump | Same |
| Class | Class II | Class II | Same |
| Indication for Use | The Wearable Breast Pump,<br>model S18 is intended to<br>express milk from lactating<br>women in order to collect milk<br>from their breasts. The device is<br>intended for a single user. | The RH-338 and RH-228<br>electric breast pump is<br>intended to be used by<br>lactating women to express<br>and collect milk from their<br>breasts. It is intended for a<br>single user. | Similar |
| Single/double pumping | Single | Single | Same |
| Media Separation<br>(backflow protection) | Yes | Yes | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Patient population | Lactating women | Lactating women | Same |
| Suction modes | 3 modes - Massage mode,<br>Stimulate mode, Expression<br>mode | 3 modes - Massage mode,<br>Stimulate mode, Expression<br>mode | Same |
| Suction Levels | 9 levels, in each mode | 9 levels, in each mode | Same |
| Vacuum strength (mmHg) | Massage mode: 101-165<br>Stimulate mode:108-180<br>Expression mode: 166-270 | Massage mode: 135-255<br>Stimulate mode: 45-165<br>Expression mode: 110-300 | Different |
| Cycle frequency (Cycles per<br>minute) | Massage mode: 40-72<br>Stimulation mode: 24-40<br>Expression mode: 31-47 | Massage mode: 56-72<br>Stimulate mode: 115-155<br>Expression mode: 25-52 | Different |
| Battery indication | Yes | Yes | Same |
| Pump Type | Diaphragm | Diaphragm | Same |
| Power Supply | Li-ion battery | Li-ion battery and USB | Different |
| User Interface | Power button, mode button,<br>Gear plus button, Gear minus<br>button | On/Off button,<br>Increase/decrease vacuum<br>button, Switch mode button; | Similar |
| Maximum Suction | 283.5 mmHg | 300 mmHg | Different |
Table 1 Comparison Between the Subject and Predicate Device
The subject and predicate devices have similar indications for use statements and the same intended use (i.e., for collection of breast milk from the breasts of lactating women).
The subject and predicate devices have similar technological features including suction modes and levels, pump type, user interface etc. The differences between the subject and predicate device include suction
{6}------------------------------------------------
strength, cycle frequency and power source. These differences do not raise new questions of safety and effectiveness and these differences in technological characteristics may be evaluated through performance testing.
## 9. Summary of Non-Clinical Performance Testing:
### Biocompatibility Testing
The biocompatibility evaluation for the Wearable Breast Pump was conducted in accordance with 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process.". The tissue-contacting components were evaluated for cytotoxicity, irritation and were found to be noncytotoxic, non-irritating, and non-sensitizing.
#### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the Wearable Breast Pump device per the following standards:
- IEC 60601-1:2005+A1:2012+A2:2020, Medical equipment-Part 1: General requirements for basic safety, and essential performance.
- IEC 60601-1-11:2015 +A1:2020. Medical electrical equipment-Part 1-11: General reguirements ● for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1-2:2014+A1:2020, Medical electrical equipment-Part 1-2: General requirements for ● basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests standard for EMC.
- IEC 62133-2 :2017 +A1:2021, Secondary cells and batteries containing alkaline or other non-● acid electrolytes - Safety requirements for portable secondary lithium cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
#### Software Verification and Validation Testing
Software verification & validation was provided in accordance with 2023 FDA guidance: Content of Premarket Submissions for Device Software Functions. The documentation level for the subject device is Basic Documentation Level.
#### Performance Testing and Use Life Verification
The subject device has a use-life of 3-years. The following performance testing were conducted at time T=0 and after fatigue testing equivalent to 3-years of device use to demonstrate that the device meets its performance specifications throughout its proposed use-life:
- Vacuum level and cycle frequency testing was performed for all modes and levels of operation to confirm that the device meets specifications for each mode and level.
- Backflow testing was conducted to verify that the fluid does not flow back into the tubing and to . the pump.
- . Battery capacity and service time testing was conducted to demonstrate that the battery meets its specified service time and the LED on the device accurately displays the battery charge status. The battery was also tested for its charging time.
- The button performance test was conducted to verify that the buttons maintain their functionality over proposed use-life of the device.
- Automatic shutdown function testing was conducted to demonstrate that the device shuts down automatically after 30 minutes of continuous operation.
#### 10. Conclusion:
The results of the performance testing described above demonstrate that the Wearable Electric Breast
{7}------------------------------------------------
Pump, Model S18 is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.