Rapid Marijuana (THC) Test Strip 20, Rapid Marijuana (THC) Test Dipcard 20, Rapid Marijuana (THC) Test Strip 50, Rapid Marijuana (THC) Test Dipcard 50
Applicant
Co-Innovation Biotech Co., Ltd.
Product Code
LDJ · Clinical Toxicology
Decision Date
Jan 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3870
Device Class
Class 2
Indications for Use
Rapid Marijuana (THC) Test Strip 20 and Rapid Marijuana (THC) Test Dipcard 20 is a rapid, screening test for the qualitative detection of Marijuana and their metabolites in human urine at the cut-off concentration of 20 ng/mL. Rapid Marijuana (THC) Test Strip 50 and Rapid Marijuana (THC) Test Dipcard 50 is a rapid, screening test for the qualitative detection of Marijuana and their metabolites in human urine at the cut-off concentration of 50 ng/mL. The tests contain two formats:1) Test Strip and 2) Test Dipcard. The tests are intended for in vitro diagnostics use. They are intended for over-the-counter use. The tests provide only a preliminary result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
Device Story
Lateral flow immunochromatographic assay for qualitative detection of Marijuana metabolites in human urine. Device formats include test strips and dipcards. Principle: competitive binding; urine sample migrates via capillary action across membrane pre-coated with Marijuana antigen; antibody-dye conjugate binds to drug if present (preventing colored band at Test Region, indicating positive) or binds to drug-protein conjugate (forming colored band, indicating negative). Performed without instrumentation by lay users in home or point-of-care settings. Results interpreted visually by user. Preliminary positive results require professional confirmation via GC/MS or LC/MS.
Clinical Evidence
No clinical diagnostic studies performed. Evidence consists of analytical performance (precision, cross-reactivity, interference, pH/specific gravity effects) and a lay user study (n=720 total, 360 per cutoff). Lay user study demonstrated high agreement with LC/MS confirmed samples across concentrations (0 to 100% of cutoff).
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Formats: test strip and dipcard. No instrumentation required. Visual readout. Materials include pre-coated membrane and antibody-dye conjugate. Qualitative detection only.
Indications for Use
Indicated for qualitative screening of Marijuana and metabolites in human urine for over-the-counter use. Cut-off concentrations are 20 ng/mL or 50 ng/mL. Provides preliminary results requiring confirmation by GC/MS or LC/MS.
Regulatory Classification
Identification
A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.
Special Controls
*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232732
B Applicant
Co-Innovation Biotech Co., Ltd.
C Proprietary and Established Names
Rapid Marijuana (THC) Test Strip 20, Rapid Marijuana (THC) Test Dipcard 20, Rapid Marijuana (THC) Test Strip 50, Rapid Marijuana (THC) Test Dipcard 50
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LDJ | Class II | 21 CFR 862.3870 - | |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Marijuana
C Type of Test:
Competitive binding, lateral flow immunochromatographic assay.
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
Rapid Marijuana (THC) Test Strip 20 and Rapid Marijuana (THC) Test Dipcard 20 is a rapid, screening test for the qualitative detection of Marijuana and their metabolites in human urine at the cut-off concentration of 20 ng/mL.
Rapid Marijuana (THC) Test Strip 50 and Rapid Marijuana (THC) Test Dipcard 50 is a rapid, screening test for the qualitative detection of Marijuana and their metabolites in human urine at the cut-off concentration of 50 ng/mL.
The tests contain two formats: 1) Test Strip and 2) Test Dipcard. The tests are intended for in vitro diagnostics use. They are intended for over-the-counter use.
The tests provide only a preliminary result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
C Special Conditions for Use Statement(s):
OTC - Over The Counter
D Special Instrument Requirements:
N/A
IV Device/System Characteristics:
A Device Description:
Rapid Marijuana (THC) Test Strip 20 (and 50) and Rapid Marijuana (THC) Test Dipcard 20 (and 50) are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Marijuana and its metabolites at or above the cut-off concentration of 20 ng/ml (and 50 ng/ml). The tests are performed without the use of an instrument.
Test Strips and Test Dipcards use identical test strips made with the same chemical formulation and manufacturing procedures.
Content of the kits: Test Strip (one test in one pouch), Desiccant (for storage purposes only), Leaflet with instructions for use, Urine Cup (optional), Labeled Vial (optional), Plastic Bag (optional), Mailing Box (optional)
B Principle of Operation:
Rapid Marijuana (THC) Test Strip 20 (and 50) and Rapid Marijuana (THC) Test Dipcard 20 (and 50) are competitive binding, lateral flow immunochromatographic assays that are used to screen for the presence of Marijuana and its metabolites at or above the cut-off concentration of 20 (or 50) ng/mL in urine. The tests are performed without the use of an instrument. When the absorbent end of the test device is immersed into the urine sample, the urine is absorbed into the device by capillary action, mixes with the antibody-dye conjugate, and flows across the pre-coated membrane, in which Marijuana antigen is coated respectively.
K232732 - Page 2 of 14
{2}
When Marijuana and its metabolites level are at or above 20ng/ml or 50ng/ml (the cutoff), the drugs in the sample bind to the antibody-dye conjugate preventing the antibody-dye conjugate from binding to the drug-protein conjugate immobilized in the Test Regions (T) of the device. This prevents the development of a distinct colored band in the test region (T), indicating a positive result.
When Marijuana and its metabolites levels are below 20ng/ml or 50ng/ml, antibody-dye conjugate binds to the drug-protein conjugate immobilized in the Test Region (T) of the device. This produces a colored band in the test region (T) which indicates a negative result.
To serve as a procedure control, a colored line will appear at the Control Region (C), if the test has been performed properly.
V Substantial Equivalence Information:
A Predicate Device Name(s):
BIOEASY Marijuana Test Dip Card 40, BIOEASY Marijuana Test Dip Card 20, BIOEASY Marijuana Test Strip 40, BIOEASY Marijuana Test Strip 20
B Predicate 510(k) Number(s):
K192515
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K232732 | K192515 |
| --- | --- | --- |
| Device Trade Name | Rapid Marijuana (THC)
Test Strip 20
Rapid Marijuana (THC)
Test Dipcard 20
Rapid Marijuana (THC)
Test Strip 50
Rapid Marijuana (THC)
Test Dipcard 50 | BIOEASY Marijuana
Test Dip Card 20
BIOEASY Marijuana
Test Strip 20
BIOEASY Marijuana
Test Dip Card 40
BIOEASY Marijuana
Test Strip 40 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Qualitative detection of Marijuana in urine | Same |
| Specimen type | Human urine | Same |
| Methodology | Competitive binding, lateral flow | Same |
K232732 - Page 3 of 14
{3}
K232732 - Page 4 of 14
| | immunochromatographic assays based on the principle of antigen antibody immunochemistry | |
| --- | --- | --- |
| General Device Characteristic Differences | | |
| Calibrator and Cutoff Values | Marijuana (THC) 20 ng/mL or 50 ng/mL | Marijuana (THC) 20 ng/ml or 40 ng/ml |
| End user environment | Over-the-counter use | Prescription use |
## VI Standards/Guidance Documents Referenced:
None
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
Precision studies were performed using 3 lots each of test Strip and test Dipcard. Drug free specimens were spiked with target drug (-)-11-nor-9-Carboxy-Δ9-THC at 0, ±75% cutoff, ±50% cutoff, ±25% cutoff and +100% cutoff of drug. The concentrations of the target drugs were confirmed with LC/MS. Each concentration of the urine specimen was divided into aliquots. Separate sets of blinded coded samples were assigned and randomized prior to testing. The study was conducted by 6 operators at 3 Point-of-Care sites. Two operators per location tested 3 aliquots at each concentration for each lot per day (3 runs/day) for 10 non-consecutive days using one device lot per location. One operator tested the test strip format and the second operator tested the test dipcard format. There were 1620 observations by 3 sites at 9 concentrations.
Rapid Marijuana (THC) Test Strip 20:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Result | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Lot 1 | | Lot 2 | | Lot 3 | |
| | | | Positive | Negative | Positive | Negative | Positive | Negative |
| 0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 5ng/ml | -75%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 10ng/ml | -50%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 15ng/ml | -25%cutoff | 60 | 6 | 54 | 4 | 56 | 4 | 56 |
| 20ng/ml | cutoff | 60 | 34 | 26 | 36 | 24 | 32 | 28 |
{4}
Rapid Marijuana (THC) Test Dipcard 20:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Result | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Lot 1 | | Lot 2 | | Lot 3 | |
| | | | Positive | Negative | Positive | Negative | Positive | Negative |
| 0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 5ng/ml | -75%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 10ng/ml | -50%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 15ng/ml | -25%cutoff | 60 | 4 | 56 | 6 | 54 | 4 | 56 |
| 20ng/ml | cutoff | 60 | 36 | 24 | 38 | 22 | 34 | 26 |
| 25ng/ml | +25%cutoff | 60 | 56 | 4 | 56 | 4 | 58 | 2 |
| 30ng/ml | +50%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 35ng/ml | +75%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 40ng/ml | +100% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
Rapid Marijuana (THC) Test Strip 50:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Result | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Lot 1 | | Lot 2 | | Lot 3 | |
| | | | Positive | Negative | Positive | Negative | Positive | Negative |
| 0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 12.5ng/ml | -75%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 25ng/ml | -50%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 37.5ng/ml | -25%cutoff | 60 | 4 | 56 | 6 | 54 | 2 | 58 |
| 50ng/ml | cutoff | 60 | 36 | 24 | 38 | 22 | 34 | 26 |
| 62.5ng/ml | +25%cutoff | 6 | 56 | 4 | 56 | 4 | 58 | 2 |
| 75ng/ml | +50%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 87.5ng/ml | +75%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 100ng/ml | +100% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
K232732 - Page 5 of 14
{5}
Rapid Marijuana (THC) Test Dipcard 50:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Result | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Lot 1 | | Lot 2 | | Lot 3 | |
| | | | Positive | Negative | Positive | Negative | Positive | Negative |
| 0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 12.5ng/ml | -75%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 25ng/ml | -50%cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 37.5ng/ml | -25%cutoff | 60 | 6 | 54 | 2 | 58 | 6 | 54 |
| 50ng/ml | cutoff | 60 | 38 | 22 | 34 | 26 | 36 | 24 |
| 62.5ng/ml | +25%cutoff | 60 | 58 | 2 | 56 | 4 | 54 | 6 |
| 75ng/ml | +50%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 87.5ng/ml | +75%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 100ng/ml | +100%cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
2. Linearity:
Not applicable, this device is intended for qualitative use only.
3. Analytical Specificity/Interference:
Cross-Reactivity:
To determine cross-reactivity, drug metabolites and structurally similar compounds were tested. All the components were added to drug-free normal human urine. Each sample was tested in 3 replicates using 3 lots of the test Strip and Dip Cards (for the 20ng/ml and 50ng/ml cut-off concentrations). If any positive result was observed, the compounds were further diluted with known drug-free urine specimen sequentially to different concentrations and tested in quintuplicate, until the highest concentration that generated a negative result was reached. The cross-reacting substances with the lowest concentration that produced a positive result for each cutoff was identified and is listed in the tables below (results were the same for test strip and dipcard device formats):
Rapid Marijuana (THC) Test Strip 20 and Rapid Marijuana (THC) Dip Card 20
| Compound | Quantity equivalent to cutoff(ng/mL) | % Cross-Reactivity |
| --- | --- | --- |
| 11-nor-Δ9-THC-9-COOH | 20 | 100% |
| 11-nor-Δ8-THC-9-COOH | 20 | 100% |
K232732 - Page 6 of 14
{6}
K232732 - Page 7 of 14
Rapid Marijuana (THC) Test Strip 50 and Rapid Marijuana (THC) Dip Card 50
| Compound | Quantity equivalent to cutoff(ng/mL) | % Cross-Reactivity |
| --- | --- | --- |
| 11-nor-Δ9-THC-9-COOH | 50 | 100% |
| 11-nor-Δ^{8}-THC-9-COOH | 50 | 100% |
| Δ^{9}-THC | 10,000 | 0.3% |
| Δ^{8}-THC | 15,000 | 0.5% |
| Cannabinol | 20,000 | 0.3% |
| Cannabidiol | >100000 | Not detected |
| 11-Nor- Δ 9-THC-carboxy glucuronide | 75 | 66.7% |
| 11-Hydroxy- Δ 9-Tetrahydrocannabinol | 50 | 100% |
| (-)-11-nor-9-carboxy-Δ 9-THC | 50 | 100% |
## Interfering substances:
To determine the interference, potentially interfering substances were added to drug-free urine and target drugs urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three batches of each of the four devices (20ng/ml cut-off Strip and Dip Card & 50ng/ml Strip and Dip Card). Compounds that showed no interference at a concentration of 100μg/mL (albumin was tested at 100 mg/dL and ethanol was tested at 1%) are summarized in the list below. There were no differences observed for both the strip and dip card formats at the two different cut-offs.
{7}
Acetaminophen, Acetophenetidin, Amoxicillin, Ampicillin, Aspirin, Atenolol, Atorvastatin, Azlocillin, Benzilic acid, Benzylpenicillin, Benzoic acid, Bilirubin, Benzydamine, Caffeine, Carbamazepine, Cephalexin, Chloralhydrate, Chloramphenicol, Chlorothiazide, Chlorpheniramine, d,l-Chlorpromazine, Cholesterol, Clonidine, Cimetidine, Citalopram, Cortisone, Creatinine, Deoxycorticosterone, Dexamethasone, Dextromethorphan, Diclofenac, Diflunisal, Digoxin, Diphenhydramine, Ephedrine, $\beta$-Estradiol, Estrone-3-sulfate, Ethyl-p-aminobenzoate, Erythromycin, Fenoprofen, Flucloxacillin, Fluoxetine, Furosemide, Gentisic acid, Hemoglobin, Hydralazine, Hydrochlorothiazide, Hydrocortisone, o-Hydroxyhippuric acid, p-Hydroxytyramine, Ibuprofen, Indomethacin, Iproniazid, d,l-Isoproterenol, Isoxsuprine, Ketamine, Ketoprofen, Labetalol, Lisinopril, Loperamide, Meperidine, Meprobamate, Methoxyphenamine, Methylphenidate, Nadolol, Nalidixic acid, Naproxen, Niacinamide, Nicotine, Nifedipine, Norethindrone, Noscapine, d,l-Octopamine, Oxalic acid, Oxolinic acid, Oxymetazoline, Oxytetracycline, Papaverine, Penicillin-G, Pentazocine, Perphenazine, Phenelzine, Prednisolone, Prednisone, d,l-Propanolol, d-Pseudoephedrine, Quinacrine, Quinine, Quindine, Ranitidine, Salicylic acid, Serotonin, Sulfamethazine, Sulindac, Tetracycline, Tetrahydrozoline, Thiamine, Thioridazine, d,l-Thyroxine, Tolbutamine, Tolbutamide, Trifluoperazine, Tryptamine, Uric acid, Verapamil, Zomepirac
The results demonstrate that all above compounds show no interference with the Rapid Marijuana Test devices of the two different cut-offs and the two different formats.
## Effect of Urinary Specific Gravity
The specific gravity studies were conducted on different specific gravity including 1.000, 1.010, 1.020, 1.030, 1.040 specimens with drug free urine containing (-)-11-nor-9-Carboxy- $\Delta$9-THC at $25\mathrm{ng/ml}$ (50% below cutoff) and $75\mathrm{ng/ml}$ (50% above cutoff) and $10\mathrm{ng/ml}$ (50% below cutoff) and $30\mathrm{ng/ml}$ (50% above cutoff). All concentrations were confirmed with LC/MS. Each sample was tested using two lots for each of the four devices (20ng/ml cut-off Strip and Dip Card & 50ng/ml Strip and Dip Card). The results demonstrate that the tested range of specific gravity does not affect the test result.
## Effect of Urinary pH
The pH of an aliquot negative urine pool is adjusted to a pH range of 3 to 9 in 1 pH unit increments and spiked with (-)-11-nor-9-Carboxy- $\Delta$9-THC at $25\mathrm{ng/ml}$ (50% below cutoff) and $75\mathrm{ng/ml}$ (50% above cutoff) and $10\mathrm{ng/ml}$ (50% below cutoff) and $30\mathrm{ng/ml}$ (50% above cutoff). All concentrations were confirmed with LC/MS. Each sample was tested using two lots of the for each of the four devices (20ng/ml cut-off Strip and Dip Card & 50ng/ml Strip and Dip Card). The result demonstrate that the tested pH range does not interfere with the performance of the test.
## 4. Assay Reportable Range:
Not applicable, this device is intended for qualitative use only.
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The device is traceable to commercially available reference materials.
K232732 - Page 8 of 14
{8}
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
Analytical performance of the device around the claimed cutoff is described in the precision section VII.A1. above.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Eighty (80) clinical urine samples were collected from three hospitals/drug relief reformatories. The study was conducted by 6 nurses at three Point-of-Care sites and each site conducted one lot test Strip and Dip Card for both the Rapid Marijuana (THC) Test 50 cutoff and the Rapid Marijuana (THC) Test 20 cutoff. Each sample in both studies was divided into 6 individuals. All aliquots were blindly labeled by a nonparticipant. Samples were also randomized prior to testing.
Rapid Marijuana (THC) Test 50 cut-off
Each participant only tested one format of Rapid Marijuana (THC) Test 50, one tested with strip format another tested with Dip Card. All samples were unaltered clinical specimens, and the concentrations were confirmed with LC/MS.
Rapid Marijuana (THC) Test 20 cut-off
Each participant only tested one format of Rapid Marijuana (THC) Test 20, one tested with strip format another tested with Dip Card. All the samples were unaltered clinical specimens, and the concentrations were confirmed with LC/MS.
Comparison data of Rapid Marijuana (THC) Test Strip 20
| Test Strip | LC/MS | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. (drug free) | Neg. (<50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) | | Pos. (>+50% cutoff) |
| Lot 1 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Lot 2 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Lot 3 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
K232732 - Page 9 of 14
{9}
Comparison data of Rapid Marijuana (THC) Test Dipcard 20
| Dip Card | LC/MS | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. (drug free) | Neg. (<50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) | | Pos. (>+50% cutoff) |
| Lot 1 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Lot 2 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Lot 3 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
Comparison data of Rapid Marijuana (THC) Test Strip 50
| Test Strip | LC/MS | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. (drug free) | Neg. (<50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) | | Pos. (>+50% cutoff) |
| Lot 1 | Positive | 0 | 0 | 1 | 9 | 30 | 80 |
| | Negative | 21 | 10 | 8 | 1 | 0 | |
| Lot 2 | Positive | 0 | 0 | 1 | 9 | 30 | 80 |
| | Negative | 21 | 10 | 8 | 1 | 0 | |
| Lot 3 | Positive | 0 | 0 | 1 | 9 | 30 | 80 |
| | Negative | 21 | 10 | 8 | 1 | 0 | |
Comparison data of Rapid Marijuana (THC) Test Dipcard 50
| Dip Card | LC/MS | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. (drug free) | Neg. (<50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) | | Pos. (>+50% cutoff) |
| Lot 1 | Positive | 0 | 0 | 1 | 9 | 30 | 80 |
| | Negative | 21 | 10 | 8 | 1 | 0 | |
| Lot 2 | Positive | 0 | 0 | 1 | 9 | 30 | |
K232732 - Page 10 of 14
{10}
K232732 - Page 11 of 14
| | Negative | 21 | 10 | 8 | 1 | 0 | 80 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 3 | Positive | 0 | 0 | 1 | 9 | 30 | 80 |
| | Negative | 21 | 10 | 8 | 1 | 0 | |
## 2. Lay User Study (Home Use Consumer Study)
### Specimens:
**THC 20ng/ml cut-off**: Drug free specimens spiked with the (-)-11-nor-9-Carboxy-Δ9-THC at different concentrations containing 0 ng/ml, 10 ng/ml, 15 ng/ml, 25 ng/ml, 30 ng/ml, 40 ng/ml. All concentrations were confirmed with LC/MS.
**THC 50ng/ml cut-off**: Drug free specimens spiked with the (-)-11-nor-9-Carboxy-Δ9-THC at different concentrations containing 0 ng/ml, 25 ng/ml, 37.5 ng/ml, 62.5 ng/ml, 75 ng/ml, 100 ng/ml. All concentrations were confirmed with LC/MS.
For both the THC20 and THC 50, each concentration urine specimens were divided into 60 individual containers for a total of 360 aliquots. All concentrations were confirmed with LC/MS. All aliquots were blindly labeled by a nonparticipant. Samples were also randomized prior to testing. The study was conducted at three sites. Ages from 18 to 65 participated in the study. None of the subjects had experience with a point of care drug testing product and were untrained operators. The study included two cohorts of 360-lay-users (a total of 720 lay-users) with no experience using this product. One cohort tested the 20ng/mL device and the other cohort of 360 tested the 50ng.mL device. Both cohorts of 360 lay-users were further divided into two groups; one tested with strip format, the other tested with dipcard format. Each participant was provided with the English package insert, 1 blind labeled sample, and a device. Each participant performed only 1 test on provided specimen with one format of Rapid Marijuana (THC) Test 20ng/mL or 50ng/mL (strip or dipcard) using the English package insert as guide to perform the test. They were asked to fill out an English questionnaire after finishing the test, 360 test results for a total of 6 drug concentration levels were obtained from the Home Use Consumer Study for each test (20ng/mL and 50ng/mL).
### Results for both THC 20 and THC 50 are shown below:
Home Use Consumer Studies date of Rapid Marijuana (THC) Test Strip 20:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Strip | | Agreement (%) |
| --- | --- | --- | --- | --- | --- |
| | | | Positive | Negative | |
| 0ng/ml | Negative | 30 | 0 | 30 | 100% |
| 10ng/ml | -50%cutoff | 30 | 0 | 30 | 100% |
| 15ng/ml | -25%cutoff | 30 | 2 | 28 | 93% |
| 25ng/ml | +25%cutoff | 30 | 27 | 3 | 90% |
| 30ng/ml | +50%cutoff | 30 | 30 | 0 | 100% |
| 40ng/ml | +100%cutoff | 30 | 30 | 0 | 100% |
{11}
Home Use Consumer Studies date of Rapid Marijuana (THC) Test Dipcard 20:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Dipcard | | Agreement (%) |
| --- | --- | --- | --- | --- | --- |
| | | | Positive | Negative | |
| 0ng/ml | Negative | 30 | 0 | 30 | 100% |
| 10ng/ml | -50%cutoff | 30 | 0 | 30 | 100% |
| 15ng/ml | -25%cutoff | 30 | 3 | 27 | 90% |
| 25ng/ml | +25%cutoff | 30 | 28 | 2 | 93% |
| 30ng/ml | +50%cutoff | 30 | 30 | 0 | 100% |
| 40ng/ml | +100%cutoff | 30 | 30 | 0 | 100% |
Home Use Consumer Studies date of Rapid Marijuana (THC) Test Strip 50:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Strip | | Agreement (%) |
| --- | --- | --- | --- | --- | --- |
| | | | Positive | Negative | |
| 0ng/ml | Negative | 30 | 0 | 30 | 100% |
| 25ng/ml | -50%cutoff | 30 | 0 | 30 | 100% |
| 37.5ng/ml | -25%cutoff | 30 | 1 | 29 | 97% |
| 62.5ng/ml | +25%cutoff | 30 | 28 | 2 | 93% |
| 75ng/ml | +50%cutoff | 30 | 30 | 0 | 100% |
| 100ng/ml | +100%cutoff | 30 | 30 | 0 | 100% |
Home Use Consumer Studies date of Rapid Marijuana (THC) Test Dipcard 50:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Dipcard | | Agreement (%) |
| --- | --- | --- | --- | --- | --- |
| | | | Positive | Negative | |
| 0ng/ml | Negative | 30 | 0 | 30 | 100% |
| 25ng/ml | -50%cutoff | 30 | 0 | 30 | 100% |
| 37.5ng/ml | -25%cutoff | 30 | 2 | 28 | 93% |
| 62.5ng/ml | +25%cutoff | 30 | 28 | 2 | 93% |
| 75ng/ml | +50%cutoff | 30 | 30 | 0 | 100% |
| 100ng/ml | +100%cutoff | 30 | 30 | 0 | 100% |
K232732 - Page 12 of 14
{12}
3. Stability of the test line (Read time)
## Rapid Marijuana (THC) Test Strip 20 and Test Dipcard 20
Three drug free specimens were spiked with the (-)-11-nor-9-Carboxy-Δ9-THC concentrations at 10ng/ml, 30ng/ml and 40ng/ml. All concentrations were confirmed with LC/MS. Spiked specimens and five drug free specimens (N1-N5) were tested in replicates of ten following the procedures in the package inserts.
## Rapid Marijuana (THC) Test Strip 50 and Test Dipcard 50
Three drug free specimens spiked with the (-)-11-nor-9-Carboxy-Δ9-THC concentrations at 25ng/ml, 75ng/ml and 100ng/ml. All concentrations were confirmed with LC/MS. Spiked specimens and five drug free specimens (N1-N5) were tested in replicates of ten following the procedures in the package inserts.
The study supported the sponsor’s claimed read time of 5 to 30 minutes.
4. Matrix Comparison:
Not applicable. This device is intended to be used with urine samples only.
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
Not applicable.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
K232732 - Page 13 of 14
{13}
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K232732 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.