Q-Switched Nd:Yag laser

K232716 · Hebei Zhemai Technology Co., Ltd. · GEX · Nov 30, 2023 · General, Plastic Surgery

Device Facts

Record IDK232716
Device NameQ-Switched Nd:Yag laser
ApplicantHebei Zhemai Technology Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateNov 30, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Q-Switched Nd: Yag laser is indicated for the treatment of: benign cutaneous lessons, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and the removal of black or blue tattoos.

Device Story

Q-Switched Nd:YAG laser system delivers 1064nm or 532nm wavelength pulses to target tissue. Device consists of main unit, articulated arm with treatment probe, and foot switch. Physician operates device in clinical setting. Laser energy absorption causes rapid, localized thermal increase in target tissue, inducing cellular changes for therapeutic effect. Output is controlled via foot switch; healthcare provider directs beam to treatment zone. Benefits include non-invasive treatment of pigmented lesions and tattoos. Device includes closed water circulation for skin cooling.

Clinical Evidence

No clinical data. Bench testing only. Compliance verified with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization).

Technological Characteristics

Nd:YAG laser; 1064nm/532nm wavelengths; 6ns pulse width; 1-10Hz repetition rate; 10-1000mJ (1064nm) / 50-500mJ (532nm) output energy. Articulated 7-arm delivery system. Class 4 laser. Closed water circulation cooling. Disinfection via 75% medical alcohol. Complies with IEC 60601-1, 60601-1-2, 60601-2-22, 60825-1.

Indications for Use

Indicated for treatment of benign cutaneous lesions (warts, scars, striae, psoriasis), benign pigmented lesions (lentigines, nevus, birthmarks), and removal of black or blue tattoos in patients requiring laser therapy.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". November 30, 2023 Hebei Zhemai Technology Co., Ltd Rav Wav Official Correspondent 1st FL, N of No.1 Warehouse, W of Shuichang Rd, N of Beiyi Rd, Yanjiao Development Zone, Sanhe Langfang, Hebei 065201 China Re: K232716 Trade/Device Name: O-Switched Nd:Yag laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 4, 2023 Received: September 5, 2023 Dear Ray Way: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Tanisha L. Hithe -Tanisha L.s 2023.11.30 Hithe -S 16:03:30 -05'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K232716 Device Name Q-Switched Nd:Yag laser Indications for Use (Describe) The Q-Switched Nd: Yag laser is indicated for the treatment of: benign cutaneous lessons, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and the removal of black or blue tattoos. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ The assigned 510(k) Number: _________________________________________________________________________________________________________________________________________________ # 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. - 1. Date of Preparation:2023/11/30 - 2. Sponsor Identification #### Hebei Zhemai Technology Co.,Ltd 1st Floor, North of No.1 Warehouse, West of Shuichang Road, North of Beiyi Road, Yanjiao Development Zone, Sanhe, Langfang, Hebei, China, 065201 Contact Person: Yuting Qu Position: Sales Manager Tel: +86-18678802036 Email: feijiakeji@zoho.com.cn - 3. Designated Submission Correspondent Mr. Ray Wang Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com {5}------------------------------------------------ - Identification of Proposed Device 4. Trade Name: Q-Switched Nd:Yag laser Common Name: Powered Laser Surgical Instrument Regulatory Information Classification Name: Powered Laser Surgical Instrument Classification: II Product Code: GEX Regulation Number: 878.4810 Review Panel: General & Plastic Surgery - న. Indication For Use Statement: The Q-Switched Nd:Yag laser is indicated for the treatment of: benign cutaneous lesions, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and birthmark; and the removal of black or blue tattoos. - 6. Device Description The Q-switched laser therapy device is laser system which delivers laser at a wavelength 1064nm or 532nm and the pulse mode is single pulse. Model: FQ-L08 In laser handpiece, there is one optical cavity containing the Nd: YAG crystal. The laser beam is directed to the treatment zone. When the laser beam contacts human tissue, the energy in the beam is absorbed, resulting in a very rapidly, highly localized temperature increase to the target tissue. The instantaneous temperature increase (thermal effect) causes the cells change of target tissue to achieve laser treatment effect. | No. | Component name | Function | |-----|------------------------|------------------------------------------------| | 1 | Main unit | Main Interface | | 2 | Smart Optics 7 Arms | Articulated arm for holding of Treatment Probe | | 3 | Treatment head | Laser Deliver | | 4 | Foot switch | Control pulse light output | | 5 | Patient goggles | Eye protection | | 6 | Doctor goggles | Eye protection | | 7 | Treatment head bracket | Support and fix the treatment head | | 8 | Key | Start device | | 9 | Power cord | Supply power | | 10 | Damping | It has a shock absorption effect | | | Table 1 Main Components introduction | | |--|--------------------------------------|--| | | | | {6}------------------------------------------------ - 7. Identification of Predicate Device(s) 510(k) Number: K161926 Product Name: ND YAG Q-switch Laser Therapy Machine Manufacturer: Beijing ADSS Development Co., Ltd - Non-Clinical Test Conclusion 8. Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - A ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - > EC 60601-1-2 Edition 4.1 2020-09, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests - > IEC 60601-2-22 2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment - > IEC 60825-1 2014 Safety of laser products - Part 1: Equipment classification, and requirements - > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity - > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. - > Software Validation & Verification Test | ITEM | Proposed Device | Predicate Device 1<br>K161926 | Remark | |----------------|-----------------|-------------------------------|--------| | Product Code | GEX | GEX | SE | | Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SE | | Class | 2 | 2 | SE | Table 2 General Comparison {7}------------------------------------------------ | Intended Use | | SE | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----| | The Q-switched laser therapy device<br>is indicated for the treatment of:<br>benign cutaneous lesions, such as<br>Warts, Scars, Striae and Psoriasis;<br>benign pigmented lesions, such as<br>Lentigines, nevus, and birthmark; and<br>the removal of black or blue tattoos. | The ND YAG Q-switch Laser<br>Therapy Machine is indicated for the<br>treatment of:<br>benign cutaneous lesions, such as<br>Warts, Scars, Striae and Psoriasis;<br>benign pigmented lesions, such as<br>Lentigines, nevus, and birthmark;<br>and the removal of black or blue<br>tattoos. | | ### Table 3 Performance Comparison | ITEM | Proposed Device | Predicate Device 1<br>K161926 | Remark | |-----------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|------------| | Laser Medium | Nd:YAG | Nd:YAG | SE | | Wavelength | 1064nm 532nm | 1064 nm<br>532 nm | SE | | Output energy | 10-1000mJ for 1064nm<br>50-500mJ for 532nm | 100-1000mJ for 1064nm<br>50-500mJ for 532nm | SE | | Max. Energy Density | 31.8J/cm² for 1064nm<br>15.9 J/cm² for 532nm | 31.8J/cm² for 1064nm<br>15.9 J/cm² for 532nm | SE | | Spot Size | 2-10mm | 2-10mm | SE | | Pulse Width | 6ns | 5ns-8ns | Analysis 1 | | Repetition Rate | 1-10Hz | 1-10 Hz | SE | | Disinfection | Arms without focusing lenses and<br>probes can be disinfected with a 75%<br>medical grade alcohol wipe. | Disinfect the handpiece before and<br>after every treatment by 75%<br>medicinal alcohol | SE | | Laser Class | Class 4 | Class 4 | SE | | Cooling method for<br>treated skin area | Closed water circulation | N.A. | SE | {8}------------------------------------------------ | Aiming Beam | <6mW, semiconductor aiming light | Red Laser, <6mW | SE | |-------------|----------------------------------|-----------------|----| |-------------|----------------------------------|-----------------|----| #### Table 4 Safety Comparison | Item | Proposed Device | Predicate Device 1 | Remark | |------------------------------------------------|--------------------------------------------|--------------------------------------------|---------------------------| | | | K161926 | | | Patient Contact Materials and Biocompatibility | | | | | Patient Contact<br>Materials | Treatment Probe | Treatment Probe (Stainless Steel) | SE | | | Cytotoxicity | No Cytotoxicity | No Cytotoxicity | | Sensitization | No evidence of sensitization | No evidence of sensitization | SE | | Irritation | No evidence of irritation | No evidence of irritation | | | EMC, Electrical and Laser Safety | | | | | Electrical Safety | Comply with IEC 60601-1, IEC<br>60601-2-22 | Comply with IEC 60601-1, IEC<br>60601-2-22 | SE | | | EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | | Laser Safety | Comply with IEC 60601-2-22, IEC<br>60825 | Comply with IEC 60601-2-22, IEC<br>60825 | SE | #### Analysis 1: The proposed device has different Pulse Width from the predicate device. Pulse Width of proposed device is within the predicate device. And the proposed device has passed the IEC 60601-1 test, IEC 60601-1-2 test, IEC 60601-2-22 test, IEC 60825-1 test, the safety and performance of the product can be ensured. - 9. Clinical Test Conclusion No clinical study is included in this submission. - 10. Substantially Equivalent (SE) Conclusion Based on the nonclinical tests performed, the subject device is as effective, and performs as well as the legally marketed predicate device, ND YAG Q-switch Laser Therapy Machine cleared under K161926.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...