uMI Panorama

K232712 · Shanghai United Imaging Healthcare Co., Ltd. · KPS · Oct 3, 2023 · Radiology

Device Facts

Record IDK232712
Device NameuMI Panorama
ApplicantShanghai United Imaging Healthcare Co., Ltd.
Product CodeKPS · Radiology
Decision DateOct 3, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1200
Device ClassClass 2
AttributesAI/ML

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
PET image motion correctionDeep learning technology—Reduced respiratory motion effects and improved accuracy of SUV and lesion volume.——Clinical image evaluation: blind comparison of images with/without uExcel Focus.2 (American board-certified nuclear medicine physicians)
Parametric PET imaging——Provided quantitatively accurate parametric images.————

Indications for Use

The uMI Panorama is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images. This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging. This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

Device Story

uMI Panorama combines PET and 160-slice CT imaging; provides functional and anatomical data. PET uses LYSO scintillators; CT provides tomographic/attenuation data. Operated by qualified healthcare professionals in clinical settings. New features: uExcel Focus (OncoFocus) uses deep learning for motion correction/artifact reduction; uKinetics generates parametric images (direct/indirect). System includes patient table, power unit, host/reconstruction computers, and vital signal module. Output: fused PET/CT images for diagnosis, staging, and treatment planning. Benefits: improved SUV accuracy, reduced respiratory motion effects, and better PET-CT alignment. Healthcare providers use output to assess metabolic/physiological functions and anatomical findings.

Clinical Evidence

Bench testing and clinical image evaluation performed. Bench testing verified image performance (SUVmax, SUVmean, volume, AUC, Ki, CoV) for uExcel Focus and uKinetics. Clinical evaluation involved a blind comparison of images with/without uExcel Focus by two board-certified nuclear medicine physicians; results confirmed images were sufficient for clinical diagnosis, with improved PET-CT alignment and reduced attenuation correction artifacts.

Technological Characteristics

PET/CT system; PET uses LYSO scintillators; 160-slice CT. Axial FOV: 280mm or 350mm. Patient bore: 760mm. Max table load: 318kg. Connectivity: DICOM. Software: uExcel Focus (deep learning-based motion correction) and uKinetics (parametric imaging). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-44, IEC 60825-1, IEC 60601-1-6, IEC 62304, NEMA XR 25/28/29, NEMA NU 2-2018, ISO 10993, ISO 14971.

Indications for Use

Indicated for patients requiring diagnostic imaging for detection, localization, diagnosis, staging, treatment planning, and response evaluation of diseases, inflammation, infection, and disorders (e.g., oncology, cardiology, neurology). Includes low-dose CT lung cancer screening for early detection of lung nodules.

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 3, 2023 Shanghai United Imaging Healthcare Co., Ltd. Xin Gao RA Manager No. 2258 Chengbei Rd. Jiading District Shanghai. 201807 China Re: K232712 Trade/Device Name: uMI Panorama Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS, JAK Dated: September 5, 2023 Received: September 5, 2023 Dear Xin Gao: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality {2}------------------------------------------------ # Indications for Use Submission Number (if known) k232712 Device Name uMI Panorama Indications for Use (Describe) The uMI Panorama is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images. This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging. This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the alphanumeric string "K232712" at the top. Below that is the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING" in a bold, sans-serif font. To the right of the text is a stylized "U" shape, possibly representing the company's initial. # 510 (K) SUMMARY - 1. Date of Preparation September 5, 2023 #### Sponsor Identification 2. Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China Contact Person: Xin GAO Position: Regulatory Affair Manager Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com #### 3. Identification of Proposed Device Device Name: uMI Panorama Common Name: Positron Emission Tomography and Computed Tomography Systems Model(s): uMI Panorama 28, uMI Panorama 35 Regulatory Information Regulation Number: 21 CFR 892.1200, 21 CFR 892.1750 Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK Review Panel: Radiology #### Identification of Primary/Reference Device(s) 4. Predicate Device 510(k) Number: K223325 Device Name: uMI Panorama Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK Review Panel: Radiology Reference Device 510(k) Number: K173578 Device Name: Biograph mCT and MCT Flow PET/CT Scanners {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is dark blue with a white vertical line running through the center. The logo is simple and modern, and the use of bold text and a strong color palette makes it easily recognizable. Regulation Name: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK Review Panel: Radiology #### 5. Device Description: The proposed device uMI Panorama combines a 280 or 350 mm axial field of view (FOV) PET and 160-slice CT system to provide high quality functional and anatomical images, fast PET/CT imaging and better patient experience. The system includes PET system. CT system, patient table, power distribution unit, control and reconstruction system (host, monitor, and reconstruction computer, system software, reconstruction software), vital signal module and other accessories. The uMI Panorama has been previously cleared by FDA via K223325. The mainly modifications performed on the uMI Panorama (K223325) in this submission are due to the algorithm update of uExcel Focus (also named OncoFocus) and the addition of uKinetics. This time, uExcel Focus adds deep learning technology based on the original technology and could be performed as intended. uKinetics is a new function. which is used to generate both indirect and direct parametric images. ## 6. Indications for Use The uMI Panorama is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images. This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging. This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of two parts: the text "UNITED IMAGING" in a bold, sans-serif font, and a stylized letter "U" to the right. The letter "U" is formed by two thick vertical lines with a horizontal line connecting them at the top, creating a shape that resembles a shield or emblem. * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. ## 7. Comparison of Technological Characteristics with the Predicate Device uMI Panorama employs the same basic operating principles and fundamental technologies, and has the same indications for use as the predicate device. A comparison between the technological characteristics of proposed and predicate devices is provided as below. | ITEM | Proposed Device<br>uMI Panorama | Predicate Device<br>uMI Panorama (K223325) | |--------------------|---------------------------------|--------------------------------------------| | Patient bore size | 760mm | 760mm | | PET System | Scintillator material: LYSO | Scintillator material: LYSO | | | Number of detector rings: | Number of detector rings: | | | • 96 (uMI Panorama 28) | • 96 (uMI Panorama 28) | | | • 120 (uMI Panorama 35) | • 120 (uMI Panorama 35) | | | Axial FOV: | Axial FOV: | | | • 280mm (uMI Panorama 28) | • 280mm (uMI Panorama 28) | | | • 350mm (uMI Panorama 35) | • 350mm (uMI Panorama 35) | | CT System | uCT ATLAS Astound (K223028) | uCT ATLAS Astound (K223028) | | Maximum table load | 318kg | 318kg | | Advanced Feature | | | | uExcel Focus | Yes | Yes | | uKinetics | Yes | No | uMI Panorama's technological characteristics do not raise new safety and effectiveness concerns. #### 8. Performance Data The following performance data were provided in support of the substantial equivalence determination. ## Non-Clinical Testing Non-clinical testing including dosimetry and image performance tests were conducted for the uMI Panorama to verify that the proposed device met all design specifications as it is Substantially Equivalent (SE) to the predicate device. UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance: ## Electrical Safety and Electromagnetic Compatibility (EMC) {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal. - ANSI/AAMIES60601-1: 2005/ (R) > 2012+A1:2012+C1:2009/(R)2012+A2:2010/(R)2012)[IncludingAmendment2(2 021)]Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC60601-1-2:2014+A1:2020, Medical electrical equipment Part 1-2: General A requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests - A IEC 60601-1-3: 2008+AMD1:2013+A2:2021, Edition 2.2, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment. - A IEC 60601-2-44:2009+A1:2012+A2:2016 Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-rav equipment for computed tomography - A IEC 60825-1: 2014, Edition 3.0, Safety of laser products - Part 1: Equipment classification and requirements. - IEC 60601-1-6:2010+A1:2013+A2:2020, Edition 3.2, Medical electrical A equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. - A IEC 62304:2006+AMD1:2015 CSV Consolidated version, Medical device software - Software life cycle processes - A NEMA XR 25-2019, Computed Tomography Dose Check - A NEMA XR 28-2018, Supplemental Requirements For User Information And System Function Related To Dose In CT - A NEMA XR 29-2013, Standard Attributes on CT Equipment Related to Dose Optimization and Management - A IEC 61223-3-5 2004 Edition 1.0, Evaluation And Routine Testing In Medical Imaging Departments - Part 3-5: Acceptance Tests - Imaging Performance Of Computed Tomography X-ray Equipment [Including: Technical Corrigendum 1 (2006)] - A NEMA NU 2-2018, Performance Measurements of Positron Emission Tomographs # Software - NEMA PS 3.1-3.20(2016): Digital Imaging and Communications in Medicine A (DICOM) - A Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - A Content of Premarket Submissions for Management of Cybersecurity in Medical Devices # Biocompatibility - A ISO 10993-5: 2009, Edition 3.0, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. - A ISO 10993-10: 2010, Edition 3.0, Biological evaluation of medical devices - Part {7}------------------------------------------------ Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold, sans-serif font. The color of the logo is a dark teal. 10: Tests for irritation and skin sensitization. ### Other Standards and Guidance - > ISO 14971: 2019, Edition 3.0, Medical Devices - Application of risk management to medical devices - Code of Federal Regulations, Title 21. Part 820 Quality System Regulation A - A Code of Federal Regulations, Title 21, Subchapter J - Radiological Health - A Provision for Alternate Measure of the Computed Tomography Dose Index (CTDI) to Assure Compliance with the Dose Information Requirements of the Federal Performance Standard for Computed Tomography ### Performance Verification Engineering bench testing was performed on uMI Panorama with uExcel Focus and uMI Panorama with uKinetics to support substantial equivalence and the product performance claims. uExcel Focus The images were compared for: - A Volume relative to no motion correction (△Volume) - A Maximum standardized uptake value relative to no motion correction (ASUVmax) - A Mean standardized uptake value relative to no motion correction (△SUVmean). The performance evaluation showed that uExcel Focus could reduce respiratory motion effects and improve the accuracy of SUV and lesion volume in comparison with no motion correction. ### uKinetics The images were compared for: - A Bias of area under curve (AUC) of the estimated image-derived input function (IDIF). - A Bias of Ki and intercept in the hot spheres in the IO phantom. A Coefficient of variation (CoV) of the background region in the IQ phantom. Bench test showed that uKinetics can provide quantitatively accurate parametric images. ### Clinical Image Evaluation A blind comparison was conducted between uMI Panorama with/without uExcel Focus. Two American board-certified nuclear medicine physicians were invited to evaluate the images independently. Clinical evaluation shows that all images produced by uMI Panorama with uExcel Focus are sufficient for clinical diagnosis and uExcel Focus can reduce attenuation correction artifacts and improve PET-CT alignment. ## Summary {8}------------------------------------------------ Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font. To the right of the text is a stylized icon that appears to be a capital letter 'U' with a horizontal line through the middle. The features described in this premarket submission are supported with the results of the testing mentioned above, the uMI Panorama was found to have a safety and effectiveness profile that is similar to the predicate device. #### 9. Conclusions Based on the comparison and analysis above, the proposed device has similar intended use, performance, safety equivalence, and effectiveness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety, and effectiveness. And no issues are raised regarding to safety and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...