Disposable Bipolar Forceps are single use devices sold sterile and are intended for use in electrosurgery for coaqulation of tissue.
Device Story
Disposable Bipolar Forceps are single-use surgical instruments used for tissue hemostasis via bipolar electrocoagulation. The device consists of two forceps pieces with silver-plated brass tips, a holder, and an input wire. In operation, a clinician grasps tissue between the forceps tips; high-frequency current from an electrosurgical generator passes through the tissue to desiccate and coagulate it. The device is intended for use by physicians in surgical settings. The forceps provide a mechanical rebound action for ease of use. The device is supplied sterile and is designed to be compatible with standard bipolar electrosurgical equipment.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity), sterility validation, mechanical strength, and ex-vivo experimental studies on thermal effects.
Indicated for use in electrosurgery for coagulation of tissue in patients requiring hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Synergetics Disposable Spetzler-Mails Standard Bipolar Forceps (K121426)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
March 6, 2024
Jiangsu Hope Biomedical Science&Technology Co., Ltd. % Kitty Zhang Regulation Affair Staff Shanghai Jiushun Enterprise Management Technology Service Co 15 floor, 25 floor,Bao An Tower,No.800 Dongfang Road Shanghai, Shanghai 200122 China
Re: K232703
Trade/Device Name: Disposable Bipolar Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: January 31, 2024 Received: February 2, 2024
Dear Kitty Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Mark Digitally signed by Mark Trumbore -S Trumbore -S Date: 2024.03.06 13:48:41
> -05'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### K232703
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Submission Number (if known)
K232703
Device Name
Disposable Bipolar Forceps
Indications for Use (Describe)
Disposable Bipolar Forceps are single use devices sold sterile and are intended for use in electrosurgery for coaqulation of tissue.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
pe of Use (Select one or both, as applicable)
Prescription Use (Part 21 CER 801 Subn
- Over-The-Counter Use (21 CFR 801 Subpart C
{3}------------------------------------------------
## 510(k) Summary
# March 6, 2024
#### 1. Submitter Information
Company Name: Jiangsu Hope Biomedical Science&Technology Co.,Ltd.
Address: No.3 Building, Hi-tech Innovation Service Center, Jiangdu District,
Yangzhou City, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA.
Applicant Contact: Shaote Geng
Applicant Contact Telephone: +86-514-87185186
Applicant Contact Email: shaote.geng(@hoboat.com
Correspondent Name: Shanghai Jiushun Enterprise Management Technology Service Co., Ltd.
Correspondent Address: 15 floor,25 floor,Bao An Tower,No.800,Dongfang
Road, Shanghai, China. Shanghai Shanghai 200122 China
Correspondent Contact: Mrs. Kitty Zhang
Correspondent Contact Telephone: +86-21-50931939
Correspondent Contact Email: kitty-zhang@jisosh.com
Correspondent Contact Telephone: +86-21-50931939
#### Subject Device Information 2.
Common Name: Electrosurgical cutting and coagulation device and accessories
Trade Name: Disposable Bipolar Forceps
Model: ZKN-3CWZ
Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories
{4}------------------------------------------------
Review Panel: General and Plastic Surgery
Product Code: GEI
Regulation Number: 21 CFR 878.4400
Regulation Class: II
# 3. Predicate Device Information
Sponsor: Synergetics, Inc.
Common Name: Electrosurgical Instrumentation
Trade Name: Synergetics Disposable Spetzler-Mails Standard Bipolar Forceps
510(k) number: K121426
Classification Name: Electrosurgical Cutting and Coagulation Device and
Accessories
Review Panel: General and Plastic Surgery
Product Code: GEI
Regulation Number: 21 CFR 878.4400
Regulation Class: II
### 4. Device Description Summary
1) The mechanism of action
The mechanism of action of Disposable Bipolar Forceps are to hemostasis of tissue by bipolar electrocoagulation. The Disposable Bipolar Forceps can carry out electrocoagulation of tissues. The tissue is grasped between the forceps tips, each of which acts as an electrode, and current passes to desiccate and coagulate the tissue.
2) Mechanical structure
Disposable Bipolar Forceps consists of forceps pieces, forceps holder, wire. The tail end of the forceps pieces are connected with the forceps holder. The forceps holder is
{5}------------------------------------------------
provided with a high frequency input wire, and the wire can be connected with the high frequency surgical equipment through a plug. The outer surface of the forceps pieces are provided with an insulation layer.
Disposable Bipolar Forceps materials: forceps tips material: silver plated brass; forceps pieces material: brass; forceps pieces coating: epoxy resin; packaging: PETG tray, Tyvek lid.
3) Principle of operation and technological characteristics
The structure of Disposable Bipolar Forceps is double forceps pieces. Two fingers pinch the left and right forceps pieces, so that the forceps tips gently clamp the tissues, through the high frequency current to achieve electrocoagulation hemostasis. The rebound action is realized by the rebound force of the forceps pieces.
4) Operating procedures
Disposable Bipolar forceps are connected with the bipolar output of an electrosurgical generator. Disposable Bipolar forceps must only be used with bipolar coagulation current. The bipolar forceps must be operated with the following parameters: Rated Accessory Voltage is 500Vpk.
# 5. Indications for Use
Disposable Bipolar Forceps are single use devices sold sterile and are intended for use in electrosurgery for coagulation of tissue.
# 6. Substantial Equivalence Discussion
| Elements of<br>Comparison | Subject Device | Predicate Device |
|---------------------------|-----------------------------------------------------------|-------------------|
| Company | Jiangsu Hope Biomedical<br>Science&Technology<br>Co.,Ltd. | Synergetics, Inc. |
1)General Comparison
{6}------------------------------------------------
| Device Name | Disposable Bipolar Forceps | Synergetics Disposable<br>Spetzler-Mails Standard<br>Bipolar Forceps |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | ZKN-3CWZ | Disposable Spetzler-Malis<br>Standard Bipolar Forceps |
| Prescription/OTC | Prescription | Prescription |
| Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Code | GEI | GEI |
| Class | II | II |
| Indication for Use | Disposable Bipolar Forceps<br>are single use devices sold<br>sterile and are intended for<br>use in electrosurgery for<br>coagulation of tissue. | The Synergetics Disposable<br>Spetzler-Mails Standard<br>Bipolar Forceps are single use<br>devices sold sterile and are<br>intended for use in<br>electrosurgery for coagulation<br>of tissue. |
| Single Use | Yes | Yes |
| Device design | Bayonet Style | Bayonet Style |
| Length | 205mm | 180mm, 200mm, 230mm |
| Tip Size | 1.2mm | 0.5 mm, 1.0 mm, 1.5 mm |
| Color | Yellow | Yellow |
| Tips material | Silver plated | Silver plated |
| Electrical Safety<br>Testing | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 |
| Biocompatibility | Cytotoxicity<br>Sensitization<br>Intracutaneous Reactivity<br>Acute systemic toxicity test<br>Pyrogenicity test | Biocompatible Silver plated<br>Aluminum base with PVDF<br>insulation |
| Sterility | Yes | Yes |
| Sterilization<br>Method | Ethylene oxide | Ethylene oxide |
| Shelf Life | 3 years | 5years |
| Packaging | Rigid PETG Tray, Tyvek lid | Rigid PETG Tray, Tyvek lid |
{7}------------------------------------------------
# 2) Indication for Use Comparison
Indication for use of the subject device: Disposable Bipolar Forceps are single use devices sold sterile and are intended for use in electrosurgery for coagulation of tissue.
Indication for use of the Equivalence: The Synergetics Disposable Speztler-Malis Standard Forceps are single use devices sold sterile and are intended for use in electrosurgery for coagulation of tissue.
The indication for use of subject device and the Equivalence is same.
3) Technological Comparison
Based on available 510(k) information provided herein, Disposable Bipolar Forceps is similar to the predicate devices in terms of material, technology, design and performance specifications.
# 7. Non-Clinical Tests Summary
EMC, electrical safety, biocompatibility and sterility were evaluated according to recognized standards listed above.
To demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified, we follow the Premarket Notification (510(k) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff.; Ex-vivo experimental study on Thermal Effects, was conducted follow Premarket Notification (510(k) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff which refer to Exvivo experimental study on Thermal Effects of Disposable Bipolar Forceps; and Mechanical Strength.
{8}------------------------------------------------
# Conclusion:
The subject device Disposable Bipolar Forceps is similar by technological
specifications and intended use to the predicate device. Performance testing results
shows that differences do not affect the safety and effectiveness of the subject device.
Thus, the subject device is substantially equivalent to the predicate device for
requested indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.