Cove Strip, OsteoCove Strip

K232668 · SeaSpine Orthopedics Corporation · MQV · Sep 27, 2023 · Orthopedic

Device Facts

Record IDK232668
Device NameCove Strip, OsteoCove Strip
ApplicantSeaSpine Orthopedics Corporation
Product CodeMQV · Orthopedic
Decision DateSep 27, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Cove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cove Strip resorbs and is replaced with bone during the healing process. Cove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis.

Device Story

Cove Strip and OsteoCove Strip are osteoconductive bone void fillers; composed of 70:30 Beta-Tricalcium Phosphate (β-TCP) and Hydroxyapatite (HA) ceramic granules within a Type-1 collagen scaffold. Used by surgeons in the OR to fill non-structural bony voids or gaps in the posterolateral spine and pelvis. Implanted in a 1:1 ratio with autograft; acts as a bone graft extender. Resorbs over time and is replaced by new bone during the healing process. Provided sterile and non-pyrogenic for single use.

Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical bench testing, including design verification, packaging shipping/shelf-life validation, and sterilization validation (ISO 11135).

Technological Characteristics

Materials: 70:30 β-TCP/HA ceramic granules and Type-1 collagen. Form factor: Strip. Sterilization: Ethylene Oxide (ISO 11135) to SAL 10^-6. Packaging: Double blister tray.

Indications for Use

Indicated for patients requiring bone void filling in the posterolateral spine and pelvis due to surgically created defects or traumatic injury. Must be used with autograft as a bone graft extender. Not for use in defects intrinsic to bony structure stability.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 27, 2023 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. SeaSpine Orthopedics Corporation Cindy Toyama Regulatory Affairs Specialist 5770 Armada Drive Carlsbad, California 92008 Re: K232668 Trade/Device Name: Cove Strip, OsteoCove Strip Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: August 31, 2023 Received: September 1, 2023 Dear Cindy Toyama: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Sara S. Thompson -S For Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232668 Device Name Cove Strip #### Indications for Use (Describe) Cove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cove Strip resorbs and is replaced with bone during the healing process. Cove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------|------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232668 Device Name OsteoCove Strip #### Indications for Use (Describe) OsteoCove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. OsteoCove Strip resorbs and is replaced with bone during the healing process. OsteoCove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # 510(k) Summary # K232668 # Contact Details | Applicant Name: | SeaSpine Orthopedics Corporation | |-----------------|---------------------------------------------| | Address: | 5770 Armada Drive, Carlsbad CA | | Phone number: | (949) 855-7175 | | Fax number: | (760) 683-6874 | | Contact Person: | Cindy Toyama, Regulatory Affairs Specialist | | Date Prepared: | September 26, 2023 | Device Name | Device/Trade Name: | Cove Strip, OsteoCove Strip | |----------------------|-------------------------------------------------------| | Common Name: | Bone Void Filler | | Classification Name: | Filler, Bone Void, Calcium Compound (21 CFR 888.3045) | | Class: | II | | Product Code: | MQV | ## Legally Marketed Predicate Device | 510(k) Number | Product Code | Trade Name | Manufacturer | |--------------------------|--------------|-----------------------------|-----------------------------| | Primary Predicate Device | | | | | K230486 | MQV | Cove Strip, OsteoCove Strip | IsoTis OrthoBiologics, Inc. | # Device Description ## Cove Strip: Cove Strip is comprised of a ceramic granule consisting of a ratio of 70:30 Beta-Tricalcium Phosphate (β-TCP): Hydroxyapatite (HA) combined with highly purified Type-1 collagen. The collagen scaffold contains ceramic granules throughout, providing osteoconductive substrates that support new bone formation. Cove Strip combined with autograft in a 1:1 ratio, is intended {6}------------------------------------------------ to be placed in the posterolateral spine. The implant is provided sterile and non-pyrogenic for single use in a double blister tray configuration. ## OsteoCove Strip: OsteoCove Strip is comprised of a ceramic granule consisting of a ratio of 70:30 Beta-Tricalcium Phosphate (β-TCP): Hydroxyapatite (HA) combined with highly purified Type-1 collagen. The collagen scaffold contains ceramic granules throughout, providing osteoconductive substrates that support new bone formation. OsteoCove Strip combined with autograft in a 1:1 ratio, is intended to be placed in the posterolateral spine. The implant is provided sterile and nonpyrogenic for single use in a double blister tray configuration. # Indications for Use ## Cove Strip: Cove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cove Strip resorbs and is replaced with bone during the healing process. Cove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis. # OsteoCove Strip: OsteoCove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. OsteoCove Strip resorbs and is replaced with bone during the healing process. OsteoCove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis. # Summary of Technological Characteristics The subject device is similar to the cited predicate device in regard to components, device description, indications for use, device characteristics (i.e., design, materials, sterility, manufacturing, packaging, sterility, etc.), and performance. Compared to the predicate device, the subject device is longer; as such, the packaging was scaled accordingly. All materials, designs, and configurations are equivalent. # Summary of Non-Clinical Testing to Support Substantial Equivalence The subject device is similar to the predicate device in terms of materials, manufacturing process, and intended use. Non-clinical testing was not performed on the subject device, as the subject device does not introduce a new worst case, for the following: biocompatibility and in vivo (animal) safety and performance. Equivalency was established with the predicate device consisting of similar product sizing and packaging that allowed the adoption of the sterilization validation. Sterilization complies with ISO 11135, Sterilization of health care products-Ethylene Oxide-Requirements to ensure a sterility assurance level (SAL) of 10-6. The following verification and validation activities were evaluated for the subject device: - 1. Design Verification {7}------------------------------------------------ - 2. Ethylene Oxide (EO) Adoption - 3. Limulus Amebocyte Lysate (LAL) Kinetic Turbidimetric Validation - 4. Packaging Shipping Validation - 5. Packaging Shelf Life Validation - 6. Design (User) Validation - 7. Process Performance Qualification All activities were executed successfully to demonstrate substantial equivalence. # Clinical Testing Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data. # Conclusions The submitted data demonstrate that the subject device is substantially equivalent to the cited legally marketed predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...