Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl)
K232654 · Nephron Nitrile, LLC · LZA · Mar 20, 2024 · General Hospital
Device Facts
Record ID
K232654
Device Name
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl)
Applicant
Nephron Nitrile, LLC
Product Code
LZA · General Hospital
Decision Date
Mar 20, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, ambidextrous, powder-free nitrile examination gloves; black color; fingertip textured; beaded cuff. Used by healthcare professionals in clinical settings to prevent cross-contamination between patient and examiner. Provides barrier protection against biological contaminants and specific chemotherapy drugs and fentanyl. Input: physical barrier requirement. Output: physical protection. Benefit: reduces risk of contamination and exposure to hazardous drugs during medical examinations.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Material: Nitrile (NBR) latex. Form factor: Ambidextrous, beaded cuff, fingertip textured. Sizes: XS, S, M, L, XL, XXL. Standards: ASTM D6319-19 (physical properties/dimensions), ASTM D5151-19 (holes), ASTM D6124-06 (residual powder), ASTM D6978-05 (chemotherapy permeation). Biocompatibility: ISO 10993-23 (irritation), ISO 10993-10 (sensitization), ISO 10993-11 (systemic toxicity). Non-sterile, single-use.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for resistance to permeation by specific chemotherapy drugs and fentanyl.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) (K231349)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
March 20, 2024
Nephron Nitrile, LLC. Lou Kennedy Chief Executive Officer 4777 12th Street Extension West Columbia, South Carolina 29172
Re: K232654
Trade/Device Name: Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, ODO Dated: March 18, 2024 Received: March 18, 2024
Dear Lou Kennedy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image contains the text "Allan Guan -S". The text is written in a simple, sans-serif font and is horizontally aligned. The background of the image is plain white, with a faint watermark or design visible behind the text.
For Bifeng Qian, M.D., Ph.D. Assistant Director
{2}------------------------------------------------
DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232654
Device Name
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentany()
Indications for Use (Describe)
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:
Bleomycin Sulfate (15.0 mg/ml) Busulfan (6.0 mg/ml) Carboplatin (10.0 mg/ml) Cisplatin (1.0 mg/ml) Cyclophosphamide (20.0 mg/ml) Cytarabine (100.0 mg/ml) Dacarbazine (10.0 mg/ml) Daunorubicin HCl (5.0 mg/ml) Docetaxel (10.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Epirubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fludarabine (25.0 mg/ml) Fluorouracil (50.0 mg/ml) Gemcitabine (38.0 mg/ml) Idarubicin HCl (1.0 mg/ml) Ifosfamide (50.0 mg/ml) Irinotecan (20.0 mg/ml) Mechlorethamine HCl (1.0 mg/ml) Melphalan (5.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone HCl (2.0 mg/ml) Paclitaxel (6.0 mg/ml) Rituximab (10.0 mg/ml) Trisenox (1.0 mg/ml) Vincristine Sulfate (1.0 mg/ml)
The tested Opioid is: Fentanyl Citrate Injection (100 mcg/2 mL) Permeation: no breakthrough up to 240 minutes
The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 26.3 minutes Thiotepa (10.0 mg/ml) : 49.1 minutes Warning: Not for Use with: Carmustine, Thiotepa
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(K) SUMMARY K232654 AS REQUIRED BY: 21CFR§807.92
### A. APPLICANT INFORMATION
| 510(K) Owner's Name | Nephron Nitrile, LLC. |
|---------------------|------------------------------------------------------|
| 510(K) Number | K232654 |
| Address | 4777 12th Street Extension, West Columbia, SC 29172. |
| Phone | 844-937-3888 |
| Fax | 1-803-926-9853 |
| E-mail | NitrileRegulatory@nephronnitrile.com |
| Contact Person #1 | Edmundo M. Garcia y Garcia |
| Designation | Vice President of Quality Operations |
| Contact Number | 1-803-908-5156 |
| Contact Email | NitrileRegulatory@nephronnitrile.com |
| Contact Person #2 | Lou Kennedy |
| Designation | Chief Executive Officer |
| Contact Number | 1-803-569-3110 |
| Contact Email | lkennedy@nephronpharm.com |
| Date Submitted | 18 March 2024 |
| Date Prepared | Revised 20 March 2024 |
### B. DEVICE IDENTIFICATION
| Trade name of the device | Nephron Nitrile™, Nitrile Powder-Free Examination<br>Gloves Black (Tested For Use With Chemotherapy<br>Drugs and Fentanyl) |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Common or usual name | Nitrile Powder-Free Examination Gloves Black (Tested<br>For Use With Chemotherapy Drugs and Fentanyl) |
| Classification name | Patient Examination Glove, Specialty |
| Device Classification | Class-1 |
| Product Code | LZA, LZC, OPJ, QDO |
| Regulation Number | 21 CFR 880.6250 |
| Review Panel | General Hospital |
### C. PREDICATE DEVICE
| Predicate Device | Nephron Nitrile Powder-Free Nitrile Examination<br>Gloves (Tested For Use With Chemotherapy<br>Drugs and Fentanyl) |
|------------------|--------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K231349 |
| Regulatory Class | Class-1 |
| Product code | LZA, LZC, OPJ, QDO |
| Manufacturer | Nephron Nitrile, LLC. |
{6}------------------------------------------------
### 510(K) SUMMARY K232654 AS REQUIRED BY: 21CFR §807.92 D. DESCRIPTION OF THE DEVICE:
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) are Class I patient examination gloves, bearing the product codes LZA, LZC, OPJ, QDO (21 CFR 880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). The powder-free gloves are made from Nitrile (NBR) latex and are black in color. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff, and single use only.
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) include glove sizes: X-Small, Medium, Large, X-Large and XX-Large.
### E. DEVICE MODIFICATIONS
The modification(s) within this 510(k) propose to revise the color additive, from a blue to black pigment. In addition to the color change, this 510(k) also intends to modify the available size range, to include size small and extra-small. All other aspects, including Intended Use, Indications for Use, and technological characteristics of the existing device remain unchanged.
### F. INDICATION FOR USE OF THE DEVICE:
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
| The following chemotherapy drugs and concentration had NO breakthrough detected up to<br>240 minutes: | | | The tested<br>Opioid is: |
|-------------------------------------------------------------------------------------------------------|-----------------------------|--------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Bleomycin Sulfate (15.0 mg/ml) | Doxorubicin HCl (2.0 mg/ml) | Mechlorethamine HCl (1.0 mg/ml) | Fentanyl<br>Citrate<br>Injection<br>(100 mcg/2 mL)<br><br>Permeation: no<br>breakthrough<br>up to 240<br>minutes |
| Busulfan (6.0 mg/ml) | Epirubicin HCl (2.0 mg/ml) | Melphalan (5.0 mg/ml) | |
| Carboplatin (10.0 mg/ml) | Etoposide (20.0 mg/ml) | Methotrexate (25.0 mg/ml) | |
| Cisplatin (1.0 mg/ml) | Fludarabine (25.0 mg/ml) | Mitomycin C (0.5 mg/ml) | |
| Cyclophosphamide (20.0 mg/ml) | Fluorouracil (50.0 mg/ml) | Mitoxantrone HCl (2.0 mg/ml) | |
| Cytarabine (100.0 mg/ml) | Gemcitabine (38.0 mg/ml) | Paclitaxel (6.0 mg/ml) | |
| Dacarbazine (10.0 mg/ml) | Idarubicin HCl (1.0 mg/ml) | Rituximab (10.0 mg/ml) | |
| Daunorubicin HCl (5.0 mg/ml) | Ifosfamide (50.0 mg/ml) | Trisenox (1.0 mg/ml) | |
| Docetaxel (10.0 mg/ml) | Irinotecan (20.0 mg/ml) | Vincristine Sulfate (1.0 mg/ml) | |
| The following chemotherapy drugs have low permeation times: | | | |
| Carmustine (3.3 mg/ml) : 26.3 minutes | | Thiotepa (10.0 mg/ml) : 49.1 minutes | |
### Warning: Not for Use with: Carmustine, Thiotepa
{7}------------------------------------------------
## 510(K) SUMMARY K232654 AS REQUIRED BY: 21CFR§807.92
### G. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS BETWEEN THE SUBJECT DEVICE AND PREDICATE DEVICE
| | CHARACTERISTICS<br>STANDARDS | DEVICE PERFORMANCE | | | |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|------------|
| | | PREDICATE | SUBJECT | Comparison | |
| 510(K) Number | --- | K231349 | K232654 | | |
| Name of device | --- | Nephron Nitrile Powder-Free<br>Nitrile Examination Gloves<br>(Tested For Use With<br>Chemotherapy Drugs and<br>Fentanyl) | Nephron Nitrile™, Nitrile<br>Powder-Free Examination<br>Gloves Black (Tested For Use<br>With Chemotherapy Drugs<br>and Fentanyl) | Similar | |
| Product Code | --- | LZA, LZC, OPJ, QDO | LZA, LZC, OPJ, QDO | Identical | |
| Indication for use | --- | Nephron Nitrile Powder-Free<br>Nitrile Examination Gloves<br>(Tested For Use With<br>Chemotherapy Drugs and<br>Fentanyl) is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs and<br>fentanyl in accordance with<br>ASTM D6978-05 (2019)<br>Standard Practice for<br>Assessment of Resistance of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs. | Nephron Nitrile™, Nitrile<br>Powder-Free Examination<br>Gloves Black (Tested For Use<br>With Chemotherapy Drugs<br>and Fentanyl) is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs and<br>fentanyl in accordance with<br>ASTM D6978-05 (2019)<br>Standard Practice for<br>Assessment of Resistance of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs. | Identical | |
| Regulation Number | --- | 21 CFR 880.6250 | 21 CFR 880.6250 | Identical | |
| Material | --- | Nitrile | Nitrile | Identical | |
| Color | --- | Blue | Black | Different | |
| AS REQUIRED BY: 21CFR§807.92 | | | | | |
| | CHARACTERISTICS STANDARDS | DEVICE PERFORMANCE | | | |
| | | PREDICATE | SUBJECT | Comparison | |
| 510(K) Number | --- | K231349 | K232654 | | |
| | | | | safety or<br>performance. | |
| | | | | Similar | |
| Size | --- | M, L, XL, XXL | XS, S, M, L, XL, XXL | Adding size<br>XS and S | |
| Single Use | --- | Single-use | Single-use | Identical | |
| Sterile/Non-sterile | --- | Non-Sterile | Non-Sterile | Identical | |
| Rx Only or OTC | --- | OTC | OTC | Identical | |
| | | Minimum 230 mm<br>(sizes M - XXL) | Minimum 220 mm<br>(Sizes XS-S)<br>Minimum 230 mm<br>(sizes M - XXL) | Similar | |
| Dimensions - Length | ASTM<br>D6319-19 | | | Adding size<br>XS and S | |
| | | M: 95±10 mm<br>L: 110±10 mm<br>XL: 120±10 mm<br>XXL: 130±10 mm | XS: 70±10 mm<br>S: 80±10 mm<br>M: 95±10 mm<br>L: 110±10 mm<br>XL: 120±10 mm<br>XXL: 130±10 mm | Similar | |
| Dimensions - Width | ASTM<br>D6319-19 | | | Adding size<br>XS and S | |
| | | Before aging<br>14MPa, min | Before aging<br>14MPa, min | Identical | |
| Physical Properties-<br>Tensile Strength | ASTM<br>D6319-19 | | | | |
| | | After aging<br>14MPa, min | After aging<br>14MPa, min | Identical | |
| | | Before aging<br>500%, min | Before aging<br>500%, min | Identical | |
| Physical Properties-<br>Ultimate Elongation | ASTM<br>D6319-19 | | | | |
| | | After aging<br>400%, min | After aging<br>400%, min | Identical | |
| | | Palm: Minimum<br>0.05 mm<br>Finger: Minimum<br>0.05 mm | Palm: Minimum<br>0.05 mm<br>Finger: Minimum<br>0.05 mm | Identical | |
| Thickness | ASTM<br>D6319-19 | | | | |
| | ASTM D6319-<br>19 | ≤ 2 mg per glove | ≤ 2 mg per glove | Identical | |
| Powder Free Residue | | | | | |
| | | In accordance with<br>ASTM D 5151-19,<br>following ASTM D6319- 19,<br>G-I, AQL 2.5 | In accordance with<br>ASTM D 5151-19,<br>following ASTM D6319- 19,<br>G-I, AQL 2.5 | Identical | |
| Freedom from holes | ASTM D5151-<br>2019 | | | | |
| Chemotherapy Drugs<br>Tested with Minimum<br>Breakthrough Detection | ASTM<br>D6978-<br>05 (2019) | Bleomycin Sulfate<br>15 mg/ml (15,000 ppm)<br>>240 Minutes | Bleomycin Sulfate<br>15 mg/ml (15,000 ppm)<br>>240 Minutes | Identical | |
| AS REQUIRED BY 21CFR807.92 | | | | | |
| CHARACTERISTICS | | STANDARDS | DEVICE PERFORMANCE | | Comparison |
| | 510(K) Number | --- | PREDICATE | SUBJECT | |
| | | | K231349 | K232654 | |
| | Time | | Busulfan | Busulfan | |
| | | | 6 mg/ml (6,000 ppm)<br>>240 Minutes | 6 mg/ml (6,000 ppm)<br>>240 Minutes | Identical |
| | | | Carboplatin<br>10 mg/ml (10,000 ppm)<br>>240 Minutes | Carboplatin<br>10 mg/ml (10,000 ppm)<br>>240 Minutes | Identical |
| | | | Carmustine<br>3.3 mg/ml (3,300 ppm)<br>33.8 Minutes | Carmustine<br>3.3 mg/ml (3,300 ppm)<br>26.3 Minutes | Similar |
| | | | Cisplatin<br>1 mg/ml (1,000 ppm)<br>>240 Minutes | Cisplatin<br>1 mg/ml (1,000 ppm)<br>>240 Minutes | Identical |
| | | | Cyclophosphamide<br>20 mg/ml (20,000 ppm)<br>>240 Minutes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.