The HotDog Temperature Management System is intended to prevent or treat hypothermia and to provide warmth to patients. The System should be used in circumstances in which patients may not mantain a state of normothermia. The System can be used with adult patients. The System is intended primarily for use in hospitals and surgical centers including, without limitation, operating rooms, recovery rooms, emergency rooms, burn units and on other medical/surgical floors. The HotDog Return Electrode Mattress is intended to conduct monopolar electrosurgical energy from the target tissue of a patient back to one or two electrosurgical units (ESU) or generators in monopolar surgery. The HotDog Return Electrode Mattress is restricted to use with isolated monopolar electrosurgical generators. The HotDog Return Electrode Mattress is intended for use with adult patients only.
Device Story
HotDog Warming Mattress + Return Electrode is a thermal regulating system for adult patients. Device uses flexible semi-conductive polymer fabric in urethane shells; powered by low-voltage floating isolated DC current via controller/sensor feedback loop. Mattress functions as capacitive return electrode for monopolar electrosurgery. Operated in hospitals/surgical centers by clinical staff. Device connects to one or two electrosurgical units (ESU) via specialized cable. Output is controlled warming and safe return path for electrosurgical energy. Benefits include normothermia maintenance and integrated electrosurgical return functionality. Healthcare providers use device to manage patient temperature and facilitate surgery; output informs clinical management of patient thermal state.
Clinical Evidence
Bench testing only. Evaluated electrical safety (ANSI AAMI ES60601-1), EMC (IEC 60601-1-2), and performance standards (IEC 80601-2-35, IEC 60601-2-2). Testing confirmed temperature rise performance and patient leakage safety under worst-case dual-generator conditions.
Indicated for adult patients to prevent or treat hypothermia and provide warmth in clinical settings (OR, recovery, ER, burn units). Also indicated to conduct monopolar electrosurgical energy back to one or two isolated electrosurgical generators.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Mega 2000 Soft Dual Cord Patient Return Electrode Pad (K031285)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
April 18, 2024
Augustine Temperature Management, LLC Garrett Augustine VP R&D 6581 City West Parkway Eden Prairie, Minnesota 55344
Re: K232627
Trade/Device Name: HotDog Warming Mattress + Return Electrode Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ, GEI Dated: March 19, 2024 Received: March 19, 2024
Dear Garrett Augustine:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232627
Device Name HotDog Warming Mattress + Return Electrode
#### Indications for Use (Describe)
The HotDog Temperature Management System is intended to prevent or treat hypothermia and to provide warmth to patients. The System should be used in circumstances in which patients may not mantain a state of normothermia. The System can be used with adult patients.
The System is intended primarily for use in hospitals and surgical centers including, without limitation, operating rooms, recovery rooms, emergency rooms, burn units and on other medical/surgical floors.
The HotDog Return Electrode Mattress is intended to conduct monopolar electrosurgical energy from the target tissue of a patient back to one or two electrosurgical units (ESU) or generators in monopolar surgery. The HotDog Return Electrode Mattress is restricted to use with isolated monopolar electrosurgical generators. The HotDog Return Electrode Mattress is intended for use with adult patients only.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### SECTION 5 Premarket 510k Summary
| Submitter<br>Information: | Augustine Temperature Management, LLC<br>15305 Minnetonka Blvd<br>Minnetonka, MN 55345<br>952.465.3529 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Garrett Augustine, VP R&D |
| Date Prepared: | 03/19/2024 |
| Trade Name | HotDog Warming Mattress + Return Electrode |
| Primary Code | DWJ -- Thermal Regulating System (21 CFR §870.5900) |
| Reference Code | GEI -- Electrosurgical Cutting and Coagulation Device and<br>Accessories (21 CFR § 878.4400) |
| Common Name | Warming Mattress + Return Electrode |
| Predicate Device | HotDog Warming Mattress + Return Electrode (K230866) |
| Reference Device | Mega 2000 Soft Dual Cord Patient Return Electrode Pad (K031285) |
| Device Description | The HotDog Warming Mattress + Return Electrode (“Mattress") is<br>part of a thermal regulating system, indicated for controlling patient<br>temperature in adult patients.<br><br>Mattresses utilize a flexible semi-conductive polymer fabric which<br>warms the patient effectively within safe limits (controlled<br>temperature, low watt density, low thermal mass) from a controller and<br>sensor feedback loop. They are RF sealed in durable urethane shells<br>designed to eliminate uncleanable crevices. Via a blue cable, they are<br>powered with a low voltage floating isolated DC current, designed to<br>safely operate in the most demanding clinical settings.<br><br>Mattresses contain a green cable which enables the mattress to function<br>as a capacitive return electrode for electrosurgery. |
| Indication for Use | The HotDog Temperature Management System is intended to prevent<br>or treat hypothermia and to provide warmth to patients. The System<br>should be used in circumstances in which patients may not maintain a<br>state of normothermia. The System can be used with adult patients. |
| | The System is intended primarily for use in hospitals and surgical<br>centers including, without limitation, operating rooms, recovery<br>rooms, emergency rooms, burn units and on other medical/surgical<br>floors. |
| | The HotDog Return Electrode Mattress is intended to conduct<br>monopolar electrosurgical energy from the target tissue of a patient<br>back to one or two electrosurgical units (ESU) or generators in<br>monopolar surgery. The HotDog Return Electrode Mattress is<br>restricted to use with isolated monopolar electrosurgical generators.<br>The Hotdog Return Electrode Mattress is intended for use with adult<br>patients only. |
| Technological<br>Characteristics | The subject and predicate devices are identical in every respect but<br>one. While the predicate device can function as a return electrode for a<br>single ESU, the subject device can function as a return electrode for<br>one or two ESUs, using a “Y” version of the predicate's cable.<br>This technological characteristic – one or two ESU's -- is well<br>established in the FDA-cleared devices in the capacitive-style return<br>electrode category going back to May of 2003. Megadyne (K031285),<br>which functions in a substantially equivalent manner to the submission<br>device, indicates for one or two ESU's. |
| | The reference Megadyne device, from their 510(k) summary, supports<br>scientific methodology / standard reference values as it relates to the<br>expanded technology of this Special 510(k) submission. Their<br>summary defines temperature rise with two ESUs and impedance<br>performance as bench testing to establish substantial equivalence. For<br>Megadyne, in K031285 both of these values were found to be within<br>acceptable limits to the standard, and substantially equivalent to their |
| Performance Data | Bench testing was performed to demonstrate that the proposed<br>Warming Mattress + Return Electrode is compliant with FDA<br>recognized consensus standards for use with one ESU. These tests<br>were underwritten by Intertek, using pass/fail criteria. |
| | Electrical safety and electromagnetic compatibility (EMC)<br>The HotDog Warming Mattress + Return Electrode is designed and<br>verified to meet the following performance standards: |
| | ANSI AAMI ES60601-1, Medical Electrical Equipment - Part 1:<br>General Requirements for Basic Safety and Essential Performance,<br>A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 FDA recognition<br>#19-4 |
| | IEC 60601-1-2, Medical Electrical Equipment - Part 1-2: General<br>Requirements for Safety - Collateral standard: Electromagnetic<br>Compatibility - Requirements and Tests, edition: 4.0 FDA recognition<br># 19-8 |
| | Particular Standards |
| | The HotDog Warming Mattress + Return Electrode is designed and<br>verified to meet the following particular standards: |
| | IEC 80601-2-35, Particular requirements for the safety of blankets, pad<br>and mattresses intended for heating in medical use, edition: 2.1 FDA<br>recognition # 6-390 |
| | IEC 60601-2-2, Medical electrical equipment - Part 2-2: Particular<br>requirements for the basic safety and essential performance of high<br>frequency surgical equipment and high frequency surgical accessories<br>edition 6.0, FDA recognition # 6-389 |
| | Risk Management & Human Factors |
| | Risk management was applied throughout the development process of<br>the device. Risks related to use with two ESUs compared to one were<br>analyzed. Risk mitigations were verified. Residual risks were<br>assessed and an overall benefit-risk was established. The benefits of<br>the warming mattress + return electrode outweigh the risks. |
| SECTION 5 | |
| Premarket 510k Summary | |
| | ISO 14971, Medical Device - Application of Risk Analysis to<br>Medical Devices, third edition FDA recognition # 5-125 |
| | Human factors and usability engineering was applied to the<br>HotDog Warming Mattress + Return Electrode, on the design<br>side, and throughout risk management. |
| | IEC 62366-1, Medical Device - Application of usability<br>engineering to medical devices, Edition: 1.0 FDA<br>recognition # 5-114 |
| Further Bench Testing | Additional bench testing was completed to assess the<br>impact on performance or safety of use of two generators<br>vs one. Utilizing the reference device to support well<br>established methodology, the temperature rise performance<br>of the HotDog Warming Mattress + Return Electrode is<br>substantially equivalent to the predicate device when<br>utilizing two or one electrosurgical generators. Impedance<br>results are theoretically unaffected by one or two<br>generators, and therefore additional bench testing on<br>impedance was not necessary. Patient Leakage testing was<br>conducted under worst case conditions, demonstrating<br>safety in dual generator use. |
| Summary | Bench testing was performed to demonstrate that the<br>proposed HotDog Warming Mattress + Return Electrode is<br>compliant with FDA recognized consensus standards related<br>to this submission. This and further bench testing<br>demonstrate that the device is substantially equivalent to the<br>predicate device. No substantial new risks exist from the<br>ability to use one or two generators in electrosurgery with<br>this device. |
| Clinical Data | Not required |
| Conclusion | The HotDog Warming Mattress + Return Electrode was found to be<br>substantially equivalent to the predicate HotDog Warming Mattress +<br>Return Electrode. Reference device Mega 2000 Soft Dual Cord Patien<br>Return Electrode Pad presents clear methodology for addressing<br>differences in technology. The expanded technology of this<br>submission, when considering the reference device capabilities and<br>testing to FDA recognized consensus standards, demonstrate<br>substantial equivalence to the predicate. The overall intended use of<br>the device is identical to the predicate. In sum, the Warming Mattress<br>+ Return Electrode this submission is as safe, as effective, and |
{4}------------------------------------------------
# SECTION 5 SECTION 5
Premarket 510k Summary
{5}------------------------------------------------
# SECTION 5 SECTION 5
Premarket 510k Summary
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.