The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the Volta AF-Xplorer software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
Device Story
Volta AF-Xplorer is a machine/deep learning-based software system assisting electrophysiologists in identifying dispersed electrograms during atrial fibrillation or tachycardia. Input: intra-cardiac multipolar signals (analog or digital) from compatible catheters and EP recording systems (LabSystem Pro, MacLab CardioLab) or 3D mapping systems (EnSite X). Operation: system analyzes 1.5-second electrogram segments to detect spatiotemporal dispersion; computes cycle lengths. Output: real-time visual cues on a display; manual or automatic tagging of regions of interest on 3D atrial maps. Used in clinical electrophysiology labs; operated by physicians/technicians. Benefits: assists in identifying potential ablation targets; clinical significance for ablation outcomes not established.
Clinical Evidence
Bench testing only. Software verification and validation performed; processing performance demonstrated to be equivalent to predicate devices VX1 and VX1+ using established acceptance criteria. No clinical data provided.
Technological Characteristics
Non-sterile reusable system comprising a computing platform (Intel Core i7-7700, 32GB RAM, Debian-based Linux) and software. Connectivity: unidirectional analog (via Advantech PCI-1713U converter) or bidirectional digital (Ethernet) to EP recording/mapping systems. Sensing: intra-cardiac multipolar signals. Algorithm: machine/deep learning-based for dispersion detection. Output: real-time visual display and 3D map tagging.
Indications for Use
Indicated for operators performing 3D anatomical and electrical mapping of human atria during atrial fibrillation or atrial tachycardia to identify multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 27, 2023
Volta Medical % Kristin Duggan Partner Hogan Lovells US LPP 555 13th St. NW Washington, District of Columbia 20004
Re: K232616
Trade/Device Name: Volta AF-Xplorer Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: August 28, 2023 Received: August 28, 2023
Dear Kristin Duggan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Hetal B. Odobasic -S
for
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| Indications for Use Statement |
|-------------------------------|
|-------------------------------|
| Indications for Use Statement | |
|-----------------------------------------|--|
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
| Food and Drug Administration | |
| Indications for Use | |
| Form Approved: OMB No. 0910-0120 |
|----------------------------------|
| Expiration Date: 06/30/2023 |
| See PRA Statement on last page |
| 510(k) Number (if known) | K232616 |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Volta AF-Xplorer |
| Indications for Use (Describe) | The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the Volta AF-Xplorer software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations. |
| Type of Use (Select one or both, as applicable) | |
| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY | |
|------------------------------------------------------------------------------|--|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data
including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
#### K232616
#### 510(k) SUMMARY
#### VOLTA MEDICAL's Volta AF-Xplorer
#### Submitter
Volta Medical 65 Avenue Jules Cantini 13006 Marseille France
Phone: +33 7 68 02 54 99 Contact Person: Paola MILPIED
Date Prepared: August 28, 2023
Name of Device: Volta AF-Xplorer
Common or Usual Name: Cardiac Mapping System
Classification Name: Programmable Diagnostic Computer
Regulatory Class: 21 C.F.R § 870.1425
Product Code: DQK
#### Primary Predicate Device
Volta Medical, VX1+ (K223516)
#### Predicate Device
Volta Medical, VX1 (K201298)
#### Purpose of the Special 510(k) notice
The Volta AF-Xplorer is a modification to the VX1+ (K223516) device which incorporates the preprocessing from the VX1 (K201298).
#### Intended Use / Indications for Use
The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.
The clinical significance of utilizing the Volta AF-Xplorer software to help identify areas with intracardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
{5}------------------------------------------------
## Device Description
The Volta AF-Xplorer is a machine and deep learning based-algorithm designed to assist operators in the real-time manual or automation of 3D anatomical and electrical maps of the human heart for the presence of electrograms exhibiting spatio-temporal dispersion, i.e., dispersed electrograms.
The Volta AF-Xplorer device is a non-sterile reusable medical device, composed of a computing platform and a software application. Volta AF-Xplorer works with all existing 510(k)-cleared catheters that meet specific dimension requirements and with one of the three specific data acquisition systems:
- two compatible EP recording systems (identical to VX1 and VX1+ (Volta Medical (K201298, । K223516)): the LabSystem Pro EP Recording System (Boston Scientific) (K141185) or the MacLab CardioLab EP Recording System (General Electric) (K130626),
- a 3D mapping system (new compared to VX1 and identical to VX1+): EnSite X 3D mapping system -(Abbott) (K221213).
A connection cable is used to connect the corresponding data acquisition system to the Volta AF-Xplorer system, depending on the type of communication used:
- -Unidirectional analog communication with the EP recording systems via a custom-made cable (two diferent variants: DSUB, Octopus) and an Advantech PCI-1713U analog-to-digital converter, which acquires analog data, digitizes it, and transmits the digital signals to the computer that hosts the Volta AF-Xplorer software.
- -Bidirectional digital communication with the EnSite 3D mapping system via an ethernet cable (four different lengths: 20.10. 5 or 2m) which transmits the digital signals directly to the computer.
The computer and its attached display are located outside the sterile operating room area. The Volta AF-Xplorer software analyzes the patient's electrograms to cue operators in real-time to intra-cardiac electrograms of interest for atrial regions harboring dispersed electrograms as well as a cycle length estimation from electrograms recorded with the mapping and the coronary sinus catheters. The results of the analysis are graphically presented on the attached computer display and/or on a secondary medical screen or on an operating room widescreen. The identified regions of interest are either manually (all configurations) or automatically (only available in digital bidirectional communication with the EnSite X 3D mapping system) tagged in the corresponding 3D mapping system.
### Technological Characteristics
The Volta AF-Xplorer device is a modified device manufactured by Volta Medical, based on the same concept (identification of electrograms dispersion) than predicates VX1+ and VX1.
Volta AF-Xplorer, VX1+ and VX1 are software programs that work with standard electrophysiology catheters to aid in mapping the heart. All three devices aid operators by assisting in annotating complex electrical maps of the heart, and process and output information via a computer and display that are operated by use of a keyboard / mouse. Volta AF-Xplorer, VX1+ and VX1 have the same input (intra-cardiac multipolar signals) and the same output (associated dispersion), with the addition of the 3D position of the corresponding electrodes available in Volta AF-Xplorer and VX1+.
{6}------------------------------------------------
Volta AF-Xplorer, VX1+ and VX1 support Electrophysiologists in the manual annotation of dispersed areas using a unidirectional analog communication. Both Volta AF-Xplorer and VX1+ have the ability to connect to a specific 3D mapping system through a bidirectional digital communication, which enables the operator to use the automatic tagging function.
The Volta AF-Xplorer indications for use are identical to the VX1+ and the VX1. The Volta AF-Xplorer displays an analysis of dispersed electrograms, just as VX1 and therefore, the intended use of the Volta AF-Xplorer and the VX1+ and the VX1 are identical.
{7}------------------------------------------------
| | Volta Medical<br>Volta AF-Xplorer<br>(subject) | Volta Medical<br>VX1+<br>(K223516) | Volta Medical<br>VX1<br>(K201298) |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 C.F.R. § 870.1425 | 21 C.F.R. § 870.1425 | 21 C.F.R. § 870.1425 |
| Classification Name | Programmable Diagnostic Computer | Programmable Diagnostic Computer | Programmable Diagnostic Computer |
| Product Code | DQK | DQK | DQK |
| | The Volta AF-Xplorer assists operators<br>in the real-time manual or automatic<br>annotation of 3D anatomical and<br>electrical maps of human atria for the<br>presence of multipolar intra-cardiac<br>atrial electrograms exhibiting<br>spatiotemporal dispersion during atrial<br>fibrillation or atrial tachycardia. | The VX1+ assists operators in the<br>real-time manual or automatic<br>annotation of 3D anatomical and<br>electrical maps of human atria for<br>the presence of multipolar intra-<br>cardiac atrial electrograms<br>exhibiting spatiotemporal dispersion<br>during atrial fibrillation or atrial<br>tachycardia. | The VX1 assists operators in the real-time<br>manual annotation of 3D anatomical and<br>electrical maps of human atria for the<br>presence of multipolar intra-cardiac atrial<br>electrograms exhibiting spatiotemporal<br>dispersion during atrial fibrillation or atrial<br>tachycardia. |
| Indications for Use | The clinical significance of utilizing the<br>Volta AF-Xplorer software to help<br>identify areas with intra-cardiac atrial<br>electrograms exhibiting spatiotemporal<br>dispersion for catheter ablation of atrial<br>arrhythmias, such as atrial fibrillation,<br>has not been established by clinical<br>investigations. | The clinical significance of utilizing<br>the VX1+ software to help identify<br>areas with intra-cardiac atrial<br>electrograms exhibiting<br>spatiotemporal dispersion for<br>catheter ablation of atrial<br>arrhythmias, such as atrial<br>fibrillation, has not been established<br>by clinical investigations. | The clinical significance of utilizing the VX1<br>software to help identify areas with intra-<br>cardiac atrial electrograms exhibiting<br>spatiotemporal dispersion for catheter<br>ablation of atrial arrhythmias, such as atria<br>fibrillation, has not been established by<br>clinical investigations. |
| System Type | Signal processing based atrial<br>mapping system | Signal processing based atrial<br>mapping system | Signal processing based atrial mapping<br>system |
| Primary Feature | Displays and analyzes electrical maps<br>such as intra-cardiac electrograms in<br>real-time using machine learning and<br>signal processing techniques | Displays and analyzes electrical<br>maps such as intra-cardiac<br>electrograms in real-time using<br>machine learning and signal<br>processing techniques | Displays and analyzes electrical maps<br>such as intra-cardiac electrograms in real-<br>time using machine learning and signal<br>processing techniques |
| | Volta Medical<br>Volta AF-Xplorer<br>(subject) | Volta Medical<br>VX1+<br>(K223516) | Volta Medical<br>VX1<br>(K201298) |
| 3D Location<br>Technology | Electroanatomic location is performed<br>by another commercially available<br>navigation system.<br>In bidirectional digital communication,<br>3D Location is shared by the 3D<br>Mapping System with Volta AF-<br>Xplorer. | Electroanatomic location is<br>performed by another commercially<br>available navigation system.<br>In bidirectional digital<br>communication, 3D Location is<br>shared by the 3D Mapping System<br>with VX1+. | Electroanatomic location is performed by<br>another commercially available navigation<br>system. |
| Compatible<br>Acquisition<br>Systems | LabSystem Pro EP Recording<br>System (Boston Scientific) CardioLab EP Recording System<br>(GE) EnSite X 3D Mapping System<br>(Abbott) | LabSystem Pro EP Recording<br>System (Boston Scientific) CardioLab EP Recording<br>System (GE) EnSite X 3D Mapping System<br>(Abbott) | LabSystem Pro EP Recording<br>System (Boston Scientific) CardioLab EP Recording System (GE) |
| Compatible<br>Catheters | Any compatible mapping and ablation<br>catheter | Any compatible mapping and<br>ablation catheter | Any compatible mapping and ablation<br>catheter |
| Display(s) | Color monitor | Color monitor | Color monitor |
| Multi-Display<br>Support | Yes, duplicate display on a secondary<br>medical screen or on an operating<br>room widescreen | Yes, duplicate display on a<br>secondary medical screen or on an<br>operating room widescreen | Yes, duplicate display on a secondary<br>medical screen or on an operating room<br>widescreen |
| Control | Standard keyboard / mouse | Standard keyboard / mouse | Standard keyboard / mouse |
| Display Timing | Real-time | Real-time | Real-time |
| Inputs Required | Analog or digital Intra-cardiac signals<br>In digital mode, 3D locations of<br>corresponding electrodes bipoles | Analog or digital Intra-cardiac<br>signals<br>In digital mode, 3D locations of<br>corresponding electrodes bipoles | Analog Intra-cardiac signals |
| Output | Presence or absence of electrogram<br>dispersion at each electrode bipole<br>under consideration<br>In digital mode, 3D locations of<br>corresponding electrodes bipoles | Presence or absence of<br>electrogram dispersion at each<br>electrode bipole under<br>consideration | Presence or absence of electrogram<br>dispersion at each electrode bipole under<br>consideration |
| | Volta Medical<br>Volta AF-Xplorer<br>(subject) | Volta Medical<br>VX1+<br>(K223516) | Volta Medical<br>VX1<br>(K201298) |
| | Computed values of mapping and<br>reference cycle length | In digital mode, 3D locations of<br>corresponding electrodes bipoles<br><br>Computed values of mapping and<br>reference cycle length | Computed values of mapping and<br>reference cycle length |
| Duration of<br>Electrogram<br>Recordings | 1.5 Seconds | 1.5 Seconds | 1.5 Seconds |
| Ouput Display | The system generates color coded<br>symbol(s) that indicates to the operator<br>that the area under investigation is one<br>exhibiting dispersion<br>In bidirectional digital communication,<br>validated dispersion area can also be<br>automatically displayed in the 3D<br>mapping system as tags in the 3D<br>atrial shell | The system generates color coded<br>symbol(s) that indicates to the<br>operator that the area under<br>investigation is one exhibiting<br>dispersion<br>In bidirectional digital<br>communication, validated<br>dispersion area can also be<br>automatically displayed in the 3D<br>mapping system as tags in the 3D<br>atrial shell | The system generates color coded<br>symbol(s) that indicates to the operator<br>that the area under investigation is one<br>exhibiting dispersion |
| Signal Information<br>Displayed | Acquired patient signals, including<br>body surface ECG and intra-cardiac<br>EGMs. | Acquired patient signals, including<br>body surface ECG and intra-cardiac<br>EGMs. | Acquired patient signals, including body<br>surface ECG and intra-cardiac EGMs. |
| Computing Platform | Computer with Intel Core i7-7700 CPU<br>(8MB Cache, up to 4.20 GHz, RAM 32<br>GB),<br>with integrated analog/digital converter<br>PCI card and TPM (Trusted Platform<br>Module)<br>Debian-based Linux OS | Computer with Intel Core i7-7700<br>CPU (8MB Cache, up to 4.20 GHz,<br>RAM 32 GB),<br>with integrated analog/digital<br>converter PCI card and TPM<br>(Trusted Platform Module)<br>Debian-based Linux OS | Computer with Intel Core i5-6500 CPU<br>(6MB Cache, up to 3.60 GHz, RAM 32<br>GB),<br>with integrated analog/digital converter PC<br>card<br>Windows 10 or higher OS |
| | Volta Medical<br>Volta AF-Xplorer<br>(subject) | Volta Medical<br>VX1+<br>(K223516) | Volta Medical<br>VX1<br>(K201298) |
| Hardware Design<br>and Materials | Computing platform, proprietary<br>software algorithm, monitor,<br>mouse/keyboard, custom-made analog<br>connection cable, ethernet cable,<br>acquisition system | Computing platform, proprietary<br>software algorithm, monitor,<br>mouse/keyboard, custom-made<br>analog connection cable, ethernet<br>cable, acquisition system | Computing platform, proprietary software<br>algorithm, monitor, mouse/keyboard,<br>custom-made analog connection cable,<br>acquisition system |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
The hardware and software differences to the subject device do not introduce new questions of safety or effectiveness.
# Performance Data
Software design verfication was performed on the subject Volta AF-Xplorer that included the same processing as cleared in VX1 and the same post-processing specifications as cleared in VX1+. The testing and acceptance criteria are the same as those in the predicate VX1+ and VX1 devices. Bench testing was previously conducted in the VX1+ (K201298) to demonstrate rigorous software verification testing including unitary testing of the main algorithm modules. The processing was evaluated and demonstrated equivalent performance with the acceptance criteria from the VX1+ and VX1 studies.
# Conclusions
ilə qalınmışdır. Bu mərkəzi vahidində bir mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının m
The Volta AF-Xplorer is as safe and effective as the VX1. The Volta AF-Xplorer has the same intended uses and same indications, and substantially similar technological characteristics, and principles of operation as its predicate devices VX1+ and VX1. The modification to the same pre-processing specification as cleared in VX1 and the same post-processing specifications as cleared in VX1+ does not alter the intended use of the device as an electrophysiological evaluation tool and effectiveness when used as labeled. Performance data, as described above, demonstrate that the Volta AF-Xplorer device is as she VX1+ and the VX1. Thus, the Volta AF-Xplorer device is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.