4CIS SARA SPINE SYSTEM, 4CIS VERTU SPINE SYSTEM, 4CIS WILL PEDICLE SCREW SYSTEM, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM
K232607 · Solco Biomedical Copany India Private Limited · NKB · Feb 27, 2024 · Orthopedic
Device Facts
Record ID
K232607
Device Name
4CIS SARA SPINE SYSTEM, 4CIS VERTU SPINE SYSTEM, 4CIS WILL PEDICLE SCREW SYSTEM, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM
Applicant
Solco Biomedical Copany India Private Limited
Product Code
NKB · Orthopedic
Decision Date
Feb 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are the pedicle screw systems indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-Sl vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis ). aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to used for Minimal Invasive Surgery (MIS).
Device Story
Top-loading posterior spinal fixation system; consists of pedicle screws, rods, nuts, and transverse links; used by surgeons in clinical settings to stabilize spinal segments and promote fusion; implants inserted into vertebral bodies via posterior approach; provides rigid fixation; components locked into configurations matching patient pathology; supports both open and minimally invasive surgery (MIS); benefits patients by immobilizing unstable segments to facilitate bone graft fusion; removed after solid fusion achieved.
Clinical Evidence
Bench testing only. Mechanical testing (static/dynamic compression bending, static tension bending, static torsion) conducted per ASTM F1717 on worst-case constructs. Results met performance requirements against predicates. No clinical data provided.
Technological Characteristics
Top-loading posterior spinal fixation system. Materials: Titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136. Components: pedicle screws, rods, nuts, transverse links. Supplied non-sterile for single use. Mechanical testing performed per ASTM F1717.
Indications for Use
Indicated for skeletally mature patients requiring spinal fusion for severe Spondylolisthesis (Grade 3-4, L5-S1) or acute/chronic instabilities/deformities (thoracic, lumbar, sacral) including degenerative Spondylolisthesis, fracture, dislocation, scoliosis, spinal tumor, and pseudarthrosis. aBle Xt and 4CIS WILL systems are indicated for MIS.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
February 27, 2024
Solco Biomedical Copany India Private Limited Darshak Shah Director Survey No.1540, Beside Torrent Pharma, Village-Rajpur Ahmedabad Mehsana Highway Kadi, Gujarat/ Mehasana 382715 India
Re: K232607
Trade/Device Name: 4CIS SARA SPINE SYSTEM, 4CIS VERTU SPINE SYSTEM, 4CIS WILL PEDICLE SCREW SYSTEM, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: January 29, 2024 Received: January 29, 2024
Dear Darshak Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Colin .. O'neill -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232607
Device Name 4CIS SARA Spine System, 4CIS VERTU Spine System 4CIS WILL PEDICLE SCREW SYSTEM aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM
Indications for Use (Describe)
The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are the pedicle screw systems indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-Sl vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis ).
aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW SYSTEM are intended to used for Minimal Invasive Surgery (MIS).
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| <div><span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
PAGE NO 1 of 4
Image /page/4/Picture/1 description: The image shows the word "SOLCO." in a bold, sans-serif font. The letters are a dark blue color, and there is a red dot to the right of the "O". The logo is simple and modern.
NO : 1.0 TITLE: 510(k) Summary
## "510(K) SUMMARY"
This summary of 510(k) substantial equivalence information is being submitted in accordance with requirement of 21 CFR 807.92.
| | | SOLCO Biomedical Company India Private Limited | | |
|-----------------------------|-------------------------------------------------------|---------------------------------------------------|-----------------------|--|
| | Survey No.1540, Beside Torrent Pharma, Village-Rajpur | | | |
| Submitter | Ahmedabad Mehsana Highway, Ta-Kadi, Dist-Mehsana | | | |
| | Gujarat- 382715, INDIA. | | | |
| | Tel : +91 851112597, Email: darshak@solco.co.in | | | |
| | Darshak Shah- Director | | | |
| Contact Person | SOLCO Biomedical Company India Private Limited | | | |
| | Phone : +91 98252 06091, Email: darshak@solco.co.in | | | |
| Submission Date | 24 FEB 2024. | | | |
| | | 4CIS SARA Spine System | | |
| Trade /<br>Proprietary name | 4CIS VERTU Spine System | | | |
| | 4CIS WILL PEDICLE SCREW SYSTEM | | | |
| | aBle SPINAL FIXATION SYSTEM | | | |
| | aBle Xt SPINAL FIXATION SYSTEM | | | |
| Classification | | Thoracolumbosacral Pedicle screw system | | |
| Name | NKB | | | |
| Classification | Class II | | | |
| Code | 888.3070 | | | |
| Regulatory Class | | | | |
| Regulation | | | | |
| Number | | | | |
| | 510K | Trade or Property or | Manufacturer | |
| | Number | model Name | SOLCO Biomedical | |
| Primary<br>Predicate Device | K202498 | 4CIS SARA Spine System<br>4CIS VERTU Spine System | | |
| | | | Company India Private | |
| | | | Limited | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "SOLCO." in a bold, sans-serif font. The letters are a dark blue color. There is a red dot to the right of the "O".
DOC NO: K202498/Summary NO : 1.0
PAGE NO 2 of 4
TITLE: 510(k) Summary
| | The Spinal Fixation System is a top-loading posterior spinal fixation system |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of<br>Device | which consists of pedicle screws, rods, and nuts, transverse (cross) link |
| | and associated instruments. Rigid fixation is provided by pedicle screws |
| | inserted into the vertebral body through pedicle of the lumbar spine via |
| | posterior approach. This system will allow surgeons to build a spinal |
| | implant construct to stabilize and promote spinal fusion through open |
| | surgery. Implant components can be rigidly locked into a variety of different |
| | configurations to suit the individual pathology and anatomical conditions of |
| | the mature patient. The implant components are supplied non-sterile single |
| | use and are fabricated from titanium alloy (Ti-6A1-4V ELI) and Cobalt |
| | Chromium ally that conforms to ASTM F136 and ASTM F1537 respectively. |
| | Also, Specialized instruments are available for the application |
| | and removal of the Spinal Fixation System. |
| Indication for Use | The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL |
| | FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL |
| | PEDICLE SCREW SYSTEM are the pedicle screw systems indicated for |
| | the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-SI |
| | vertebra in skeletally mature patients receiving fusion by autogenous bone |
| | graft having implants attached to the lumbar and sacral spine (L3 to |
| | sacrum) with removal of the implants after the attainment of a solid fusion. |
| | The 4CIS SARA Spine System , 4CIS VERTU Spine System, aBle SPINAL |
| | FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL |
| | PEDICLE SCREW SYSTEM are intended to provide immobilization and |
| | stabilization of spinal segments in skeletally mature patients as an adjunct |
| | to fusion in the treatment of the following acute and chronic instabilities or |
| TITLE: 510(k) Summary | |
| | deformities of the thoracic, lumbar and sacral spine: degenerative<br>Spondylolisthesis with objective evidence of neurological impairment,<br>fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous<br>fusion (pseudarthrosis ).<br>aBle Xt SPINAL FIXATION SYSTEM and 4CIS WILL PEDICLE SCREW<br>SYSTEM are intended to used for Minimal Invasive Surgery (MIS). |
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate Devices | The 4CIS WILL PEDICLE SCREW SYSTEM, aBle Xt SPINAL FIXATION<br>SYSTEM & aBle SPINAL FIXATION SYSTEM have similar indications as<br>compare to cleared 4CIS SARA Spine System and 4CIS VERTU Spine<br>System. The system is composed of the same material as the predicate-<br>K202498 devices conforming to recognized industry standards for<br>permanent implants (titanium alloy (Ti-6A1-4V ELI) that conforms to ASTM<br>F136) and surgical orthopedic instruments. aBle SPINAL FIXATION<br>SYSTEM is modified version of 4CIS SARA Spine System and 4CIS<br>VERTU Spine System which does not raise any concern of safety and<br>performance. 4CIS WILL PEDICLE SCREW SYSTEM is intended to used<br>for Minimal Invasive Surgery (MIS). Also they are provided non-sterile for<br>single use only. |
| Performance Data | Mechanical testing (static and dynamic compression bending, static<br>tension bending, static torsion) is conducted in accordance with ASTM<br>F1717 on the worst case construct. Static Compression Bending,<br>Dynamic Compression Bending, Static Torsion, and Static Tension tests<br>results meet performance requirement against predicates. |
| Biocompatibility | The subject implants are permanent implants (> 30 days) and are classified<br>as body contacting devices according to FDA's Draft Guidance for Industry |
| | and FDA Staff "Use of International Standard ISO-10993, Biological |
| | Evaluation of Medical Devices Part 1: Evaluation and Testing.” The subject |
| | implants are manufactured from identical materials as the predicate |
| | devices, in accordance with the ASTM F136 – Standard Specification for |
| | Wrought Titanium – 6Aluminum – 4Vanadium ELI (Extra-Low-Interstitial) |
| | Alloy for Surgical Implants The materials used for manufacturing the |
| | subject device have a long history of safe and effective use identical |
| | to predicate devices and biocompatibility testing is not required. |
| Conclusion | The overall technology characteristics, material of construction , design |
| | characteristics and performance data lead to the conclusion that our Spine |
| | Systems (4CIS SARA Spine System, 4CIS VERTU Spine System, aBle |
| | SPINAL FIXATION SYSTEM, aBle Xt SPINAL FIXATION SYSTEM and |
| | 4CIS WILL PEDICLE SCREW SYSTEM) are substantially equivalent to our |
| | FDA previously cleared devices of K202498 for intended use, material |
| | composition, principles of operation, and overall design. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for SOLCO. The text "SOLCO" is written in a bold, dark blue font. There is a red dot to the right of the text.
DOC NO: K202498/Summary NO : 1.0
TITLE: 510(k) Summary
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "SOLCO." in a bold, sans-serif font. The letters are a dark blue color. There is a red dot to the right of the word.
DOC NO: K202498/Summary NO : 1.0
PAGE NO 4 of 4
TITLE: 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.