Horos MD™ is a software device intended for viewing images acquired from computed tomography (CT), computed radiography (CR), magnetic resonance (MR), ultrasound (US) and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be stored, communicated, processed, and displayed within the system. It is intended for use as a diagnostic and review tool by trained healthcare professionals. This device is not to be used for mammography. It is the User's responsibility to operate the device in accordance with the software and hardware requirements listed in the instructions for use, in particular ensuring that monitor (display) quality, ambient light conditions, and image compression ratios are consistent with the clinical application.
Device Story
Horos MD™ is a diagnostic image viewing and navigation software for radiologists and clinicians. It ingests DICOM-compliant medical images (CT, MR, CR, US) from PACS or local storage. The software runs on macOS, utilizing Apple's METAL framework for 3D graphics acceleration. It provides 2D/3D visualization tools, including windowing, MPR, volume rendering, and measurement functions (distance, angle, area). The device does not operate imaging hardware or generate original images. Healthcare providers use the interface to review scans, perform measurements, and compare studies to support clinical decision-making. It facilitates diagnostic workflows by enabling interactive navigation and analysis of radiological data. The device is intended for use in clinical environments by trained professionals.
Clinical Evidence
Bench testing only. Validation of measurement accuracy (distance, area, angle, point) performed using 75 Digital Reference Objects representing clinical ranges (1-190 mm). Results showed high accuracy (99.19% to 100%) for all functions. Repeatability and reproducibility were assessed. No clinical data required.
Technological Characteristics
Software-based medical image management system running on macOS. Utilizes Apple METAL framework for 3D graphics. Compliant with DICOM (NEMA PS3.1 2023b) and JPEG (IEC/ISO 10918-1) standards. Features include 2D/3D visualization, MPR, volume rendering, and measurement tools. Cybersecurity controls implemented per FDA guidance.
Indications for Use
Indicated for viewing DICOM-compliant medical images (CT, CR, MR, US) by trained healthcare professionals for diagnostic and review purposes. Contraindicated for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
iCat Solutions Ltd % Bhoomika Joyappa Senior Regulatory Consultant Medical Device Academy, Inc. 345 Lincoln Hill Rd Shrewsbury, Vermont 05738
January 3, 2024
Re: K232589
Trade/Device Name: Horos MD™M Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: December 13, 2023 Received: December 13, 2023
Dear Bhoomika Joyappa:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
Device Name
Horos MD TM
Indications for Use (Describe)
Horos MD™ is a software device intended for viewing images acquired from computed tomography (CT), computed radiography (CR), magnetic resonance (MR), ultrasound (US) and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be stored, communicated, processed, and displayed within the system. It is intended for use as a diagnostic and review tool by trained healthcare professionals.
This device is not to be used for mammography.
It is the User's responsibility to operate the device in accordance with the software and hardware requirements listed in the instructions for use, in particular ensuring that monitor (display) quality, ambient light conditions, and image compression ratios are consistent with the clinical application.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | | |
|--------------------|------------------|---------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 |
### 1. DESCRIPTION
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
# 2. SUBMITTER
| Company Name | iCat Solutions Ltd |
|----------------|---------------------------------------------------|
| Address | 51-59 Rose Lane, NR1 1BY, Norwich, United Kingdom |
| tel | +44(0)7517456306 |
| Contact Person | Georgios Michalopoulos |
## 3. DATE PREPARED
2023/07/21
# 4. DEVICE
| Device Trade Name | Horos MD™ |
|-----------------------------|------------------------------------------------|
| Classification Name | Medical Image Management and Processing System |
| Regulation | 21 CFR 892.2050 |
| Regulatory Class | Class II |
| Device Panel | Radiology |
| Product Classification Code | LLZ |
{4}------------------------------------------------
| 510(k) Summary | | |
|--------------------|------------------|---------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 |
# 5. PREDICATE DEVICE
| Predicate Manufacturer | Pixmeo, Sarl |
|------------------------|--------------|
| Predicate Trade Name | Osirix MD™ |
| Predicate 510(k) | K101342 |
No reference devices were used in this submission.
# 6. DEVICE DESCRIPTION
The Horos MD™ is an interactive image display and navigation software device for diagnosis of medical images. It provides both 2D and 3D image visualization tools for CT and MRI scans from different makes and models of image acquisition hardware. It does not produce any original medical images and does not contain controls for the direct operation of a diagnostic imaging system. Horos MD™ conforms to the DICOM standard to allow the sharing of medical images with other digital imaging systems such as PACS (Picture Archiving and Communication System).
The Horos MD™ software device runs on the macOS X platform, taking advantage of its optimized 3D graphic capabilities, which are provided by the METAL framework developed and maintained by Apple Inc.
The user interface of the software follows Apple's Human Interface Guidelines (HIG) in order to create a user interface that is intuitive and easy to use for users who are familiar with other Apple products. Typical users of this device are radiologists and clinicians who are familiar with 2D scan images.
### 7. INDICATIONS FOR USE
Horos MD™ is a software device intended for viewing images acquired tomography (CT), computed radiography (CR), magnetic resonance (MR), ultrasound (US) and other DICOM compliant medical imaging systems when institude commercial standard hardware. Images and data can be stored, communicated, processed, and displayed within the system. It is intended for use as a diagnostic and review tool by trained healthcare professionals. This device is not to be used for mammography.
{5}------------------------------------------------
| 510(k) Summary | | |
|--------------------|------------------|---------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 |
It is the User's responsibility to operate the device in accordance with the software requirements listed in the instructions for use, in particular ensuring that monitor (display) quality, ambient light conditions, and image compression ratios are consistent with the clinical application.
### 8. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
| Characteristics | Subject device<br>Horos MD™ | Primary Predicate<br>Osirix MD™ | Justification for<br>differences |
|------------------------|------------------------------------------------------------------------|-------------------------------------------------------------------|--------------------------------------------------------------|
| Submitter | iCat Solutions Ltd | Pixmeo, Sarl | - |
| 510(k) Number | K232589 | K101342 | - |
| Product classification | Class II | Class II | |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Product code | LLZ | LLZ | Same |
| Type of use | Prescription (RX Only) | Prescription (RX Only) | Same |
| Class | II | II | |
| Regulations number | 21 CFR 892.2050 | 21 CFR 892.2050 | |
| Regulation Name^ | Medical Image Management and<br>Processing System | Picture archiving and<br>communications system | 21 CFR<br>892.2050 name has<br>been amended in<br>April 2021 |
| Classification Name | System, image processing, radiological | System, image processing,<br>radiological | Same |
| Product code | LLZ | LLZ | Same |
| Indications for Use | Horos MD™ is a software device<br>intended for viewing images acquired | OsiriX MD"M isa software device<br>intended for viewing of images | Same. |
Table 1: Comparison of Horos MD™ with Osirix MD™ (K101342).
{6}------------------------------------------------
| 510(k) Summary | | |
|--------------------|------------------|---------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 |
| | from computed tomography (CT),<br>computed radiography (CR), magnetic<br>resonance (MR), ultrasound (US) and<br>other DICOM compliant medical imaging<br>systems when installed on suitable<br>commercial standard hardware. Images<br>and data can be stored, communicated,<br>processed, and displayed within the<br>system.<br>It is intended for use as a diagnostic and<br>review tool by trained healthcare<br>professionals.<br>This device is not to be used for<br>mammography.<br>It is the User's responsibility to operate<br>the device in accordance with the<br>software and hardware requirements<br>listed in the instructions for use, in<br>particular ensuring that monitor (display)<br>quality, ambient light conditions, and<br>image compression ratios are consistent<br>with the clinical application. | acquired from CT, MR, CR, DR, US<br>and other DICOM compliant medical<br>imaging systems when installed on<br>suitable commercial standard<br>hardware.<br>Images and data can be captured,<br>stored, communicated, processed,<br>and displayed within the system and<br>or across computer networks at<br>distributed locations.<br>Lossy compressed mammographic<br>images and digitized film screen<br>images must not be reviewed for<br>primary diagnosis or image<br>interpretation. For primary diagnosis,<br>post process<br>DICOM "for presentation" images<br>must be used. Mammographic<br>images should only be viewed with a<br>monitor approved by FDA for viewing<br>mammographic images.<br>It is the User's responsibility to<br>ensure monitor quality, ambient light<br>conditions, and image compression<br>ratios are consistent with the clinical<br>application. | The subject device<br>is contraindicated<br>for Mammography.<br>Therefore, the<br>subject device<br>eliminates the risks<br>associated with the<br>analysis of<br>mammographic<br>images. |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Users | Trained healthcare professionals such as<br>radiologists, clinicians and other qualified<br>physicians | Trained healthcare professionals | Same |
| Mammography diagnostic use | No | Yes (Mammographic images should<br>only be viewed with a monitor<br>approved by FDA for viewing<br>mammographic images) | The subject device<br>is contraindicated<br>for Mammography.<br>Therefore, the |
{7}------------------------------------------------
| 510(k) Summary | | | |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 | |
| | | subject device | |
| | | | subject device<br>eliminates the risks<br>associated with the<br>diagnosis and<br>analysis of<br>mammographic<br>images. |
| Imaging modalities | • CT<br>• MR<br>• CR<br>• DR<br>• US | • CT<br>• MR<br>• CR<br>• DR<br>• US | Same |
| Communications | DICOM | DICOM | Same |
| Installation | Local computers | Local computers | Same |
| Operating system for<br>diagnostic viewing | MacOS (Mac) | MacOS (Mac) | Same |
| User authentication (unique<br>login) | Yes | Yes | Same |
| Access to DICOM studies | • Connect to existing onsite or<br>remote PACS<br>• Local computer/server<br>• USB drives / CD / DVD | • Connect to existing onsite or<br>remote PACS<br>• Local computer/server<br>• USB drives / CD / DVD | Same |
| Basic visualization tools | • Grayscale display<br>• Rotate<br>• Pan<br>• Zoom<br>• Magnification<br>• Fit to screen.<br>• Scroll<br>• Flip horizontal / vertical<br>• Axial / coronal / sagittal views<br>• Cross-hair | • Grayscale display<br>• Rotate<br>• Pan<br>• Zoom<br>• Magnification<br>• Fit to screen<br>• Scroll<br>• Flip horizontal / vertical<br>• Axial / coronal / sagittal views<br>• Cross-hair | Same |
{8}------------------------------------------------
| 510(k) Summary | | | | | |
|--------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Horos MD (HOROSMD) | | Document: DOC-21 | | Date: 2023/08/25 14:27:46 | |
| | Adjustable windowing (WW/WL) HU (Hounsfield Unit) Slice indicator Reset | Adjustable windowing (WW/WL) HU (Hounsfield Unit) Slice indicator Reset | Adjustable windowing (WW/WL) HU (Hounsfield Unit) Slice indicator Reset | | |
| Advanced tools | Volume Rendering (3D) MPR | Volume Rendering (3D) MPR | Volume Rendering (3D) MPR SSD | | The subject device<br>does not include<br>SSD. This limitation<br>does not alter the<br>device's intended<br>use or indications<br>for use, nor does it<br>introduce any new<br>safety risks.<br>Therefore, there is<br>no risk to the<br>device's safety or<br>performance. |
| Measurements | Line Angle Area (circular, irregular shape, polygonal) Point Cobb Angle CTR | Line Angle Area (circular, irregular shape, polygonal) Point | Line Angle Area (circular, irregular shape, polygonal) Point | | The subject device<br>has additional<br>measuring<br>functions. Subject<br>device features and<br>functions are<br>described in the<br>user's manual and<br>tested with intended<br>users for clinical<br>use. Additional<br>measurement<br>functions do not<br>change the device's<br>intended use or |
{9}------------------------------------------------
| 510(k) Summary | | | | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 | | |
| Annotations | yes | yes | Same | |
| Anonymization function | yes | yes | Same | |
| Studies comparison | yes | yes | Same | |
| Share function | yes | yes | Same | |
| Export functions | Yes (JPEG, .dcm, .zip) | Yes (JPEG, PNG,.dcm, .zip) | The subject device exports files in JPEG, .dcm, and .zip formats but not in PNG format. This limitation does not alter the device's intended use or indications for use, nor does it introduce any new safety risks. Therefore, there is no risk to the device's safety or performance. | |
| Report generation | No | No | Same | |
| Device type/ form | Software device that operates on off-the-shelf hardware | Software device that operates on off-the-shelf hardware | Same | |
| 510(k) Summary | | | | |
| Horos MD (HOROSMD) | | Document: DOC-21 | | Date: 2023/08/25 14:27:46 |
| DICOM standard for data<br>exchange | NEMA PS3.1 2023b (2023) Digital<br>Imaging and Communications In<br>Medicine (Dicom) Set | | NEMA PS 3.1 - 3.20 (2016) Digital<br>Imaging and Communications In<br>Medicine (Dicom) Set | Same |
| JPEG standard for image<br>compression | IEC/ ISO 10918-1, Edition 1 (1994-02-<br>15), Information Technology – Digital<br>Compression and Coding of Continuous-<br>Tone Still Images: Requirements And<br>Guidelines [Including: Technical<br>Corrigendum 1 (2005) | | IEC/ ISO 10918-1, Edition 1 (1994-<br>02-15), Information Technology -<br>Digital Compression and Coding of<br>Continuous-Tone Still Images:<br>Requirements And Guidelines<br>[Including: Technical Corrigendum 1<br>(2005) | Same |
{10}------------------------------------------------
#### 9. EVALUATION OF SIMILARITIES AND DIFFERENCES
The main differences between the subject device and the primary predicate is in the system installation via the Apple AppStore versus installation from a .dmg binary file for the predicate device) and the support of mammography images (Horos MD™ must not be used for mammography diagnostic use)
Horos MD™ and the primary predicate device have the same technical characteristics, general function, intended use, and does not raise any new potential safety risks and is equivalent in performance. In terms of use and functions, both systems access, upload, and display DICOM images and metadata, and resources to trained healthcare professionals for study review and analysis.
### 10. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Non-Clinical Data
#### Software Verification and Validation Testing •
All of the different components of Horos MD™ software have been tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use. Software verification testing were conducted, and documentation was provided as detailed in FDA's Guidance for Industry and FDA Staff: "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices."
{11}------------------------------------------------
| 510(k) Summary | | |
|--------------------|------------------|---------------------------|
| Horos MD (HOROSMD) | Document: DOC-21 | Date: 2023/08/25 14:27:46 |
iCat Solutions Ltd has implemented security features for the device and data protection. Cybersecurity requirements, risk analysis, and mitigation was addressed in accordance Content of Premarket Submission for Management of Cybersecurity in Medical Devices (Oct 2014).
#### Bench Performance Testing .
No clinical testing was required to demonstrate safety or effectiveness for the subject device as the device's non-clinical (bench) testing was sufficient to support the intended use of the device.
The measurement accuracy for the distance (length), and area features was validated using Digital Reference Objects compared against the known values. The Digital Reference Objects (n=75 test cases) created were representative of the clinical range typically encountered in radiology practice (1-190 mm). Inter-operator error (reproducibility), and intra-operator error (repeatability) were assessed.
| Function | Accuracy |
|-------------------|----------|
| Distance (1-10mm) | 99.68% |
| Distance (>10mm) | 99.82% |
| Area (1-10mm) | 99.19% |
| Area (>10mm) | 99.66% |
| Angle (1-190mm) | 100% |
| Point (1-190mm) | 100% |
### 11. CONCLUSIONS
There is no difference between the subject device (Horos MD™ ) and the predicate device in terms of its intentions for use and fundamental scientific technology. Neithers in usage, safety, or effectiveness, thus no new questions regarding safety or effectiveness are raised. The non-clinical performance test data and validation demonstrate that the subject device performs comparably to and is as safe and effective as the predicate with the 21 CFR Part 807 and based on the information provided in this premarket notification, Horos MD™ is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.