K232581 · Suzhou Haishen Medical Device Associates Co., Ltd. · GXZ · Dec 4, 2023 · Neurology
Device Facts
Record ID
K232581
Device Name
Medical Disposable Sterile Needle Electrode
Applicant
Suzhou Haishen Medical Device Associates Co., Ltd.
Product Code
GXZ · Neurology
Decision Date
Dec 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1350
Device Class
Class 2
Attributes
Pediatric
Indications for Use
This product is intended for use with recording, monitoring equipment for the recording of bioelectrical signals, such as electromyograph (EMG). The electrodes are sterile and for single patient use only.
Device Story
Medical disposable sterile needle electrodes; invasive, single-use devices; penetrate skin to contact extracellular fluid/muscle/nerves. Collect bioelectrical signals (myoelectric currents) for transmission to EMG diagnostic instruments. Used in hospital settings under physician supervision. Configurations include monopolar, concentric, spiral, and subdermal designs; some include lead wires. Device acts as passive interface between tissue and recording equipment; no active processing or stimulation. Healthcare providers use output to assess muscle electrical activity; aids in clinical diagnosis of neuromuscular conditions. Benefits include reliable signal acquisition for diagnostic monitoring.
Clinical Evidence
Bench testing only. Testing included surface quality, sharpness index, toughness, connection strength, electrical conductivity, insulation, and sterility. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity) confirmed safety. Sterilization validated per ISO 11135.
Technological Characteristics
Stainless steel needle; PVC-insulated tin-plated copper lead wires. Ethylene Oxide sterilization (SAL 10^-6). Dimensions: 20-75mm length, 0.25-0.65mm diameter depending on configuration. Passive sensing; no energy source. Complies with ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-23, and ISO 11135.
Indications for Use
Indicated for adults, children, and infants requiring recording of bioelectrical signals (EMG) via subdermal or muscle tissue needle electrodes. Contraindications: Not for stimulation.
Regulatory Classification
Identification
A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.
{0}------------------------------------------------
December 4, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Suzhou Haishen Medical Device Associates Co., Ltd % Jie Gao Consultant Sinow Medical AS Høyteknologisenteret Thormøhlens Gate 55 Bergen, 5008 Norway
Re: K232581
Trade/Device Name: Medical Disposable Sterile Needle Electrode Regulation Number: 21 CFR 882.1350 Regulation Name: Needle Electrode Regulatory Class: Class II Product Code: GXZ Dated: August 23, 2023 Received: August 25, 2023
Dear Jie Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/8 description: The image shows the name "Patrick Antkowiak -S" in black text. The name "Patrick" is on the first line, and "Antkowiak -S" is on the second line. To the left of the name is a faded FDA logo.
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical,
{2}------------------------------------------------
Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Medical Disposable Sterile Needle Electrode
Indications for Use (Describe)
This product is intended for use with recording, monitoring equipment for the recording of bioelectrical signals, such as electromyograph (EMG). The electrodes are sterile and for single patient use only.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Submitter: Suzhou Haishen United Medical Device Associates Co., Ltd
Address: 3F, 4F 401, 4F 402, 5F, Building 10, No. 168 Majian Road, Suzhou, Jiangsu Province, China
Telephone: +86 18921623557
Email: 785445240@qq.com
Contact Person: Leyi Dai
Date Prepared: 01.07.2023
II. Consultant
Company: Sinow Medical AS
Address: Høyteknologisenteret Thormøhlens Gate 55, Bergen, Norway
Contact Person: Jie Gao
Telephone: +47 99127913
Email: gao@bergemed.com
III. DEVICE
Trade Name: Medical Disposable Sterile Needle Electrode
Classification Name: Needle Electrode
Regulatory Class: II
Product Code: GXZ
IV. PREDICATE DEVICE
Product Name: Subdermal Needle Electrodes
510(k) Number: K072276
Product Code: GXZ
Manufacture: Xian Friendship Electronics Co., Ltd.
V. DEVICE DESCRIPTION
Medical disposable sterile needle electrodes are sterile, single use products that penetrate the skin and directly contact the extracellular fluid during use, forming a good electrolyte interface. The pointed part of the needle electrode makes contact with the nerves, muscles, or
{5}------------------------------------------------
other tissues under examination to collect and transmit the corresponding bioelectrical signals (myoelectric currents) to the electromyography (EMG) recorder. Needle electrodes are used in conjunction with EMG diagnostic instruments and serve as accessories to the EMG machine for detecting myoelectric potentials. Medical disposable sterile needle electrodes are invasive since they are positioned subcutaneously and are used under the supervision of a licensed physician.
The subject device, available in configurations with and wires, which encompasses subdermal, monopolar, spiral, and concentric designs. The subject device's configurations of monopolar and concentric needles correspond to those without lead wires, while the configurations involving spiral and subdermal designs include lead wires.
### VI. INDICATIONS FOR USE
This product is intended for use with recording, monitoring equipment for the recording of bioelectrical signals, such as electromyograph (EMG). The electrodes are sterile and for single patient use only.
## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The technological characteristics comparing to the predicate device are outlined in the Substantial Equivalent table.
| Attribute | Medical Disposable<br>Sterile Needle Electrodes<br>(Subject Device) | Friendship Subdermal<br>Needle Electrodes<br>(Predicate device<br>K072276) | Comment |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Regulation<br>Number | 21 CFR 882.1350 | 21 CFR 882.1350 | SE |
| Regulation Name | Needle Electrode | Needle Electrode | SE |
| Regulatory Class | Class II | Class II | SE |
| Product Code | GXZ | GXZ | SE |
| Indications for<br>Use | This product is intended for<br>use with recording,<br>monitoring equipment for<br>the recording of<br>bioelectrical signals, such<br>as electromyograph<br>(EMG). The electrodes are<br>sterile and for single patient<br>use only. | Subdermal Needle<br>Electrodes are intended for<br>use with recording,<br>monitoring and<br>stimulation/recording<br>equipment for the recording<br>of biopotential signals<br>including<br>electroencephalograph<br>(EEG), electromyograph<br>(EMG) and nerve potential<br>signals and for stimulation<br>during the intraoperative<br>diagnosis of acute<br>dysfunction in corticospinal | Subject<br>Device<br>without<br>stimulation<br>function;<br>Subject<br>Device only<br>records in<br>EMG |
| | | axonal conduction. The<br>electrodes are sterile and for<br>single patient use only. | |
| Environment of<br>Use | Hospital | Hospital | SE |
| Provided Sterile | Yes | Yes | SE |
| Method of<br>Sterilization | Ethylene Oxide | Ethylene Oxide | SE |
| Sterility<br>assurance level<br>(SAL) | $10^{-6}$ | $10^{-6}$ | SE |
| Single Use,<br>Disposable | Yes | Yes | SE |
| Basic Principle | A needle electrode is an<br>electrode that requires<br>penetration through the<br>skin to directly contact<br>extracellular fluid. It forms<br>a good electrode-electrolyte<br>interface and is used in<br>conjunction with<br>electromyography (EMG)<br>diagnostic instruments as<br>an electrode for detecting<br>muscle electrical activity. | A needle electrode is an<br>electrode that requires<br>penetration through the skin<br>to directly contact<br>extracellular fluid. It forms<br>a good electrode-electrolyte<br>interface and is used in<br>conjunction with recording,<br>monitoring and<br>stimulation/recording<br>equipment for the recording<br>of biopotential signals<br>including<br>electroencephalograph<br>(EEG), electromyograph<br>(EMG) and nerve potential<br>signals and for stimulation<br>during the intraoperative<br>diagnosis of acute<br>dysfunction in corticospinal<br>axonal conduction. | Same basic<br>principle but<br>Subject<br>Device has<br>less clinical<br>usage than<br>Predicate<br>Device |
| Target Population | Adults, children and infants | Adults, children and infants | SE |
| Anatomical Sites | Subdermal, muscle tissue | Subdermal, nerve or muscle<br>tissue | Subject<br>Device has<br>less<br>Anatomical<br>Sites |
| Needle material | Stainless steel | Stainless steel | SE |
| Needle length | Spiral needle:<br>$20 \pm 1.5mm$<br>Concentric needle:<br>$(20 - 75) \pm 1.5mm$<br>Subdermal needle: | Spiral needle:<br>NA<br>Concentric needle:<br>20 - 75mm<br>Subdermal needle: | Equivalent |
| | | | |
| | (6, 7, 13, 19, 20, 22) ±<br>1.5mm | 7, 13, 17, 23mm | |
| | Monopolar needle:<br>(20, 25, 30, 37, 50, 75) ±<br>1.5mm | Monopolar needle:<br>20, 25, 30, 38, 50, 75mm | |
| Needle diameter | Spiral needle:<br>0.60 ± 0.15mm | Spiral needle:<br>0.60 mm | Equivalent |
| | Concentric needle:<br>(0.25 - 0.65) ± 0.15mm | Concentric needle:<br>0.25 - 0.65mm | |
| | Subdermal needle:<br>(0.35, 0.40, 0.45) ±<br>0.15mm | Subdermal needle:<br>0.35, 0.40, 0.45mm | |
| | Monopolar needle:<br>(0.25, 0.30, 0.35, 0.40,<br>0.45, 0.50, 0.65) ± 0.15mm | Monopolar needle:<br>0.25, 0.30, 0.35, 0.40, 0.45,<br>0.50, 0.65mm | |
| Lead wire<br>material | PVC insulated tin plated<br>with copper | PVC insulated tin plated<br>with copper | SE |
| Lead wire length | (1000 - 2500) ± 25mm | 1200 - 2500mm | Equivalent |
{6}------------------------------------------------
{7}------------------------------------------------
#### VIII. NON-CLINCAL TESTING
Suzhou Haishen United Medical Device Associates Co., Ltd has performed the following nonclinical laboratory testing to determine substantial equivalence.
| Test | Result |
|-------------------------------------------------|--------|
| Surface quality | Pass |
| Sharpness index | Pass |
| Toughness | Pass |
| Connection strength | Pass |
| Electrical conductivity | Pass |
| Conductive terminals and needle body insulation | Pass |
| Sterility testing | Pass |
## IX. BIOCOMPATIBILITY TESTING
The contact classification for the Medical Disposable Sterile Needle Electrode component of the subject device is a subdermal communicating device with muscle. The biocompatibility evaluation for the subject device was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and in accordance with FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk
{8}------------------------------------------------
management process", Guidance for Industry and Food and Drug Administration Staff. The results of biocompatibility testing included in the table below demonstrate that the device meets biological safety requirements per ISO 10993-1 for subdermal communicating devices with muscle.
| Test | Standard | Result |
|---------------------------|--------------|-----------------------------|
| In vitro cytotoxicity | ISO 10993-5 | Non-cytotoxic |
| Skin sensitization | ISO 10993-10 | Non-sensitive |
| Intracutaneous reactivity | ISO 10993-23 | Non- irritation |
| Acute systemic toxicity | ISO 10993-11 | Non-acute systemic toxicity |
| Pyrogen | ISO 10993-11 | Non-pyrogenic |
## X. STERILIZATION AND SHELF-LIFE TESTING
The subject device is sterilized using Ethylene Oxide. The sterilization validation has been performed in accordance with the principles of ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. A sterility assurance level (SAL) of 100 has been demonstrated. The device meets EO residuals per ISO 10993-7.
A shelf-life of 3 years has been established based on accelerated and real-time aging.
#### XI. CONCLUSIONS
The subject and predicate device share intended use - for temporary use with recording and monitoring equipment for the recording and monitoring of bioelectrical signals from the corresponding muscle parts of the human body, such as electromyography (EMG). The subject device does not have stimulation function as the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness, and the nonclinical performance data submitted in the 510(k) demonstrate that the subject device is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.