This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
Device Story
Steerable guide wire; 0.014-inch diameter; 190cm or 300cm lengths. Used in peripheral vascular procedures to facilitate placement/exchange of diagnostic/therapeutic devices. Features solid core, hydrophilic-coated coil-type distal end, and PTFE-coated core shaft. Radiopaque distal end allows fluoroscopic visualization for vessel selection and positioning. Includes detachable extension wire and torque device. Operated by physicians in clinical settings. Provides mechanical support and guidance for interventional devices; benefits patient by enabling access to peripheral vasculature.
Clinical Evidence
Bench testing only. Tests included dimensional verification, visual inspection, tensile strength/tip pull, torque strength, torqueability, coating adhesion/integrity, catheter compatibility/lubricity, corrosion resistance, kink resistance, tip flexibility, and radiopacity. Results demonstrated the device met all acceptance criteria and performed similarly to the predicate.
Technological Characteristics
Steerable guide wire; 0.014" (0.36mm) OD. Materials: Platinum-Nickel outer coil, stainless steel tapered core wire. Coatings: PTFE (shaft), hydrophilic (distal end). Sterilization: Ethylene Oxide (validated per AAMI TIR28). Form factor: 190cm/300cm lengths. Mechanical sensing/actuation via manual torque and navigation.
Indications for Use
Indicated for patients undergoing intravascular procedures requiring the placement and exchange of diagnostic and therapeutic devices in the peripheral vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2023
Filmec Co. Ltd. Meagan Fagan Principal Consultant CardioMed Device Consultants 1783 Forest Drive Suite 254 Annapolis. Maryland 21401
#### Re: K232574
Trade/Device Name: Vassallo GT Crossing14 Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: August 24, 2023 Received: August 25, 2023
#### Dear Meagan Fagan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Image /page/1/Picture/5 description: The image shows the end of a letter. The text "Sincerely," is at the top, followed by the name "Lydia S. Glaw -S" in a larger font. The name is split across three lines.
Digitally signed by Lydia S. Glaw -S Date: 2023.10.11 14:38:35 -04'00'
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232574
Device Name VASSALLO® GT Crossing14
Indications for Use (Describe)
This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
Image /page/4/Picture/1 description: The image shows the logo for "Filmecc". The logo consists of a red abstract shape on the left, resembling a stylized letter "A" or an incomplete circle. To the right of this shape is the word "Filmecc" in black, with a small red square above the "i". The font is bold and sans-serif, giving the logo a modern and professional appearance.
# VASSALLO® GT G14
| Date Prepared: | August 24, 2023 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | FILMECC CO., LTD.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya-shi, Aichi 463-0024<br>Japan<br>TEL: +81-52-768-1212, FAX: +81-52-768-1222 |
| Contact: | Noriaki Tsuzuki<br>Regulatory Affairs / Vigilance Section<br>FILMECC CO., LTD.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya-shi, Aichi 463-0024<br>Japan<br>TEL: +81-52-768-1212, FAX: +81-52-768-1222<br>e-mail: noriaki.tsuzuki@filmecc.com |
| Trade Name: | VASSALLO® GT Crossing14 |
| Device Classification: | Class 2 per 21 CFR §870.1330 |
| Classification Name: | Catheter, Guide, Wire |
| Product Code: | DQX - Catheter Guide Wire |
| Predicate Devices: | Filmecc VASSALLO GT (K203533) |
#### INTENDED USE/INDICATIONS FOR USE:
#### VASSALLO® GT Crossing14
This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
### DEVICE DESCRIPTION:
The VASSALLO® GT Crossing14 Peripheral Guide Wire in this submission is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) in available lengths of 190cm and 300cm. The outer coil wire diameter has been modified to 0.072 mm.
{5}------------------------------------------------
The devices have a solid core with a hvdrophilic coated coil-type distal end. The coil is radiopaque to facilitate selection of the blood vessel and confirmation of the position of the guide wire's distal end by fluoroscopy.
The core shaft surface is coated with Polytetrafluoroethylene (PTFE). About 2 cm of the distal end can be shaped. A detachable extension wire (hereafter "extension wire") is available to connect with the proximal end of the quide wire with a length of less than 300cm. A Torque device is included in the same package.
## COMPARISON WITH PREDICATE DEVICE:
Comparisons of the VASSALLO® GT Crossing14 and predicate device show that the technological characteristics of the subject device such as the components, design, materials, sterilization method, shelf life and operating principle are similar, if not identical, to the currently marketed predicate device. The intended use of the subject device and its predicate are the same.
| Name of Devices | VASSALLO® GT Crossing14 | VASSALLO GT |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| | Subject | Predicate |
| 510(k) | K232574 | K203533 |
| Intended Use and<br>Indications | This product is intended to facilitate the placement and exchange of diagnostic and<br>therapeutic devices during intravascular procedures. This device is intended for<br>peripheral vascular use only. | |
| Nominal OD | 0.36mm<br>(0.014in) | 0.36mm<br>(0.014in) |
| Overall Length | 190, 300cm | |
| Outer Coil | Platinum-Nickel | Platinum or Platinum-Nickel and<br>Stainless Steel |
| Outer Coil Diameter | 0.072 mm | 0.060 mm - 0.080 mm |
| Tapered Core Wire | Stainless Steel | |
| Tip Shape | Straight | |
| Coating | Hydrophilic, Hydrophobic | |
| Sterilization | Ethylene Oxide | |
#### NON-CLINICAL TESTING/PERFORMANCE DATA:
Non-clinical laboratory testing was performed on the VASSALLO® GT Crossing14 to determine substantial equivalence to the predicate device. The following testing/assessments were performed:
- Dimensional Verification ●
- Visual Inspection .
- Tensile Strength / Tip Pull
- Torque Strength
- Torqueability ●
- . Coating Adhesion/Integrity
- Catheter Compatibility / ● Lubricity
- Corrosion Resistance
- Kink Resistance ●
- Tip Flexibility ●
- Radiopacity ●
{6}------------------------------------------------
The in vitro bench tests demonstrated that the VASSALLO GT Crossing14 met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate devices.
### BIOCOMPATIBILITY:
Testing was performed to assess biocompatibility of the predicate VASSALLO GT. The following tests were performed:
- Cytotoxicity ●
- Sensitization
- Intracutaneous Irritation ●
- Acute Systemic Toxicity ●
- Material Mediated ● Pyrogenicity
- Hemolysis
- Partial Thromboplastin Time
- In Vivo Thromboresistance
- Complement Activation (SC5b-9)
The results from this testing are being leveraged for the subject device and support the biocompatibility of the VASSALLO GT Crossing14.
#### STERILIZATION:
The minor differences of the subject device do not represent a greater sterilization challenge than the predicate VASSALLO GT quide wires. As such, no additional sterilization validation was required, and the subject device was adopted into the product family for which a sterilization cycle was validated in accordance with AAMI TIR28.
#### CONCLUSION:
The VASSALLO GT Crossing14 has the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate device. Performance data demonstrates that the device functions as intended. The conclusions drawn from the nonclinical tests demonstrate that the VASSALLO GT Crossing14 is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Therefore, the VASSALLO GT Crossing14 is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.