Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crown, bridges, or over-dentures.
Device Story
Angled Abutment is a dental implant component used to support prosthetic restorations like crowns, bridges, or over-dentures. It is manufactured from Ti-6Al-4V Eli titanium alloy with a partial TiN coating. The device features HEX or Non-HEX connections and is compatible with specific Arum Dentistry implant systems (NB 1 SA and NB). It is supplied non-sterile and requires end-user sterilization before clinical use. The abutment is secured to the implant using an abutment screw. It is intended for use by dental professionals in a clinical setting. The device provides the necessary structural interface between the endosseous implant and the final dental prosthesis, facilitating patient oral rehabilitation.
Clinical Evidence
Bench testing only. Mechanical performance validated via static compression and compression fatigue testing according to ISO 14801 on worst-case constructs. Biocompatibility leveraged from K213506. Sterilization validated per ISO 17655-1 and FDA guidance. MRI safety evaluated via scientific rationale and literature.
Technological Characteristics
Material: Ti-6Al-4V Eli (ASTM F136). Surface: TiN coating. Connection: HEX and Non-HEX. Dimensions: Ø 4.5, 5.5 mm; 8 mm post height; 17° angle. Sterilization: Non-sterile, requires end-user sterilization. Connectivity: Mechanical interface for dental implants.
Indications for Use
Indicated for patients requiring prosthetic restoration support (crowns, bridges, or over-dentures) via dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2023
Arum Dentistry Co., Ltd. Won-Yi Choi Assistant Manager 23, Gukjegwahak 11-ro, Yuseong-gu Daejeon, 34002 REPUBLIC OF KOREA
Re: K232560
Trade/Device Name: Angled Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 21, 2023 Received: November 21, 2023
Dear Won-Yi Choi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K232560
Device Name Angled Abutment
Indications for Use (Describe)
Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crown. bridges, or over-dentures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle on top of a smaller, light orange triangle. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in bold, dark brown letters, with "ARUM" on top of "DENTISTRY".
# 510(k) Summary
## Submitter
ARUM DENTISTRY Co., Ltd. Won-Yi Choi 23, Gukjegwahak 11-ro, Yuseong-gu Daejeon, 34002 Republic of Korea Email: arum_ra@arumdentistry.com Tel. +82-42-935-3644 Fax. +82-42-935-3633
Date prepared: December 14, 2023
## Device Information
- . Trade Name: Angled Abutment
- Common Name: Endosseous Dental Implant Abutment
- Classification Name: Abutment, Implant, Dental, Endosseous ●
- Primary Product Code: NHA
- Panel: Dental ●
- . Regulation Number: 21 CFR 872.3630
- Device Class: Class II
- . Date Prepared: 11/20/2023
#### Predicate Devices
The subject device is substantially equivalent to the following predicate devices:
#### Primary Predicate
K182091, Osstem Abutment System by OSSTEM Implant Co., Ltd.
#### Reference Device
K213506, NB 1 SA Implant System by ARUM DENTISTRY Co., Ltd. K230725, NB Implant System by ARUM DENTISTRY Co., Ltd.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle and a yellow triangle on the left side of the image. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in a bold, dark brown font.
# General Description
# Abutment
- . Angled Abutment
The Angled Abutment is made from Ti-6Al-4V Eli (conforming to ASTM F136). Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crown, bridges, or over-dentures. The Angled Abutment has two types of connection (HEX and Non-HEX). The Angled Abutment connections compatible with NB 1 SA Implant System and NB Implant System. The Angled Abutment surfaces are partially TiN coated. Angled Abutments are supplied with an abutment screw previous cleared device as K213506.
The dimension ranges of the subject device are below:
| No. | Device Name | Dimension |
|-----|----------------------------------------|---------------------------------------------|
| 1 | Angled Abutment | $Ø$ 4.5, 5.5 (D) x 8 mm (Post Height) x 17° |
| 2 | Abutment Screw<br>(Cleared in K213506) | $Ø$ 2.35 (D) x 8.4 mm (L) |
Angled Abutment and Abutment Screw (Cleared in K213506) are provided non-sterile. The abutment should be sterilized before use by end user sterilization. These devices are intended for single use only.
The Angled Abutment is compatible with the following implant systems.
| Manufacturer | 510(k) No. | Implant system<br>compatibility | Dimension |
|--------------------------------|------------|---------------------------------|-------------------------------------|
| ARUM<br>DENTISTRY<br>Co., Ltd. | K213506 | NB 1 SA Implant System | Ø 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 |
| ARUM<br>DENTISTRY<br>Co., Ltd. | K230725 | NB Implant System | Ø 3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 |
# Indication for Use
Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crown, bridges, or over-dentures.
# Materials:
| No. | Component Name | Materials | Surface treatment |
|-----|-----------------|----------------------------|-------------------|
| 1 | Angled Abutment | Ti-6Al-4V Eli of ASTM F136 | TiN coating |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
| | Abutment Screw<br>(Cleared in K213506) | Ti-6Al-4V Eli of ASTM F136 | Non coating |
|---|----------------------------------------|----------------------------|-------------|
| 2 | | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font, with "ARUM" on top and "DENTISTRY" on the bottom.
# Summaries of Technology Characteristics
# 1) Angled Abutment
| | Subject Device | Primary Predicate |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Device Name | Angled Abutment | Osstem Abutment System |
| 510(k) Number | N/A | K182091 |
| Intended Use/<br>Indications for use | Angled Abutment is intended for use with a dental<br>implant to provide support for prosthetic<br>restorations such as crown, bridges, or over-<br>dentures. | Osstem Abutment System is intended for use<br>with a dental implant to provide support for<br>prosthetic restorations such as crowns, bridges,<br>or overdentures. |
| Design | Image: Angled Abutment | Image: Osstem Abutment System |
| Material | Ti-6Al-4V Eli (ASTM F136) | Ti-6Al-4V Eli (ASTM F136) |
| Anti-Rotational<br>Feature | Hex, Non-Hex | Hex, Non-Hex |
| Range of<br>Diameters (ø) | 4.5, 5.5 | 4.0, 4.5, 5.0, 6.0 |
| Rang of Post<br>(mm) | 8 | 8 |
| Angle (°) | 17 | 17 |
| Sterilization | Non-Sterilization | Non-Sterilization |
| Surface<br>Treatment | TiN Coating | TiN Coating |
| Substantial<br>Equivalent<br>Discussion | <b>1. Similarities</b><br>The Angled Abutment has the same intended use for, technological characteristics to the K182091.The<br>Angled Abutment have same device characteristics with the reference device predicate such as<br>material, anti-rotational feature, range of post, angle, sterilization intended use, functions, general<br>shape (Design), structure and applied production method. | |
| | <b>2. Differences</b><br>Compared to the reference device predicate, the subject device's diameter is different. However, except<br>for the diameters, range of lengths (post, angle, intended use, material, functions and general shape<br>(Design) are the same. Although the diameters are slightly different but it doesn't impact product's<br>safety and effectiveness because the Subject Device is included in the range of reference device<br>predicate's diameters. | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of a red triangle on top of a smaller, light orange triangle. The words "ARUM DENTISTRY" are in a bold, dark brown font, with "ARUM" on top of "DENTISTRY".
{9}------------------------------------------------
K232560
Image /page/9/Picture/1 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle and a yellow triangle on the left side of the image. To the right of the triangles, the words "ARUM" and "DENTISTRY" are written in a bold, brown font.
## Performance Data
Non-clinical testing data submitted, referenced or relied on in this submission support demonstrating substantial equivalence.
## Biocompatibility
Biocompatibility of Ti-6AI-4V Eli (ASTM F136) demonstrated by the reference ARUM DENTISTRY submission, K213506, using the same materials and manufacturing processes as the subject device.
# Sterilization validation
The Angled Abutment delivered non-sterile to be end-user sterilized, the recommended sterilization has been validated according to ISO 17655-1 and to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015". The predicate devices may be leveraged for the subject devices because of using the same materials, manufacturing methods, and sterilization procedures. The sterilization methods were leveraged from predicated K213506. The worst-case construct was tested, and results demonstrated equivalence to the predicate devices.
#### Non-Clinical Data
Mechanical performance testing was performed according to ISO 14801. For each compatible implant line, worst-case constructs were subjected to static compression and compression fatique testing. Minor differences in the designs, dimensions, sizes, or compatible implant lines among the subject device, the primary predicate devices, and the reference devices do not affect substantial equivalence. These minor differences do not impact substantial equivalence because these differences are related to the compatible implant designs, or are mitigated by the mechanical performance testing.
Non-clinical performance data submitted to demonstrate substantial equivalence included:
- Static and fatigue testing according to ISO 14801.
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the primary predicate.
# MR Environment Condition
Non-Clinical worst-case MRI review was performed to evaluated the Angled Abutment devices in the MRI environment using scientific rationale and published literature (e.g., Terry O. Woods, Jana Delfino, & Sunder Rajan. (2019). Assessment of Magnetically Induced Displacement
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of a red triangle on top and a smaller, light orange triangle on the bottom. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
K232560
Force and Torque on Metal Alloys Used in Medical Devices. Journal of Testing and Evaluation 49.2, 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
## Conclusion
The Indications for Use statements are highly similar. Overall, the Technological Characteristics of the Subject device are highly similar to the Predicate devices. The Subject device, the Predicate devices, and the Reference devices have the same intended use, have similar technological characteristics, and are made of the same materials. The Subject device, the Predicate devices, and Reference devices encompass the same range of physical dimensions, and are to be sterilized using similar methods. The data included in this premarket notification demonstrate substantial equivalence to the Predicate devices listed above. Overall, the Subject device is substantially equivalent to the Predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.