NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment. The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
Device Story
NemoCast is a software system for orthodontic treatment planning and management. It processes 3D scanned dental models (from intra-oral scanners or STL/PLY/OBJ files) and DICOM CT image data. The software enables dental professionals to perform systematic inspection, 2D/3D measurements, and orthodontic analysis (e.g., arch shape, Bolton analysis, space analysis, overjet/overbite). It supports virtual treatment simulation, tooth movement planning, and virtual placement of orthodontic brackets. The device produces STL files for dental casts or aligner printing and designs for Indirect Bonding Transfer Trays. Used in clinical settings by trained dental professionals, the software assists in visualizing treatment progress and evaluating outcomes against planned objectives. It does not have patient contact.
Clinical Evidence
No clinical data. Bench testing only. The device underwent design verification and validation testing in accordance with a documented test plan. Performance testing was leveraged from the reference device (K193003).
Technological Characteristics
Software-based orthodontic planning tool; C++ language; Windows OS; supports STL, PLY, OBJ, and DICOM formats; provides 2D/3D measurement, segmentation, and virtual appliance design; standalone software application.
Indications for Use
Indicated for patients requiring orthodontic treatment. Used by dental professionals for management of orthodontic models, inspection, measurement, analysis, treatment simulation, and virtual appliance design.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
November 21, 2023
Software Nemotec S.L. % Kevin Walls Principal Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, Colorado 80127
Re: K232549
Trade/Device Name: NemoCast Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN, LLZ Dated: August 23, 2023 Received: August 23, 2023
Dear Kevin Walls:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K232549
Device Name
NemoCast
Indications for Use (Describe)
NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned dectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment.
The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
즈 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the letters in a dark blue color. Below the word "NEMOTEC" is the phrase "A BIOTECH DENTAL COMPANY" in a smaller, lighter blue font. To the right of the text is a circular graphic with a dark blue circle in the center and three light blue segments around it.
## 510(K) Summary - K232549
This summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Applicant Information:
Software Nemotec S.L.
Av. Juan Caramuel, №1
Leganés, Madrid, Spain, 28919
### Submission Correspondent:
Kevin Walls Principal Consultant, Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, Colorado 80127
#### Device Information:
| Trade Name: | NemoCast |
|----------------------------|-----------------------------|
| Common Name: | Orthodontic Software |
| Classification Name: | Orthodontic Plastic Bracket |
| Classification Regulation: | 21CFR 872.5470 |
| Device Class: | II |
| Product Code: | PNN, LLZ |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the "NEMO" in a darker blue and the "TEC" in a lighter blue. Below the name is the text "A BIOTECH DENTAL COMPANY" in a smaller, lighter font. To the right of the name is a circular graphic divided into three sections of varying shades of blue, with a dark blue circle in the center.
#### Reference Device:
NemoCast, Software Nemotec, K193003
#### Primary Predicate:
3Shape Ortho System™, 3Shape A/S, K152086
#### Date Prepared:
Aug 08, 2023
#### Predicate Device:
The NemoCast Software has the same intended uses and technical characteristics as primary predicate 3Shape Ortho System™ (K152086) from 3Shape A/S and reference predicate NemoCast from Software Nemotec (K193003) as listed in the table below:
Based on the information and supporting documentation provided, the NemoCast Software and the primary predicate (3Shape Ortho System™) have the same intended use.
Both software devices are used by Dental Professionals in orthodontic treatment planning (before, during, after treatment) covering management of patients and models, inspection, 2D and 3D measurement and orthodontic analysis of models, 2D & 3D treatment simulation, as well as virtual appliance preparation, handling and export, and they are both providing device output. Additionally, the indirect bonding functionality of both systems is intended for use with commercially available brackets and wires. Therefore, the NemoCast Software and the primary predicate (3Shape) are found to be similar in their intended use, supported anatomic areas and the majority of the available features and functionalities.
#### Device Description:
NemoCast is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analyzing, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes STL Models (also called dental casts) for thermoforming aligners, STL files for direct printing aligners and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). The device has no patient contact.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in a sans-serif font, with the first three letters in a darker blue and the last four letters in a lighter blue. Below the word "NEMOTEC" is the phrase "A BIOTECH DENTAL COMPANY" in a smaller, sans-serif font. To the right of the word "NEMOTEC" is a circular graphic divided into three sections, with a smaller circle in the center.
### Indications for Use:
NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment.
The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for NEMOTEC, a biotech dental company. The word "NEMOTEC" is written in large, bold, dark blue letters. Below the word "NEMOTEC" is the phrase "A BIOTECH DENTAL COMPANY" in smaller, light blue letters. To the right of the text is a circular graphic with three sections in different shades of blue and a dark blue circle in the center.
# Comparison of Intended Use and Technological Characteristics with the reference Device:
| Substantial Equivalence Table | | | | |
|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature name | SUBJECT<br>DEVICE<br>NemoCast | REFERENCE<br>DEVICE<br>NemoCast<br>(K193003) | PRIMARY<br>PREDICATE<br>3Shape Ortho<br>System ™,<br>K152086 | Differences |
| Product Code | PNN, LLZ | PNN, LLZ | PNN, LLZ | None |
| Common Name | Orthodontic<br>Software | Orthodontic<br>Software | Orthodontic<br>Software | None |
| Classification Name | Orthodontic<br>Plastic<br>Bracket | Orthodontic<br>Plastic Bracket | Orthodontic<br>Plastic Bracket | None |
| Regulation Number | 21 CFR<br>872.5470 | 21 CFR<br>872.5470 | 21 CFR<br>872.5470 | None |
| Supported anatomic areas | Maxilla and<br>Mandible | Maxilla and<br>Mandible | Maxilla and<br>Mandible | None |
| Intended Use | | | | |
| Use by dental professionals in<br>orthodontic treatment<br>planning (before, during, after<br>treatment). | Yes | Yes | Yes | Reference device<br>NemoCast was<br>used by dental<br>professionals<br>only before<br>treatment. |
| Management of patients and<br>models. | Yes | Yes | Yes | None |
| Inspection, measurement and<br>analysis of orthodontic<br>models. | Yes | Yes | Yes | None |
| Treatment simulation. | Yes | Yes | Yes | None |
| Virtual appliance preparation<br>(including dental casts),<br>handling and export. | Yes | Yes | Yes | None |
| Provide digital file and device<br>output. | Yes | Yes | Yes | All 3 devices<br>output STL files<br>for fabrication of<br>dental casts.<br>The subject<br>Device and<br>predicate Device |
| | | | | both additionally<br>output indirect<br>Bonding Transfer<br>Media. |
| Supported PC formats | Windows | Windows | Windows | None |
| Managing patient and case base data: | | | | |
| Creating, editing, deleting<br>and copying patient data | Yes | Yes | Yes | None |
| Creating, editing, deleting<br>and copying case data | Yes | Yes | Yes | None |
| Collection of study material | | | | |
| Surface scan from<br>intra-oral scanner | Yes | Yes | Yes | None |
| Surface scan from STL, PLY,<br>OBJ file | Yes | No | Yes, only STL | Although<br>predicate device<br>only import STL<br>format, the OBJ<br>and PLY formats<br>are other<br>equivalent<br>formats which do<br>not affect the<br>security and<br>safety of the<br>device. |
| CT image data | DICOM | DICOM | DICOM | None |
| 2D overlay | PNG, JPG,<br>BMP | PNG, JPG,<br>BMP | PNG, JPG,<br>BMP | None |
| Alignment of study material | | | | |
| Aligning surface scan and<br>CT image | Yes | Yes | Yes | None |
| Aligning cephalometric<br>images | Yes | Yes | Yes | None |
| Alignment of surface scan<br>with 2D overlays | Yes | Yes | Yes | None |
| Ability to check/adjust<br>DICOM visibility | Yes | Yes | Yes | None |
| DICOM scan segmentation | Yes | Yes | Yes | None |
| Occlusal Orientation | Yes | Yes | No | Although the<br>predicate device<br>does not perform<br>occlusal |
| | | | | orientation, the |
| | | | | reference device |
| | | | | contains this |
| Segmenting teeth roots | Yes | Yes | No | option.<br>Although the |
| | | | | predicate device |
| | | | | does not |
| | | | | segment teeth |
| | | | | roots, the |
| | | | | reference device |
| | | | | contains this |
| | | | | option. |
| DICOM orientation | Yes | Yes | Yes | None |
| Measuring study material | | | | |
| 2D measurement toolbox | Yes | Yes | Yes | None |
| 3D measurement toolbox | Yes | Yes | Yes | None |
| Analyzing study material | | | | |
| Arch shape | Yes | Yes | Yes | None |
| Wire length | Yes | Yes | Yes | None |
| Tooth width | Yes | Yes | Yes | None |
| Bolton | Yes | No | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.