Meta G UCLA Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
Device Story
Meta G UCLA Abutment is a dental abutment used to support prosthetic restorations; consists of abutment and screw. Used in dental clinics by dentists. Device is non-sterile; requires user steam sterilization before use. Hex-type used for single-unit restorations; Non-Hex-type for multi-unit restorations. No angulation permitted; not for use on angled implants. Post height must be 4-7mm. Provides structural support for crowns, bridges, or overdentures, enabling functional and aesthetic dental restoration.
Clinical Evidence
Bench testing only. Includes sterilization validation (ISO 17665-1/2), biocompatibility (ISO 10993-3, 5, 10, 11), and non-clinical MRI safety assessment based on scientific literature and displacement force/torque analysis.
Technological Characteristics
Materials: CCM (ASTM F1537) and POM (ASTM F1855); Abutment screw: Ti-6Al-4V. Connection types: Hex and Non-Hex. Dimensions: 4.5mm diameter, 1-3mm cuff heights. Energy source: None (mechanical). Sterilization: User steam sterilization. Connectivity: None.
Indications for Use
Indicated for use with dental implants to support prosthetic restorations (crowns, bridges, overdentures) in patients requiring dental reconstruction.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
INNO SLA Submerged Narrow Implant System (K231395)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
Cowellmedi Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave, Suite 160 Irvine. California 92612
Re: K232546
Trade/Device Name: Meta G UCLA Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 26, 2023 Received: August 24, 2023
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K232546
Device Name
Meta G UCLA Abutment
Indications for Use (Describe)
Meta G UCLA Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
(K232546)
This summary of 510(k) substantial equivalence and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date:
#### 1. Applicant / Submitter:
Cowellmedi Co., Ltd. 48, Hakgam-daero 221beon-gil, Sasang-gu, Busan, 46986, Republic of Korea Tel. + 82-51-314-2028
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine CA 92612 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### 3. Device:
| Proprietary Name: | Meta G UCLA Abutment |
|------------------------------|------------------------------------|
| Common Name: | Dental Abutment |
| Classification Name: | Endosseous Dental Implant Abutment |
| Classification: | Class II, 21 CFR 872.3630 |
| Classification Product Code: | NHA |
#### Predicate and Reference Devices: 4.
- Primary Predicate Device: UCLA CCM Abutment (K192263) by DIO Corporation
- · Reference Device: INNO SLA Submerged Implant System (K201323) by Cowellmedi Co., Ltd.
- · Reference Device: INNO SLA Submerged Implant System (K132242) by Cowellmedi Co., Ltd.
{5}------------------------------------------------
- · Reference Device: INNO SLA Submerged Narrow Implant System (K231395) by Cowellmedi Co., Ltd.
#### 5. Device Description:
The Meta G UCLA Abutment is used for prosthetic restoration. The Meta G UCLA Abutment consists of Meta G UCLA Abutment and abutment screws. The Meta G UCLA Abutment is made of CCM (ASTM F1537) and POM (ASTM F1855). The casting material to make prosthesis is also CCM. When cast a prosthesis with the Meta G UCLA Abutment, the post height above the transmucosal collar of the Meta G UCLA Abutment has to be taller than 4mm and maximum 7mm.
The Meta G UCLA Abutment has Hex, Non Hex connection. Hex-type abutment should be used for single unit restorations and is not recommended for multiple tooth restorations. Non Hex-type abutment is for multiunit restorations only. No angulation is to be incorporated into the cast final device and abutments should not be used on implants placed at an angle.
It is provided non-sterile, this should be user steam sterilized before use. The Meta G UCLA abutment is compatible with our own implant system.
- · INNO SLA Submerged Implant System(K132242) by Cowellmedi Co., Ltd.
- INNO SLA Submerged Narrow Implant System (K231395) by Cowellmedi Co., Ltd.
The design envelope for the Meta G UCLA Abutment is as follows.
<Design limits - Meta G UCLA Abutment> Minimum Diameter: 4.5 mm (0.18 in) Minimum Thickness : 0.8mm Maximum Total Cuff Height : 4mm Minimum Total Cuff Height : 1mm Maximum Cuff Height of Casting CCM part : 3mm Minimum Cuff Height of Casting CCM part : 0mm Post height: minimum of 4mm and maximum of 7mm
<Design limits - Meta G UCLA Abutment (Narrow)> Minimum Diameter: 4.5 mm (0.18 in) Minimum Thickness : 0.93mm Maximum Total Cuff Height : 4mm Minimum Total Cuff Height : 1mm Maximum Cuff Height of Casting CCM part : 3mm Minimum Cuff Height of Casting CCM part : 0mm Post height: minimum of 4mm and maximum of 7mm
#### 6. Indications for Use:
Meta G UCLA Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
{6}------------------------------------------------
#### 7. Performance Data (Non-Clinical):
The following tests were performed on the subject device or previously provided in a reference device, and the test results support that the subject device is substantially equivalent to the predicate devices.
- . Sterilization validating testing has been performed in accordance with ISO 17665-1 and ISO 17665-2 for steam sterilization.
- . Biocompatibility testing for the subject device has been performed in accordance with ISO 10993-3, 5, 10, 11.
- . Biocompatibility of the abutment screw (Ti-6A1-4V) was included in a previously cleared submission identified as reference device (K201323).
- . MR Environment Condition
Non-clinical worst-case MRI review was performed to evaluate the metallic Meta G UCLA abutment in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry 0., Jana G. Delfino, and Sunder Rajan. 'Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodied, dental abutments and, fixation screws) and material composition. Rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
#### 8. Substantial Equivalence
### 8.1. Comparison Chart
| | Subject Device | Predicate Device |
|-----------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| 510(K) No. | K232546 | K192263 |
| Applicant | Cowellmedi Co., Ltd. | DIO Corporation |
| Trade Name | Meta G UCLA Abutment | UCLA CCM Abutment |
| Classificatio<br>n Name | Endosseous Dental Implant Abutment<br>(872.3630) | Endosseous Dental Implant Abutment<br>(872.3630) |
| Product<br>Code | NHA | NHA |
| Class | II | II |
| Material | CCM & POM | CCM & POM |
| Indications<br>For Use/<br>Intended Use | Meta G UCLA Abutment is intended for<br>use with a dental implant to provide | UCLA CCM Abutment is intended for use<br>with a dental implant to provide support for |
{7}------------------------------------------------
| | Subject Device | Predicate Device |
|-------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| | support for prosthetic restorations such as<br>crowns, bridges, or overdentures. | prosthetic restorations such as crowns,<br>bridges, or overdentures. |
| Design<br>Limits | Image: Subject Device | Image: Predicate Device |
| Diameters<br>(mm) | 4.5 | 4.0 / 4.5 |
| Cuff (mm) | 1, 2, 3 mm | 1, 3 mm |
| Angulation | 0° | 0° |
| Sterile | Non-sterile | Non-sterile |
### 8.2. Substantial Equivalence Discussion
Meta G UCLA Abutment is substantially equivalent in indications for use, material, dimension, sterilization and similar design, technological characteristics as primary predicate device (K192263). The only difference is in design, but it is very minor and does not raise new questions. We identified refence devices (K13242 & K231395) which is used with the subject device, and another device (K201323) to leverage biocompatibility testing.
Based on the information provided herein, we determine that the subject device is substantially equivalent to the predicate device.
#### 9. Conclusion:
Based on the testing result and a comparison of technological characteristics, Cowellmedi Co., Ltd. concludes that the Meta G UCLA Abutment is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.