The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Device Story
VenusX is a medical linear accelerator delivering therapeutic radiation for CRT/IMRT treatment techniques. It utilizes a 6 MV photon therapeutic X-ray beam and a kV X-ray beam for diagnostic imaging. The system is installed in a radiation-shielded vault with a control console located outside the treatment room. It integrates a Cone Beam CT (CBCT) imager for patient positioning and treatment field verification relative to anatomical or fiducial landmarks. The device is operated by healthcare professionals to deliver radiation according to a physician's prescription. It interfaces with third-party Radiation Treatment Planning Systems (TPS) and Oncology Information Systems (OIS) using DICOM RT/3.0 standards. By providing precise radiation delivery and image-guided verification, the device assists in the treatment of lesions and tumors, potentially improving clinical outcomes through accurate dose distribution.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1-2), performance verification (IEC 60976, IEC 61217), cybersecurity assessment, and biocompatibility (ISO 10993-5, ISO 10993-10).
Indicated for patients requiring precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
ARIA Radiation Therapy Management System (K173838)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 12, 2024
LinaTech LLC Jonathan Yao President 1294 Kifer Road #705 Sunnyvale, California 94086
Re: K232489
Trade/Device Name: VenusX Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: March 12, 2024 Received: March 12, 2024
Dear Jonathan Yao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.G.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K232489
Device Name VenusX
Indications for Use (Describe)
The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| | Department of Health and Human Services |
| | Food and Drug Administration |
| | Office of Chief Information Officer |
| | Paperwork Reduction Act (PRA) Staff |
| | <a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a> |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
#### 510(k) SUMMARY - K232489
As required by 21 CFR 807.92
#### SUBMITTER l.
| Submitter's Name: | LinaTech<br>1294 Kifer Road #705<br>Sunnyvale, CA 94086, USA |
|----------------------------|--------------------------------------------------------------------------------------------------|
| Primary Contact<br>Person: | Jonathan Yao<br>Phone: (408) 733 2051<br>Fax: (408) 733 2045<br>E-mail: Jonathanyao@LinaTech.com |
| Date Prepared: | 11 April 2024 |
#### II. DEVICE
lll.
| Proprietary Name: | VenusX |
|----------------------|---------------------------------------------------|
| Common/Usual Name: | Medical Linear Accelerator |
| Classification Name: | Medical charged-particle radiation therapy system |
| Regulation: | 21 CFR §892.5050 |
| Regulatory Class: | Class II |
| Product Code: | IYE |
| PREDICATE DEVICE | |
| Predicate Devices: | Varian UNIQUE (K101751) |
| Reference Device: | Varian On Board Imaging (OBI) (K042720) |
#### IV. DEVICE DESCRIPTION
Device Description: The VenusX Radiotherapy System is a medical linear accelerator that delivers therapeutic radiation to patient in accordance with the physician's prescription. It supports CRT /IMRT Treatment Techniques (Mode). The system consists of a photon therapeutic 6 MV X-ray radiation beam producing component with a photon diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located
outside the treatment room.
{4}------------------------------------------------
# V. INTENDED USE AND INDICATIONS FOR USE
| Intended Use: | The intended use is the same with precision radiotherapy as the predicate device. VenusX does not support stereotactic radiosurgery.<br>The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The VenusX radiotherapy delivery system is intended to provide precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
# VI. TECHNOLOGICAL CHARACTERISTICS
| Summary<br>Of Technological<br>Characteristics: | Both the VenusX and the predicate device contain similar<br>technological characteristics and functional scientific technology<br>to deliver radiation therapy by authorized medical practitioners.<br>A subset of technological characteristics and features of the<br>current device is different from the predicate. The<br>biocompatibility of patient-contacting components remains the<br>same as the predicate device. The results of the verification,<br>validation and safety standards testing demonstrate that there<br>are no changes to the safety profile of the device. |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
The feature comparison chart below notes the similarities and differences between predicate, reference, and subject device.
{5}------------------------------------------------
The feature comparison with predicate device:
| Attribute/ Feature | Predicate Device<br>Varian UNIQUE<br>K101751 | Subject Device<br>LinaTech VenusX |
|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The UNIQUE is intended to<br>provide stereotactic<br>radiosurgery and precision<br>radiotherapy for lesions,<br>tumors and conditions<br>anywhere in the body when<br>radiation treatment is<br>indicated. | The VenusX radiotherapy<br>delivery system is intended<br>to provide precision<br>radiotherapy for lesions,<br>tumors, and conditions<br>anywhere in the body where<br>radiation treatment is<br>indicated. |
| Indications for Use | The UNIQUE is indicated for<br>stereotactic radiosurgery and<br>precision radiotherapy for<br>lesions, tumors and<br>conditions anywhere in the<br>body when radiation<br>treatment is indicated. | The VenusX radiotherapy<br>delivery system is intended<br>to provide precision<br>radiotherapy for lesions,<br>tumors, and conditions<br>anywhere in the body where<br>radiation treatment is<br>indicated. |
| Photon Energy (MV)<br>available | 6 | 6x (FFF) |
| Dose Rates | 100 - 600 cGy/min<br>for FF mode | 600、800、1000 cGy/min<br>for FFF mode |
| SAD (cm) | 100 | 90 |
| Maximum standard<br>treatment field size | 40cmx40cm | 40cmx40cm |
| Number of MLC leaves | 120 | 101 leaves in each of two<br>orthogonal layers |
| Maximum leaf over<br>travel distance (cm) | 14 | Upper layer 20<br>Lower layer 14 |
| Average leaf<br>transmission (%) | ≤ 2.0 | ≤ 0.05 |
| Leaf end physical<br>position accuracy (cm) | ≤ ± 0.1 | ≤ ± 0.1 |
| Attribute/ Feature | Predicate Device<br>Varian UNIQUE<br>K101751 | Subject Device<br>LinaTech VenusX |
| Maximum leaf speed<br>(cm/sec) is greater than<br>2.5 | Yes | Yes |
| Electron<br>Treatment<br>Mode | No | No |
| Non-arc photon<br>treatment mode<br>programmable MU<br>maximum dose is larger<br>than 1999 MU | Yes | Yes |
| Patient support surface:<br>couch pedestal with<br>carbon fiber couch top<br>for improved MV<br>imaging | Yes | Yes |
| Microwave Source is<br>Magnetron | Yes | Yes |
| Interface for External<br>System Gating | Yes | Yes |
| MV Detector size | 30cm x 40cm | 43cm x 43cm |
| MV Pixel | 1024 x 768 | 2816×2816 |
| Radiation dose for MV<br>Image is less than 1MU | Yes | Yes |
| Image registration with<br>planning image | Yes | Yes |
| Basis of image<br>registration are for<br>minimum of | soft tissue, bony anatomy,<br>fiducial markers, digital<br>representation of treatment<br>apertures. | soft tissue, bony anatomy,<br>fiducial markers, digital<br>representation of treatment<br>apertures. |
| Data Interface is using<br>DICOM RT/3.0 data and<br>with image | Yes | Yes |
| Attribute/ Feature | Predicate Device<br>Varian UNIQUE<br>K101751 | Subject Device<br>LinaTech VenusX |
| Record treatment<br>delivery results | Yes | Yes |
| Treatment Techniques<br>(Mode) | CRT, IMRT | CRT, IMRT |
| Compatible TPS and OIS<br>for the VenusX and<br>compare with predicate. | UNIQUE is compatible with a<br>commercially available<br>independent third-party<br>Radiation Treatment<br>Planning System (TPS), Such<br>as Eclips TPS, and also to<br>compatible with the ARIA<br>Radiation Therapy<br>Management System (FDA<br>No.: K173838) (OIS). | VenusX is compatible with a<br>commercially available<br>independent third-party<br>compatible with TiGRT TPS<br>(FDA No.: K090893)<br>( TPS ) and also ARIA<br>(FDA No.: K173838) (OIS)<br>system. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The feature comparison with reference device:
| Attribute / Feature | Reference Device<br>Varian OBI K042720 | Subject Device<br>LinaTech VenusX |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Varian On-Board<br>Imager device (OBI) is used<br>for verification of correct<br>patient position in relation<br>to isocenter and<br>verification of the<br>treatment fields in relation<br>to anatomical and/or<br>fiducial landmarks. | The VenusX's Cone Beam CT<br>Imager device (CBCT) is<br>used for verification of<br>correct patient position in<br>relation to isocenter and<br>verification of the<br>treatment fields in relation<br>to anatomical and/or<br>fiducial landmarks. |
| kV imaging | Yes | Yes |
| Cone-beam CT image<br>Reconstruction | Yes | Yes |
| General type of<br>Reconstruction<br>method/algorithm | FDK (Feldkamp-Davis-<br>Kress) algorithm | FDK (Feldkamp-Davis-Kress)<br>algorithm |
| Dose per CBCT<br>acquisition | ≤ 1.4cGy | ≤ 2cGy |
| CBCT acquisition mode<br>(pixels / degrees) | 512×512<br>reconstruction<br>matrix / 200°<br>(head) or<br>360° (body) | 512×512<br>reconstruction<br>matrix / 360° |
{9}------------------------------------------------
## VII. SUMMARY OF PERFORMANCE TESTING
The following performance data were provided in support of the substantial equivalence determination.
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electromagnetic compatibility (EMC) testing was conducted on the subject device. The system complies with the following standards:
- IEC 60601-1-2:2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
### Bench Test
The VenusX was thoroughly tested on bench to evaluate and verify that it meets the required performance specifications. The bench testing plan was developed based on the risk assessment of the device, and the recommendations outlined in the applicable FDA guidance documents, ISO and IEC standards. VenusX conforms to the following standards:
- ANSI AAMIES/IEC60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical eguipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-3:2008+AMD1:2013+AMD2:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-1:2020 Medical electrical equipment - Part 2-1: Requirements for the basic safety and essential performance of electron accelerators in the range 1 MeV to 50 MeV
- IEC 60601-2-68:2014 Medical electrical equipment - Part 2-68: Requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam therapy equipment
- IEC 62274:2005 Medical electrical equipment— Safety of radiotherapy record and verify systems
- IEC 62366-1:2015+AMD1:2020 Medical devices - Part 1: Application of usability engineering to medical devices
- IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60976:2007 Medical electrical equipment - Medical electron accelerators - Functional performance characteristics
- IEC 61217:2011 Radiotherapy equipment - Coordinates, movements, and scales
Test results met all the pre-determined acceptance criteria.
{10}------------------------------------------------
## Cybersecurity
In addition to the bench testing, the VenusX was also thoroughly tested to verify that its cybersecurity objectives are fully met, and security risk control measures are effective. Specifically, the following cybersecurity testing was conducted:
— Security requirement testing that demonstrates requirement was implemented correctly.
— Threat mitigation testing demonstrating effective risk control measures according to the threat modeling.
— A closed box testing of known vulnerability scanning by qualified external testers, with reference to Section 9.4 of "IEC 62443-4-1 Security for industrial automation and control systems Part 4-1: Secure product development lifecycle requirements ".
- Penetration testing by qualified external testers.
The results of the cybersecurity testing showed that VenusX meets the cybersecurity requirements under Section524B(b) of FD&C Act.
### Hardware and Software Verification and Validation Testing
Hardware and software verification and validation process were conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards.
Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions," issued June 2023. The software for this device is considered as a "enhanced" level of concern.
### Biocompatibility Testing
Biocompatibility testing was conducted on the Carbon fiber couch top in accordance with ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The Biocompatibility testing conducted for the VenusX included the following:
- ISO 10993-5 Cytotoxicity Test
- ISO 10993-10 Skin Sensitization Test
- ISO 10993-10 Skin Irritation Test
Test results met all the pre-determined acceptance criteria.
#### CONCLUSION OF NON-CLINICAL TESTING VIII.
The results of testing showed that VenusX conforms to the defined user needs and intended uses and that there were no software DRs (Discrepancy Reports) remaining which had a priority of Safety Intolerable or Customer Intolerable.
{11}------------------------------------------------
## IX. CONCLUSION FOR SUBSTANTIAL EQUIVALENCE TO THE PREDICATE AND REFERENCE DEVICE
The VenusX is substantially equivalent to the UNIQUE (K101751) (predicate device). The intended use is the same with precision radiotherapy as the predicate device. VenusX does not support stereotactic radiosurgery. A subset of technological characteristics and features of the current device is different to the predicate. However, the Verification and Validation demonstrate that the device is as safe and effective as the predicate and reference device. Based on the comparison and analysis above, the VenusX (subject device) is substantially equivalent to the UNIQUE (K101751) (predicate device).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.