ProVecta 3D Prime and ProVecta 3D Prime Ceph are computed tomography x-ray units intended to generate 3D, panoramic and cephalometric (ProVecta 3D Prime Ceph Model) X-ray images in dental radiography for adult and pediatric patients. They provide diagnostic details of the maxillofacial areas for a dental treatment. The devices are operated and used by physicians, dentists, and x-ray technicians. Not intended for mammography use.
Device Story
Cone beam CT (CBCT) dental imaging system; X-ray tube and sensor rotate 180°-540° around patient; acquires series of 2D images; reconstruction computer calculates 3D volume graphic (voxels) for sectional tomograms and 3D views. Used in dental clinics by physicians, dentists, or technicians. Provides diagnostic maxillofacial images for treatment planning. Includes bite block, chin rest, and headrest for patient positioning. Subject device update includes new biocompatible bite block materials (standard and comfort variants).
Clinical Evidence
No clinical data required. Bench testing only: biocompatibility evaluation (cytotoxicity, chemical analysis) of new bite block materials per EN ISO 10993 standards.
Technological Characteristics
CBCT system; 60-99 kV tube voltage; 4-16 mA tube current; 0.5 mm focal spot; 2.5 mm Al filtration. Imaging receptor: Xmaru1404CF (CT/Panoramic), Xmaru 2602CF (Cephalometric). Connectivity: DICOM 3.0 compatible. Software: VisionX 3.0. Biocompatible bite block materials.
Indications for Use
Indicated for adult and pediatric patients requiring 3D, panoramic, or cephalometric X-ray imaging of maxillofacial areas for dental treatment. Not for mammography.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 14, 2023
Durr Dental SE % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K232487
Trade/Device Name: Provecta 3D Prime and Provecta 3D Prime Ceph Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: August 10, 2023 Received: August 17, 2023
## Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232487
Device Name ProVecta 3D Prime and ProVecta 3D Prime Ceph
### Indications for Use (Describe)
Pro Vecta 3D Prime and ProVecta 3D Prime Ceph are computed tomography x-ray units intended to generate 3D, panoramic and cephalometric (ProVecta 3D Prime Ceph Model) X-ray images in dental radiography for adult and pediatric patients. They provide diagnostic details of the maxillofacial areas for a dental treatment. The devices are operated and used by physicians, dentists, and x-ray technicians. Not intended for mammography use.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary, DÜRR DENTAL SE,
# ProVecta 3D Prime and ProVecta 3D Prime Ceph, K232487
This 510(k) is being submitted in accordance with the requirements of 21 CFR §807.92. Date Summary Prepared: September 14, 2023
#### Submitter's Identification: (Below) 1.
| Submitter's Identification: | DÜRR DENTAL SE<br>Höpfigheimer Str. 17<br>74321 Bietigheim-Bissingen<br>Germany<br>Phone: + 49 (0) 7142 70 5-0<br>Fax: + 49 (0) 7142 705-500<br>E-Mail: info@duerr.de<br>www.duerrdental.com | Establishment Registration<br>Name in FURLS:<br>Duerr DENTAL SE |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Establishment Registration Number: | 3015509619 | |
| Submitter's Contact: | Mr. Oliver Lange<br>Director of Quality Management<br>DÜRR DENTAL SE<br>Höpfigheimer Str. 17<br>74321 Bietigheim-Bissingen, Germany<br>Phone: + 49 (0) 7142 70 5-190<br>Email: oliver.lange@duerrdental.com | |
| U.S. Agent & Contact: | Mr. Joseph Latkowski<br>Director of Quality and Regulatory<br>Air Techniques, Inc.<br>1295 Walt Whitman Road<br>Melville, NY 11747, USA<br>U.S. Phone: 516-214-5574<br>E-Mail: Joseph.Latkowski@airtechniques.com | |
#### 2. Device:
| Trade /Proprietary Name: | ProVecta 3D Prime and ProVecta 3D Prime Ceph |
|-------------------------------|----------------------------------------------|
| Device: | X-Ray, Tomography, Computed, Dental |
| Regulation Description: | Computed tomography x-ray system |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | OAS |
| Regulation Number: | 892.1750 |
| Device Class: | 2 |
{4}------------------------------------------------
#### 3. Predicate Device:
| 510(k) Number: | K193139 |
|-------------------------------|-------------------------------------|
| Manufacturer: | DÜRR DENTAL SE |
| Trade /Proprietary Name: | ProVecta 3D Prime Ceph |
| Device: | X-Ray, Tomography, Computed, Dental |
| Regulation Description: | Computed tomography x-ray system |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | OAS |
| Regulation Number: | 892.1750 |
| Device Class: | 2 |
- 4. Reference Device: (Software used with the proposed device)
| 510(k) Number: | K213326 |
|-------------------------------|------------------------------------------------|
| Manufacturer: | DÜRR DENTAL SE |
| Trade /Proprietary Name: | VisionX 3.0 |
| Device: | Imaging Software |
| Regulation Description: | Medical image management and processing system |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | LLZ |
| Regulation Number: | 892.2050 |
| Device Class: | 2 |
#### 5. Device Description:
This device is a cone beam CT x-ray device for the acquisition of dental images. Similar to computer tomography or magnetic resonance tomography, sectional images can be generated with CBCT. With CBCT, an X-ray tube and an imaging sensor opposite it rotate around a seated or standing patient. The X-ray tube rotates through 180°-540° and emits a conical X-ray beam. The X-rays pass through the region under investigation and are measured for image generation by a detector as an attenuated grey scale X-ray image. Here, a large series of two-dimensional images is acquired during the revolution of the X-ray tube. Using a mathematical calculation on the rotating image series via a reconstruction computer, a grey value coordinate image is generated in the three spatial dimensions. This three-dimensional coordinate model corresponds to a volume graphic that is made up of individual voxels. This volume can be used to generate sectional images (tomograms) in all spatial dimensions as well as 3D views. The system complies with US Radiation Safety Performance Standard. The ProVecta 3D Prime model does not have the CEPH function.
This premarket notification is because of the biological evaluation of medical devices documentation according to EN ISO 10993-1:2020. The revision of the document is the inclusion of the Comfort Bite Foam for the bite block with direct patient contact. The relevant documents regarding biological safety were included and evaluated in this biological evaluation. Furthermore, the Biological Evaluation has been updated to the latest Version of the standard. The name, application and biocompatibility-relevant materials of the product have not changed
{5}------------------------------------------------
since the last version. In addition to the current bite block (REF: 2210200100), two new, more comfortable variants were developed:
- standard bite block: an optimized version of the existing bite block
- comfort bite block: an extension of the existing bite block
The image management software was recently updated in K213326.
#### 6. Indications for use:
ProVecta 3D Prime and ProVecta 3D Prime Ceph are computed tomography x-ray units intended to generate 3D, panoramic and cephalometric (ProVecta 3D Prime Ceph Model) X-ray images in dental radiography for adult and pediatric patients. They provide diagnostic details of the maxillofacial areas for a dental treatment. The devices are operated and used by physicians, dentists, and x-ray technicians. Not intended for mammography use.
#### 7. Summary of the technological characteristics of the device compared to the predicate device:
| Descriptive Information | ProVecta 3D Prime Ceph<br>K193139 | ProVecta 3D Prime | ProVecta 3D Prime Ceph | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Indications for Use | ProVecta 3D Prime Ceph is a<br>computed tomography x-ray<br>unit intended to generate 3D,<br>panoramic and cephalometric<br>X-ray images in dental<br>radiography for adult and<br>pediatric patients. It provides<br>diagnostic details of the<br>maxillofacial areas for a dental<br>treatment. The device is<br>operated and used by<br>physicians, dentists, and x-ray<br>technicians. Not intended for<br>mammography use. | ProVecta 3D Prime and ProVecta 3D Prime Ceph are computed<br>tomography x-ray units intended to generate 3D, panoramic anc<br>cephalometric (ProVecta 3D Prime Ceph Model) X-ray images in<br>dental radiography for adult and pediatric patients. They<br>provide diagnostic details of the maxillofacial areas for a dental<br>treatment. The devices are operated and used by physicians,<br>dentists, and x-ray technicians. Not intended for<br>mammography use. | ProVecta 3D Prime and ProVecta 3D Prime Ceph are computed<br>tomography x-ray units intended to generate 3D, panoramic anc<br>cephalometric (ProVecta 3D Prime Ceph Model) X-ray images in<br>dental radiography for adult and pediatric patients. They<br>provide diagnostic details of the maxillofacial areas for a dental<br>treatment. The devices are operated and used by physicians,<br>dentists, and x-ray technicians. Not intended for<br>mammography use. | |
| Image Acquisition Modes | Panoramic, cephalometric,<br>and computed tomography | SAME except without<br>cephalometric | SAME as predicate | |
| Imaging Software | VisionX 3, includes 2D and 3D.<br>Cleared in K192743 | VisionX 3.0, includes 2D and 3D,<br>updated in K213326 | VisionX 3.0, includes 2D and<br>3D, updated in K213326 | |
| Input Voltage | AC 200-240V | SAME | SAME | |
| Tube Voltage | 60-99 kV | SAME | SAME | |
| Tube Current | 4~16mA | SAME | SAME | |
| Focal Spot Size | 0,5 mm | SAME | SAME | |
| Exposure Time | Max. 20 s | SAME | SAME | |
| Slice Width | 0.1 mm min. | SAME | SAME | |
| Total Filtration | 2.5 mm Al | SAME | SAME | |
| Chin Rest | Bite block, chin rest and<br>headrest | SAME | SAME | |
| Mechanical | Compact design | SAME | SAME | |
| Descriptive Information | | ProVecta 3D Prime Ceph<br>K193139 | ProVecta 3D Prime | ProVecta 3D Prime Ceph |
| Electrical | | LDCP logic circuit (Low<br>Dark Current<br>Processing) | SAME | SAME |
| Software | | VistaSoft, DICOM 3.0<br>compatible | SAME | SAME |
| 2D Image Viewing<br>Program | | VisionX | SAME | SAME |
| 3D Image Viewing<br>Program | | VisionX | SAME | SAME |
| Anatomical Sites | | Maxillofacial | SAME | SAME |
| Image<br>Receptor | Computed<br>Tomography | Xmaru1404CF | SAME | SAME |
| | Panoramic | Xmaru1404CF | SAME | SAME |
| | Cephalometric | Xmaru 2602CF | N/A | SAME |
| Size of Imaging<br>Volume (cm) | | Xmaru1404CF: Max.<br>10x8.5 | SAME | SAME |
| Pixel<br>Resolution | Computed<br>Tomography | 2.5 lp/mm - 4x4 binning<br>Does not support:<br>5.0 lp/mm - 2x2 binning | SAME | SAME |
| | Panoramic | 2.5 lp/mm - 4x4 binning<br>Does not support:<br>5.0 lp/mm - 2x2 binning | SAME | SAME |
| | Cephalometric | Xmaru 2602CF | N/A | SAME |
| Pixel Size<br>Computed | Computed<br>Tomography | Xmaru1404CF :<br>99 µm - 2x2 binning<br>198 µm- 4x4 binning | SAME | SAME |
| | Panoramic | Xmaru1404CF :<br>99 µm - 2x2 binning<br>198 µm- 4x4 binning | SAME | SAME |
| | Cephalometric | Xmaru 2602CF<br>200 x 200 µm | N/A | SAME as Predicate |
| Descriptive Information | ProVecta 3D Prime Ceph<br>K193139 | ProVecta 3D Prime | ProVecta 3D Prime Ceph | |
| Photograph | Image: ProVecta 3D Prime Ceph K193139 | Image: ProVecta 3D Prime<br>Without CEPH Arm. | Image: ProVecta 3D Prime Ceph<br>Identical appearance, no<br>change. | |
{6}------------------------------------------------
{7}------------------------------------------------
- 8. Discussion of Similarities and Differences: This premarket notification is specifically for a change in the bite-block material. Because of the change in material, biocompatibility issues had to be addressed.
- 9. Non-Clinical Data and Performance Testing (New Bite Foam) Cytotoxicity, L 929-Proliferation, EN ISO 10993-1, -5, -12. Result: Not cytotoxic.
The following new documents have been added to our technical file: Test Report Chemical Analysis Comfort Bite Foam Test Report Cytotoxicity Comfort Bite Foam Test Report Cytotoxicity Packaging Comfort Bite Foam Test Report Bite Block Manufacturer Report Cytotoxicity-Datasheet, O.C.-PE RG 45 KB-D-No 5607267 Safety Information
- 10. Clinical Data Clinical data is not required for a finding of substantial equivalence.
- 11. Conclusion: In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate device in terms of technology, performance and indications for use, Dürr Dental SE concludes that the ProVecta 3D Prime Ceph is substantially equivalent to the predicate device as described herein. The differences between the new device and the predicate device shown in the comparison table above do not raise any new questions about safety and effectiveness and so we consider it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.