← Product Code [FBK](/productcode/FBK) · K232442

# Disposable Endoscope Injection Needles (K232442)

_Zhejiang Soudon Medical Technology Co., Ltd. · FBK · Nov 9, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232442

## Device Facts

- **Applicant:** Zhejiang Soudon Medical Technology Co., Ltd.
- **Product Code:** [FBK](/productcode/FBK.md)
- **Decision Date:** Nov 9, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Disposable Endoscope Injection Needles are used with endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

## Device Story

Disposable Endoscope Injection Needles are manual surgical instruments used during endoscopic procedures. The device consists of a sliding handle, handle, cap, sheath, inner tube, guide tube, connecting tube, and tip. It is inserted through the working channel of an endoscope to deliver therapeutic agents via injection into the GI tract vasculature or submucosa. The device is operated by a clinician (physician) in a clinical setting. It functions as a mechanical delivery system for fluids; it does not involve electronic processing, AI, or software. The device benefits patients by enabling targeted local delivery of therapeutic substances during endoscopic interventions.

## Clinical Evidence

No clinical data. Bench testing only, including appearance and physical properties, demonstrated compliance with ISO 10993 (biocompatibility) and ISO 11135-1/10993-7 (EO sterilization).

## Technological Characteristics

Manual endoscopic injection needle. Components: sliding handle, handle, cap, sheath, inner tube, guide tube, connecting tube, tip. Dimensions: 2.3mm outer diameter; 21G/23G/25G needle gauges; 1800mm/2300mm working lengths. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Biocompatibility: ISO 10993 compliant.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Disposable Sclerotherapy Needle ([K190032](/device/K190032.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

November 9, 2023

Zhejiang Soudon Medical Technology Co.,Ltd % Nick Wang RA Specialist Shanghai Vanhe Consulting Co., Ltd 2F, Building No.1, 3938 Huqingping Road, Qingpu District Shanghai, 201703 China

Re: K232442

Trade/Device Name: Disposable Endoscope Injection Needles Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBK Dated: August 11, 2023 Received: August 14, 2023

Dear Nick Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232442

Device Name Disposable Endoscope Injection Needles

Indications for Use (Describe) Disposable Endoscope Injection Needles are used with endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for Soudan Medical. The logo is blue and features a stylized design with the words "soudan medical" written in a smaller font below the main design. The main design consists of two vertical bars connected by a horizontal bar, with a flame-like symbol above it.

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

## 1. Submitter

| Submitted by:      | Zhejiang Soudon Medical Technology Co.,Ltd<br>Address:Room 302-1,Floor 3, Building 4, No.1 Nangonghe<br>Road, Donghu Street, Linping District, Hangzhou 311100<br>Zhejiang, China                                           |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact<br>Person: | Nick Wang<br>RA Specialist<br><br>Shanghai Vanhe Consulting Co., Ltd<br>Address: 2F, Building No.1, 3938 Huqingping Road, Qingpu<br>District, Shanghai, China.<br>Phone: 0086-13585860297<br>Email: vanheconsulting@126.com |
| Date<br>Prepared:  | August 11, 2023                                                                                                                                                                                                             |

# Device

| Device Name:         | Disposable Endoscope Injection Needles                |
|----------------------|-------------------------------------------------------|
| Classification Name: | Endoscopic Injection Needle, Gastroenterology-Urology |
| Regulatory Class:    | II                                                    |
| Regulation Number:   | 21 CFR 876.1500                                       |
| Regulation Name:     | Endoscope and Accessories                             |
| Product Code:        | FBK                                                   |

#### Predicate Device

| Device Name:         | Disposable Sclerotherapy Needle, K190032                 |
|----------------------|----------------------------------------------------------|
| Manufacturer:        | Hangzhou AGS MedTech CO., Ltd                            |
| Classification Name: | Endoscopic Injection Needle,<br>Gastroenterology-Urology |
| Regulatory Class:    | II                                                       |
| Regulation Number:   | 21 CFR 876.1500                                          |
| Regulation Name:     | Endoscope and Accessories                                |
| Product Code:        | FBK                                                      |

# 2. Device Description

The Disposable Endoscope Injection Needles consist of a Sliding handle, Handle, Cap,

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for Soudon Medical. The logo is primarily blue and features a stylized design. The top part of the logo includes a symbol that resembles a flame or a stylized bird in flight. Below this symbol, there is a geometric shape that looks like two pillars connected by a horizontal bar. Underneath the geometric shape, the words "soudon medical" are written in a smaller, sans-serif font.

Sheath, Inner Tube, Guide Tube, Connecting tube, and Tip (only Type B). It is available in a variety of configurations with varying needle lengths, gauges and working lengths. Type A and Type B are included in this submission.

# 3. Indication for Use:

Disposable Endoscope Injection Needles are used with endoscope to perform endoscopic vascular or submucosal injection in the GI tract.

# 4. Comparison of Technological Characteristics

The Disposable Endoscope Injection Needles has substantially equivalent device design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Hangzhou AGS MedTech CO., Ltd 's Disposable Sclerotherapy Needle, K190032. The differences between the proposed device and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below.

| Item                    | Disposable Endoscope<br>Injection Needles(Proposed<br>Device)                                                                                         | Disposable<br>Sclerotherapy Needle,<br>K190032                                                                         | Discussion |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for<br>Use   | Disposable Endoscope<br>Injection Needles are used<br>with endoscope to perform<br>endoscopic vascular or<br>submucosal injection in the<br>GI tract. | The Disposable<br>Sclerotherapy Needle is<br>intended for endoscopic<br>injection into the<br>gastrointestinal mucosa. | Same       |
| Outer Diameter          | 2.3mm                                                                                                                                                 | 2.4mm                                                                                                                  | Same       |
| Needle Gauge            | 21G, 23G, 25G                                                                                                                                         | 21G, 22G, 23G, 24G,<br>25G                                                                                             | Similar    |
| Working Length          | 1800mm, 2300mm                                                                                                                                        | 1600mm, 2000mm,<br>2300mm                                                                                              | Similar    |
| SAL                     | 10-6                                                                                                                                                  | 10-6                                                                                                                   | Same       |
| Biocompatibility        | Comply with ISO 10993-1                                                                                                                               | Comply with ISO 10993-1                                                                                                |            |
| Sterilization<br>Method | EO Sterilization                                                                                                                                      | EO Sterilization                                                                                                       | Same       |

# 5. Non-clinical Performance Data

The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing", ISO 11135-1 "Sterilization of Health Care products Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image features a logo for "Soudon Medical". The logo is predominantly blue and consists of two distinct parts. The upper part of the logo contains a stylized image that resembles a flame or a stylized letter 'S'. The lower part of the logo features the text "soudon medical" in a smaller, sans-serif font, positioned beneath a larger, more abstract design that could be interpreted as stylized letters.

ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals" .

The following bench tests were performed on Disposable Endoscope Injection Needles: Appearance, Physical properties. The results of all testing were passing.

#### 6. Clinical Test Data

No Clinical Study is included in this submission.

#### 7. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Zhejiang Soudon Medical Technology Co.,Ltd has demonstrated that proposed device Disposable Endoscope Injection Needles is substantially equivalent to Hangzhou AGS MedTech CO., Ltd 's currently marketed Disposable Sclerotherapy Needle, K190032.

---

**Source:** [https://fda-staging.innolitics.com/device/K232442](https://fda-staging.innolitics.com/device/K232442)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com