The ENDOFIX EXO is indicated for use in interventions where a rigid endoscope is indicated for use. The device is intended to hold and position a rigid endoscope / exoscope during general and neurological diagnostic and therapeutic procedures.
Device Story
ENDOFIX EXO is a manually operated, table-mounted surgical arm for holding and positioning rigid endoscopes/exoscopes. Device features 6 axes of movement with electromagnetic brakes. Surgeon operates device by pressing a sterilizable CONTROL button to unlock all axes for manual positioning; releasing button locks axes. Integrated sensor in CONTROL detects magnetic bracket on endoscope clamp to enable operation, ensuring safety. Device is covered with sterile drape during use. Provides stable, rocksteady field of vision for one-handed or two-handed surgery. Quick-release fastener allows rapid removal for emergency open surgery. Used in OR by surgeons. Benefits include stable visualization and reduced surgeon fatigue. Powered by 100-240 VAC; no internal battery. Includes counterbalance for weight management.
Clinical Evidence
No clinical data. Bench testing performed: lifetime testing (device and control), displacement after fixation testing, package drop testing, and usability testing with 15 American physicians. Electrical safety verified per IEC 60601-1:2005 + A1:2012.
Technological Characteristics
Electromagnetic brake system; 6-axis manual positioning; 100-240 VAC, 50-60Hz power input; 24V DC output. Table-mounted via quick-acting clamp. Sterilizable control unit with magnetic sensor. No software, biologics, or coatings. Tested per IEC 60601-1.
Indications for Use
Indicated for use in general and neurological diagnostic and therapeutic procedures requiring a rigid endoscope or exoscope. Suitable for all patients within the scope of indications for use. Intended for use by trained surgeons in professional healthcare environments.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
WalterLorenz Surgical Assist Arm Scope Holder (K190576)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 25, 2024
AKTORmed GmbH Hanna Kafurke Regulatory Affairs/Quality Management Neugablonzer Strasse 13 Neutraubling, BY 93073 Germany
Re: K232405
Trade/Device Name: Endofix Exo Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCV Dated: August 1, 2023 Received: August 10, 2023
#### Dear Hanna Kafurke:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.04.25 16:18:53
> -04'00' Mark Trumbore Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K232405
Device Name
ENDOFIX EXO (171943)
#### Indications for Use (Describe)
The ENDOFIX EXO is indicated for use in interventions where a rigid endoscope is indicated for use. The device is intended to hold and position a rigid endoscope / exoscope during general and neurological diagnostic and therapeutic procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" in white, followed by the word "AKTORmed" in gray, and the words "ROBOTIC SURGERY" in a smaller font below. The letters "TM" are in the upper right corner of the word "AKTORmed".
Dokumentennummer / Document Number Q002-244988 Version / Version 00 Ersteller / Created by Erstellt Datum / Created Date 22.04.2024 Dokumentenname / Document name Ablageort / File
Hanna Kafurke 510(k) Summary 1DA9486B9E2CC68
# freigegeben / released 22.04.2024
| Funktion | Name | Datum | Signatur | Vermerk |
|--------------------------------------|---------------|------------|-------------------------------------------------------------------------------|----------------------|
| QM / Regulatory<br>Affairs | Hanna Kafurke | 22.04.2024 | Hanna Kafurke<br>Q102-244968-00 | erstellt/created |
| Geschäftsführer<br>Managing Director | Robert Geiger | 22.04.2024 | 22.04.2024 15:20:27<br>Robert Geiger<br>Q002-244988-00<br>22.04.2024 15:20:36 | freigegeben/released |
| Version /<br>version | Änderung / Changes | Name /<br>name | Datum / Date |
|----------------------|----------------------|----------------|--------------|
| 00 | Erstellung / created | HK | 22.04.2024 |
# 510(k) Summary
#### Inhaltsverzeichnis / contents
| 1. | 510(k) Summary | |
|----|-------------------------------------------------------------|--|
| 2. | Contact Details | |
| 3. | Device Name | |
| 4. | Legally Marketed Predicate Devices | |
| 5. | Device Description Summary | |
| 6. | Intended use/Indications for Use | |
| 7. | Indications for Use Comparison | |
| 8. | Technological Comparison | |
| | 9. Non-Clinical and/or Clinical Tests Summary & Conclusions | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" in white on the left side. To the right of the circle is the word "AKTORmed" in gray, with the words "ROBOTIC SURGERY" in a smaller font below it. The letters "TM" are in superscript to the right of the word "AKTORmed".
## 1. 510(k) Summary
510(k) Number: K232405
#### 2. Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | AKTORmed GmbH |
|-----------------------------|-----------------------------------------------------------|
| Applicant Address | Neugablonzer Strasse 13, Neutraubling, BY, 93073, Germany |
| Applicant Contact Telephone | +49 9401 9320 150 |
| Applicant Contact | Mrs. Hanna Kafurke |
| Applicant Contact Email | Hanna.kafurke@aktormed.com |
#### 3. Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | ENDOFIX EXO (171943) |
|---------------------|---------------------------|
| Common Name | Endoscope and accessories |
| Classification Name | Endoscope Holder |
| Regulation Number | 876.1500 |
| Product Code | OCV |
#### 4. Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | K190576 |
|----------------------------------------------------------------|-----------------------------------------------|
| Predicate Trade Name<br>(Primary Predicate is listed<br>first) | WalterLorenz Surgical Assist Arm Scope Holder |
| Product Code | OCV |
#### 5. Device Description Summary
21 CFR 807.92(a)(4)
The ENDOFIX EXO is a manually operated surgical device intended to hold endoscopes during general and neurological diagnostic and therapeutic procedures. It emulates an arm which is capable of guiding and positioning an endoscopic camera with several degrees of movement.
lt ensures stable and sensitive positioning of the endoscope and allows the surgeon to work one-handed and/or two-handed. The system provides surgeons with a stable and rocksteady field of vision, even in extreme endoscopic positions. The wide scope of movement provides surgeons with full visibility of the situs. The ENDOFIX exo can be used with both, an endoscope and exoscope.
An integrated DEACTIVATION PUSHBUTTON permits manual movement of the ENDOFIX EXO at the push of a button. In order to reposition a guided endoscopic camera, the ENDOFIX EXO is unlocked by a sterilizable CONTROL at the push of a button. The ENDOFIX EXO is operated by unlocking all 6 axes and subsequent manual positioning using the CONTROL. As soon as the button on the CONTROL is released, all 6 axes are locked again.
Despite its large scope of movement, the ENDOFIX EXO is light-weight and compact and is fastened directly to the OR table by means of a quick-acting clamp. Thus, repositioning with reference to the patient is not necessary even if the OR table is moved in the meantime.
On the rigid endoscope, there is a camera linked that sends a video to a monitor. Endoscope and camera as well as the monitor are not part of the ENDOFIX EXO.
To allow the surgeon to convert to an open surgery in case of an emergency, the ENDOFIX EXO can be completely mounted or removed by a quick fastener.
| Dokumentennummer / document number | Q002-244988 |
|------------------------------------|-------------|
| Version / version | 00 |
| Datum / date | 22.04.2024 |
| Seite / page | 2 / 4 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AKTORmed Robotic Surgery. The logo consists of a blue circle with the letters "am" in white inside of it. To the right of the circle is the word "AKTORmed" in gray, with the words "ROBOTIC SURGERY" in a smaller font below it.
During use, the ENDOFIX EXO is covered with a STERILE COVER. The ENDOFIX EXO can be removed from the OR table and stored on the TROLLEY.
#### 6. Intended use/Indications for Use
21 CFR 807.92(a)(5)
The ENDOFIX EXO is indicated for use in interventions where a rigid endoscope is indicated for use. The device is intended to hold and position a rigid endoscope / exoscope during general and neurological diagnostic and therapeutic procedures.
# 7. Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use are very similar for both devices have the same intended use, which is the fixation of an endoscope in a position. The ENDOFIX EXO is for use during general and neurological diagnostic and therapeutic procedures. The WalterLorenz Surgical Assist Arm can be used in general and neurological, diagnostic and therapeutic procedures. However, this does not create a new intended use, since the is identical for both devices. For WalterLorenz Surgical Assist Arm Scope Holder the user is a trained medical professional, for ENDOFIX EXO the user is defined as a trained surgeon performing the operation. Both statements mean the same thing, they are just worded differently. The Patient group for the ENDOFIX EXO is all patients, who are suitable within the scope of indications for use. There are no known patient limitations for the WalterLorenz Surgical Assist
Arm Scope Holder. Both devices should only be used in professional healthcare environment.
## 8. Technological Comparison
21 CFR 807.92(a)(6)
Both products have for the most part the same technological characteristics:
Both devices have a permanently installed unlock button, which releases all locks so that the arm can move freely. The unlock button of the ENDOFIX exo is called DEACTIVATION and is located at the distal end of the arm. In contrast, the button of the WalterLorenz Surgical Assist arm is located on the Control Box with User Interface Keypad. However, both devices have a second control panel with an unlock button that is connected to the base unit of the arm via a cable. The control unit of the WalterLorenz Surgical Assist Arm is a sterile, disposable accessory that is supplied separately and can be attached to the distal end of the arm as a release button. The CONTROL of the ENDOFIX EXO is slid over the magnetic bracket of the endoscope clamp and secured from slipping off by a magnetic attraction. When the status LED next to the DEACTIVATION PUSHBUTTON indicates, the CONTROL is ready to operate. This does not happen until the CONTROL is on the magnetic bracket of the endoscope clamp, because there is a sensor in the CONTROL that detects the magnet and only then activates the CONTROL. This increases safety for the patient, as the locks cannot be released if the CONTROL is not on the endoscope clamp. For both devices the Energy Source is Electricity and both do not emit energy. The control units are in the devices. The ENDOFIX exo as well as the WalterLorenz Surgical Assist Arm can be fixed on U.S. table rails. They have no software and do not include biologics, drugs, coatings and additives. For both devices, the sterilizable accessories can be sterilized by steam sterilization and cleaned by wiping disinfection. Regarding the power supply, both devices have the identical input voltage of 100-240 VAC; 50-60Hz and the operating mode is continuous.
Nevertheless, there are few differences of technological characteristics:
The first difference is the mechanism of the two devices. While the WalterLorenz Surgical Assist Arm uses electromechanical brakes in the axes, the ENDOFIX EXO uses electromagnetic brakes to lock movement in all axes. However, there are no safety or effectiveness implications or other safety issues arising from this difference, as the electromagnetic locks have been mechanically tested according to IEC 60601-1:2005 + A1:2012. In addition, the displacement after fixation of the arm has also been tested. From both tests, there are no risks regarding the safety of the arm.
| Dokumentennummer / document number | Q002-244988 |
|------------------------------------|-------------|
| Version / version | 00 |
| Datum / date | 22.04.2024 |
| Seite / page | 3 / 4 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Aktormed logo. The logo consists of a blue circle with the letters "am" in white, followed by the word "AKTORmed" in gray. Below the word "AKTORmed" is the phrase "ROBOTIC SURGERY" in a smaller, gray font. The logo is clean and modern, and the colors are professional and trustworthy.
The ENDOFIX EXO is compatible with endoscopes. For the WalterLorenz Surgical Assist Arm Scope Holder, no exact specification of which scopes are compatible could be found. By specifying exactly which scopes are compatible, the ENDOFIX EXO has at least an equivalent level of safety in terms of compatibility with devices as the WalterLorenz Surgical Assist Arm Scope Holder.
The ENDOFIX EXO's safe working load is specified as 1kg, it was even tested with 5kg and no problems occurred. No defined safe working load was found for the WalterLorenz Surgical Assist Arm.
The WalterLorenz Surgical Assist Arm is battery powered. The ENDOFIX EXO, on the other hand, does not have a battery, which means that there is no risk of the battery failing during an operation and the safety of continuous power operation is ensured. However, this means that the ENDOFIX EXO can no longer be moved in the event of a power failure, but it has been assured that the arm will continue to hold the endoscope in a fixed position. In the event that power cannot be restored by reconnection, both the endoscope itself and the device can be safely and quickly removed from the patient using quick-release fasteners as described in the instructions for use.
The weights of the two devices differ significantly. The ENDOFIX EXO weighs 26.5lbs, the WalterLorenz Surgical Assist Arm without the Scope Holder 8lbs. However, the higher weight does not lead to an increased safety risk, as the ENDOFIX EXO is equipped with a counterbalance to compensate for the endoscope and the camera head. In addition, the ENDOFIX EXO is mounted to the operating table so that no weight has to be carried by one person. If the ENDOFIX EXO has to be transported over a long distance, it can be mounted on the trolley provided for this purpose and then only needs to be pushed. A usability test was also conducted with 15 American physicians, where weight was not mentioned negatively.
The output voltage of the two devices varies slightly. The ENDOFIX EXO has an output voltage of 24V DC and the WalterLorenz Surgical Assist Arm has an output voltage of 18V DC. However, the ENDOFIX EXO has been sufficiently tested according to IEC 60601-1 for electrical safety to state that it poses no danger to the user.
The last difference is that the distal unlock button of the WalterLorenz Surgical Assist Arm is a single-use product and comes sterile packed. The control unit of the ENDOFIX EXO, on the other hand, can be used more often, but must be cleaned and sterilized before the first use. The complete validation of the reprocessing and sterilization process confirms that there is no increased risk.
# 9. Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The following tests were performed:
- 1. Lifetime test
- 2. Lifetime test CONTROL
- 3. Moving after fixation test
- 4. Usability test
- 5. Package drop test
Nonclinical performance tests of the substantial device are unknown. Therefore own nonclinical performance tests were successfully performed to demonstrate the safety and efficiency of the ENDOFIX EXO.
The non-clinical tests have shown that the ENDOFIX EXO device is as safe and effective as the predicate device. For the WalterLorenz Surgical Assist Arm there are unfortunately no which non-clinical performance tests have been performed. However, since the WalterLorenz Surgical Assist Arm is only a Class 1 device, it can be assumed that all non-clinical performance tests for the ENDOFIX EXO are sufficient to ensure at least equivalent safety.
| Dokumentennummer / document number | Q002-244988 |
|------------------------------------|-------------|
| Version / version | 00 |
| Datum / date | 22.04.2024 |
| Seite / page | 4 / 4 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.