K232385 · Iss Medical, Inc. · MUD · Mar 1, 2024 · Cardiovascular
Device Facts
Record ID
K232385
Device Name
OxiplexTS200
Applicant
Iss Medical, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Mar 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The OxiplexTS200 tissue oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. It is indicated for monitoring patients during circulatory or perfusion examinations of skeletal muscle or when there is a suspicion of compromised circulation. The device can be used for monitoring in Medical Clinics, laboratories and outpatient facilities. The OxiplexTS200 should not be used as the sole basis for diagnosis or therapy. The value of these measurements in disease states has not been demonstrated.
Device Story
OxiplexTS200 is a non-invasive tissue oximeter using near-infrared spectroscopy (NIRS) to estimate percent oxygen saturation (StO2) in skeletal muscle tissue. The device emits light at 690nm and 830nm into tissue via a fiber optic bundle sensor; it collects returning light at set distances to determine optical parameters and calculate oxy- and deoxy-hemoglobin concentrations. The system includes a main unit, reusable sensors, and optional analog input/output modules. It is operated by healthcare professionals in clinics, laboratories, and outpatient facilities for spot-checking or continuous monitoring. The device provides real-time StO2 measurements and trends, alerting users via visual alarms when values fall below custom thresholds or if error conditions (e.g., signal strength, sensor placement) occur. By providing hemodynamic data, the device assists clinicians in assessing tissue oxygenation status, though it is not intended as the sole basis for clinical diagnosis or therapy.
Clinical Evidence
Clinical testing involved 25 human subjects comparing the OxiplexTS200 to the ViOptix ODISsey. Measurements were taken on the brachioradialis and thenar muscles at baseline and during venous/arterial occlusion using a blood pressure cuff. Results showed similar rates of change in StO2 for both devices during occlusion. While absolute estimates differed slightly at baseline and after two minutes of occlusion, the study concluded the device provides sensitive and reliable measurements consistent with the predicate. Human factors/usability testing confirmed that users could successfully perform measurements on first use.
Technological Characteristics
NIRS-based tissue oximeter; dual-wavelength (690nm, 830nm) laser diodes (Class I). Fiber optic bundle sensor (reusable). Dimensions/form factor: instrument system with sensors, phantom block, and optional analog I/O modules. Connectivity: USB for analog modules. Software: PEMS compliant with IEC 62304. Electrical safety: IEC 60601-1. Biocompatibility: ISO 10993-5.
Indications for Use
Indicated for monitoring patients during circulatory or perfusion examinations of skeletal muscle or when there is a suspicion of compromised circulation in medical clinics, laboratories, and outpatient facilities. Not for home use or out-of-hospital transport. Not for use as sole basis for diagnosis or therapy.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
InSpectra StO2Tissue Oxygenation Monitor, Model 650 (k100915)
ViOptix ODISsey Tissue Oximeter (k042657)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 1, 2024
ISS Medical, Inc. Beniamino Barbieri President 1602 Newton Drive Champaign, Illinois 61822
Re: K232385
Trade/Device Name: OxiplexTS200 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: August 4, 2023 Received: August 9, 2023
Dear Beniamino Barbieri:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Tanisha L. Hithe -S 2024 03 01 Hithe -S 14:49:37 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232385
Device Name OxiplexTS200
#### Indications for Use (Describe)
The OxiplexTS200 tissue oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. It is indicated for monitoring patients during circulatory or perfusion examinations of skeletal muscle or when there is a suspicion of compromised circulation. The device can be used for monitoring in Medical Clinics, laboratories and outpatient facilities.
The OxiplexTS200 should not be used as the sole basis for diagnosis or therapy. The value of these measurements in disease states has not been demonstrated.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
### Section 5: 510(k) Summary
### 510(k) Summary and Certification
### 1. Submitter's Name/Contact Person
| Submitted by: | Contact Person: |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| ISS Medical, Inc.<br>1602 Newton Drive<br>Champaign, IL 61822 | Beniamino Barbieri<br>President<br>Tel: 217.359.8681 x 11<br>Fax: 217.359.7879<br>Email: beniamino.barbieri@iss.com |
#### Summary Date: July 31, 2017
#### 2. General Information
- a. Proprietary Name: OxiplexTS200
- b. Common/Usual Name: Tissue Oximeter
- Classification Name: Oximeter C.
- d. Class: Il
- e. Product Code: 74MUD
- CFR Reference: 21 CFR 870.2700 f.
- g. Identification of Equivalent Devices:
InSpectra StO2Tissue Oxygenation Monitor, Model 650 (k100915) ViOptix ODISsey Tissue Oximeter (k042657)
#### 3. Device Description
The ISS Medical non-invasive tissue oximeter. OxiplexTS200. is intended for use as an adjunct monitor of regional oxy- and deoxy-hemoglobin concentration, and of hemoglobin oxygen saturation of blood in tissues during circulatory or perfusion examinations of skeletal muscle or when there is a suspicion of compromised circulation.
The OxiplexTS200 is a device for use in spot-checking as well as continuous monitoring of patients in healthcare facilities. It is not intended for home use or to be used for "out of hospital" transport.
The OxiplexTS200 is an instrument system and includes the following components:
- OxiplexTS200 Unit with two sensors and phantom block ●
- Power cable ●
- Accessories (straps for sensors; strap for phantom block; dark cloth) ●
- ISS flash memory
- Operation Manual
{5}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
- Analoq Input module with USB cable (optional)
- . Analog Output module with USB cable (optional)
The OxiplexTS200 uses near-Infrared Spectroscopy (NIRS), a noninvasive diagnostic tool, to offer features for the assessment and monitoring of oxygenation in tissues such as brain and muscle. Near-infrared light (in the wavelength range 700-900 nm) can penetrate several centimeters into numerous body tissues. NIRS enables continuous real time measurements of changes in the hemoglobin oxygenation state, thus providing information on tissue oxygenation and hemodynamics.
#### 4. Indications for Use
The OxiplexTS200 tissue oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. It is indicated for monitoring patients during circulatory or perfusion examinations of skeletal muscle or when there is a suspicion of compromised circulation. The device can be used for monitoring in Medical Clinics, laboratories and outpatient facilities. The OxiplexTS200 should not be used as the sole basis for diagnosis or therapy. The value of these measurements in disease states has not been demonstrated.
### 5. Comparison with Predicate Devices
ISS Medical's OxiplexTS200 System is compared to the InSpectra StO2Tissue Oxygenation Monitor, Model 650 and the ViOptix ODISsey Tissue Oximeter in Table 1 below.
| | OxiplexTS200 | InSpectra<br>StO2Tissue<br>Oxygenation<br>Monitor, Model<br>650 (k100915) | ViOptix ODISsey<br>Tissue Oximeter<br>(k042657) | Similar/<br>Same |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Indications for<br>Use | The OxiplexTS200<br>tissue oximeter is<br>intended to non-<br>invasively estimate<br>the percent oxygen<br>saturation (StO2) in<br>a volume of tissue.<br>It is indicated for<br>monitoring patients<br>during circulatory or<br>when there is a<br>suspicion of | Hutchinson<br>Technology<br>Incorporated's<br>InSpectra™<br>StO2Tissue<br>Oxygenation<br>Monitor is intended<br>for use as a<br>noninvasive<br>monitoring system<br>that measures an<br>approximated value | The ViOptix<br>ODISsey Tissue<br>Oximeter is<br>intended to<br>noninvasively<br>estimate the<br>percent oxygen<br>saturation (StO2) in<br>a volume of tissue. | |
Table 5-1: Comparison to Predicate Devices
{6}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
| | OxiplexTS200 | InSpectra<br>StO2Tissue<br>Oxygenation<br>Monitor, Model<br>650 (k100915) | ViOptix ODISsey<br>Tissue Oximeter<br>(k042657) | Similar/<br>Same |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Indications for<br>Use | compromised<br>circulation. The<br>device can be used<br>for monitoring in<br>Medical Clinics,<br>laboratories and<br>outpatient facilities.<br>The OxiplexTS200<br>should not be used<br>as the sole basis<br>for diagnosis or<br>therapy. The value<br>of these<br>measurements in<br>disease states has<br>not been<br>demonstrated. | of percent<br>hemoglobin oxygen<br>saturation in tissue<br>(StO2).<br>The InSpectra™<br>StO2Tissue<br>Oxygenation<br>Monitor is indicated<br>for use in<br>monitoring patients<br>during circulatory or<br>perfusion<br>examinations of<br>skeletal muscle or<br>when there is a<br>suspicion of<br>compromised<br>circulation. | This is performed in<br>medical<br>environments<br>including physician<br>offices, hospitals,<br>ambulatory care,<br>and Emergency<br>Medical Services.<br>The ODISsey<br>Tissue Oximeter is<br>indicated for use in<br>monitoring patients<br>during circulatory or<br>perfusion<br>examinations of<br>skeletal muscle or<br>when there is a<br>suspicion of<br>compromised<br>circulation. | Similar |
| | OxiplexTS200 | InSpectra<br>StO2Tissue<br>Oxygenation<br>Monitor, Model<br>650 (k100915) | ViOptix ODISsey<br>Tissue Oximeter<br>(k042657) | Similar/<br>Same |
| Principle of<br>Operation | Light is used to<br>probe the region of<br>the tissue under the<br>sensor. Light is<br>sent through the<br>issue at four<br>different locations; a<br>fraction of it returns<br>to the surface and it<br>is collected at set<br>distances from the<br>points of insertion.<br>Optical parameters<br>of the tissue are<br>determined and<br>hence the<br>concentration of<br>oxy- and deoxy-<br>hemoglobin. The<br>tissue oxygen<br>saturation is<br>derived. | Diffuse reflectance<br>spectroscopy. Light<br>is used to probe a<br>cross section of the<br>microvasculature of<br>tissue (mixed bed of<br>arterioles,<br>capillaries, and<br>venuoles). The light<br>returning from<br>tissue, after having<br>passed through<br>tissues, for<br>hemoglobin in its<br>oxygenated and<br>deoxygenated<br>forms in the<br>optically sampled<br>region. The tissue<br>oxygen saturation<br>(%StO2), a value<br>reflecting<br>percentage of<br>oxygenated<br>hemoglobin in the<br>sampled<br>microvasculature of<br>tissue is calculated. | Diffuse reflectance<br>spectroscopy. Light<br>is used to probe a<br>cross section of the<br>microvasculature of<br>tissue (mixed bed of<br>arterioles,<br>capillaries and<br>venuoles). The<br>light returning from<br>tissue, after having<br>passed through<br>tissues, for<br>hemoglobin in its<br>oxygenated and<br>deoxygenated<br>forms in the<br>optically sampled<br>region. The tissue<br>oxygen saturation<br>(%StO2), a value<br>reflecting<br>percentage of<br>oxygenated<br>hemoglobin in the<br>sampled<br>microvasculature of<br>tissue, is calculated. | Similar |
| Range (StO2) | 0-100% | 0-99% | 0-100% | Similar |
| Accuracy<br>[parameter<br>StO2] | 50-100%: 2%<br>26-49%: 3%<br>0-25%: 10% | 70-99%; 1.63 StO2<br>units<br>0-70%; 2.94 StO2<br>units | 80-100%: 2.7%<br>60-80%: 2.3%<br>40-60%: 5.2%<br>10-40%: 13.3% | Similar |
| Alarm Limit<br>Range | Visual alarms when<br>value is below<br>custom set<br>threshold. | Visual alarms for<br>error conditions | Audio and visual<br>alarms when<br>range limit is<br>exceeded | Same for<br>ViOptix;<br>different<br>for<br>InSpectra |
| Measures<br>Near-infrared<br>Wavelength | Yes | Yes | Yes | Same |
| | OxiplexTS200 | InSpectra<br>StO2Tissue<br>Oxygenation<br>Monitor, Model<br>650 (k100915) | ViOptix ODISsey<br>Tissue Oximeter<br>(k042657) | Similar/<br>Same |
| Sensor | Re-useable | Single use;<br>disposable | Single use;<br>disposable | Different |
| Sample<br>Acquisition | Fiber optic bundle<br>placed in contact<br>with skin. | Fiber optic probe<br>placed on intact<br>skin | Fiber optic probe<br>placed on intact<br>skin | Same |
| Features (e.g.,<br>alarms, display<br>and indicators,<br>modes) | Several errors and<br>alarms are reported<br>by the software<br>during the operation<br>of the instrument:<br>a. Overload<br>Warning (when<br>the signal from<br>the environment<br>saturates the<br>detectors);<br>b. Signal Strength<br>Warning<br>(regarding the<br>placement of the<br>sensors on the<br>patient;<br>c. Visual Alarm for<br>set thresholds of<br>[THC] and<br>oxygen<br>saturation;<br>d. Overvoltage<br>warning.<br>e. Warnings<br>resulting from<br>Error condition<br>tests (AC slope,<br>regression<br>coefficient,<br>phase standard<br>deviation, out of<br>range optical<br>parameters). | Audiovisual<br>•<br>high/low StO2<br>error condition,<br>• Defective LED<br>or cable sensor<br>• Communication<br>error<br>• Over<br>temperature<br>• Cable not<br>connected<br>Customizable StO2<br>alarm limits and<br>notification of error<br>conditions | System will perform<br>self-test when<br>power is turned on.<br>Visual and audible<br>alarms include: low<br>StO2, high StO2,<br>StO2 drop rate and<br>low battery. Visual<br>alarms include:<br>inadequate patient<br>contact, sensor<br>laser ON, and AC<br>power ON. Alarm<br>volume 50 to 80<br>dbA at 30cm (1<br>foot). Single-<br>channel (default) or<br>dual channel<br>(applies to consoles<br>with dual channel<br>option when 2<br>sensors are<br>attached.) 3-cell<br>Lithium Ion Battery<br>with 30 minute<br>minimum. | Similar |
| Time to first<br>reading | Continuous reading<br>with up to 8 hours<br>trended display.<br>Max display update<br>rate 0.5 seconds | Continuous reading<br>with up to 8 hours<br>trended display.<br>Display update rate<br>2 seconds. | ~20 seconds | Same for<br>InSpectra;<br>different<br>for<br>ViOptix |
| | OxiplexTS200 | InSpectra<br>StO2Tissue<br>Oxygenation<br>Monitor, Model<br>650 (k100915) | ViOptix ODISsey<br>Tissue Oximeter<br>(k042657) | Similar/<br>Same |
| Number of<br>different<br>wavelengths of<br>light emitted | Two (2); 690nm and<br>830nm | Four (4); 680, 720,<br>760, and 800nm | Two (2); 690nm and<br>830nm | Same for<br>ViOptix;<br>different<br>for<br>InSpectra |
| Battery | Not available,<br>(the instrument<br>does not operate<br>with a battery) | Lithium battery;<br>capacity 2 hours<br>(new/full charge);<br>Life 300 full<br>charge/discharge<br>cycles | Lithium battery | Different |
| Product Laser<br>Rating | Class I | Class I | Class 3R | Same for<br>InSpectra;<br>different<br>from<br>ViOptix |
{7}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
{8}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
{9}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
Similarities between the OxiplexTS200 and the predicates include:
- 1. The intended use for all three devices is similar.
- 2. The principles of operation are similar.
- 3. The range and accuracy are similar.
- 4. All three devices measure near-infrared wavelength.
- 5. The sample acquisition is the same for all three devices.
- 6. The safety specifications are the same.
- 7. The features of the three devices are similar.
Differences between the OxiplexTS200 and the predicates include:
- 1. The OxiplexTS200 and ViOptix have both audio and visual alarms; the InSpectra has only visual alarms.
- 2. The InSpectra and ViOptix have disposable sensors; the OxiplexTS200 has reusable sensors.
- 3. The OxiplexTS200 and InSpectra have continuous readings with up to 8 hours trending displayed; the ViOptix has 20 second time to read.
- 4. The OxiplexTS200 and ViOptix have two (2) different wavelengths of light emitted; InSpectra has four (4).
- 5. The InSpectra and ViOptix have battery option: OxiplexTS200 does not have a battery option.
- 6. OxiplexTS200 is considered to be a Class I laser product (as the InSpectra), while the ViOptix is classified as a Class 3R laser product.
{10}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
### 6. Performance Test - Bench
The differences in the technological character between the OxiplexTS200 and the predicates do not raise new types of question regarding safety and effectiveness. In response to the reviewer's request that ISS Medical conducts a direct comparison of the OxiplexTS200 with a predicate instrument, it was arranged to compare it to the ODISey Tissue Oximeter model OXY-2 by Vioptix Inc. The comparison measurements were acquired in the Biomedical Optics Research Laboratory directed by Professor Martin Wolf at the Department of Neonatology at University Hospital Zürich in Zürich, Switzerland.
The staff and equipment there enabled us to conduct a series of simultaneous SO2 measurements using the Vioptix predicate device and the OxiplexTS200 on a phantom containing human blood diluted in an optically scattering medium to simulate a physiologically appropriate level of tissue hemoglobin concentration [tHb] and a physiologically realistic optical scattering and absorption properties. Measurements were made at three levels; from physiological levels to increasingly higher concentrations. The primary comparison of interest is the relation of the OxiplexTS200 percent oxygen saturation [SO2%] estimates with the Vioptix SO2% estimates. These are presented in section 19 performance bench testing where it can be seen that for physiologically realistic tissue hemoglobin concentrations [tHb] both instruments report nearly identical values over the range of 20% to 98% SO2. The presence of the pigmented "skin" filter had no discernable effect on either instrument.
#### 7. Performance Testing- Clinical
We conducted testing using the OxiplexTS200 and the Vioptix devices to compare StO2 readings from a total of 25 human subjects.
Measurements were acquired over the brachioradialis, the largest muscle in the forearm, and over the thenar of the thumb, a complex region of small muscles and connective tissue. In addition to baseline measures of StO2, we varied StO2 by occluding blood flow to the arm both completely (venous and arterial occlusion) and partially (venous occlusion only), using a blood pressure cuff on the upper arm.
We found that the rate of change of StO2 after both types of occlusions was very similar for both instruments, and baseline measurements on the thenar were not statistically different, while the two instruments gave different estimates of StO2 after two minutes of occlusion at both locations, and on the arm at baseline.
These results are consistent with the results of our previous in vitro comparison of these two instruments which demonstrated that at physiologically realistic levels of [tHb] the OxiplexTS200™ and ViOptix ODISsey instruments provided essentially identical sensitive, reliable and accurate measurements of StO2.
{11}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
ISS Medical Inc. has also conducted human factors testing to verify and demonstrate the safety and effectiveness of the OxiplexTS200 for its intended users, use and environment.
Most of the tests mimic the actual daily use of the system. The primary goal of these studies was to evaluate if the device operators are able to accomplish routine tasks with the provided instructions and to assess the clarity of the Instructions For Use (IFU).
Based on the results of the usability studies, the majority of users would be able to make successful and accurate measurements on first use of the device. We conclude that the device could be used effectively without any changes to the user interface.
### 8. Performance Testing
Bench testing and biocompatibility testing was conducted to demonstrate that the OxiplexTS200 functions as intended. The OxiplexTS200 is designed to conform to the applicable requirements of electrical safety, laser safety and electromagnetic compatibility requirements as summarized below:
- . Evaluation of Risk Management in accordance with IEC 60601-1 3rd Ed. and ISO 14971.
- . Evaluation to 60601-1 (3.1 Edition) IEC, EN, ANSI/AAMI, and CSA. Includes European, US, and Canadian National differences.
- . Evaluation of Programmable Electrical Medical Systems (PEMS) compliance, using IEC 62304:2006 - (Software)
- . Laser classification testing according to IEC 60825-1 2nd Ed. for two laser diodes (830nm and 690nm)
- Evaluation of Usability compliance, using IEC 60601-1-6 :2010 . Collateral standard.
- Full EMC evaluation for mains powered device using EN/IEC 60601-. 1-2:2007.
- Testing for biocompatibility (ISO 10993-5:2009) ●
All testing met specified requirements. Test reports were provided in the submission as Appendices 4, 6 and 7.
#### 9. Conclusions
The OxiplexTS200 device that is the subject of this 510k pre-market notification, is the same or similar in design, intended use, principles of operation, and technological
{12}------------------------------------------------
ISS Medical 510(k) Premarket Notification OxiplexTS200
characteristics, compared to the predicate devices. Differences of the subject device's output parameters compared to those of the predicate devices' output parameters do not raise new types of questions regarding safety and effectiveness, and performance testing provided in this 510k supports that the device can be used safely and effectively for the proposed indications for use. The OxiplexTS200 device that is the subject of this 510K is considered to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.