The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs specimens from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test and followed with a molecular test. The test does not differentiate between SARS-CoV and SARS-CoV-2. A negative test result is presumptive and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions. Positive results do not rule out co-infection with other respiratory pathogens. Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.
Device Story
Lateral flow immunochromatographic assay; detects SARS-CoV-2 nucleocapsid protein antigen in anterior nasal swabs. HCP collects specimen; transfers to extraction tube for lysis; applies 4 drops to test cassette. Colloidal gold-conjugated monoclonal antibodies bind target; captured by second antibody on reaction membrane. Visual readout of colored lines at test (T) and control (C) regions after 15 minutes. Used in point-of-care settings; aids COVID-19 diagnosis. Negative results presumptive; require molecular confirmation. Benefits include rapid, qualitative detection to inform clinical management.
Clinical Evidence
Prospective multi-center study (May-July 2022) with 806 symptomatic subjects (age 2+). Compared against composite RT-PCR comparator. Results: PPA 85.4% (95% CI: 79.1-90.0%); NPA 99.7% (95% CI: 98.9-99.9%).
Indicated for symptomatic individuals aged 2 years or older with signs of upper respiratory infection within 6 days of symptom onset. For prescription use only.
Regulatory Classification
Identification
The Sofia 2 SARS Antigen+ FIA is a lateral flow immunofluorescent sandwich assay used with the Sofia 2 instrument for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from symptomatic individuals. It is intended as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when tested at least twice over three days with at least 48 hours between tests.
Predicate Devices
Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set (DEN220039)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232377
B Applicant
Healgen Scientific LLC
C Proprietary and Established Names
Healgen Rapid COVID-19 Antigen Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QVF | Class II | 21 CFR 866.3982 - Simple Point-Of-Care Device To Directly Detect SARS-Cov-2 Viral Targets From Clinical Specimens In Near-Patient Settings | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain 510(k) clearance for the Healgen Rapid COVID-19 Antigen Test.
B Measurand:
Nucleocapsid protein antigen from SARS-Coronavirus 2 (SARS-CoV-2)
C Type of Test:
Qualitative lateral flow immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test and followed with a molecular test.
The test does not differentiate between SARS-CoV and SARS-CoV-2.
A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.
Positive results do not rule out co-infection with other respiratory pathogens.
Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
IVD - For In Vitro Diagnostic Use Only
D Special Instrument Requirements:
N/A
IV Device/System Characteristics:
A Device Description:
The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic membrane assay that uses monoclonal antibodies to detect nucleocapsid protein from SARS-CoV-2 virus in anterior nasal swab specimens. The test strip is composed of a sample pad, reagent pad, reaction membrane, and absorbing pad housed within a test cassette. The reagent pad contains a colloidal gold-conjugated monoclonal antibody, which recognizes and binds to the nucleocapsid protein of SARS-CoV-2; the reaction membrane in the test line (T) contains the second antibody that recognizes another epitope of the nucleocapsid protein. The whole strip is fixed inside a plastic cassette. Figure 1 shows the test principle.
K232377 - Page 2 of 14
{2}

Figure 1. Diagram of Device Principle
External quality controls are required but not included with the test kit and are sold separately as the COVID-19 Antigen Control Kit. The control swabs should be processed according to the Instructions for Use (IFU) and are intended to be used as quality control samples to demonstrate that the test is performing and is being performed correctly. The Negative Control Swab is composed of negative control buffer dried onto a swab, with a blue shaft and containing $0.1\%$ sodium azide preservative. The Positive Control Swab also contains SARS-CoV-2 recombinant antigen extract dried onto a swab at a concentration of $0.1\mathrm{ng / mL}$ , with a red shaft and containing $0.1\%$ sodium azide preservative.
# B Principle of Operation:
The Healgen Rapid COVID-19 Antigen Test uses immunochromatographic-based technology in a lateral flow design to detect nucleocapsid protein from SARS-CoV-2 virus. Anterior nasal swab specimens are collected by a healthcare professional (HCP). Following sample collection, the nasal swab is transferred to the prefilled extraction tube to lyse the sample and solubilize the viral nucleoproteins. Four (4) drops of the lysed sample is loaded onto the test cassette sample well, and the test result is read by the HCP after 15 minutes. If antigen is present, the colloidal gold monoclonal antibody conjugates bind to the nucleocapsid protein, and the antigen-antibody complex then migrate up along the reaction membrane, which will be captured by the second specific anti-SARS-2 antibody coated on the T line. The sample migrates along the test strip across two distinct areas: the test line and the internal procedural control line. If SARS-CoV-2 viral antigens are present, they will be captured by antibodies and bound to the test line of the test strip, and a colored line will be visible in the test window of the cassette. A red line will always appear in the control line (C), which is coated with secondary antibody recognizing the primary antibody in the migrating sample.
The Healgen Rapid COVID-19 Antigen Test gives three possible results: positive, negative or invalid. A positive result will present two colored lines beside the "C" and "T" regions of the test window and indicates that SARS-CoV-2 antigen was detected. A negative result will present only one colored line next to the "C" region of the test window and indicates that SARS-CoV-2 antigen was not detected. If no line is seen in the "C" region after 20 minutes, then the result is invalid and should be repeated with new test materials.
K232377 - Page 3 of 14
{3}
V Substantial Equivalence Information:
A Predicate Device Name(s):
Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set
B Predicate 510(k) Number(s):
DEN220039
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device K232377 | Predicate DEN220039 |
| --- | --- | --- |
| Device Trade Name | Healgen Rapid COVID-19 Antigen Test, COVID-19 Antigen Control Kit | Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set |
| Intended Use/Indications For Use | The Healgen Rapid COVID-19 Antigen Test is a lateral flow immunochromatographic assay intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swabs from individuals with signs and symptoms of upper respiratory infection within the first six (6) days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Healgen Rapid COVID-19 Antigen Test and followed with a molecular test.
The test does not differentiate between SARS-CoV and SARS-CoV-2.
A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.
Positive results do not rule out co-infection with other respiratory pathogens.
Performance characteristics for | The Sofia 2 SARS Antigen+ FIA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia 2 instrument for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 6 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when tested at least twice over three days with at least 48 hours between tests.
The test does not differentiate between SARS-CoV and SARS-CoV-2.
A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Negative results do not preclude SARS-CoV-2 infections and should not be used as the sole basis for treatment or other patient management decisions.
Positive results do not rule out co-infection with other respiratory pathogens.
Performance characteristics for |
K232377 - Page 4 of 14
{4}
K232377 - Page 5 of 14
| Device & Predicate Device(s): | Device
K232377 | Predicate
DEN220039 |
| --- | --- | --- |
| | pathogens.
Performance characteristics for SARS-CoV-2 were established from May 2022 to July 2022 when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variants are emerging, performance characteristics may vary. | SARS-CoV-2 were established during the 2021-2022 SARSCoV-2 pandemic when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variant are emerging, performance characteristics may vary.
This test is intended for prescription use only and can be used in Point-of-Care settings. |
| **General Device Characteristic Similarities** | | |
| Regulation | 21 CFR 866.3982 | Same |
| Target Analyte | Nucleocapsid protein antigen from SARS-CoV-2 | Same |
| Specimen Type | Direct anterior nasal (nares) swabs | Same |
| Intended Use Population | Symptomatic individuals within 6 days of symptom onset | Same |
| Test Time | 15 minutes after the extracted sample is added to the test cassette sample well | Same |
| Serial Testing | Yes | Same |
| Results Reported | Qualitative (positive, negative, invalid) | Same |
| Quality Controls | Built-in procedural control (control line region (C)) in the test strip and external quality controls (separately packaged kit). | Built-in procedural control in the test strip and external quality controls (included in kit). |
| **General Device Characteristic Differences** | | |
| Instrumentation | None | Sofia 2 Instrument |
| Principle of Operation | Lateral flow immunochromatographic assay | Lateral flow immunofluorescence sandwich assay |
| Result Interpretation | Visually read - visual interpretation of the presence or absence of colored line(s) on the control and test line(s) of the test strip is used to determine Positive, Negative, or Invalid results. | Instrument read - the Sofia 2 instrument scans the test strip and measures the fluorescent signal by processing the results using method specific algorithms. Sofia 2 displays the test results (Positive, Negative, or Invalid) on the screen. |
| Storage | Room temperature or refrigerated (2 to 30°C, 36 to 86°F) | Room temperature, 59°F to 86°F (15°C to 30°C) |
{5}
VI Standards/Guidance Documents Referenced:
No standard/guidance documents referenced.
VII Performance Characteristics (if/when applicable):
A. Analytical Performance:
1. Precision/Reproducibility:
A multisite precision study was performed at three external, CLIA-waived testing sites to evaluate reproducibility of the Healgen Rapid COVID-19 Antigen Test. Testing consisted of three replicates each of positive (prepared at 3x LoD), low positive (prepared at 1x LoD), and negative samples tested by three (3) untrained operators per site over 5 days, i.e., 3 replicates × 3 operators × 3 sites × 5 days = 135 replicates per concentration for a total of 405 total data points collected. Three (3) test lots were used in this study, and lot-to-lot variability was also assessed (see Table 2 below).
Fifty (50) µL of the prepared sample were applied to kit swabs, shipped and stored frozen at -20°C until testing. The results were ≥ 90% agreement between expected and read result within run, by lot, by operator, by day, between sites and overall.
Table 1. Summary Results of Multisite Precision Study (Reproducibility)
| Site | Negative | | Weak Positive | | Positive | |
| --- | --- | --- | --- | --- | --- | --- |
| | Correct Reads/Total | NPA | Correct Read /Total | PPA | Correct Reads/Total | PPA |
| 1 | 44/45 | 97.8% | 45/45 | 100.0% | 45/45 | 100.0% |
| 2 | 45/45 | 100.0% | 45/45 | 100.0% | 45/45 | 100.0% |
| 3 | 45/45 | 100.0% | 45/45 | 100.0% | 45/45 | 100.0% |
| Total | 134/135 | 99.2% | 135/135 | 100.0% | 135/135 | 100.0% |
Table 2. Summary Results of Lot-to Lot Precision Study
| Lot | Negative | | Weak Positive | | Positive | |
| --- | --- | --- | --- | --- | --- | --- |
| | Correct Reads/Total | NPA | Correct Read /Total | PPA | Correct Reads/Total | PPA |
| 1 | 45/45 | 100.0% | 45/45 | 100.0% | 45/45 | 100.0% |
| 2 | 44/45 | 97.8% | 45/45 | 100.0% | 45/45 | 100.0% |
| 3 | 45/45 | 100.0% | 45/45 | 100.0% | 45/45 | 100.0% |
| Total | 134/135 | 99.2% | 135/135 | 100.0% | 135/135 | 100.0% |
2. Linearity:
Not Applicable; the Healgen Rapid COVID-19 Antigen Test device is a qualitative assay.
K232377 - Page 6 of 14
{6}
# 3. Analytical Specificity/Interference:
# a. Cross-Reactivity and Microbial Interference (Analytical Specificity)
The purpose of the cross-reactivity and microbial interference studies is to evaluate potential cross-reactants and microbial interferents for their impact on the Healgen Rapid COVID-19 Antigen Test performance. For this testing, live, commercial sources of the organisms were obtained, excluding MERS which used inactivated virus. Three replicates each of microorganism prepared in negative clinical matrix were tested in the absence and presence of $2\mathrm{x}$ LoD inactivated SARS-CoV-2 virus (USA-WA1/2020) on 3 test lots. Fifty $(50)\mu \mathrm{L}$ of the prepared sample were applied to the kit swabs and eluted into the extraction buffer tube per the IFU. No cross-reactivity or microbial interference was observed with these organisms at the tested concentration.
Table 3. Cross-reactivity and Microbial Interference Study Results
| Virus/Microorganism | Concentration | Cross-Reactivity (% Negative Agreement) | Interference (% Positive Agreement) |
| --- | --- | --- | --- |
| Human coronavirus 229E | 8.00 × 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Human coronavirus OC43 | 7.00 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Human coronavirus NL63 | 2.93 x 104TCID50/mL | 100 (3/3) | 100 (3/3) |
| MERS-coronavirus | 7.00 × 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Adenovirus 10 | 2.39 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Adenovirus 21 | 1.14 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Human metapneumovirus | 3.95 x 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Parainfluenza virus Type 1 | 2.23 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Parainfluenza virus Type 2 | 2.23 x 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Parainfluenza virus Type 3 | 4.00 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Parainfluenza virus Type 4a | 7.00 × 104TCID50/mL | 100 (3/3) | 100 (3/3) |
| Influenza A (H1N1) | 4.00 x 108CEID50/mL | 100 (3/3) | 100 (3/3) |
| Influenza A (H3N2) | 7.00 x 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Influenza B | 7.00 x 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Enterovirus 68 | 2.23 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Enterovirus 71 | 4.00 x 107TCID50/mL | 100 (3/3) | 100 (3/3) |
| Respiratory syncytial virus | 2.23 x 106TCID50/mL | 100 (3/3) | 100 (3/3) |
| Rhinovirus 60 | 8.00 x 105TCID50/mL | 100 (3/3) | 100 (3/3) |
| Haemophilus influenzae | 1.74 x 108CFU/mL | 100 (3/3) | 100 (3/3) |
| Streptococcus pneumoniae | 3.35 x 108CFU/mL | 100 (3/3) | 100 (3/3) |
| Streptococcus pyogenes | 5.98 x 108CFU/mL | 100 (3/3) | 100 (3/3) |
| Candida albicans | 1.19 × 108CFU/mL | 100 (3/3) | 100 (3/3) |
| Bordetella pertussis | 4.90 x 109CFU/mL | 100 (3/3) | 100 (3/3) |
| Mycoplasma pneumoniae | 6.75 x 107CCU/mL | 100 (3/3) | 100 (3/3) |
| Chlamydia pneumoniae | 4.25 x 107CFU/mL | 100 (3/3) | 100 (3/3) |
| Legionella pneumophila | 9.20 x 109CFU/mL | 100 (3/3) | 100 (3/3) |
| Staphylococcus aureus | 5.00 × 106CFU/mL | 100 (3/3) | 100 (3/3) |
| Staphylococcus epidermidis | 1.75 × 108CFU/mL | 100 (3/3) | 100 (3/3) |
| Pooled human nasal wash | NA | 100 (3/3) | 100 (3/3) |
In silico evaluation of SARS-CoV-1 coronavirus, Human coronavirus HKU1, Mycobacterium tuberculosis, and Pneumocystis jirovecii was also conducted. The N protein sequence (GenBank ID: UUL70282.1) derived from an Omicron variant of SARS-CoV-2, belonging to the sublineages BA.1, BA.1.1, BA.2 (Accession Number OP160218) was used for the analysis.
K232377 - Page 7 of 14
{7}
There was no significant sequence homology with M. tuberculosis and P. jirovecii genomes, suggesting cross-reactivity would not occur; however, cross-reactivity cannot be ruled out. Homologous N protein sequences were identified in SARS-coronavirus, Human coronavirus HKU1, and MERS coronavirus. Human coronavirus HKU1 shares 41.18% to 49.00% homology or identity with the SARS-CoV-2 Omicron N protein sequence, and MERS coronavirus shares 52.87% to 54.04% identity. Homology is relatively low; however, cross-reactivity cannot be ruled out. The N protein sequence of SARS coronavirus shares 79.01% to 97.61% sequence identity with that of SARS CoV-2 Omicron variant indicating that cross-reactivity is likely. Therefore, the intended use states that it cannot distinguish SARS-CoV-2 and SARS-CoV.
## b. Endogenous/Exogenous Interference
An endogenous/exogenous substances interference study was conducted in which samples of 2x LoD inactivated SARS-CoV-2 virus (USA-WA1/2020) were prepared in negative clinical matrix and tested in the presence of potentially interfering substances on 3 test lots, to evaluate endogenous/exogenous interference on the Healgen Rapid COVID-19 Antigen Test. Fifty (50) µL of the prepared sample were applied to the kit swabs and eluted into the extraction buffer tube per the IFU. No interference was observed with the tested interfering substances at the tested concentration.
Table 4. Interfering Substances Study Results
| Substance | Active Ingredient | Concentration | % Positive Agreement | % Negative Agreement |
| --- | --- | --- | --- | --- |
| Whole Blood | | 4% | 100 (3/3) | 100 (3/3) |
| Human Leukocytes | | 1 × 10^{5} cells/mL | 100 (3/3) | 100 (3/3) |
| Mucin | | 0.5% | 100 (3/3) | 100 (3/3) |
| Chloroseptic | Menthol, Benzocaine | 3 mg/mL | 100 (3/3) | 100 (3/3) |
| NasoGEL (NeilMed) | Sodium Chloride, Sodium Bicarbonate, Sodium Hyaluronate | 5% | 100 (3/3) | 100 (3/3) |
| CVS Nasal Drops | Phenylephrine | 15% | 100 (3/3) | 100 (3/3) |
| Afrin Naal Spray | Oxymetazoline | 15% | 100 (3/3) | 100 (3/3) |
| CVS Nasal Spray | Cromolyn | 15% | 100 (3/3) | 100 (3/3) |
| ZICAM Cold Remedy + Multi-Symptom Relief | Galphimia glauca, luffa operculata, sabadilla | 5% | 100 (3/3) | 100 (3/3) |
| Homeopathic (Alkalol) | Galphima glauca, Luffa operculata, Sabadila 6X | 15% | 100 (3/3) | 100 (3/3) |
| Sore throat phenol spray | Phenol | 15% | 100 (3/3) | 100 (3/3) |
| Hand soap | | 1% | 100 (3/3) | 100 (3/3) |
| Hand sanitizer | | 1% | 100 (3/3) | 100 (3/3) |
| Tobramycin | Tobramycin | 4 µg/mL | 100 (3/3) | 100 (3/3) |
| Mupirocin | Mupirocin | 10 mg/mL | 100 (3/3) | 100 (3/3) |
| Fluticasone Propionate | Glucocorticoid | 15% | 100 (3/3) | 100 (3/3) |
| Tamiflu | Oseltamivir Phosphate | 5 mg/mL | 100 (3/3) | 100 (3/3) |
## c. High-dose Hook Effect Study
The purpose of the hook effect study is to evaluate if a hook effect occurs when high levels of the target analyte are present in the test sample. For this, 1:1, 1:10 and 1:100 dilutions of heat inactivated SARS-CoV-2 virus (USA-WA1/2020) were prepared in negative clinical matrix and tested on the Healgen Rapid COVID-19 Antigen Test. Fifty (50) µL of the
K232377 - Page 8 of 14
{8}
prepared sample were applied to the kit swabs and eluted into the extraction buffer tube per the IFU. The highest concentration tested was $5.75 \times 10^{6} \mathrm{TCID}_{50} / \mathrm{mL}$ , and no high-dose hook effect was observed.
Table 5. High-Dose Hook Effect Study Results
| Stock Dilution | Concentration | % Expected Agreement |
| --- | --- | --- |
| 1:1 | 5.75 × 106TCID50/mL | 100 (3/3) |
| 1:10 | 1.15 × 106TCID50/mL | 100 (3/3) |
| 1:100 | 1.15 × 105TCID50/mL | 100 (3/3) |
| Negative | 0 | 100 (3/3) |
# 4. Assay Reportable Range:
Not Applicable; the Healgen Rapid COVID-19 Antigen Test is a qualitative assay.
# 5. Traceability, Stability and Expected Values (Controls, Calibrators, or Methods):
# a. Internal Control
The Healgen Rapid COVID-19 Antigen Test has an internal, built-in procedural control. When sample is added and migrates up the test strip, a red line will appear in the control line region (C), which is coated with secondary antibody recognizing the primary antibody in the migrating sample. The control line should always appear in a test that functions accurately when the test procedure is followed.
# b. External Controls
The Healgen Rapid COVID-19 Antigen Test is required to be run with external quality controls as specified under the CLIA regulations. External quality controls are sold separately as the COVID-19 Antigen Control Kit. The Negative Control Swab is composed of negative control buffer dried onto a swab, and the Positive Control Swab also contains SARS-CoV-2 recombinant antigen dried onto a swab at a concentration of $0.1\mathrm{ng / mL}$ . Lot-to-lot reproducibility of the external positive and negative control swabs was evaluated with three (3) external control lots as follows: 4 replicates per lot of positive or negative external control swabs were tested on 3 different test kit lots for a total of 36 positive or negative external control results. All results were as expected.
Table 6. Summary Results of External Controls Validation
| Control | Control Lot # | Test Lot # | n | # of Neg | # of Pos | % Agreement |
| --- | --- | --- | --- | --- | --- | --- |
| Negative Control Swab | 1 | 1 | 4 | 4 | 0 | 100 |
| | | 2 | 4 | 4 | 0 | 100 |
| | | 3 | 4 | 4 | 0 | 100 |
| | 2 | 1 | 4 | 4 | 0 | 100 |
| | | 2 | 4 | 4 | 0 | 100 |
| | | 3 | 4 | 4 | 0 | 100 |
| | 3 | 1 | 4 | 4 | 0 | 100 |
| | | 2 | 4 | 4 | 0 | 100 |
| | | 3 | 4 | 4 | 0 | 100 |
K232377 - Page 9 of 14
{9}
| Control | Control Lot # | Test Lot # | n | # of Neg | # of Pos | % Agreement |
| --- | --- | --- | --- | --- | --- | --- |
| Positive Control Swab | 1 | 1 | 4 | 0 | 4 | 100 |
| | | 2 | 4 | 0 | 4 | 100 |
| | | 3 | 4 | 0 | 4 | 100 |
| | 2 | 1 | 4 | 0 | 4 | 100 |
| | | 2 | 4 | 0 | 4 | 100 |
| | | 3 | 4 | 0 | 4 | 100 |
| | 3 | 1 | 4 | 0 | 4 | 100 |
| | | 2 | 4 | 0 | 4 | 100 |
| | | 3 | 4 | 0 | 4 | 100 |
## c. Reagent Stability/Shelf-Life
**Healgen Rapid COVID-19 Antigen Test:** A real-time reagent stability study was conducted to support a shelf-life of 18 months. Three lots of the Healgen Rapid COVID-19 Antigen Test were stored at 2 to 8°C and 30 ± 3°C. At each timepoint test devices were visually inspected and tested with 5 replicates of a low positive sample (prepared at 2x LoD) and one replicate each of a high negative (prepared at 0.1x LoD) and a negative sample. Testing was performed at 0, 3, 6, 9, 12, 15, 18 and 20 months. All samples at all timepoints gave the expected results, supporting a shelf-life of the test device for 18 months under the intended storage condition.
**COVID-19 Antigen Control Kit:** Three lots of the COVID-19 Antigen Control Kit were stored at 2 ± 3°C and 30 ± 3°C and then each lot was tested in duplicate on 2 test kit lots at Time 0 and every 3 months to 27 months. All samples at all timepoints and storage conditions gave the expected results, supporting a shelf-life of the control kit for 24 months under the intended storage condition.
## d. Specimen Stability
A specimen stability study in which 3 replicates each of low positive (prepared at 2x LoD) and negative samples (NCM) were prepared and tested on 3 test lots. Fifty (50) µL of the prepared sample were applied to kit swab and incubated at room temperature (RT). Swabs were tested at 0, 2, 4 and 6 hours. All samples gave the expected result, demonstrating that swab specimens are stable up to 4 hours after collection when stored at RT.
Table 7. Results of Sample Stability Study
| | % Agreement (Correct Read /Total) | | | |
| --- | --- | --- | --- | --- |
| Sample | 0 hours | 2 hours | 4 hours | 6 hours |
| Low positive (2x LoD) | 100 (9/9) | 100 (9/9) | 100 (9/9) | 100 (9/9) |
| Negative (NCM) | 100 (9/9) | 100 (9/9) | 100 (9/9) | 100 (9) |
## 6. Detection Limit:
### a. Limit of Detection (Analytical Sensitivity)
For the limit of detection study, heat-inactivated SARS-CoV-2 virus, isolate 2019-nCoV/USA-WA1/2020, was prepared in negative clinical matrix (NCM) composed of leftover, RT-PCR negative clinical swab specimens collected from healthy volunteers, eluted
K232377 - Page 10 of 14
{10}
in 0.5 to 1mL saline or PBS, and pooled together. Testing was performed on 3 different test kit lots. A preliminary LoD was first established in a range-finding study using 3 replicates per concentration per lot of the Healgen Rapid COVID-19 Antigen Test. Thereafter, the LoD was confirmed using 20 replicates per lot of 1:1,000, 1:2,000, and 1:4,000 serial dilution preparations. For both studies, fifty (50) μL of the prepared sample were applied to kit swab and eluted into the extraction buffer tube prior to testing with the Healgen Rapid COVID-19 Antigen Test device per the IFU. A 98.3% (58/60) positive concurrence rate was observed at 1:2,000 sample, and the LoD was determined as 5.75 × 10³ TCID₅₀/mL.
Table 8. Summary Results of Limit of Detection Study
| | | | Range Finding | | Confirmatory | |
| --- | --- | --- | --- | --- | --- | --- |
| Dilution | Analyte Concentration (TCID50/mL) | Analyte per Swab (TCID50/swab) | # Positive | % Positive | # Positive | % Positive |
| 1:500 | 2.3×104 | 1.15×103 | 9/9 | 100.0% | | |
| 1:1000 | 1.15×104 | 5.75×102 | 9/9 | 100.0% | | |
| 1:2000 | 5.75×103 | 2.875×102 | 9/9 | 100.0% | | |
| 1:4000 | 2.875×103 | 1.438×102 | 4/9 | 44.4% | | |
| 1:1000 | 1.15×104 | 5.75×102 | | | 60/60 | 100.0% |
| 1:2000 | 5.75×103 | 2.875×102 | | | 58/60 | 96.7% |
| 1:4000 | 2.875×103 | 1.438×102 | | | 51/60 | 85.0% |
## b. WHO International Standard LoD
A study was performed to determine the LoD of the Healgen Rapid COVID-19 Antigen Test device with the WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in real clinical matrix. The results are shown below for range finding and confirmatory LoD studies, and the LoD for the WHO International Standard was determined to be 250 IU.
Table 9. Results of LoD Study with WHO SARS-CoV-2 Standard
| Dilution | Analyte Concentration (IU) | Analyte per Swab (IU/swab) | # Positive | % Positive | # Positive | % Positive |
| --- | --- | --- | --- | --- | --- | --- |
| 1:20 | 1000 | 50 | 3/3 | 100.0% | | |
| 1:40 | 500 | 25 | 3/3 | 100.0% | | |
| 1:80 | 250 | 12.5 | 3/3 | 100.0% | | |
| 1:160 | 125 | 6.25 | 2/3 | 66.6% | | |
| 1:40 | 500 | 25 | | | 20/20 | 100.0% |
| 1:80 | 250 | 12.5 | | | 20/20 | 100.0% |
| 1:160 | 125 | 6.25 | | | 13/20 | 65.0% |
## c. Inclusivity (Analytical Reactivity)
For the inclusivity study, commercially obtained sources of alpha (B.1.1.7), delta (B.1.617.2), omicron (B.1.1.529), beta (B.1.351), gamma (P1) and kappa (B.1.617.1) SARS-CoV-2 virus were tested to evaluate the ability of the Healgen Rapid COVID-19 Antigen Test device to detect a variety of SARS-CoV-2 variants. The study utilized a dilution series for each lineage with testing of each dilution in triplicate on 3 test lots. The lowest concentration that gave 100% positive results for each of the variants using the Healgen Rapid COVID-19 Antigen Test is provided below.
K232377 - Page 11 of 14
{11}
Table 10. Summary Results of Inclusivity Testing
| SARS-CoV-2 Variant | Lowest Concentration with 9/9 Positive Results (TCID_{50}/mL) |
| --- | --- |
| B.1.1.7 (Alpha) | 1.00×10^{2} |
| B.1.351 (Beta) | 3.83×10^{2} |
| B.1.617.2 (Delta) | 1.10×10^{2} |
| P1 (Gamma) | 6.30×10^{2} |
| B.1.617.1 (Kappa) | 1.90×10^{2} |
| B.1.1.529 (Omicron) | 2.51×10^{2} |
7. Assay Cut-Off:
Not Applicable; the Healgen Rapid COVID-19 Antigen Test device gives a non-numerical, visually read qualitative result and does not use an instrument.
B. Comparison Studies:
1. Method Comparison with Predicate Device:
Please refer to Section VII.C (Clinical Studies) below for the clinical validation.
2. Matrix Comparison:
Not Applicable; the Healgen Rapid COVID-19 Antigen Test device is only intended for use with direct, anterior nasal swabs, and no other sample type is claimed or presented.
C. Clinical Studies:
The clinical performance of the Healgen Rapid COVID-19 Antigen Test was evaluated in a multi-center, prospective study conducted from May 2022 to July 2022 using 32 untrained operators at six different CLIA-waived sites.
Testing is described as being performed in a simulated home environment. Enrolled subjects were aged 2 years or older symptomatic subjects within six (6) days post symptom onset (DPSO) and exhibiting symptoms at the time of collection. Symptomatic individuals were defined as exhibiting symptoms of fever or 2 or more of the following: chills, cough, shortness of breath/difficulty breathing, fatigue, muscle/body aches, headache, new loss of taste or smell, sore throat, congestion or runny nose, nausea or vomiting, and diarrhea. A total of 806 evaluable subjects were collected and tested. Patient demographics and specimen positivity are summarized in the following tables.
Table 11. Overview of Patient Demographics
| Age Group | Total # | Total Positive | Prevalence (%) |
| --- | --- | --- | --- |
| 2 to 13 years | 75 | 13 | 17.3 |
| 14 to 24 years | 118 | 19 | 16.1 |
| 25 to 64 years | 514 | 110 | 21.4 |
| ≥ 65 years | 99 | 22 | 22.2 |
K232377 - Page 12 of 14
{12}
Table 12. Overview of Specimen Positivity Based on DPSO
| DPSO | # Samples | # Positive Samples | % Positive |
| --- | --- | --- | --- |
| 0 | 26 | 3 | 11.5 |
| 1 | 92 | 16 | 17.4 |
| 2 | 213 | 46 | 21.6 |
| 3 | 219 | 39 | 17.8 |
| 4 | 150 | 29 | 19.3 |
| 5 | 75 | 19 | 25.3 |
| 6 | 31 | 12 | 38.7 |
Two swabs were collected from each subject, one anterior nares (AN) swab was tested on the candidate device, and one nasopharyngeal (NP) swab was collected into 3 mL of viral transport media (VTM), stored at 2-8°C and tested within 120 hours using a composite comparator method (or frozen at -70°C for long-term storage). For comparator testing, samples were tested on two highly sensitive RT-PCR comparator assays; testing on a third comparator would then be performed if discordant results were observed between the first two tests. The comparator result for those samples was determined based on a 2 out of 3 rule. Results obtained on the candidate device were compared to the composite comparator method to determine a clinical sensitivity/positive percent agreement (PPA) of 85.4% and a specificity/negative percent agreement (NPA) of 99.7%.
Table 13. Summary of Clinical Performance
| Candidate device | Composite Comparator Method | | |
| --- | --- | --- | --- |
| | Detected | Not Detected | Total |
| Reactive | 140 | 2 | 142 |
| Non-reactive | 24 | 640 | 664 |
| Total | 164 | 642 | 806 |
| Positive Percent Agreement: 85.4% (95% CI: 79.1 to 90.0%) | | | |
| Negative Percent Agreement: 99.7% (95% CI: 98.9 to 99.9%) | | | |
Table 14. Summary of Clinical Performance by DPSO
| DPSO | Number of Samples | Candidate Positive | Comparator Positive | PPA (%) |
| --- | --- | --- | --- | --- |
| Day 0 | 26 | 3 | 3 | 100.0 |
| Day 1 | 92 | 14 | 16 | 87.5 |
| Day 2 | 213 | 36 | 46 | 78.3 |
| Day 3 | 219 | 31 | 39 | 79.5 |
| Day 4 | 150 | 27 | 29 | 93.1 |
| Day 5 | 75 | 17 | 19 | 89.5 |
| Day 6 | 31 | 12 | 12 | 100 |
3. Clinical Sensitivity:
Please refer to Section VII.C (Clinical Studies) above for the clinical validation, regarding the test sensitivity/PPA. The PPA for the test is 85.4% (140/164; 95% CI: 79.1% - 90.0%).
K232377 - Page 13 of 14
{13}
4. Clinical Specificity:
Please refer to Section VII.C (Clinical Studies) above for the clinical validation, regarding the test specificity/NPA. The NPA for the test is 99.7% (640/642; 95% CI: 98.9% - 99.9%).
D. Clinical Cut-Off:
Not Applicable; there is no clinical cutoff for the presence of SARS-CoV-2 in patient samples.
E. Expected Values/Reference Range:
Not Applicable; a patient sample is expected to test negative for SARS-CoV-2.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K232377 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.