K232373 · Sofradim Production · FTL · Jan 18, 2024 · General, Plastic Surgery
Device Facts
Record ID
K232373
Device Name
Progrip™ Self-Gripping Polypropylene Mesh
Applicant
Sofradim Production
Product Code
FTL · General, Plastic Surgery
Decision Date
Jan 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ProGrip™ Self-Gripping Polypropylene Mesh is intended to be used for the reinforcement of abdominal wall soft tissues where weakness exists in procedures involving inguinal and ventral hernia repair by open approach.
Device Story
ProGrip™ Self-Gripping Polypropylene Mesh is a sterile, single-use surgical mesh for open inguinal and ventral hernia repair. Device consists of non-absorbable knitted monofilament polypropylene textile with resorbable polylactic acid (PLA) monofilament grips on one side. Grips facilitate mesh positioning and provide temporary fixation to surrounding tissue for at least eight weeks; PLA material resorbs via hydrolysis over 36-50 months. Surgeon places mesh extraperitoneally; self-gripping feature allows positioning without additional fixation, though suture/tacks may be used at surgeon discretion. Device provides long-term soft tissue reinforcement. Packaged in double sterile Tyvek® pouches and terminally sterilized via Ethylene Oxide.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), mechanical performance, sterilization validation (ISO 11135, AAMI TIR28), shelf-life testing, and shipping validation (ASTM D4169). Human factors evaluation confirmed no new critical tasks.
Technological Characteristics
Materials: Polypropylene (non-absorbable) and polylactic acid (resorbable) monofilament yarns. Construction: Knitted monofilament textile with one-sided grips. Pore size: 1.6±0.1 mm and 0.6±0.1 mm. Sterilization: Ethylene Oxide. Packaging: Double sterile barrier (Tyvek®/polyester/polyethylene). Form factor: Rectangular/square sheets (various sizes).
Indications for Use
Indicated for reinforcement of abdominal wall soft tissue where weakness exists, specifically for inguinal and ventral hernia repair in patients undergoing open surgical procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 18, 2024
Sofradim Production % Nancy Sauer Regulatory Affairs Senior Director Covidien Ilc 200 Medtronic Drive Lafayette, Colorado 80026
Re: K232373
Trade/Device Name: Progrip™ Self-Gripping Polypropylene Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: December 18, 2023 Received: December 18, 2023
Dear Nancy Sauer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Tek N.<br>Lamichhane -S | Digitally signed by Tek<br>N. Lamichhane -S<br>Date: 2024.01.18<br>15:36:39 -05'00' |
|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Tek Lamichhane, Ph.D. | |
| Assistant Director | |
| DHT4B: Division of Infection Control<br>and Plastic and Reconstructive Surgery Devices | |
| OHT4: Office of Surgical<br>and Infection Control Devices | |
| Office of Product Evaluation and Quality | |
| Center for Devices and Radiological Health | |
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K232373
Device Name
ProGrip™ Self-Gripping Polypropylene Mesh
Indications for Use (Describe)
ProGrip™ Self-Gripping Polypropylene Mesh is intended to be used for the reinforcement of abdominal wall soft tissues where weakness exists in procedures involving inguinal and ventral hernia repair by open approach.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right:5px">☒</span>Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="padding-left:15px; padding-right:5px">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{4}------------------------------------------------
# 510(k) Summary
| Date Prepared: | Jan 17, 2024 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Sofradim Production (subsidiary of Covidien Ilc<br>116, avenue du Formans<br>01600 Trevoux, France<br>Telephone: +33 (0)4 74 08 90 00<br>Fax: +33 (0) 4 74 08 90 02 |
| Contact: | Nancy Sauer<br>Regulatory Affairs Senior Director<br>200 Medtronic Dr.<br>Lafayette, CO 80026<br>Phone: 720-361-5290<br>Email: nancy.k.sauer@medtronic.com |
| Name of device: | |
| Trade/Proprietary name: | Progrip™ self-gripping polypropylene mesh |
| Common name: | Surgical Mesh |
| Classification name: | Mesh, Surgical, Polymeric<br>Product code: FTL<br>Regulation number: 21 CFR 878.3300 |
| Predicate Device: | |
| Trade/Proprietary name: | Progrip™ self-gripping polyester mesh |
| Common name: | Surgical Mesh |
| Classification name: | Mesh, Surgical, Polymeric<br>Product code: FTL<br>Regulation number: 21 CFR 878.3300 |
| 510(k) Number: | K220586 |
| Manufacturer: | Sofradim Production (subsidiary of Covidien Ilc)<br>116, avenue du Formans<br>01600 Trevoux, France |
| Reference Device: | |
| Trade/Proprietary name: | PROLENE® Soft Polypropylene Mesh |
| Common name: | Surgical Mesh |
| Classification name: | Mesh, Surgical, Polymeric<br>Product code: FTL<br>Regulation number: 21 CFR 878.3300 |
| 510(k) Number: | K172089 |
| Trade/Proprietary name: | Progrip™ self-gripping polypropylene mesh |
| Common name: | Surgical Mesh |
| Classification name: | Mesh, Surgical, Polymeric<br>Product code: FTL<br>Regulation number: 21 CFR 878.3300 |
| 510(k) Number: | K220540 |
| Manufacturer: | Sofradim Production (subsidiary of Covidien Ilc)<br>116, avenue du Formans<br>01600 Trevoux France |
{5}------------------------------------------------
Manufacturer: Ethicon, Inc. a Johnson & Johnson company P.O. Box 151 Route 22 West Somerville, NJ 08876-0151
### Reference Device:
Progrip™ self-gripping polypropylene mesh is designed to allow Device Description: extraperitoneal mesh placement for the repair of inquinal and ventral hernias by open approach.
Progrip™ self-gripping polypropylene mesh is a sterile non-pyrogenic device made of a non-absorbable knitted monofilament polypropylene textile with resorbable polylactic acid (PLA) monofilament grips on one side.
Progrip™ self-gripping polypropylene mesh is available in different shapes and sizes.
The non-absorbable textile is designed to ensure long term reinforcement of soft tissues.
The monofilament polylactic acid grips facilitate placing and positioning the mesh, and they contribute to fixation of the mesh to the surrounding tissue for at least eight (8) weeks. The polylactic acid grips are bioresorbable. Over the time, they resorb in vivo by hydrolysis and are metabolized by the body into CO2 and H2O. Preclinical studies showed that the polylactic acid material is essentially resorbed in 36 to 50 months post-implantation. However, the resorption period depends on numerous factors including patient-related factors.
Progrip™ self-gripping polypropylene mesh is a single use device, presented in a double sterile barrier packaging (two Tyvek® pouches). The packed device is terminally sterilized by Ethylene Oxide (EtO) and placed into a commercial envelope with the Instructions for Use (IFU) and Patient Implant Card (PIC). All the devices are packaged unitary in a commercial envelope (single pack configuration: 1 unit per commercial envelope).
Intended Use: Reinforcement of soft tissue where weakness exists.
{6}------------------------------------------------
#### Indications for use:
Progrip™ self-gripping polypropylene mesh is intended for use in reinforcement of abdominal wall soft tissue where weakness exists, in procedures involving inguinal and ventral hernia repair by open approach.
Summary comparing the technological characteristics of the subject and predicate device:
The subject device is substantially equivalent to Progrip™ self-gripping polyester mesh (K220586), which is the primary predicate device.
| | Progrip™ self-<br>gripping<br>polypropylene mesh<br>K232373<br>(subject device) | Progrip™ self-<br>gripping polyester<br>mesh (K220586)<br>(predicate device) | Comparison |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use | Reinforcement of soft<br>tissue where a<br>weakness exists | Reinforcement of soft<br>tissue where<br>weakness exists. | Same |
| Classification | FTL, per 21 CFR<br>878.3300 | FTL, per 21 CFR<br>878.3300 | Same |
| Indications for<br>use | Inguinal and ventral<br>hernias repair | Inguinal and incisional<br>hernias repair | Different |
| Method of<br>Insertion | Open | Open and<br>Laparoscopic<br>(conventional and<br>robotically assisted<br>approach) | Equivalent |
| Location of<br>placement | Extraperitoneal | Extraperitoneal | Same |
| Defect Closure | Recommended | Recommended | Same |
| Materials | Surgical mesh, textile<br>based, knitted,<br>monofilament,<br>synthetic made of<br>polypropylene and<br>polylactic acid yarns | Surgical mesh, textile<br>based, knitted,<br>monofilament,<br>synthetic made of<br>polyester and<br>polylactic acid yarns<br>only for elliptic<br>(precut) codes:<br>polyester yarn marker<br>dyed with D&C blue<br>no. 6. | Different |
| Shape | Rectangular and<br>square | Rectangular and<br>square<br>Elliptic | Equivalent |
| Sizes | Rectangular and<br>square:<br>15x15cm<br>20x15cm<br>30x20 cm<br>30x30cm<br>40x30cm<br>The mesh may be<br>recut as needed | - Rectangular:<br>15x9 cm<br>15x15 cm<br>20x15 cm<br>30x15 cm<br><br>- Elliptic, slit with<br>overlapping flap left<br>side/right side: 12x8<br>cm<br>The mesh may be<br>recut as needed | Different |
| | | | |
| Texture | Two-dimensional<br>monofilament textile<br>with monofilament<br>absorbable grips on<br>one side of the mesh | Two-dimensional<br>monofilament textile<br>with monofilament<br>absorbable grips on<br>one side of the mesh | Same |
| Pore size | $1.6\pm0,1$ mm and<br>$0.6\pm0,0$ mm | $1.6\pm0,1$ mm and<br>$0.6\pm0,1$ mm | Same |
| Fixation<br>instructions | The technique used to fixate the mesh (suture<br>and/or tacks) is left up to the surgeon.<br>Compatibility with other fixation devices may<br>not have been established. Using means of<br>fixation other than those for which compatibility<br>is established may lead to mesh damage. If<br>tacks are used to fixate the mesh, the use of<br>CovidienTM fixation devices is recommended. It<br>is suggested to fixate the mesh at a distance of<br>approximately 1 cm from the edge of the mesh.<br>The textile self-gripping feature makes it<br>possible to position the mesh without fixation,<br>depending on the size of the defect, the hernia<br>position and the quality of the anatomical<br>structures. | The technique used to fixate the mesh (suture<br>and/or tacks) is left up to the surgeon.<br>Compatibility with other fixation devices may<br>not have been established. Using means of<br>fixation other than those for which compatibility<br>is established may lead to mesh damage. If<br>tacks are used to fixate the mesh, the use of<br>CovidienTM fixation devices is recommended. It<br>is suggested to fixate the mesh at a distance of<br>approximately 1 cm from the edge of the mesh.<br>The textile self-gripping feature makes it<br>possible to position the mesh without fixation,<br>depending on the size of the defect, the hernia<br>position and the quality of the anatomical<br>structures. | Same |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Same |
| Packaging | Sealed Tyvek®//<br>polyester/polyethylene<br>pouch (double sterile<br>barrier system) | Sealed Tyvek®//<br>polyester/polyethylene<br>pouch (double sterile<br>barrier system) | Same |
{7}------------------------------------------------
The subject device and predicate device have the same intended use and nearly identical indications for use, differing only in the addition of primary ventral hernia repair for the subject device. Differences in the technological characteristics between the subject and predicate devices do not raise new questions of safety or effectiveness. Performance testing demonstrated substantial equivalence for biocompatibility, has mechanical properties such as strength, human factors evaluation, sterilization, packaging, transport and shelf life.
This 510(k) also references the PROLENE® Soft Polypropylene Mesh (K172089) and the Progrip™ self-gripping polypropylene mesh (K220540). PROLENE® Soft Polypropylene Mesh (K172089) was selected because this legally marketed mesh includes large sizes (up to 50x50cm). Progrip™ self-gripping polypropylene mesh (K220540) was selected because this legally marketed mesh presents the same base materials as the subject device.
Performance data: The following performance data is provided in support of substantial equivalence demonstration:
- . Sterilization: the sterilization process has been validated in accordance with ISO 11135 (2014) and AAMI TIR28 (2016) standards.
{8}------------------------------------------------
- Shelf-life: the 5-year shelf life has been demonstrated for the subject ● device.
- Shipping test: shipping test was performed in accordance with the . ASTM D4169 (2022) standard.
- Biocompatibility: biocompatibility evaluation was performed for the ● subject Progrip™ self-gripping polypropylene mesh in accordance with the FDA quidance "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued September 4, 2020, and international standard ISO 10993-1 (2018). The subject device meets the criteria for biocompatibility set forth in the FDA guidance and ISO standard.
- Performance testing In vitro (bench) tests has been performed to . compare the subject and predicate in accordance with the FDA Guidance "Guidance for the Preparation of a Premarket Notification Application of a Surgical Mesh" issued March 2, 1999. Results demonstrate that physical and mechanical performance of the subject are substantially equivalent to the predicate.
- Human factors evaluation was conducted. A comparison of the critical tasks of the subject device and the predicate device was performed. The critical tasks are the same and there are no critical tasks introduced and no existing critical tasks have been impacted. Therefore, human factors data are not needed to demonstrate substantial equivalence.
This premarket submission does not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
#### Conclusion:
Comparison of the subject and predicate devices and labeling as well as the results of performance testing demonstrate that the subject Progrip™ self-gripping polypropylene mesh is substantially equivalent to the predicate device Progrip™ self-aripping polvester mesh (K220586).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.