Copan Universal Transport Medium (UTM-RT®) System is intended for the collection and transport of clinical specimen containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT® can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. UTM-RT® is intended for the stabilization and transportation of an unprocessed upper respiratory clinical specimen suspected of containing respiratory viruses' nucleic acids. UTM-RT® is intended for use with compatible molecular assays.
Device Story
System consists of tube containing 3mL transport medium, optionally supplied with flocked nylon swab; used for collection/transport of clinical specimens (viruses, chlamydiae, mycoplasma, ureaplasma) from collection site to laboratory. Medium maintains microorganism viability for culture and preserves nucleic acid integrity for molecular assays. Used in clinical settings; operated by healthcare professionals. Output is stabilized specimen in transport medium; processed by laboratory staff using standard culture or molecular diagnostic procedures. Benefits include maintained specimen viability and nucleic acid stability during transport, facilitating accurate downstream diagnostic testing.
Clinical Evidence
No clinical data. Bench testing only. Stability studies evaluated Influenza A, Influenza B, and RSV nucleic acid recovery in simulated nasopharyngeal/nasal fluid matrix. Samples stored at 2–8°C and 22–28°C for up to 96 hours. Performance measured by cycle threshold (Ct) values using an FDA-cleared molecular assay. Acceptance criteria (ΔCt < 3) met for all tested lots, including new, middle-aged, and expired (up to 18 months) media.
Technological Characteristics
Polypropylene tube with screw cap; 3mL transport medium (Modified Hank's Balanced Salts, HEPES, BSA, antibiotics/antifungals, L-cysteine, gelatin, phenol red, sucrose, L-glutamic acid). Nylon-flocked swab on plastic shaft with break point. Optional glass beads. Non-automated, standalone collection/transport device. Sterilized. Compatible with molecular assays and culture.
Indications for Use
Indicated for collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma, or ureaplasma, and stabilization of unprocessed upper respiratory specimens suspected of containing respiratory viral nucleic acids for use with compatible molecular assays.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
Merit Medical Systems Cultura Collection and Transport System (K201674)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232357
B Applicant
Copan Italia S.p.A.
C Proprietary and Established Names
Copan Universal Transport Medium (UTM-RT) System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM, LIO | Class I, reserved | 21 CFR 866.2390 – Transport Culture Medium | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for the Copan Universal Transport Medium (UTM-RT) for a new Intended Use claim of stabilization and transportation of an unprocessed upper respiratory clinical specimen suspected of containing respiratory viruses' nucleic acids (to add on to the existing Intended Use of the previously cleared device) under the same device name.
B Measurand:
Not applicable
C Type of Test:
Non-propagating Transport Device with culture medium
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimen containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture.
UTM-RT is intended for the stabilization and transportation of an unprocessed upper respiratory clinical specimen suspected of containing respiratory viruses' nucleic acids. UTM-RT is intended for use with compatible molecular assays.
C Special Conditions for Use Statement(s):
Rx – For Prescription Use Only
Do not freeze specimens intended to be processed for nucleic acids investigations. Stability of nucleic acids when frozen in UTM-RT has not been evaluated and validated.
D Special Instrument Requirements:
Not Applicable
IV Device/System Characteristics:
A Device Description:
The candidate device, Copan UTM-RT, is provided in a screw capped, conical bottom, polypropylene tube filled with 3mL of UTM-RT transport medium. Copan UTM-RT is supplied in either of two (2) configurations, namely, medium tube alone in a paper-film peel pouch (bulk configuration), or medium tube accompanied by sterile flocked swab in a peel pouch for specimen collection (kit configuration). In the kit configuration, the flocked swab is available in three (3) shapes—i.e., regular, minitip, and minitip flexible—that are suitable for specimen collection from various sites on patient, including the upper respiratory cavity. Further, the medium tube for both the swab configurations above can be provided with or without glass beads inside. The medium tube with beads will be distinguished from tube without beads by the wording “with beads” written in the product description on tube and external labels, as well as in other labeling materials.
The screw cap of the UTM-RT medium tube has a unique internal molded design that is able to capture a swab shaft when the shaft is broken off into the tube and cap is closed. The flocked specimen collection swab (provided sterile in a paper-peel pouch with UTM-RT medium in the kit configuration) has a tip flocked with nylon fiber and a solid plastic shaft with a molded breaking point. The nylon flocked swab along with its plastic shaft is the only component that may come in transient (typically < 1 minute) contact with patient tissue (e.g., mucosal surfaces) during use in specimen collection. After the specimen is collected, the swab shaft is broken off at the breaking point such that the swab with the collected specimen remains within the medium tube.
B Principle of Operation:
The UTM-RT medium is designed to maintain the integrity and viability of viruses, chlamydiae, mycoplasma, or ureaplasma and the integrity of nucleic acids for upper respiratory viruses.
K232357 - Page 2 of 10
{2}
Composition of the UTM-RT medium:
Modified Hank's Balanced Salts
BEPEs Buffer 1M
Bovine Serum Albumin
Vancomycin
L-Cysteine Hydrochloride
Amphotericin B
Sodium Bicarbonate
Colistin
Gelatin
Phenol Red
Sucrose
Distilled Water
L-Glutamic Acid
Specimens preserved in UTM-RT transport media are suitable for testing with subsequent culture procedure (all clinical specimens) or molecular amplification (limited to upper respiratory clinical specimens suspected of containing nucleic acids from respiratory viruses).
V Substantial Equivalence Information:
A Predicate Device Name(s):
Merit Medical Systems Cultura Collection and Transport System
B Predicate 510(k) Number(s):
K201674
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device (K232357) | Predicate (K201674) |
| --- | --- | --- |
| Device Trade Name | Copan Universal Transport Medium (UTM-RT) System | Merit Medical Systems Cultura Collection and Transport System |
| Manufacturer | Copan Italia S.p.A. | Merit Medical Systems, Inc. |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimen containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. UTM-RT is intended for the stabilization and transportation of an unprocessed upper respiratory clinical specimen suspected of containing respiratory viruses' nucleic acids. UTM-RT is intended for use with compatible molecular assays. | The Merit Cultura Collection and Transport System is intended for collection and transport of clinical specimens to the laboratory for standard diagnostic/identification techniques. The Merit Cultura Collection and Transport System is a culture-based media that can be used for upper respiratory viral diagnostic assays including Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), Influenza A, Influenza B, Respiratory Syncytial Virus (RSV), and Rhinovirus. |
K232357 - Page 3 of 10
{3}
| Device & Predicate Device(s): | Device (K232357) | Predicate (K201674) |
| --- | --- | --- |
| Tube and cap material | Plastic | Same |
| Swab material | Various flocked nylon applicators on a plastic shaft with break point | Same |
| Media Volume | 3 mL | Same |
| Single use device | Yes | Same |
| Principle of operation | Culture-based medium for maintenance of target microorganism viability and integrity. The specimen can be tested with culture- based technique and/or with PCR-based technique. | Same |
| General Device Characteristic Differences | | |
| Shelf-life | 18 months | 12 months |
| Media list of ingredients | Hank’s Balanced Salt Solution, bovine serum albumin, L-cysteine, gelatin, L-glutamic acid, HEPES buffer, phenol red, sucrose, vancomycin, amphotericin B and colistin | Hank’s Balanced Salt Solution (HBSS), Fetal Bovine Serum (FBS), D-glucose, Phenol Red, Gentamicin sulfate and Amphotericin B |
| Specimen transport conditions and storage time | Up to 96 hours at 2–25°C. | Up to 120 hours at 2–25°C. |
VI Standards/Guidance Documents Referenced:
Not applicable to the current studies.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility; Linearity; Analytical Specificity/Interference; Assay Reportable Range; Traceability; Expected Values (Controls, Calibrators, or Methods); Assay Cut-off:
Not Applicable
2. Stability/Shelf life:
The previously cleared device (K042970) has a shelf-life of 18-months. For the candidate device, no changes were made to the manufacturing of the UTM-RT transport medium or kit components, and no extension of the shelf-life was claimed. The existing shelf-life stability validation data from the original clearance remains appropriate to support an 18-month shelf life claim for the candidate device.
K232357 - Page 4 of 10
{4}
3. Detection Limit:
# Performance Testing - Culture Recovery Studies:
Performance of the Copan Universal Transport Medium (UTM-RT) System was evaluated by culture-based recovery studies for viral and bacterial test strains under and the results were deemed acceptable during the clearance of the original device, K042970. Since the medium tube (container), UTM-RT media composition, and other kit components remain unchanged, and no extension of the specimen stability is being claimed, a new set of performance evaluation for specimen stability via culture recovery is not required for the current candidate device (K232357).
# Performance Testing - Nucleic Acid Stability Studies:
To support the new intended use claim, "UTM-RT is intended for the stabilization and transportation of an unprocessed upper respiratory clinical specimen suspected of containing respiratory viruses' nucleic acids. UTM-RT is intended for use with compatible molecular assays", the sponsor conducted fresh studies to determine the LoD of molecular detection of three (3) representative target viral nucleic acids—namely, Influenza A, Influenza B, and Respiratory Syncytial Virus (RSV)—followed by specimen stability studies in the candidate transport medium system. These stability studies, performed both in cold $(2 - 8^{\circ}\mathrm{C})$ and at room temperature $(22 - 28^{\circ}\mathrm{C})$ , utilized variously aged lots (as available) of UTM-RT provided in medium tubes with and without glass beads as shown below in Table 1.
Table 1: Use of UTM-RT media lots in stability studies
| UTM-RT | Number of lots used (as available) | Lot age designation | Lot age (in months from production) at testing |
| --- | --- | --- | --- |
| With Beads | Four (4) | New | 1–4.2 |
| | Two (2) | Middle-Aged | 12.9–13.5 |
| | Nine (9) | Old / Expired | 20.3–22.7 |
| Without Beads | Three (3) | New | 0.8–1.4 |
| | Three (3) | Middle-Aged | 11.1–13.3 |
| | Three (3) | Old / Expired | 18.7–19.6 |
Further, the LoD and stability studies with the candidate device were performed in an artificial nasopharyngeal/nasal (referred to as "NPS/NS") fluid matrix. This simulated matrix was previously qualified as an appropriate matrix when used with Copan UTM/VTM since this was shown to be equivalent to clinical nasopharyngeal and anterior nasal matrices in Sample Matrix and Transport Media Equivalency studies performed to support clearance of the Cepheid Xpert Xpress CoV-2/Flu/RSV plus Assay (K231481). Therefore, the use of the same simulated matrix to support this transport medium candidate device was deemed appropriate, and the studies were designed and conducted accordingly.
As shown in the following summary Tables 2-4, the freshly performed stability tests with representative viral targets (as indicated above) demonstrated specimen stability for up to 96 hours during storage and/or transport at a temperature range of $2 - 25^{\circ}\mathrm{C}$ in UTM-RT tubes with or without glass beads. Samples were tested using a compatible, representative, FDA-cleared molecular assay (i.e., Cepheid Xpert Xpress CoV-2/Flu/RSV plus Assay in combination with GeneXpert Dx) following the molecular assay's package insert. The cycle threshold (Ct) values were used to assess performance and calculate the difference in Ct values $(\Delta \mathrm{Ct})$ at tested time
K232357 - Page 5 of 10
{5}
points relative to baseline (Time 0). Results were considered acceptable if a $\Delta \mathrm{Ct} < 3$ was observed. Additionally, these studies established acceptable performance of the candidate device for UTM-RT medium aged up to 18 months by incorporating lots of different ages at the time of testing.
Table 2: Overall stability data for Influenza A1/A2 virus in lots of UTM-RT per specimen storage time and temperature, all tests PASSED preset acceptance criteria $(\Delta Ct < 3)$ .
| Virus | Beads | Lot age of UTM-RT at testing | UTM-RT lot | Time point | ΔCt* (average Time x - T0) at 22-28°C storage | ΔCt* (average Time x - T0) at 2-8°C storage |
| --- | --- | --- | --- | --- | --- | --- |
| Flu A1 | YES | New | 2400409 LAB | 72 hrs. | 0.6 | -0.4 |
| | | | | 96 hrs. | 0.6 | 0.4 |
| | | | 2319013 | 72 hrs. | 0.6 | -0.1 |
| | | | | 96 hrs. | 0.4 | 0.7 |
| | | | 2318982 | 72 hrs. | 0.6 | 0.5 |
| | | | | 96 hrs. | 0.8 | 0.5 |
| | | | 2317477 | 72 hrs. | 0.3 | 0.3 |
| | | | | 96 hrs. | 0.9 | 0.7 |
| | | Middle-Aged | 2301907 | 72 hrs. | -0.2 | 0 |
| | | | | 96 hrs. | 0.3 | 0.2 |
| | | | 2301790 | 72 hrs. | 0.9 | 0.5 |
| | | | | 96 hrs. | 0.9 | 0.5 |
| | | Old / Expired | 2211427 | 72 hrs. | 0.2 | 0.2 |
| | | | | 96 hrs. | 0.5 | 0.1 |
| | | | 2211110 | 72 hrs. | 0.6 | 0.1 |
| | | | | 96 hrs. | 0.5 | 0.3 |
| | | | 2207355 | 72 hrs. | 0.1 | 0.1 |
| | | | | 96 hrs. | 0.4 | 0 |
| | NO | New | B400273 | 72 hrs. | 0.9 | 0.5 |
| | | | | 96 hrs. | 0.8 | 0.8 |
| | | | B400416 | 72 hrs. | 0.7 | 0.4 |
| | | | | 96 hrs. | 0.8 | 0.6 |
| | | | B400266 | 72 hrs. | 1 | 0.5 |
| | | | | 96 hrs. | 1 | 0.8 |
| | | Middle-Aged | B300767 | 72 hrs. | 0.7 | 0.4 |
| | | | | 96 hrs. | 1.4 | 0.9 |
| | | | B302525 | 72 hrs. | 0.3 | 0 |
| | | | | 96 hrs. | 0.7 | 0.2 |
| | | | B302734 | 72 hrs. | 0.7 | 0.6 |
| | | | | 96 hrs. | 1.1 | 0.8 |
| | | Old / Expired | B203103 | 72 hrs. | 0.5 | 0.4 |
| | | | | 96 hrs. | 0.6 | 0.3 |
| | | | B203226 | 72 hrs. | 0.5 | -0.1 |
| | | | | 96 hrs. | 0.6 | 0.4 |
| | | | B203262 | 72 hrs. | 0.3 | 0.1 |
| | | | | 96 hrs. | 0.4 | 0 |
| | YES | New | | 72 hrs. | 0.5 | -0.3 |
K232357 - Page 6 of 10
{6}
| Virus | Beads | Lot age of UTM-RT at testing | UTM-RT lot | Time point | ΔCt* (average Time x - T0) at 22-28°C storage | ΔCt* (average Time x - T0) at 2-8°C storage |
| --- | --- | --- | --- | --- | --- | --- |
| Flu A2 | | | 2400409 LAB | 96 hrs. | 0.4 | 0.5 |
| 2319013 | 72 hrs. | 0.4 | 0.1 |
| 96 hrs. | 0.3 | 0.6 |
| 2318982 | 72 hrs. | 0.5 | 0.7 |
| 96 hrs. | 0.6 | 0.7 |
| 2317477 | 72 hrs. | 0.2 | 0.2 |
| 96 hrs. | 0.6 | 0.7 |
| Middle-Aged | 2301907 | 72 hrs. | -0.1 | 0.1 |
| 96 hrs. | 0.3 | 0.2 |
| 2301790 | 72 hrs. | 0.7 | 0.4 |
| 96 hrs. | 0.8 | 0.5 |
| Old / Expired | 2211427 | 72 hrs. | 0.2 | 0.3 |
| 96 hrs. | 0.5 | 0.2 |
| 2211110 | 72 hrs. | 0.6 | 0.3 |
| 96 hrs. | 0.6 | 0.4 |
| 2207355 | 72 hrs. | 0.2 | 0.3 |
| 96 hrs. | 0.6 | 0.2 |
| NO | New | B400273 | 72 hrs. | 0.7 | 0.3 |
| 96 hrs. | 0.6 | 0.8 |
| B400416 | 72 hrs. | 0.5 | 0.3 |
| 96 hrs. | 0.8 | 0.7 |
| B400266 | 72 hrs. | 0.7 | 0.4 |
| 96 hrs. | 0.7 | 0.6 |
| Middle-Aged | B300767 | 72 hrs. | 0.5 | 0.4 |
| 96 hrs. | 1.1 | 0.8 |
| B302525 | 72 hrs. | 0.2 | 0.1 |
| 96 hrs. | 0.5 | 0.3 |
| B302734 | 72 hrs. | 0.6 | 0.6 |
| 96 hrs. | 1 | 0.8 |
| Old / Expired | B203103 | 72 hrs. | 0.3 | 0.4 |
| 96 hrs. | 0.5 | 0.4 |
| B203226 | 72 hrs. | 0.6 | -0.1 |
| 96 hrs. | 0.7 | 0.6 |
| B203262 | 72 hrs. | 0.3 | 0 |
| 96 hrs. | 0.3 | 0 |
Note: The average Ct value is calculated for each time point X hrs. (Time $x$ ) starting with time 0 ( $T_0$ ) and the difference ( $\Delta Ct$ ) at each Time $x$ is then estimated relative to the average Ct at $T_0$ , i.e., (Time $x - T_0$ ), as indicated by the Asterisk (*).
K232357 - Page 7 of 10
{7}
Table 3: Overall stability data for Influenza B virus in lots of UTM-RT per specimen storage time and temperature, all tests PASSED preset acceptance criteria ( $\Delta Ct < 3$ ).
| Virus | Beads | Lot age of UTM-RT at testing | UTM-RT lot | Time point | ΔCt* (average Time x - T₀) at 22–28°C storage | ΔCt* (average Time x - T₀) at 2–8°C storage |
| --- | --- | --- | --- | --- | --- | --- |
| Flu B | YES | New | 2400409
LAB | 72 hrs. | -1.2 | -0.8 |
| | | | | 96 hrs. | -0.5 | -0.6 |
| | | | 2319013 | 72 hrs. | 1 | -0.4 |
| | | | | 96 hrs. | 1 | 0.1 |
| | | | 2318982 | 72 hrs. | 0.4 | 0.4 |
| | | | | 96 hrs. | 1 | 0 |
| | | | 2317477 | 72 hrs. | 0.1 | -0.2 |
| | | | | 96 hrs. | 0.8 | -0.4 |
| | | Middle-Aged | 2301907 | 72 hrs. | 0 | -0.4 |
| | | | | 96 hrs. | 0 | 0.2 |
| | | | 2301790 | 72 hrs. | 0 | -0.5 |
| | | | | 96 hrs. | -0.8 | -0.4 |
| | | Old / Expired | 2211429 | 72 hrs. | 0.4 | -0.6 |
| | | | | 96 hrs. | -0.4 | -1 |
| | | | 2212455 | 72 hrs. | -0.2 | -0.8 |
| | | | | 96 hrs. | -0.1 | -1 |
| | | | 2208203 | 72 hrs. | 0.5 | 0.2 |
| | | | | 96 hrs. | 0.2 | -0.1 |
| | NO | New | B400273 | 72 hrs. | 1.1 | 0.3 |
| | | | | 96 hrs. | 1 | -0.2 |
| | | | B400416 | 72 hrs. | 0 | 0.3 |
| | | | | 96 hrs. | -0.2 | -0.1 |
| | | | B400266 | 72 hrs. | 0.8 | 0.3 |
| | | | | 96 hrs. | -0.4 | 0 |
| | | Middle-Aged | B300767 | 72 hrs. | 0.8 | -0.1 |
| | | | | 96 hrs. | 0.8 | 0.4 |
| | | | B302525 | 72 hrs. | 0.3 | -0.6 |
| | | | | 96 hrs. | 0.2 | -0.6 |
| | | | B302734 | 72 hrs. | 1 | 0.5 |
| | | | | 96 hrs. | 0 | -0.4 |
| | | Old / Expired | B203103 | 72 hrs. | 0.4 | 0 |
| | | | | 96 hrs. | 0.1 | 0.1 |
| | | | B203226 | 72 hrs. | 0.3 | -0.1 |
| | | | | 96 hrs. | -0.8 | -0.5 |
| | | | B203262 | 72 hrs. | 1.1 | 0.1 |
| | | | | 96 hrs. | 0.1 | -0.3 |
Note: The average Ct value is calculated for each time point X hrs. (Time x) starting with time 0 (T₀) and the difference (ΔCt) at each Time x is then estimated relative to the average Ct at T₀, i.e., (Time x - T₀), as indicated by the Asterisk (*).
K232357 - Page 8 of 10
{8}
Table 4: Overall stability data for Respiratory Syncytial virus (RSV) in lots of UTM-RT per specimen storage time and temperature, all tests PASSED preset acceptance criteria ( $\Delta Ct < 3$ ).
| Virus | Beads | Lot age of UTM-RT at testing | UTM-RT lot | Time point | ΔCt* (average Time x - T0) at 22-28°C storage | ΔCt* (average Time x - T0) at 2-8°C storage |
| --- | --- | --- | --- | --- | --- | --- |
| RSV | YES | New | 2400409 LAB | 72 hrs. | 0.2 | -0.1 |
| | | | | 96 hrs. | 0.2 | -0.4 |
| | | | 2319013 | 72 hrs. | 0.4 | 0.2 |
| | | | | 96 hrs. | 0.1 | 0.2 |
| | | | 2318982 | 72 hrs. | -0.3 | 0 |
| | | | | 96 hrs. | 0.1 | -0.2 |
| | | | 2317477 | 72 hrs. | 0.5 | 0.3 |
| | | | | 96 hrs. | 0.6 | 0.2 |
| | | Middle-Aged | 2301907 | 72 hrs. | 0.4 | 0.4 |
| | | | | 96 hrs. | 0.4 | 0.4 |
| | | | 2301790 | 72 hrs. | -0.1 | 0 |
| | | | | 96 hrs. | 0.1 | -0.3 |
| | | Old / Expired | 2211426 | 72 hrs. | 0.6 | 0.1 |
| | | | | 96 hrs. | 0.5 | 0.2 |
| | | | 2212454 | 72 hrs. | 0.4 | 0.1 |
| | | | | 96 hrs. | 0.2 | 0.1 |
| | | | 2208079 | 72 hrs. | 0.2 | 0 |
| | | | | 96 hrs. | -0.1 | -0.3 |
| | NO | New | B400273 | 72 hrs. | -0.2 | 0 |
| | | | | 96 hrs. | 0.2 | -0.2 |
| | | | B400416 | 72 hrs. | -0.2 | 0 |
| | | | | 96 hrs. | 0 | -0.3 |
| | | | B400266 | 72 hrs. | 0.1 | 0.1 |
| | | | | 96 hrs. | 0.2 | 0 |
| | | Middle-Aged | B300767 | 72 hrs. | -0.1 | 0 |
| | | | | 96 hrs. | -0.1 | 0 |
| | | | B302525 | 72 hrs. | 0.3 | -0.1 |
| | | | | 96 hrs. | 0.5 | 0.4 |
| | | | B302734 | 72 hrs. | -0.1 | 0.3 |
| | | | | 96 hrs. | 0 | 0.1 |
| | | Old / Expired | B203103 | 72 hrs. | -0.2 | -0.3 |
| | | | | 96 hrs. | -0.1 | -0.2 |
| | | | B203226 | 72 hrs. | 0.9 | 1.1 |
| | | | | 96 hrs. | 1.1 | 1 |
| B203262 | | | 72 hrs. | 0.2 | 0.1 | |
| | | | 96 hrs. | 0.4 | 0.4 | |
Note: The average Ct value is calculated for each time point X hrs. (Time $x$ ) starting with time 0 ( $T_0$ ) and the difference ( $\Delta Ct$ ) at each Time $x$ is then estimated relative to the average Ct at $T_0$ , i.e., (Time $x - T_0$ ), as indicated by the Asterisk (*).
Of note, the performance parameters established in these studies pertain exclusively to the newly introduced claim of stability of respiratory viral nucleic acids in the candidate medium
K232357 - Page 9 of 10
{9}
for subsequent testing in a compatible molecular assay. Especially, frozen storage of specimens in UTM-RT at or below $-70^{\circ}\mathrm{C}$ for testing with a molecular assay has not been validated and is not recommended. However, it is acceptable to store specimens in UTM-RT at or below $-70^{\circ}\mathrm{C}$ for subsequent processing using virus, chlamydiae, mycoplasma, and ureaplasma culture, as was previously cleared (K042970). This distinction is noted in device labeling.
## B Comparison Studies:
Method Comparison with Predicate Device; Matrix Comparison:
Not Applicable
## C Clinical Studies:
Clinical Sensitivity; Clinical Specificity; Other Clinical Supportive Data (if applicable):
Not Applicable
## D Clinical Cut-Off:
Not Applicable
## E Expected Values/Reference Range:
Not Applicable
## VIII Proposed Labeling:
The provided labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a decision of substantial equivalence.
K232357 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.