← Product Code [FDS](/productcode/FDS) · K232314

# Hood (DH-106STL, DH-116STL, DH-126STL, DH-096ST) (K232314)

_FUJIFILM Healthcare Americas Corporation · FDS · Sep 1, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232314

## Device Facts

- **Applicant:** FUJIFILM Healthcare Americas Corporation
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Sep 1, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

This hood is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection

## Device Story

Endoscopic accessory hood; attaches to distal end of compatible endoscopes. Comprised of attaching portion, distal portion, and drain portion. Used during endoscopic procedures (e.g., mucosal resection) to maintain field of view by preventing fluid accumulation on endoscope surface. Operated by clinicians in clinical settings. Provides physical barrier/spacer to ensure clear visualization. Benefits patient by facilitating procedural accuracy and safety during endoscopic interventions.

## Clinical Evidence

Bench testing only. Performance evaluated against ISO 8600-1:2015 and ISO 8600-4:2014 standards. Testing confirmed specifications for outer diameter, maximum diameter of attaching endoscope, and distance from the tip.

## Technological Characteristics

Materials: Silicone rubber (Shin-Etsu KE-2090-60-A/B) and Polycarbonate resin (Mitsubishi IUPILON S-2001R 5355). Single-use, EOG sterilized. Attaches to distal end of endoscope. Dimensions vary by model (11.0mm-13.7mm outer diameter).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Fujifilm Hood Model DH-28GR, DH-29CR, DH-30CR ([K162749](/device/K162749.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 1, 2023

Fujifilm Healthcare Americas Corporation Chaitrali Kulkarni Sr. Regulatory Affairs Specialist 81 Hartwell Avenue Suite 300 Lexington, Massachusetts 02421

Re: K232314

Trade/Device Name: Hood (DH-106STL, DH-126STL, DH-126STL, DH-096ST) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDS, FDF Dated: August 2, 2023 Received: August 2, 2023

Dear Chaitrali Kulkarni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Sivakami Venkatachalam -S

for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K232314

Device Name

Hood (DH-106STL, DH-116STL, DH-126STL, DH-096ST)

Indications for Use (Describe)

This hood is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection

Type of Use (Select one or both, as applicable)

< | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

### FUJIFILM Medical Systems U.S.A., Inc.'s FUJIFILM Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST

#### Date: August 31, 2023

#### Submitter's Information

FUJIFILM Healthcare Americas Corporation 81 Hartwell ave Lexington, MA 02421

#### Contact Person:

Chaitrali Kulkarni Sr. Regulatory Affairs Specialist Telephone: (704) 517 4886 E-Mail: chaitrali.kulkarni@fujifilm.c om

#### Identification of the Proposed Device:

Proprietary/Trade Name: Fujifilm Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST Common Name: Hood Device Class: Class II Review Panel: Gastroenterology/Urology

#### Classification Information:

| Classification Name                          | CFR Section        | Product Codes |
|----------------------------------------------|--------------------|---------------|
| Gastroscope and Accessories (Flexible/Rigid) | 21 CFR<br>876.1500 | FDS           |
| Colonoscope and Accessories (Flexible/Rigid) | 21 CFR<br>876.1500 | FDF           |

#### Predicate Devices

- . Fujifilm Hood Model DH-28GR, DH-29CR, DH-30CR (K162749)

{4}------------------------------------------------

#### Intended Use / Indications for Use

The FUJIFILM Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST are intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

#### Device Description

FUJIFILM Hood Models DH-106STL. DH-116STL. and DH-096ST are intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

The FUJIFILM Hood Models DH-106STL, DH-126STL, and DH-096ST are comprised of three main sections: an attaching portion, a distal portion, and a drain portion. The attaching portion is a wider diameter opening which is used to connect the hood to an applicable endoscope; a distal portion is the ending portion of the hood which tapers into narrower diameter opening, the drain slits on the distal portion form drain portion which prevent the fluids lodging on the surface of the endoscope.

The subject devices are used in combination with their respective applicable Fujifilm's endoscopes as shown in the table 7.1. All the applicable endoscopes marketed in USA are cleared under respective 510(k) notices.

## Technological Characteristics

A comparison of the technological characteristics between the subject and predicate devices is provided in the table below.

{5}------------------------------------------------

Table 7.1 Comparison of technological characteristics of FIOSTL, DH-116STL, DH-116STL, and DH-096ST with their predicate device DH-29GR, DH-29GR, DH-29GR, DH-29GR, DH-29CR, D 30CR (K162749)

|                                                  | Predicate Device model<br>DH-28GR, DH-29CR, DH-30CR |                                                                                                         |         | Proposed Device model<br>DH-106STL, DH-116STL, DH-126STL |                                                                                              |           | Proposed Device model<br>DH-096ST                                                                                                                                       |  | FUJIFILM Comments on Predicate Device model                                                                                                                                                       |
|--------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------|----------------------------------------------------------|----------------------------------------------------------------------------------------------|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                  | Image: DH-28GR, DH-29CR, DH-30CR devices            |                                                                                                         |         | Image: DH-106STL, DH-116STL, DH-126STL devices           |                                                                                              |           | Image: DH-096ST device                                                                                                                                                  |  |                                                                                                                                                                                                   |
|                                                  | DH-28GR                                             | DH-29CR                                                                                                 | DH-30CR | DH-106STL                                                | DH-116STL                                                                                    | DH-126STL | DH-096ST                                                                                                                                                                |  |                                                                                                                                                                                                   |
| Manufacturer                                     | DH-28GR                                             | FUJIFILM corporation                                                                                    |         | DH-106STL                                                | FUJIFILM corporation                                                                         |           | FUJIFILM corporation                                                                                                                                                    |  | Same as Predicate Device model                                                                                                                                                                    |
|                                                  | DH-29CR                                             |                                                                                                         | DH-30CR | DH-116STL                                                |                                                                                              | DH-126STL |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Outer diameter                                   | DH-28GR                                             | 11.8mm                                                                                                  |         | DH-106STL                                                | 11.6mm                                                                                       |           | 11.0mm                                                                                                                                                                  |  | The outer diameter changes slightly due to the shape<br>change of the resin                                                                                                                       |
|                                                  | DH-29CR                                             | 13.0mm                                                                                                  |         | DH-116STL                                                | 11.6mm                                                                                       |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
|                                                  | DH-30CR                                             | 14.8mm                                                                                                  |         | DH-126STL                                                | 13.7mm                                                                                       |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Maximum<br>diameter of<br>attaching<br>endoscope | DH-28GR                                             | 15.5mm                                                                                                  |         | DH-106STL                                                | 15.3mm                                                                                       |           | 14.2mm                                                                                                                                                                  |  | DH-106STL, DH-116STL, DH-126STL<br>Since the dimensions of the rubber have been changed in<br>line with the change in the shape of the resin, the<br>maximum diameter will change when installed. |
|                                                  | DH-29CR                                             | 16.5mm                                                                                                  |         | DH-116STL                                                | 15.8mm                                                                                       |           |                                                                                                                                                                         |  | DH-096ST<br>Since the size of the rubber is reduced to combine with an<br>endoscope that is thinner than the existing hood, the<br>maximum diameter changes when attached.                        |
|                                                  | DH-30CR                                             | 18.4mm                                                                                                  |         | DH-126STL                                                | 16.8mm                                                                                       |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Distance from<br>the tip                         | DH-28GR                                             |                                                                                                         |         | DH-106STL                                                |                                                                                              |           | 6.0mm                                                                                                                                                                   |  | In order to increase versatility, the protruding length was<br>shortened to secure the field of view when the endoscope<br>is attached.                                                           |
|                                                  | DH-29CR                                             | 7.0mm                                                                                                   |         | DH-116STL                                                | 6.0mm                                                                                        |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
|                                                  | DH-30CR                                             |                                                                                                         |         | DH-126STL                                                |                                                                                              |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Total length                                     | DH-28GR                                             |                                                                                                         |         | DH-106STL                                                |                                                                                              |           | 16.0mm                                                                                                                                                                  |  | The total length is shortened by changing the Distance<br>from the tip.                                                                                                                           |
|                                                  | DH-29CR                                             | 17.0mm                                                                                                  |         | DH-116STL                                                | 16.0mm                                                                                       |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
|                                                  | DH-30CR                                             |                                                                                                         |         | DH-126STL                                                |                                                                                              |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Combination<br>endoscope                         | DH-28GR                                             | EG-590WR, EG-580RD,<br>EG-600ZW, EG-760Z,<br>EC-580RD/M, EC-580RD/L,<br>EC-740T/M, EC-740T/L            |         | DH-106STL                                                | EG-760Z, EC-740T/L                                                                           |           | EG-760R                                                                                                                                                                 |  | Appropriate endoscopes are set in combination according<br>to the shape of the hood.                                                                                                              |
|                                                  | DH-29CR                                             | EG-760CT, EG-590ZW,<br>EG-530CT, EC-760P-V/M,<br>EC-760P-V/L                                            |         | DH-116STL                                                | EG-760CT, EG-530CT,<br>EC-760P-V/L                                                           |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
|                                                  | DH-30CR                                             | EC-760ZP-V/L, EC-600WL v2,<br>EC-760R-V/L, EC-530DL,<br>ES-530WE, EC-600HL,<br>EC-760Z-V/L, EC-760S-V/L |         | DH-126STL                                                | EC-600LS, EC-600WL,<br>EC-760R-V/L,<br>EC-760ZP-V/L                                          |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| How to attach<br>to the<br>endoscope             | DH-28GR                                             | Align the objective lens of<br>endoscope with the drain of                                              |         | DH-106STL                                                | Attach the hood to the distal<br>end of the endoscope.                                       |           | Align the objective lens of<br>endoscope with the drain<br>of the hood and attach the<br>hood to the distal end of<br>endoscope by pressing<br>the hood until it stops. |  | Same as Predicate Device model                                                                                                                                                                    |
|                                                  | DH-29CR                                             | the hood and attach the hood<br>to the distal end of endoscope                                          |         | DH-116STL                                                | Align the opening of the<br>distal portion of the hood                                       |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
|                                                  | DH-30CR                                             | by pressing the hood until it<br>stops.                                                                 |         | DH-126STL                                                | with the instrument channel<br>outlet of the endoscope and<br>press the hood until it stops. |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Existence of<br>Sterile                          | DH-28GR                                             |                                                                                                         |         | DH-106STL                                                |                                                                                              |           | EOG Sterilized                                                                                                                                                          |  | Same as Predicate Device model                                                                                                                                                                    |
|                                                  | DH-29CR                                             | EOG Sterilized                                                                                          |         | DH-116STL                                                | EOG Sterilized                                                                               |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
|                                                  | DH-30CR                                             |                                                                                                         |         | DH-126STL                                                |                                                                                              |           |                                                                                                                                                                         |  |                                                                                                                                                                                                   |
| Reuse or not<br>re-use                           | DH-28GR                                             | Single Use                                                                                              |         | DH-106STL                                                | Single Use                                                                                   |           | Single Use                                                                                                                                                              |  | Same as Predicate Device model                                                                                                                                                                    |
|                                                  | DH-29CR                                             |                                                                                                         | DH-30CR | DH-116STL                                                |                                                                                              | DH-126STL |                                                                                                                                                                         |  |                                                                                                                                                                                                   |

{6}------------------------------------------------

| Intended Use                            | DH-28GR<br>DH-29CR<br>DH-30CR | These hoods are intended to<br>be used in combination with<br>compatible endoscopes to<br>maintain the field of view<br>during observation of the<br>digestive tract.                                                 | DH-106STL<br>DH-116STL<br>DH-126STL | These hoods are intended to<br>be used in combination with<br>compatible endoscopes to<br>maintain the field of view<br>during observation of the<br>digestive tract.                                                | These hoods are intended<br>to be used in combination<br>with compatible<br>endoscopes to maintain<br>the field of view during<br>observation of the<br>digestive tract                                              | Same as Predicate Device model |
|-----------------------------------------|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Material                                | DH-28GR<br>DH-29CR<br>DH-30CR | Silicone rubber<br>Manufacturer: Shin-Etsu<br>Chemical Co., Ltd. Model:KE-<br>2090-60-A/B<br><br>Polycarbonate resin<br>Manufacture: Mitsubishi<br>Engineering-Plastics<br>Corporation Model: IUPILON<br>S-2001R 5355 | DH-106STL<br>DH-116STL<br>DH-126STL | Silicone rubber<br>Manufacturer: Shin-Etsu<br>Chemical Co., Ltd.<br>Model:KE-2090-60-A/B<br><br>Polycarbonate resin<br>Manufacture: Mitsubishi<br>Engineering-Plastics<br>Corporation Model: IUPILON<br>S-2001R 5355 | Silicone rubber<br>Manufacturer: Shin-Etsu<br>Chemical Co., Ltd.<br>Model:KE-2090-60-A/B<br><br>Polycarbonate resin<br>Manufacture: Mitsubishi<br>Engineering-Plastics<br>Corporation Model:<br>IUPILON S-2001R 5355 | Same as Predicate Device model |
| Transport and<br>Storage<br>Environment | DH-28GR<br>DH-29CR<br>DH-30CR | Temperature: -20 to +60°C<br>Humidity: 10 to 95% (No<br>condensation)<br>Atmosphere: 70 to 106<br>kPa(Within range of<br>atmospheric pressure)                                                                        | DH-106STL<br>DH-116STL<br>DH-126STL | Temperature: -20 to +60°C<br>Humidity: 10 to 95% (No<br>condensation)<br>Atmosphere: 70 to 106<br>kPa(Within range of<br>atmospheric pressure)                                                                       | Temperature: -20 to +60°C<br>Humidity: 10 to 95% (No<br>condensation)<br>Atmosphere: 70 to 106<br>kPa(Within range of<br>atmospheric                                                                                 | Same as Predicate Device model |

{7}------------------------------------------------

The principle of operation and intended use of the FUJIFILM Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST are identical to that of the predicate device FUJIFILM Hood Model DH-28GR, DH-29CR, DH-30CR (K162749). The modifications done to the subject device include changes in dimensions and applicable endoscopes. The distance from the tip and total length is slightly smaller in the subject device compared to the predicate device. The maximum diameter of attaching endoscope of DH-096ST and DH-126STL is slightly smaller the predicate device.

As detailed in the following sections of the 510(k) notice, these changes do not alter the intended use or fundamental technology of the subject devices neither affects their safety and effectiveness.

## Performance Data

Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2015; and ISO 8600-4:2014.

Subject devices met performance specifications in the following additional testing:

- . Outer diameter
- Maximum diameter of attaching endoscope
- Distance from the tip ●

{8}------------------------------------------------

## Substantial Equivalence

FUJIFILM Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST have substantially the same intended use and similar indications, technological characteristics, and principles of operation as their predicate device Fujifilm Hood Model DH-28GR, DH-29CR and DH-30CR (K162749) . The minor dimensional differences between the FUJIFILM Hood Models DH-106STL. DH-116STL, DH-126STL, and DH-096ST and their predicate device Fujifilm Hood Model DH-28GR, DH-29CR and DH-30CR (K162749) were made for the purpose of overall product enhancement and general technological advancement, and raise no new issues of safety or effectiveness. Performance data demonstrated that the FUJIFILM Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST have substantial equivalent performance to the predicate device Fujifilm Hood Model DH-28GR, DH-29CR and DH-30CR (K162749) .

## Conclusions

The subject devices, Hood Models DH-106STL, DH-126STL, and DH-096ST are substantially equivalent to their predicate device Hood Model DH-28GR, DH-29CR and DH-30CR (K162749) , based on intended use/indications for use and technological characteristics. The differences in the dimensions between the subject devices and its predicate device raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject devices have substantially equivalent performance to the predicate. Thus, the subject devices FUJIFILM Hood Models DH-106STL, DH-116STL, DH-126STL, and DH-096ST are as substantially equivalent as their predicate Fuiifilm Hood Model DH-28GR, DH-29CR and DH-30CR (K162749) .

---

**Source:** [https://fda-staging.innolitics.com/device/K232314](https://fda-staging.innolitics.com/device/K232314)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).