K232308 · The Monarch Company · JKA · Dec 4, 2023 · Clinical Chemistry
Device Facts
Record ID
K232308
Device Name
Monarch Blood Collection Set
Applicant
The Monarch Company
Product Code
JKA · Clinical Chemistry
Decision Date
Dec 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The Monarch Safety Blood Collection Set is a sterile, multi-sample, single use fixed winged blood collection set. The device is indicated for general medical use in Healthcare Facilities by medical professionals with the patient population to include both adolescent and adult patients. The Monarch Blood Collection Set is intended to be used with evacuated blood collection tubes and/or blood culture bottles. The Monarch Blood Collection device is to remain under direct supervision of a medical professional during use. The Monarch Blood Collection set includes a safety feature to prevent accidental needlestick injury during normal handling and disposal. The safety feature is initiated by full manual retraction of the needle from the venipuncture site to prevent accidental needlesticks during handling and disposal. The Monarch Blood Collection set is not intended for infusion, IV administration, or transfusion.
Device Story
Sterile, single-use, winged needle blood collection set; includes tubing and vacutainer holder. Operates via manual venipuncture; features finger-activated slide mechanism for needle retraction into a protective housing to prevent needlestick injuries. Used in healthcare facilities by medical professionals; device remains under direct supervision during use. Provides safe blood specimen collection into evacuated tubes or culture bottles. Benefits patient and clinician by reducing risk of accidental sharps injury during handling and disposal.
Clinical Evidence
Bench testing only. No clinical data required for substantial equivalence.
Technological Characteristics
Materials: Stainless steel needle/spring; MethylMethacrylonitrile Butadiene Styrene hub/luer; Polypropylene sliding button/protector; Polyoxymethylene door; EthylVinyl Acetate wings; PVC tubing; CLEARBLEND 165 holder. Standards: ISO 7864, ISO 9626, ISO 80369-7/20, ISO 23908, ISO 10993 series, ISO 11135, ISO 11607-1. Sterilization: Ethylene Oxide (SAL 10^-6). Form factor: Winged needle with 12-inch tubing.
Indications for Use
Indicated for general medical use in healthcare facilities by medical professionals for adolescent and adult patients; intended for blood collection using evacuated tubes or blood culture bottles; not for infusion, IV administration, or transfusion.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
SOL-GUARD™ Safety Pull-Button Blood Collection Set (K213718)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 4, 2023
The Monarch Company Pooja Kannam Director of Quality and Regulatory affairs 4000 Eagle Point Corporate Drive Birmingham, Alabama 35242
Re: K232308
Trade/Device Name: Monarch Blood Collection Set Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: November 2, 2023 Received: November 2, 2023
Dear Pooja Kannam:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K232308
Device Name Monarch Blood Collection Set
#### Indications for Use (Describe)
The Monarch Safety Blood Collection Set is a sterile, multi-sample, single use fixed winged blood collection set. The device is indicated for general medical use in Healthcare Facilities by medical professionals with the patient population to include both adolescent and adult patients. The Monarch Blood Collection Set is intended blood collection tubes and/or blood culture bottles. The Monarch Blood Collection device is to remain under direct supervision of a medical professional during use.
The Monarch Blood Collection set includes a safety feature to prevent accidental needlestick injury during normal handling and disposal. The safety feature is initiated by full manual retraction of the needle from the venipuncture site to prevent accidental needlesticks during handling and disposal. The Monarch Blood Collection set is not intended for infusion, IV administration, or transfusion.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Monarch blood collection sets. The logo features a blue butterfly on the left, followed by the word "monarch" in blue, bold letters. Below "monarch" is the phrase "blood collection sets" in a smaller, lighter blue font. The "TM" symbol is located to the right of the word "monarch".
# K232308: 510K summary For Monarch Blood Collection set
### 1. Submission date
12/01/2023
### 2. Submitter Information
The Monarch Company, LLC
4000 Eagle point Corporate Drive
Birmingham, AL, 35242-1900
Contact Information Phone Number Email Address
### 3. Device Information
Trade/Proprietary name Common Name of device
Product Code Regulatory Class Regulation Number Regulation name Review Panel
Pooja Priyanka Kannam 901-598-1429 poojak@hmg-distribution.com
Monarch Blood Collection Set Blood Collection Set
JKA II 862.1675 Blood Specimen Collection Device Clinical Chemistry
## 4. Predicate Device
K213718
SOL-GUARD™ Safety Pull-Button Blood Collection Set
## 5. Device Description
The Monarch Blood Collection Set is a sterile, single-use, winged needle with tubing and luer connector. Each Monarch Blood Collection device is individually wrapped in a blister pouch. Monarch Blood collection sets include configurations for three (3) needle gauges (21, 23, 25) each with a tubing length of 12 inches. All the models are included with the back-end needle and a vacutainer holder.
A blood collection assembly, comprises of a needle assembly fixedly coupled to a fingeractivated actuator and tubing, the needle assembly comprising a needle; a hub including a channel in a top surface of the hub, a proximal end and a distal end, the distal end
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Monarch blood collection sets. The logo features a blue butterfly on the left, followed by the word "monarch" in blue, block letters. Below "monarch" is the phrase "blood collection sets" in a smaller, blue font. The "TM" symbol is located to the right of the word "monarch".
defining a cavity and a distal passageway, the channel slidably engaging the fingeractivated actuator such that when the finger-activated actuator is moved from a first position at a distal end of the hub to a second position toward a proximal end of the hub; an active closure assembly including a door and a biasing mechanism positioned within the cavity such that when the finger-activated actuator is in the second position, the biasing mechanism operates to move the door along the cavity to contact the needle; and a support wall positioned adjacent the cavity such that, when the finger-activated actuator is in the second position, the needle contacts the support wall.
Nature of contact duration for the device is limited exposure ≤ 24 h. Nature of body contact is external communicating device: Circulating blood.
| Model Name | Configuration |
|------------|----------------------------------------------------|
| M2112 | 21g, 3/4" Needle, 12" tubing, without luer adapter |
| M2312 | 23g, 3/4" Needle, 12" tubing, without luer adapter |
| M2512 | 25g, 3/4" Needle, 12" tubing, without luer adapter |
The Monarch Blood Collection set is available in the following models,
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for "monarch blood collection sets". The word "monarch" is in a large, sans-serif font, and the words "blood collection sets" are in a smaller, sans-serif font below the word "monarch". To the left of the word "monarch" is a stylized butterfly. The letters and butterfly are all in blue.
## 6. Indications For Use
| Characteristic | Subject Device<br>Monarch Safety blood Collection Set | Predicate Device<br>SOL-GUARDTM Safety Pull-Button Blood<br>Collection Set<br>K213718 | |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Monarch Safety Blood Collection Set is<br>a sterile, multi-sample, single use fixed<br>winged blood collection set. The device is<br>indicated for general medical use in<br>Healthcare Facilities by medical<br>professionals with the patient population to<br>include both adolescent and adult patients.<br>The Monarch Blood Collection Set is<br>intended to be used with evacuated blood<br>collection tubes and/or blood culture bottles.<br>The Monarch Blood Collection device is to<br>remain under direct supervision of a medical<br>professional during use.<br>The Monarch Blood Collection set includes<br>a safety feature to prevent accidental<br>needlestick injury during normal handling<br>and disposal. The safety feature is initiated<br>by full manual retraction of the needle from<br>the venipuncture site to prevent accidental<br>needlesticks during handling and disposal.<br>The Monarch Blood Collection set is not<br>intended for infusion, IV administration, or<br>transfusion. | The SOL-GUARDTM Safety Pull- Button Blood<br>Collection Set is a sterile, multi-sample, single-<br>use fixed winged blood collection set intended<br>for venipuncture to obtain blood specimens fron<br>patients. When used without the male Luer<br>adapter, the device allows the clinician to obtain<br>blood sampling to the female hub with a syringe,<br>if necessary, or can be used for short-term (up to<br>2 hours), single infusions with consideration<br>given to patient size and appropriateness for the<br>solution being infused. The device is not to be<br>left in place and remain under the direct<br>supervision of a clinician.<br>The recommended use of the device is to<br>activate the needle safety feature prior to<br>removal the venipuncture site.<br>The retraction of the intravenous (IV) end of the<br>needle aids in the prevention of accidental<br>needlesticks<br>injury. | |
| Prescription<br>Only or Over<br>the Counter | Prescription Only | Prescription Only | |
| Specification | Subject Device<br>Monarch Safety blood<br>Collection Set | Predicate Device<br>SOL-GUARD™ Safety<br>Pull-Button Blood<br>Collection Set<br>K213718 | Comments (Same or<br>Different) |
| Indicated for<br>infusion | No | Yes | Different. Subject device should<br>not be used for infusion. The<br>subject device is intended for a<br>subset of the predicate device<br>intended use. The difference does<br>not raise new or different questions<br>of safety and effectiveness. |
| Single use | Yes | Yes | Same |
| Activation of<br>safety mechanism | Finger Tab That pulls back<br>(Slide actuated) | Pull Button | Subject device has a different<br>mechanism to retract the needle.<br>The needle safety feature was<br>tested in accordance with<br>ISO23908:2011. This difference<br>does not impact the safety and<br>effectiveness of the subject<br>device. |
| Components: | | | |
| Needle Gauge Sizes | 21G, 23G, 25G | 21G, 23G, 25G | Same |
| Needle Diameter ID | Thin Wall | Thin Wall | Same |
| Tubing Length | 12" | 7" and 12" | Different. Only one size of tubing<br>is provided by Monarch. The<br>device was adequately tested in<br>accordance with ISO 80369-7<br>:2021. This difference does not<br>impact the safety and<br>effectiveness of the subject<br>device. |
| Needle Length | 0.75" | 0.75" | Same |
| Needle Point | 3-bevel | 3-bevel | Same |
| Performance: | | | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Sterility Assurance<br>Level (SAL) | $10^{-6}$ | $10^{-6}$ | Same |
| Non-pyrogenic | Yes | Yes | Same |
| Shelf life | 5 years | 3 years | Shelf life of the subject device is<br>5 years. Device performance was<br>conducted after 5 years of aging<br>to demonstrate the device<br>performed to specification over<br>the 5-year shelf life. The<br>difference does not raise any new<br>or different questions of safety |
| | Subject Device | Predicate Device | Comments (Same or |
| Specification | Monarch Safety blood<br>Collection Set | SOL-GUARD™ Safety<br>Pull-Button Blood<br>Collection Set | Different) |
| | | K213718 | and effectiveness. |
| Materials: | | | |
| Needle | Stainless Steel | Stainless Steel | Same |
| Needle/Hub Glue | UV cure Adhesive | Epoxy Glue | The needle was tested in<br>accordance with ISO 7864:2016<br>and ISO9626:2016.<br>This difference does not impact<br>the safety and effectiveness of the<br>subject device. |
| Spring | Stainless Steel | Stainless Steel | Same |
| Needle Hub | MethylMethacrylonitrile<br>Butadiene Styrene | Low Density Polyethylene | Different material: however,<br>ISO10993-1 testing demonstrates<br>biocompatibility of the device.<br>There is no impact to safety and<br>effectiveness caused by the<br>subject device |
| Sliding Button | Polypropylene | Polycarbonate | Different material: however,<br>ISO10993-1 testing demonstrates<br>biocompatibility of the device.<br>There is no impact on the safety<br>and effective caused by the<br>different type of material used by<br>the subject device. |
| Door | Polyoxymethylene | N/A | Different. Predicate device does<br>not have a door; however, device<br>performance and biocompatibility<br>testing demonstrate this difference<br>does not impact the safety and<br>effectiveness of the subject<br>device. |
| Needle Protector | Polypropylene | Low Density Polyethylene | Slightly different for the density<br>of the material; however, device<br>performance and biocompatibility<br>testing demonstrate there is no<br>impact on the safety and<br>effectiveness of the subject due to<br>the density difference of the<br>material. |
| Wings | EthylVinyl Acetate | Polyvinyl Chloride…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.