DVT Motion Pneumatic Compression Device is a prescription device intended for prophylaxis of Deep Vein Thrombosis (DVT), stimulating venous and arterial circulation, aiding in the prevention of venous stasis ulcers, aiding in the healing of cutaneous ulcers, reducing acute/chronic edema and compartmental pressures. For use in home or hospital setting.
Device Story
Portable, battery-powered pneumatic compression device; prevents DVT by stimulating blood flow in lower extremities. Device consists of electronically controlled pump and leg garment. Pump delivers air to garment to compress calf; inflates to preset 50mmHg pressure then deflates; cycle repeats until powered off. Used in home or hospital settings; operated by patient or clinician. Integrated lithium-ion battery provides >18 hours of treatment. Output is mechanical compression of limb; healthcare provider uses device to manage venous stasis and edema. Benefits include continuous DVT prophylaxis during patient transport and home recovery.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. Safety and effectiveness were established through non-clinical performance testing, including biocompatibility, software validation, electromagnetic compatibility, electrical safety, and mechanical performance testing.
Indicated for patients requiring prophylaxis of Deep Vein Thrombosis (DVT), stimulation of venous and arterial circulation, prevention of venous stasis ulcers, healing of cutaneous ulcers, and reduction of acute/chronic edema and compartmental pressures. For use in home or hospital settings.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
March 6, 2024
Koge Micro Tech Co., Ltd. Doris Chiu Regulatory Affairs 5F., No. 6, Jiankang Rd. 1, Zhonghe Dist. New Taipei. 23586 Taiwan
Re: K232270
Trade/Device Name: DVT Motion Pneumatic Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: January 29, 2024 Received: February 7, 2024
Dear Doris Chiu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232270
Device Name
DVT Motion Pneumatic Compression Device
### Indications for Use (Describe)
DVT Motion Pneumatic Compression Device is a prescription device intended for prophylaxis of Deep Vein Thrombosis (DVT), stimulating venous and arterial circulation, aiding in the prevention of venous stasis ulcers, aiding in the healing of cutaneous ulcers, reducing acute/chronic edema and compartmental pressures. For use in home or hospital setting.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"><img src="checkbox-checked.png" style="width:20px; height:20px; margin-right:5px;"/>Research Use (Part 21 CFR 50 Subpart D)</div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><img src="checkbox.png" style="width:20px; height:20px; margin-right:5px;"/>Gene Therapy Use (21 CFR 601 Subpart E)</div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Koge Micro Tech Co., Ltd. DVT Motion Pneumatic Compression Device
Traditional 510(k)_K232270.IR1 Attachment 2 - Section 5 (revised)
# 510(k) SUMMARY
#### 5.1 Traditional Type of Submission:
- 5.2 Date of Summary: 03/04/2024
- 5.3 Submitter: Koge Micro Tech Co., Ltd.
| Address: | 5F., No. 6, Jiankang Rd. 1, Zhonghe Dist., New Taipei City<br>23586, Taiwan (R.O.C) |
|----------|-------------------------------------------------------------------------------------|
| Phone: | +886-2-77381338 |
| Contact: | Doris Chiu<br>(doris_chiu@koge.com) |
#### 5.4 Identification of the Device:
| Proprietary/Trade name: | DVT Motion Pneumatic Compression Device |
|------------------------------|-----------------------------------------|
| Classification Product Code: | JOW |
| Regulation Number: | 870.5800 |
| Regulation Description: | Sleeve, Limb, Compressible |
| Review Panel: | Cardiovascular |
| Device Class: | II |
| Basis for the Submissions | New Device |
#### 5.5 Identification of the Predicate Device:
| Predicate Device Name: | VascuEase IC-1200-WH |
|------------------------------|-------------------------------|
| Applicant: | Bio Compression Systems. Inc. |
| Classification Product Code: | JOW |
| Regulation number: | 870.5800 |
| Device Class: | II |
| 510(k) Number: | K180248 |
{4}------------------------------------------------
#### 5.6 Indications for Use / Intended Use of the Device
DVT Motion Pneumatic Compression Device is a prescription device intended for the prophylaxis of Deep Vein Thrombosis (DVT), stimulating venous and arterial circulation, aiding in the prevention of venous stasis ulcers, aiding in the healing of cutaneous ulcers, reducing acute/chronic edema and compartmental pressures. For use in home or hospital setting.
#### 5.7 Description of the Device
The DVT Motion Pneumatic Compression Device is a portable, rechargeable battery-powered device intended for prescribed home use or hospital use to help prevent Deep Vein Thrombosis (DVT) in patients by stimulating blood flow.
This is accomplished using an electronically controlled pump delivering a set amount of air to the leg garment that compresses the calf and aids blood flow through the lower extremities.
The pump will inflate to a preset pressure of 50mmHg and deflate once this pressure is reached. The cycle continues until the unit is turned off.
The tube free solution comes complete with a fully integrated lithium ion battery that supplies over 18 hours of treatment on a single charge, assuring that patients can be transported easily from the hospital to their homes without interruptions in treatment.
Package contents:
- . a pair of DVT Motion Pneumatic Compression Pumps
- a pair of leg garments
- . a charger (including an adaptor)
- a set of instructions
{5}------------------------------------------------
Koge Micro Tech Co., Ltd. DVT Motion Pneumatic Compression Device
#### 5.8 Non-clinical Testing
A series of non-clinical safety and performance studies were conducted on the subject device. The following tests and studies were according to the related recognized consensus standards.
- Shelf Life Test
- Biocompatibility
- In Virto Cytotoxicity Test
- Skin Sensitization Study i
- Skin Irritation Test i
- Software Validation
- Electromagnetic compatibility and electrical safety
- Performance
- Alert Test
- Discharge and Charge
- Inflation and Deflation Time Test -
- System Leakage Test i
- System Pressure Test i
- . Usability test
All the test results demonstrate DVT Motion Pneumatic Compression Device meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.
#### 5.9 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence. The safety and effectiveness of the finished DVT Motion Pneumatic Compression Device have been established through previous non-clinical performance testing.
{6}------------------------------------------------
#### Substantial Equivalence Determination 5.10
The DVT Motion Pneumatic Compression Device submitted in this 510(k) file is substantially equivalent in intended use, safety and performance claims to the cleared device, VascuEase IC-1200-WH (K180248). Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
{7}------------------------------------------------
| | Subject Device | Predicate Device | Substantial Equivalence<br>Discussion |
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| | Koge Micro Tech Co., Ltd.<br>DVT Motion Pneumatic Compression Device | Bio Compression Systems, Inc.<br>VascuEase IC-1200-WH | Traditional 510(k)_K232270.IR1<br>Attachment 2 - Section 5 (revised) |
| | (to be assigned) | K180248 | |
| | DVT Motion Pneumatic Compression Device<br>is a prescription device intended for the prophylaxis of Deep Vein Thrombosis (DVT), stimulating venous and arterial circulation, aiding in the prevention of venous stasis ulcers, aiding in the healing of cutaneous ulcers, reducing acute/chronic edema and compartmental pressures. For use in home or hospital setting. | VascuEase is a prescription device intended for the prophylaxis of Deep Vein Thrombosis (DVT), stimulating venous and arterial circulation, aiding in prevention of venous stasis ulcers, aiding in the healing of cutaneous ulcers, reducing acute/chronic edema and compartmental pressures. For use in home or hospital setting. | Same |
| | Intermittent Pneumatic Compression | Intermittent Pneumatic Compression | Same |
| | Prescription Use | Prescription Use | Same |
| | $260 g \pm 30 g$<br>$(0.57 \pm 0.066 pounds)$ | 0.7 pounds | Different but Equivalent |
6/9
{8}------------------------------------------------
Koge Micro Tech Co., Ltd.
DVT Motion Pneumatic Compression Devic
Traditional 510(k)_K232270.IR1
Attachment 2 - Section 5 (revised)
| Item | | | | | | | | | | | | | | | | | | |
|-------------------------------------|-------------------------------------------------------|--|-------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|------|--|--|--|--|--|--|
| | Subject Device | | Predicate Device | | Substantial Equivalence | | | | | | | | | | | | | |
| Manufacturer | Koge Micro Tech Co., Ltd. | | Bio Compression Systems, Inc. | | Discussion | | | | | | | | | | | | | |
| Trade Name | DVT Motion Pneumatic Compression Device | | VascuEase IC-1200-WH | | performances of the subject device are not affected and meet the requirements. Therefore, it would not affect the equivalence. | | | | | | | | | | | | | |
| 510(k) No. | (to be assigned) | | K180248 | | Different but the performances of the subject device are not affected and meet the requirements. Therefore, it would not affect the equivalence. | | | | | | | Same | | | | | | |
| Dimensions | 140 mm x 77 mm x 40 mm<br>(5.51 x 3.03 x 1.57 inches) | | 5.3 x 2.7 x 1.5 inches | | | | | | | | | | | | | | | |
| Number of<br>Segments in<br>garment | 1 | | 1 | | | | | | | | | | | | | | | |
{9}------------------------------------------------
Koge Micro Tech Co., Ltd.
DVT Motion Pneumatic Compression Devic
Traditional 510(k)_K232270.IR1
Attachment 2 - Section 5 (revised
| Item | Manufacturer | Trade Name | 510(k) No. | Cycle Time | Pressure | Pressure Adjustment | Pressure Display | Power Requirement | Mobility | Predicate Device | Substantial Equivalence Discussion |
|------|---------------------------|-----------------------------------------|------------------|-------------------------------------------------------------|----------------|---------------------|------------------|----------------------|----------------|-------------------------------|------------------------------------|
| | Koge Micro Tech Co., Ltd. | DVT Motion Pneumatic Compression Device | (to be assigned) | 60 ± 10 seconnds | 50 mmHg ± 20 % | N/A | No | Rechargeable battery | Portable, worn | Bio Compression Systems, Inc. | |
| | | | | (Inflation time: 15 seconds;<br>Deflation time: 45 seconds) | 50 mmHg | N/A | No | Rechargeable battery | Portable, worn | VascuEase IC-1200-WH | Same |
| | | | K180248 | Inflation Time: 15 seconds<br>Deflation Time: 45 seconds | | | | | | | Same |
| | | | | | | | | | | | Same |
| | | | | | | | | | | | Same |
| | | | | | | | | | | | Same |
{10}------------------------------------------------
#### 5.11 Similarity and Difference
The DVT Motion Pneumatic Compression Device is compared with VascuEase IC-1200-WH. The subject device has same intended use, principal of operation, type of use and performance, and similar safety with the predicate device. No specifications are significantly different between these two devices.
Furthermore, the subject device has undergone other safety and performance tests, and the results complied with the testing standards. Therefore, any differences between the subject device and the predicate device are insignificant and do not raise any problem of substantial equivalence. The subject device is substantially equivalent to the predicate device as it claims.
#### 5.12 Conclusion
After analyzing non-clinical laboratory studies, safety and performance testing data, it can be concluded that the DVT Motion Pneumatic Compression Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.