STERI-OSS Implant System is intended for use in partially or fully edentulous mandibles and maxilla, in support of single of multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. STERI-OSS Implant System is for single stage surgical procedures. This system is intended for delayed loading.
Device Story
STERI-OSS Implant System consists of titanium fixtures and titanium alloy abutments; inserted into maxillary or mandibular alveolar bone to support dental prostheses; restores masticatory function. Fixtures are root-form, threaded, and SLA surface-treated. System includes various abutment types (straight, solid, angled, temporary) and screws to connect prosthetics to fixtures. Used in dental clinics by dentists; surgical placement in alveolar bone. Output is a stable foundation for artificial teeth. Benefits include recovery of masticatory function and aesthetic restoration for edentulous patients.
Clinical Evidence
Bench testing only. Includes fatigue testing (ISO 14801), surface evaluation (SEM/EDS), biocompatibility (ISO 10993), sterilization validation (ISO 11137/17665), and shelf-life testing (ASTM standards). MR compatibility assessed via scientific rationale and literature.
Technological Characteristics
Materials: Pure Titanium Grade 4 (ASTM F67) for fixtures; Ti6Al4V ELI (ASTM F136) for abutments/screws. Design: Threaded, root-form, internal hex connection. Surface: SLA treated fixtures. Sterilization: Gamma irradiation. Connectivity: None (mechanical).
Indications for Use
Indicated for partially or fully edentulous mandibles and maxilla in patients requiring single or multiple-unit restorations (cemented, screw-retained, or overdenture) and terminal or intermediate abutment support for fixed bridgework. Intended for single-stage surgical procedures and delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Zeros Co., Ltd. Uk Joo CTO 138-13, Pyeonggang-ro 345beon-gil Gangseo-gu, Busan 46700 REPUBLIC OF KOREA
March 25, 2024
Re: K232268
Trade/Device Name: STERI-OSS Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: February 20, 2024 Received: February 20, 2024
Dear Uk Joo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232268
Device Name STERI-OSS Implant System
#### Indications for Use (Describe)
STERI-OSS Implant System is intended for use in partially or fully edentulous mandibles and maxilla, in support of single of multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. STERI-OSS Implant System is for single stage surgical procedures. This system is intended for delayed loading.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------|------------------------------------------------------------|
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Date revised : March 24, 2024
## I. SUBMITTER
Zeros Co., Ltd. 138-13, Pyeonggang-ro 345beon-gil, Gangseo-gu, Busan, 46700, Rep, of Korea TEL: +82-51-831-0094 FAX: +82-51-831-0095 Contact Name: Uk Hyon Joo Email: ecometal@nate.com
## II. DEVICE INFORMATION
- Device's Trade name: STERI-OSS Implant System ●
- Classification Name: Implant, Endosseous, Root-Form
- Device's Common Name: Endosseous Dental Implant
- . Regulation Number 872.3640
- . Class: 2
- Primary Product Code: DZE
- Secondary Product Code: NHA
### III. PREDICATE DEVICE
Primary Predicate Device:
- K121995, TS Fixture System, OSSTEM Implant Co., Ltd. 0
Reference Device:
- K120503. CMI Implant IS II Active. Neobiotech Co., Ltd. 0
- K202039, Honorst Implant System, MEDIMECCA Co., Ltd.
- K182091, Osstem Abutment System, Osstem Implant Co., Ltd.
- K161689, OSSTEM Implant System - Abutment, Osstem Implant Co., Ltd.
- . K172100, URIS OMNI System, TruAbutment Korea Co., Ltd.
- K210354, SNUCONE Bone Level Implant System, SNUCONE Co., Ltd.
- K052823, Implantium Abutment, Dentium Inc.
- K140934, HIOSSEN Implant System, HiOSSEN Inc.
- K161604, OSSTEM Implant System, Osstem Implant Co., Ltd.
# IV. DEVICE DESCRIPTION
STERI-OSS Implant System, Fixture is a medical device made of titanium that is inserted into the maxillary or mandibular alveolar bone to support prosthesis such as artificial teeth for the patient's recovery of masticatory function.
STERI-OSS Implant System. Abutment is inserted to support prosthesis such as artificial teeth, and is a medical device made of titanium alloy. It connects artificial teeth with a fixture implanted in the maxillary or mandibular alveolar bone where teeth are lost.
## V. INDICATION FOR USE
STERI-OSS Implant System is intended for use in partially or fully edentulous mandibles and maxilla. in support of single of multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. STERI-OSS Implant System is for single stage and two stage surgical procedures. This system is intended for delayed loading.
{5}------------------------------------------------
# VI. COMPARISON OF TECHNILOGICAL CHARACTERISTICS WITH THE REDICATE DEVICE
| Item | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K232268 | K121995 | K120503 | K202039 |
| Device Name | STERI-OSS<br>Implant System | TS Fixture System | CMI Implant IS II<br>Active | Honorst Implant<br>System |
| Manufacturer | Zeros Co., Ltd. | Osstem Implant Co.,<br>Ltd. | Neobiotech Co.,<br>Ltd. | MEDIMECCA Co.,<br>Ltd. |
| Indications for<br>Use Statement | STERI-OSS Implant<br>System is intended<br>for use in partially<br>or fully edentulous<br>mandibles and<br>maxilla, in support<br>of single of<br>multiple-unit<br>restorations<br>including; cemented<br>retained, screw<br>retained, or<br>overdenture<br>restorations, and<br>terminal or<br>intermediate<br>abutment support for<br>fixed bridgework.<br>STERI-OSS Implant<br>System is for single<br>stage and two stage<br>surgical procedures.<br>This system is<br>intended for delayed<br>loading | The TS Fixture<br>System is designed<br>for dental implant<br>surgery; it is placed<br>in the maxillary or<br>mandibular alveolar<br>bone through a<br>surgical procedure,<br>and after<br>osseointegration<br>with the alveolar<br>bone, it can replace a<br>lost tooth by<br>connecting the<br>abutment post.<br>The TS Fixture<br>System & TSIII SA<br>Ultra-Wide Fixture<br>are indicated for use<br>in partially or fully<br>edentulous<br>mandibles and<br>maxillae, in support<br>of single or multiple-<br>unit restorations<br>including; cemented<br>retained, or<br>overdenture<br>restorations, and<br>final or temporary<br>abutment support for<br>fixed bridgework. It<br>is intended for<br>delayed loading. | The CMI Implant IS<br>II active is intended<br>to be surgically<br>placed in the bone of<br>the upper or lower<br>jaw arches to provide<br>support for<br>prosthetics devices.<br>such as artificial<br>teeth, and to restore<br>the patient's chewing<br>function. It is<br>intended for<br>immediate loading<br>when good primary<br>stability is achieved<br>and with appropriate<br>occlusal loading. | Honorst Implant<br>System is intended<br>for use in partially or<br>fully edentulous<br>mandibles and<br>maxilla, in support of<br>single of multiple<br>unit restorations<br>including; cemented<br>retained, screw<br>retained, or<br>overdenture<br>restorations, and<br>terminal or<br>intermediate<br>abutment support for<br>fixed bridgework.<br>Honorst Implant<br>System is for single<br>stage and two stage<br>surgical procedures.<br>This system is<br>intended for delayed<br>loading |
| Substantial<br>Equivalence<br>Discussion | Similarity :<br>The subject device, "STERI-OSS Implant System", and primary predicate device,<br>'TS Fixture System' (K121995) and reference devices 'CMI Implant IS II Active'<br>(K120503)', 'Hornorst Implant System' (K202039) are dental implants that can<br>recover masticatory function when teeth are lost. | | | |
# Table 1 : Substantial Equivalence – Indication for Use Statements
{6}------------------------------------------------
Г
| <b>Difference :</b><br>Indication for use of subject device is slight differences in phrase with the<br>primary predicate device (K121995), reference device (K120503), but<br>fundamental indication is same. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Therefore, Indication for Use statement of the subject device, "STERI-OSS<br>Implant System" is identical and substantially equivalent to the identified primary<br>predicate device, K121995 and reference devices, K202039. |
# Table 2 : Substantial Equivalence - EQUIVALENCE DISCUSSION
# 1. Fixture
| Item | Subject Device | Primary<br>Predicate Device | Reference Device | Reference Device | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|-------------------------------------------|-----------------------------------------|--|--|--|--|--|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| 510(k)<br>Number | K2322685 | K121995 | K120503 | K202039 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Device Name | STERI-OSS<br>Implant System | TS Fixture<br>System | CMI Implant IS II<br>Active | Honorst Implant<br>System | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Manufacturer | Zeros Co., Ltd. | Osstem Implant<br>Co., Ltd. | Neobiotech Co.,<br>Ltd. | MEDIMECCA<br>Co., Ltd. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Design | Image: Design of STERI-OSS Implant System | Image: Design of TS Fixture System | Image: Design of CMI Implant IS II Active | Image: Design of Honorst Implant System | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Connection<br>type | Internal Hex | Internal Hex | Internal Hex | Internal Hex | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Material | Pure Titanium<br>Grade 4<br>(ASTM F67) | Pure Titanium<br>Grade 4<br>(ASTM F67) | Pure Titanium<br>Grade 4<br>(ASTM F67) | Pure Titanium<br>Grade 4<br>(ASTM F67) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Diameter<br>&<br>Length | TSII SA Fixture<br>Diameter (mm) Length(mm) 3.8 8.5, 9, 10, 11, 11.5, 12, 13, 15 4.2 7, 8, 8.5, 9, 10, 11, 11.5, 12, 13, 15 4.6 7, 8, 8.5, 9, 10, 11, 11.5, 12, 13, 15 5.1 7, 8, 8.5, 9, 10, 11, 11.5, 12, 13 5.5 7, 8, 8.5, 9, 10, 11, 11.5, 12, 13 6.0 7, 8, 8.5, 9, 10, 11, 11.5, 12 | | | | | | | | | | | | | | | TSII SA Fixture<br>Diameter (mm) Length(mm) 3.5 8.5, 10, 11.5, 13, 15 4.2 7, 8.5, 10, 11.5, 13, 15 4.4 7, 8.5, 10, 11.5, 13, 15 4.9 7, 8.5, 10, 11.5, 13, 15 TSIII SA Fixture<br>Diameter (mm) Length(mm) 3.75 10, 11.5, 13, 15 3.77 8.5 4.2 10, 11.5, 13, 15 4.25 7, 8.5 4.6 10, 11.5, 13, 15 4.63 8.5 | | | | | | | | | | | | | | | | | | | | | | | | | Diameter (mm) Length(mm) 4.5 7.3, 8.5, 10.0, 11.5, 13.0, 15.0 5 7.3, 8.5, 10.0, 11.5, 13.0, 15.0 5.5 7.3, 8.5, 10.0, 11.5, 13.0, 15.0 6 7.3, 8.5, 10.0, 11.5, 13.0 7 7.3, 8.5, 10.0, 11.5, 13.0 | | | | | | | | | | | | | Diameter (mm) Length(mm) 3.75 8.5, 10.0, 11.5, 13.0 4.2 7.0, 8.5, 10.0, 11.5, 13.0 4.6 7.0, 8.5, 10.0, 11.5, 13.0 5.1 7.0, 8.5, 10.0, 11.5, 13.0 5.5 7.0, 8.5, 10.0, 11.5, 13.0 6 7.0, 8.5, 10.0, 11.5, 12.5 | | | | | | | | | | | | | | |
ZEROS Co., Ltd.
{7}------------------------------------------------
| 5.05 10, 11.5, 13, 15 5.08 8.5 5.1 7 TSIII SA Ultra-Wide Fixture Diameter (mm) Length(mm) 5.92 11.2 5.95 9.7 6 7.2, 8.2 6.8 7.2, 8.2, 9.7, 11.2 | | | | | | | | | | | | | | | | | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|---------------------|---------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Surface Treatment | SLA | SLA | SLA | SLA | | | | | | | | | | | | | | | | | | |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | | | | | | | | | | | | | | | | | | |
| Shelf life | 5 years | 8 years | 5 years | 5 years | | | | | | | | | | | | | | | | | | |
| Substantial Equivalence Discussion | Similarity :<br>The subject device, "STERI-OSS Implant System" has almost the same<br>indications for use, raw material, and design when compared to the primary<br>predicate device, 'TS Fixture System'(K121995). The fixture surface treatment<br>method is the same as SLA, and the product is sterilized by gamma irradiation.<br>In addition, the subject device is the same as the reference devices 'CMI Implant<br>IS II Active'(K120503) and 'Hornorst Implant System'(K202039) in the raw<br>material, design, surface treatment method (SLA), and sterilization method<br>(gamma irradiation).<br>Difference :<br>In addition, although there is a slight difference between the dimensions of the<br>subject device and that of the predicate device (K121995) and reference devices<br>(K120503, K202039), it is judged that they exist within a similar dimensional<br>range overall.<br>Therefore, "STERI-OSS Implant System" is similar in that they are threaded,<br>root form implants with SLA surface treated. The subject and predicate fixtures<br>are a similar body shape design such as cylindrical body design, similar<br>dimension range. The difference is very minor not affecting substantial<br>equivalence. | | | | | | | | | | | | | | | | | | | | | |
# 2. Abutment
| Item | Subject Device | Reference Device | Reference Device |
|------------------|-----------------------------|---------------------------|--------------------------------|
| 510(k)<br>Number | K232268 | K182091 | K172100 |
| Device Name | STERI-OSS Implant<br>System | Osstem Abutment<br>System | URIS OMNI System |
| Manufacturer | Zeros Co., Ltd. | Osstem Implant Co., Ltd. | TruAbutment Korea Co.,<br>Ltd. |
ZEROS Co., Ltd.
{8}------------------------------------------------
| Indication for<br>use | STERI-OSS Implant<br>System is intended for<br>use in partially or fully<br>edentulous mandibles<br>and maxilla, in support<br>of single of multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework.<br>STERI-OSS Implant<br>System is for single stage<br>and two stage surgical<br>procedures. This system<br>is intended for delayed<br>loading. | Osstem Abutment<br>System is intended for<br>use with a dental implant<br>to provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or overdentures. | URIS OMNI System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading. |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Part Name | Straight Abutment | Transfer Abutment | D Basis Abutment-<br>Cemented Type |
| Design | Image: Straight Abutment | Image: Transfer Abutment | Image: D Basis Abutment |
| Principle of<br>operation | Cement retained<br>restoration | Cement retained<br>restoration | Cement retained<br>restoration |
| Material | Ti6Al4V ELI<br>(ASTM F136) | Ti6Al4V ELI<br>(ASTM F136) | Ti6Al4V ELI<br>(ASTM F136) |
| Diameter<br>(mm) | 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 | 4.0, 4.6, 5.0, 6.0, 7.0 | 4.0, 4.5, 5.5, 6.5 |
| Post<br>Height(mm) | 4.0, 5.5, 7.0 | 4.0, 5.5. 7.0 | 4.0, 5.5, 7.0 |
| Gingival<br>Height(mm) | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0,<br>7.0 | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 |
| Surface<br>Treatment | None | TiN Coated (partially) | None |
| Substantial<br>Equivalence<br>Discussion | Similarity :<br>- Subject device, "Straight Abutment" is generally used for cement-retained<br>restoration, and has the same function, indications for use, material compared to<br>that of the reference device, 'Transfer Abutment' in Osstem Abutment System<br>(K182091). Also, subject device has the similar design (dimension range) and<br>appearance compared to that of the reference device (K182091).<br>- Indication for use of subject device is slight differences in phrase with the<br>reference device (K182091), but fundamental indication is same.<br>- Subject device has the same function, indications for use, material, surface<br>treatment compared to that of the reference device (K172100).<br>Difference :<br>The subject device, "Straight Abutment", has a machined surface without surface<br>treatment and is similar to the reference device (K172100). The reference device (K172100) has a TiN coated surface. | | |
{9}------------------------------------------------
(K182091) was partially coated with TiN.
Therefore, the subject device, "Straight Abutment" is very similar to the reference device (K182091) in terms of function, purpose of use, raw materials (Ti alloys), design, etc., except for the different surface treatment. The difference is minor not affecting substantial equivalence.
| Item | Subject Device | Reference Device | Reference Device |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K232268 | K161689 | K172100 |
| Device Name | STERI-OSS Implant<br>System | OSSTEM Implant<br>System - Abutment | URIS OMNI System |
| Manufacturer | Zeros Co., Ltd. | Osstem Implant Co., Ltd. | TruAbutment Korea Co.,<br>Ltd. |
| Indication for<br>use | STERI-OSS Implant<br>System is intended for<br>use in partially or fully<br>edentulous mandibles<br>and maxilla, in support<br>of single of multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework.<br>STERI-OSS Implant<br>System is for single stage<br>and two stage surgical<br>procedures. This system<br>is intended for delayed<br>loading. | The OSSTEM Implant<br>System - Abutment is<br>intended for use with a<br>dental implant to provide<br>support for prosthetic<br>restorations such as<br>crowns, bridges, or<br>overdentures. | URIS OMNI System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading. |
| Part Name | Solid Abutment | Rigid Abutment | D Basis Abutment<br>- Direct Type |
| Design | Image: Solid Abutment | Image: Rigid Abutment | Image: D Basis Abutment |
| Principle of<br>operation | Cement retained<br>restoration | Cement retained<br>restoration | Cement retained<br>restoration |
| Material | Ti6Al4V ELI<br>(ASTM F136) | Ti6Al4V ELI<br>(ASTM F136) | Ti6Al4V ELI<br>(ASTM F136) |
| Diameter<br>(mm) | 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 | 4.0, 4.6, 5.0, 6.0, 7.0 | 4.0, 4.5, 5.5, 6.5 |
| Post<br>Height(mm) | 4.0, 5.5, 7.0 | 4.0, 5.5, 7.0 | 4.0, 5.5, 7.0 |
ZEROS Co., Ltd.
{10}------------------------------------------------
| Gingival<br>Height(mm) | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 | 1.0, 2.0, 3.0, 4.0, 5.0 | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.