GXR-ES/ECS Diagnostic X-Ray System, (Models GXR-ES series, GXR-ECS series)

K232178 · DRGEM Corporation · KPR · Sep 19, 2023 · Radiology

Device Facts

Record IDK232178
Device NameGXR-ES/ECS Diagnostic X-Ray System, (Models GXR-ES series, GXR-ECS series)
ApplicantDRGEM Corporation
Product CodeKPR · Radiology
Decision DateSep 19, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2

Indications for Use

The "GXR-ES/ECS" System is intended for use in generating radiographic images of human anatomy. The Diagnostic Xray System consisting of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, patient table, wall Bucky stand, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position

Device Story

Stationary X-ray system for hospital use; operated by trained medical professionals. Inputs: user-selected technique factors (kVp, mA) via control console. Operation: high-frequency inverter generator produces X-rays; beam limiting device directs radiation through anatomy to external image detector (not included). Features: APR (Anatomical Programming), optional AEC (Automatic Exposure Control), and automatic calibration. Output: radiographic images (captured by external detector) and system status/exposure data on console LCD. Remote diagnosis software allows system/error reporting via USB/internet. Benefits: provides diagnostic images for skeletal, respiratory, and urinary system assessment; modular design; real-time monitoring/self-diagnosis improves system uptime.

Clinical Evidence

Bench testing only. Validation testing confirmed the device meets predetermined acceptance criteria for input/output functions and safety. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, and ISO 14971 demonstrated safety and performance equivalence.

Technological Characteristics

Stationary X-ray system; high-frequency inverter generator; 20-50kW output. Components: HV generator, tube support, beam limiter, patient table, wall Bucky stand, X-ray tube. Connectivity: USB for remote diagnosis. Software: X-CON firmware for system control (collimation, filter, generator). Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, ISO 14971, ISO 15223-1.

Indications for Use

Indicated for generating radiographic images of human anatomy for diagnosis of skeletal, respiratory, and urinary system diseases (e.g., skull, spine, chest, abdomen, extremities). Suitable for patients in sitting, standing, or prone/supine positions.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable. September 19, 2023 DRGEM Corporation % Mr. Park Sunghee Quality Management Director 7F/13F, E-B/D Gwangmyeong Techno-Park, 60Haan-ro Gwang-Myeong-si, Gyeong-gi-do 14322 KOREA Re: K232178 Trade/Device Name: GXR-ES/ECS Diagnostic X-Ray System, (Models GXR-ES series, GXR-ECS series) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: July 24, 2023 Received: July 24, 2023 Dear Mr. Sunghee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232178 Device Name GXR-ES/ECS Diagnostic X-ray System, (Models : GXR-ES series, GXR-ECS series) #### Indications for Use (Describe) The "GXR-ES/ECS" System is intended for use in generating radiographic images of human anatomy. The Diagnostic Xray System consisting of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, patient table, wall Bucky stand, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position Type of Use (Select one or both, as applicable) | <div> <object data="checkbox_checked.svg" height="20" type="image/svg+xml" width="20"></object> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |-------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <object data="checkbox.svg" height="20" type="image/svg+xml" width="20"></object> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for DRGEM. The text "DRGEM" is displayed in a bold, sans-serif font, with "DR" in black and "GEM" in teal. Below the main text, the tagline "Your Best Healthcare" is written in a smaller, italicized font, also in black. # 510(k) Summary # K232178 This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. # I. SUBMITTER [21 CFR 807.92(a) (1)] DRGEM Corporation 7F E-B/D Gwangmyeong Techno-Park 60, Haan-ro, Gwangmyeong-si, Gyeonggi-do, 14322 Korea Email: sunghee.park@drgem.co.kr TEL: +82-2-869-8566, FAX: +82-2-869-8567 Contact Person: Mr.Sung-Hee PARK, Director | Quality Management Date Prepared: May 25, 2023 #### II. PROPOSED DEVICE INFORMATION [21 CFR 807.92(a) (2)] | Product Name: | GXR-ES/ECS Diagnostic X-Ray System, (Models GXR-ES series, GXR-ECS series) | |----------------------|----------------------------------------------------------------------------| | Common Name: | Stationary X-Ray System | | Classification Name: | Stationary x-ray system | | Product Code: | KPR | | Regulatory Class: | II | | Regulation Number: | 892.1680 | #### III. PREDICATE DEVICES INFORMATION [21 CFR 807.92(a) (3)] Product Name: GXR-series Diagnostic X-Ray System, (Models GXR-S series) Common Name: Stationarv X-Rav System Classification Name: Stationary x-ray system K202572 510(k) Number : Product Code: KPR Regulatory Class: = Requlation Number: 892.1680 #### IV. DEVICE DESCRIPTION [21 CFR 807.92(a) (4)] Device Features: The GXR-ES/ECS system offers stable operation and good performance, delivering state-of-the-art image quality with good quality images. This diagnostic x-ray system is designed to diagnose human body by providing radiographic x-ray image with anatomical structure. The flat-panel detectors (a necessary component of a fully-functional diagnostic x-ray system) are not part of the subject device. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for DRGEM. The letters "DR" are in bold black font, while "GEM" is in a teal color. Below the logo is the text "Your Best Healthcare" in a smaller, lighter font. "GXR-ES/ECS" System provides good performance and stable operation while providing good quality images. The operator control console is designed in two types to be simple and user-friendly. The first type features a large graphic LCD panel display and a soft membrane switch to allow easy selection or change of X-rav parameters. The second type can be easily operated using the Touch function through a large graphic LCD panel display and intuitive GUI configuration. The "GXR-E/EC" Series high frequency X-ray generator features excellent accuracy, reproducibility and long-term stability with UPS functionality. The APR (Anatomical Programming) and the optional AEC (Automatic Exposure Control) features qive you controlled exposure factors, automatically optimized for the radiological study selected. Automatic Calibration function of the generator will minimize calibration time and set correct calibration value. Also the control console of "GXR-ES/ECS" System offers graphic waveform and data of x-ray exposure. So, it is not needed to perform the manual calibration procedure with measurement equipment. Naturally, the generator supports Automatic, Semi-auto and Manual Calibrations. Remote Diagnosis Software can make report about information of system, user, service and error by the USB connection between PC and control console, and transfer report to manufacturer via internet. So this function enables fast and accurate diagnosis on problems and saves service cost and system downtime. The control console allows the operator to select the technique factors, image receptors, etc., and to initiate an X-ray exposure. - . Compact System for essential uses. - . Good Performance and Stability. - Real-time monitoring self-diagnosis function and Error code display. . - . Overload & HU protection and error message display. - . Automatic calibration without measurement equipment. - . Adaptable calibration keeps up accuracy through long-term usage. - The system is small, light and features a convenient modular design and user-friendly system . configuration. - . Constant dose output due to kV and mA requlation during exposures. - Large graphic LCD panel user-friendly controls for APR and technique display at a glance. . - . Time and mA / mAs selections are based on R'10 rule of ISO 497. - . Remote diagnosis software for system diagnosis via internet. - . Easy parameter setting and Firmware upgrade. - . System diagnosis, Error log and Statistical data display. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for DRGEM, a healthcare company. The logo features the company name in bold, with "DR" in black and "GEM" in teal. Below the company name is the tagline "Your Best Healthcare" in a smaller, lighter font. #### · Device Identification: The subject device consists of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, a patient table, wall Bucky stand, and an x-ray tube, that operates on a highfrequency inverter method. #### · Device Characteristics: #### Software The subject device 'GXR-ES/ECS Diagnostic X-ray System' use software (firmware). Its S/W can perform system control such as the collimation size, filter selection, Control of Generator. The software being used is identical to the predicate device "GXR-S series Diagnostic X-ray System", and its LOC (Level of Concern) is 'Moderate'. Accordingly, this software (X-CON) is based on predicate device 'GXR-S series Diagnostic system. #### Determination (Moderate Level of Concern) The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention, interprets all images and information being displayed and printed. #### · Environment of Use: This "GXR-ES/ECS" System is for use by medical professional Facility. To prevent excess radiation exposure to patient and operator from either primary or secondary radiation, this "GXR-ES/ECS" System must be operated and serviced by trained personnel who are familiar with the safety precautions required. ## · Brief Written Description of the Device: The operating principles are as follows. The irradiation conditions are 40 to 125 (150) kVp at the photographing site, and the tube current is 10 to 400 (500) mA. When X-rays generated under X-ray irradiation conditions enter the X-ray Film, the film detects X-rays incident through the incident surface during X-ray irradiation, and finally generates a radiographic image when X-ray irradiation is completed. #### · Material of Use: The "GXR-E/EC series Diagnostic X-ray System has a two patient-applied part. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains a logo for DRGEM. The text "DR" is in bold, black font, while "GEM" is in a teal color. Below the logo, the text "Your Best Healthcare" is written in a smaller, lighter font. Some device components may briefly touch the patient. Device biocompatibility has been tested according to relevant standards and it was found that critical components satisfy biocompatibility requirements as applicable to the device Indications for Use. #### · Compatible x-ray image detector: This model is an analog-based X-ray system and does not include image detectors as an essential component. It is possible to compatible with available detectors and can be used with certified detector panel sizes as long as they fit. - Available detector - Size (inch): 17X17, 17X14, 10 X12 - Products whose compatibility has been verified by the manufacturer can be used. #### V. INDICATION FOR USE [21 CFR 807.92(a) (5)] "GXR-ES/ECS" System is indicated for use in generating radiographic images of human anatomy. The Diagnostic X-ray System consisting of a high voltage (HV) generator, a tube support unit, an Xray beam limiting device, patient table, wall Bucky stand, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position #### VI. TECHNOLOGICAL CHARACTERISTICS [21 CFR 807.92(a) (6)] The subject device and the primary predicate device are stationary x-ray devices. There will be 2 main differences: First, this device, 'GXR-ES/ECS Diagnostic X-ray System has Compact System', is for only essential use in that it is composed of simplified mechanical parts. But, the predicate device 'GXR- series Diagnostic Imaging System' has Manual type as well as motorized option of Tube stand, patient table and WBS (Diagnostic X-ray Full system). And the subject device has lower high voltage output rating than predicate device for Use of small-scale hospital facility operation. Second, Difference between subject and predicate device is diversity of the power supply method of the generator Generally, the subject device 'GXR-ES, ECS' has general and capacitor type generator supplied from wall power. However, the predicate device 'GXR-series Diagnostic Imaging System' has also UPS battery type (total three type). Differences between the subject device and the predicates has no impact on safety or effectiveness of the new device and does not raise any new potential or increased safety risks and is equivalent in performance to existing legally marketed devices. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for DRGEM. The letters "DR" are in black, while "GEM" is in teal. Below the logo is the text "Your Best Healthcare" in a smaller font. # VII.SUBSTANTIAL EQUIVALENCE [21 CFR 807.92(b)] | Item | Subject Device | Predicate Device | Impact of Differences | |---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | GXR-ES/ECS Series Diagnostic<br>X-ray System | GXR-Series<br>Diagnostic X-Ray<br>System (K202572) | Not applicable | | Manufacturer | DRGEM Corporation | DRGEM Corporation | Not applicable | | Model Number | GXR-ES series,<br>GXR-ECS series | GXR-S series | Not applicable | | High Frequency X- ray Generator | | | | | Output Power | 20kW, 25kW, 32kW, 40kW,<br>50kW | 32KW, 40KW,<br>52KW, 68KW,<br>82KW | Yes, there is a difference in output<br>values but no difference in<br>generators.<br>See Difference Explanation below. | | Generator<br>models<br>(manufactured<br>by DRGEM) | GXR-E20, GXR-E25,<br>GXR-E32, GXR-E40,<br>GXR-E50 | GXR-32, GXR- 40, GXR-<br>52, GXR-68, GXR- 82 | Yes, there is a difference.<br>Models have been tested against<br>International Safety and EMC<br>Standards. Any differences<br>between the subject device and<br>predicate device do not change or<br>add new potential safety risks. It is<br>our determination that there is "No<br>negative impact on safety or<br>effectiveness" and there are no new<br>potential<br>or increased safety risks concerning<br>this difference. | | Line voltage | 100~120VAC<br>200~240VAC | 220~230VAC,<br>380/400/480VA C | Yes, there are differences in line<br>voltage depending upon the system<br>requirements. Models have been<br>tested against International Safety<br>and EMC Standards. Any<br>differences between the subject<br>device and predicate device do not<br>change or add new potential safety<br>risks. It is our determination that<br>there is "No negative impact on<br>safety or efficacy" and there are no<br>new potential or increased safety<br>risks<br>concerning this difference | | Image Acquisition System | | | | | Flat panel<br>Detector | Detector is not supplied with system | Detector is not supplied with system | Not applicable | | Patient table | | | | | Configuration model | PBT-E | PBT-4, PBT-6, PDT-1 | Yes, there is a difference.<br>Models have been tested against International Safety and EMC Standards. Any differences between the subject device and predicate device do not change or add new potential safety risks. It is our determination that there is "No negative impact on safety or effectiveness" and there are no new potential or increased safety risks concerning this difference | | Movement | Longitudinal :<br>1050mm ± 10mm | Longitudinal :<br>1000mm ± 500mm | | | Wall Bucky stand | | | | | Configuration model | WBS-E, WBS | WBS, WBS-TM | Yes, there is a difference.<br>Models have been tested against International Safety and EMC Standards. Any differences between the subject device and predicate device do not change or add new potential safety risks. It is our determination that there is "No negative impact on safety or effectiveness" and there are no new potential or increased safety risks concerning this difference | | Dimension/weight | 560(W) x 690(D) x 1,826(H)mm /<br>32.4kg(71.4lbs) (Floor Base)<br>420(W) x 256(D) x 1,826(H)mm /<br>21.4kg(47.2lbs) (Floor-Wall Mounted)<br>659(W) x 393(D) x 2,169(H)mm /<br>120kg(264lbs) | 659(W) x 445(D) x<br>1,599(H)mm / 110kg(242lbs)<br>659(W) x 445(D) x<br>1,749(H)mm / 113kg(249lbs)<br>659(W) x 445(D) x<br>1,919(H)mm / 116kg(255lbs)<br>659(W) x 445(D) x<br>2,169(H)mm / 120kg(264lbs) | | | Tube stand | | | | | Configuration model | TS-FT6, TS-FT4 | TS-FC6, TS-FC4, TS-FC2<br>TS-FM6, TS-CSA, TS-CSE | Not applicable | | Tube Rotation<br>Angle | ±135° | ±135° | | | Configuration<br>model | DXR-8M, DXT-10M, DXT-<br>11M, DXR-12M, E7239X,<br>E7242X, E7884X, | E7239X, DXT-8M,<br>E7242X, DXT-11M<br>E7843X, DXT-10M<br>E7876X, DXT-12M<br>E7884X, E7252X<br>DXT-14U, RAD-14U,<br>DXT-15U, RAD-21<br>RAD-60, E7255FX,<br>E4254FX | No Impact of difference | | Max. kV | 125kV, 150kV | 125kV, 150kV | | | Collimator | | | | | Configuration<br>model | DXC-RML | DXC-RML, DXC-RMH<br>MCR, R108, R302A,<br>R302MLP/A,<br>R302MFMLP/A | Yes, there is a difference.<br>Models have been tested against<br>International Safety and EMC<br>Standards. Any differences<br>between the subject device and<br>predicate device do not change or<br>add new potential safety risks. It is<br>our determination that there is "No<br>negative impact on safety or<br>effectiveness" and there are no new<br>potential<br>or increased safety risks concerning<br>this difference | | Lamp Type | LED lamp<br>(Over 160LUX at 100cm<br>SID) | LED and Halogen lamp<br>(Over 160LUX at 100cm<br>SID) | | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for DRGEM. The letters "DRG" are in bold black font, while the letters "EM" are in a teal font. Below the logo is the text "Your Best Healthcare" in a smaller, gray font. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image contains the logo for DRGEM, a healthcare provider. The letters "DR" are in bold, black font, while "GEM" is in a teal color. Below the logo, the text "Your Best Healthcare" is written in a smaller, lighter font. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for DRGEM. The letters "DR" are in black, and the letters "GEM" are in teal. Below the letters is the text "Your Best Healthcare" in a smaller font. #### VIII. SUMMARY OF NON-CLINICAL Data [21 CFR 807.92(b) (1)] Nonclinical Testing: The GXR-ES/ ECS Series Diagnostic X-Ray System, has been assessed and tested and has passed predetermined testing criteria. The Validation Test Plan was designed to evaluate input functions, output functions, and actions performed by the subject device and followed the process documented in the System Validation Test Plan. Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met. | Std # | Safety/EMC Standards Description | FDA Rec.<br>Standard # | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------| | IEC 60601-1-3 | Medical electrical equipment<br>Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray<br>Equipment | 12-269 | | IEC 60601-1-6 | IEC 60601-1-6 Edition 3.1 2013-10<br>Medical electrical equipment - Part 1-6: General requirements for basic safety and<br>essential performance - Collateral standard: Usability | 5-89 | | IEC 60601-2-28 | IEC 60601-2-28<br>Medical electrical equipment Part 2: Particular requirements for the safety of X-ray<br>source assemblies and X-ray tube assemblies for medical diagnosis | 12-309 | | IEC 60601-2-54 | IEC 60601-2-54<br>Medical electrical equipment<br>Part 2: Particular requirements for the basic safety and essential<br>performance of X-ray equipment for radiography and radioscopy | 12-317 | | IEC 60601-1-2<br>(EMC) | IEC 60601-1-2 Edition 4.0 2014-02. Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances Requirements and tests. | 19-8 | | IEC 62304:2006 | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION<br>Medical device software - Software life cycle processes | 13-79 | | IEC 60601-1 | Medical electrical equipment, Part 1: General requirements for basic safety and<br>essential performance | 19-4 | | ISO 14971:2019 | ISO 14971:2019 Third Edition, Medical devices - Applications of risk management to<br>medical devices. | 5-125 | | ISO 15223-1 | ISO 15223-1 Fourth Edition 2021-07, Medical devices - Symbols to be used with<br>medical device labels, labelling, and information to be supplied - Part 1: General<br>requirements. | 5-134 | | TR 60601-4-2 | TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and<br>interpretation - Electromagnetic immunity: performance of medical electrical equipment<br>and medical electrical systems | 19-19 | | FDA Guidance | Pediatric Information for X-ray Imaging Device Premarket Notifications dated November 28, 2017 | | | FDA Guidance | Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance<br>for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014 | | | FDA Guidance | Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software<br>contained in Medical Devices, Document issued on: May 11, 2005 Medical Devices, Document issued<br>on: May 11 2005 | | {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for DRGEM, a healthcare company. The letters "DR" are in bold, black font, while the letters "GEM" are in a teal color. Below the company name is the tagline "Your Best Healthcare" in a smaller, lighter font. #### Summary: Based on the performance as documented in the V&V Testing, the subject device was found to have a safe and effectiveness profile that is similar to the predicate device. The above standards were used to develop and verify electrical safety, and EMC. GXR-ES/ ECS Series Diagnostic X-Ray System device has met all the requirements listed in the Standards except for inapplicable requirements. The subject device conform to all applicable aspects of 21CFR 1020.30 # IX. CONCLUSIONS [21 CFR 807.92(b) (3)] The 510(k) Pre-Market Notification for GXR-ES/ECS series Diagnostic X-ray System, contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device and the predicate device are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices. Nonclinical tests demonstrate that the device is as safe, as effective, and performs comparably to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...