The "GXR-ES/ECS" System is intended for use in generating radiographic images of human anatomy. The Diagnostic Xray System consisting of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, patient table, wall Bucky stand, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position
Device Story
Stationary X-ray system for hospital use; operated by trained medical professionals. Inputs: user-selected technique factors (kVp, mA) via control console. Operation: high-frequency inverter generator produces X-rays; beam limiting device directs radiation through anatomy to external image detector (not included). Features: APR (Anatomical Programming), optional AEC (Automatic Exposure Control), and automatic calibration. Output: radiographic images (captured by external detector) and system status/exposure data on console LCD. Remote diagnosis software allows system/error reporting via USB/internet. Benefits: provides diagnostic images for skeletal, respiratory, and urinary system assessment; modular design; real-time monitoring/self-diagnosis improves system uptime.
Clinical Evidence
Bench testing only. Validation testing confirmed the device meets predetermined acceptance criteria for input/output functions and safety. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, and ISO 14971 demonstrated safety and performance equivalence.
Technological Characteristics
Stationary X-ray system; high-frequency inverter generator; 20-50kW output. Components: HV generator, tube support, beam limiter, patient table, wall Bucky stand, X-ray tube. Connectivity: USB for remote diagnosis. Software: X-CON firmware for system control (collimation, filter, generator). Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, ISO 14971, ISO 15223-1.
Indications for Use
Indicated for generating radiographic images of human anatomy for diagnosis of skeletal, respiratory, and urinary system diseases (e.g., skull, spine, chest, abdomen, extremities). Suitable for patients in sitting, standing, or prone/supine positions.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
September 19, 2023
DRGEM Corporation % Mr. Park Sunghee Quality Management Director 7F/13F, E-B/D Gwangmyeong Techno-Park, 60Haan-ro Gwang-Myeong-si, Gyeong-gi-do 14322 KOREA
Re: K232178
Trade/Device Name: GXR-ES/ECS Diagnostic X-Ray System, (Models GXR-ES series, GXR-ECS series) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: July 24, 2023 Received: July 24, 2023
Dear Mr. Sunghee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232178
Device Name
GXR-ES/ECS Diagnostic X-ray System, (Models : GXR-ES series, GXR-ECS series)
#### Indications for Use (Describe)
The "GXR-ES/ECS" System is intended for use in generating radiographic images of human anatomy. The Diagnostic Xray System consisting of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, patient table, wall Bucky stand, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position
Type of Use (Select one or both, as applicable)
| <div> <object data="checkbox_checked.svg" height="20" type="image/svg+xml" width="20"></object> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <object data="checkbox.svg" height="20" type="image/svg+xml" width="20"></object> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for DRGEM. The text "DRGEM" is displayed in a bold, sans-serif font, with "DR" in black and "GEM" in teal. Below the main text, the tagline "Your Best Healthcare" is written in a smaller, italicized font, also in black.
# 510(k) Summary
# K232178
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
# I. SUBMITTER [21 CFR 807.92(a) (1)]
DRGEM Corporation 7F E-B/D Gwangmyeong Techno-Park 60, Haan-ro, Gwangmyeong-si, Gyeonggi-do, 14322 Korea Email: sunghee.park@drgem.co.kr TEL: +82-2-869-8566, FAX: +82-2-869-8567
Contact Person: Mr.Sung-Hee PARK, Director | Quality
Management Date Prepared: May 25, 2023
#### II. PROPOSED DEVICE INFORMATION [21 CFR 807.92(a) (2)]
| Product Name: | GXR-ES/ECS Diagnostic X-Ray System, (Models GXR-ES series, GXR-ECS series) |
|----------------------|----------------------------------------------------------------------------|
| Common Name: | Stationary X-Ray System |
| Classification Name: | Stationary x-ray system |
| Product Code: | KPR |
| Regulatory Class: | II |
| Regulation Number: | 892.1680 |
#### III. PREDICATE DEVICES INFORMATION [21 CFR 807.92(a) (3)]
Product Name: GXR-series Diagnostic X-Ray System, (Models GXR-S series) Common Name: Stationarv X-Rav System Classification Name: Stationary x-ray system
K202572 510(k) Number : Product Code: KPR Regulatory Class: = Requlation Number: 892.1680
#### IV. DEVICE DESCRIPTION [21 CFR 807.92(a) (4)] Device Features:
The GXR-ES/ECS system offers stable operation and good performance, delivering state-of-the-art image quality with good quality images.
This diagnostic x-ray system is designed to diagnose human body by providing radiographic x-ray image with anatomical structure.
The flat-panel detectors (a necessary component of a fully-functional diagnostic x-ray system) are not part of the subject device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for DRGEM. The letters "DR" are in bold black font, while "GEM" is in a teal color. Below the logo is the text "Your Best Healthcare" in a smaller, lighter font.
"GXR-ES/ECS" System provides good performance and stable operation while providing good quality images.
The operator control console is designed in two types to be simple and user-friendly.
The first type features a large graphic LCD panel display and a soft membrane switch to allow easy selection or change of X-rav parameters.
The second type can be easily operated using the Touch function through a large graphic LCD panel display and intuitive GUI configuration.
The "GXR-E/EC" Series high frequency X-ray generator features excellent accuracy, reproducibility and long-term stability with UPS functionality.
The APR (Anatomical Programming) and the optional AEC (Automatic Exposure Control) features qive you controlled exposure factors, automatically optimized for the radiological study selected. Automatic Calibration function of the generator will minimize calibration time and set correct calibration value.
Also the control console of "GXR-ES/ECS" System offers graphic waveform and data of x-ray exposure.
So, it is not needed to perform the manual calibration procedure with measurement equipment. Naturally, the generator supports Automatic, Semi-auto and Manual Calibrations.
Remote Diagnosis Software can make report about information of system, user, service and error by the USB connection between PC and control console, and transfer report to manufacturer via internet. So this function enables fast and accurate diagnosis on problems and saves service cost and system downtime.
The control console allows the operator to select the technique factors, image receptors, etc., and to initiate an X-ray exposure.
- . Compact System for essential uses.
- . Good Performance and Stability.
- Real-time monitoring self-diagnosis function and Error code display. .
- . Overload & HU protection and error message display.
- . Automatic calibration without measurement equipment.
- . Adaptable calibration keeps up accuracy through long-term usage.
- The system is small, light and features a convenient modular design and user-friendly system . configuration.
- . Constant dose output due to kV and mA requlation during exposures.
- Large graphic LCD panel user-friendly controls for APR and technique display at a glance. .
- . Time and mA / mAs selections are based on R'10 rule of ISO 497.
- . Remote diagnosis software for system diagnosis via internet.
- . Easy parameter setting and Firmware upgrade.
- . System diagnosis, Error log and Statistical data display.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for DRGEM, a healthcare company. The logo features the company name in bold, with "DR" in black and "GEM" in teal. Below the company name is the tagline "Your Best Healthcare" in a smaller, lighter font.
#### · Device Identification:
The subject device consists of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, a patient table, wall Bucky stand, and an x-ray tube, that operates on a highfrequency inverter method.
#### · Device Characteristics:
#### Software
The subject device 'GXR-ES/ECS Diagnostic X-ray System' use software (firmware).
Its S/W can perform system control such as the collimation size, filter selection, Control of Generator.
The software being used is identical to the predicate device "GXR-S series Diagnostic X-ray System", and its LOC (Level of Concern) is 'Moderate'.
Accordingly, this software (X-CON) is based on predicate device 'GXR-S series Diagnostic system.
#### Determination
(Moderate Level of Concern)
The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention, interprets all images and information being displayed and printed.
#### · Environment of Use:
This "GXR-ES/ECS" System is for use by medical professional Facility. To prevent excess radiation exposure to patient and operator from either primary or secondary radiation, this "GXR-ES/ECS" System must be operated and serviced by trained personnel who are familiar with the safety precautions required.
## · Brief Written Description of the Device:
The operating principles are as follows.
The irradiation conditions are 40 to 125 (150) kVp at the photographing site, and the tube current is 10 to 400 (500) mA. When X-rays generated under X-ray irradiation conditions enter the X-ray Film, the film detects X-rays incident through the incident surface during X-ray irradiation, and finally generates a radiographic image when X-ray irradiation is completed.
#### · Material of Use:
The "GXR-E/EC series Diagnostic X-ray System has a two patient-applied part.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for DRGEM. The text "DR" is in bold, black font, while "GEM" is in a teal color. Below the logo, the text "Your Best Healthcare" is written in a smaller, lighter font.
Some device components may briefly touch the patient. Device biocompatibility has been tested according to relevant standards and it was found that critical components satisfy biocompatibility requirements as applicable to the device Indications for Use.
#### · Compatible x-ray image detector:
This model is an analog-based X-ray system and does not include image detectors as an essential component. It is possible to compatible with available detectors and can be used with certified detector panel sizes as long as they fit.
- Available detector
- Size (inch): 17X17, 17X14, 10 X12
- Products whose compatibility has been verified by the manufacturer can be used.
#### V. INDICATION FOR USE [21 CFR 807.92(a) (5)]
"GXR-ES/ECS" System is indicated for use in generating radiographic images of human anatomy. The Diagnostic X-ray System consisting of a high voltage (HV) generator, a tube support unit, an Xray beam limiting device, patient table, wall Bucky stand, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position
#### VI. TECHNOLOGICAL CHARACTERISTICS [21 CFR 807.92(a) (6)]
The subject device and the primary predicate device are stationary x-ray devices. There will be 2 main differences:
First, this device, 'GXR-ES/ECS Diagnostic X-ray System has Compact System', is for only essential use in that it is composed of simplified mechanical parts.
But, the predicate device 'GXR- series Diagnostic Imaging System' has Manual type as well as motorized option of Tube stand, patient table and WBS (Diagnostic X-ray Full system). And the subject device has lower high voltage output rating than predicate device for Use of small-scale hospital facility operation.
Second, Difference between subject and predicate device is diversity of the power supply method of the generator
Generally, the subject device 'GXR-ES, ECS' has general and capacitor type generator supplied from wall power. However, the predicate device 'GXR-series Diagnostic Imaging System' has also UPS battery type (total three type).
Differences between the subject device and the predicates has no impact on safety or effectiveness of the new device and does not raise any new potential or increased safety risks and is equivalent in performance to existing legally marketed devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for DRGEM. The letters "DR" are in black, while "GEM" is in teal. Below the logo is the text "Your Best Healthcare" in a smaller font.
# VII.SUBSTANTIAL EQUIVALENCE [21 CFR 807.92(b)]
| Item | Subject Device | Predicate Device | Impact of Differences |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | GXR-ES/ECS Series Diagnostic<br>X-ray System | GXR-Series<br>Diagnostic X-Ray<br>System (K202572) | Not applicable |
| Manufacturer | DRGEM Corporation | DRGEM Corporation | Not applicable |
| Model Number | GXR-ES series,<br>GXR-ECS series | GXR-S series | Not applicable |
| High Frequency X- ray Generator | | | |
| Output Power | 20kW, 25kW, 32kW, 40kW,<br>50kW | 32KW, 40KW,<br>52KW, 68KW,<br>82KW | Yes, there is a difference in output<br>values but no difference in<br>generators.<br>See Difference Explanation below. |
| Generator<br>models<br>(manufactured<br>by DRGEM) | GXR-E20, GXR-E25,<br>GXR-E32, GXR-E40,<br>GXR-E50 | GXR-32, GXR- 40, GXR-<br>52, GXR-68, GXR- 82 | Yes, there is a difference.<br>Models have been tested against<br>International Safety and EMC<br>Standards. Any differences<br>between the subject device and<br>predicate device do not change or<br>add new potential safety risks. It is<br>our determination that there is "No<br>negative impact on safety or<br>effectiveness" and there are no new<br>potential<br>or increased safety risks concerning<br>this difference. |
| Line voltage | 100~120VAC<br>200~240VAC | 220~230VAC,<br>380/400/480VA C | Yes, there are differences in line<br>voltage depending upon the system<br>requirements. Models have been<br>tested against International Safety<br>and EMC Standards. Any<br>differences between the subject<br>device and predicate device do not<br>change or add new potential safety<br>risks. It is our determination that<br>there is "No negative impact on<br>safety or efficacy" and there are no<br>new potential or increased safety<br>risks<br>concerning this difference |
| Image Acquisition System | | | |
| Flat panel<br>Detector | Detector is not supplied with system | Detector is not supplied with system | Not applicable |
| Patient table | | | |
| Configuration model | PBT-E | PBT-4, PBT-6, PDT-1 | Yes, there is a difference.<br>Models have been tested against International Safety and EMC Standards. Any differences between the subject device and predicate device do not change or add new potential safety risks. It is our determination that there is "No negative impact on safety or effectiveness" and there are no new potential or increased safety risks concerning this difference |
| Movement | Longitudinal :<br>1050mm ± 10mm | Longitudinal :<br>1000mm ± 500mm | |
| Wall Bucky stand | | | |
| Configuration model | WBS-E, WBS | WBS, WBS-TM | Yes, there is a difference.<br>Models have been tested against International Safety and EMC Standards. Any differences between the subject device and predicate device do not change or add new potential safety risks. It is our determination that there is "No negative impact on safety or effectiveness" and there are no new potential or increased safety risks concerning this difference |
| Dimension/weight | 560(W) x 690(D) x 1,826(H)mm /<br>32.4kg(71.4lbs) (Floor Base)<br>420(W) x 256(D) x 1,826(H)mm /<br>21.4kg(47.2lbs) (Floor-Wall Mounted)<br>659(W) x 393(D) x 2,169(H)mm /<br>120kg(264lbs) | 659(W) x 445(D) x<br>1,599(H)mm / 110kg(242lbs)<br>659(W) x 445(D) x<br>1,749(H)mm / 113kg(249lbs)<br>659(W) x 445(D) x<br>1,919(H)mm / 116kg(255lbs)<br>659(W) x 445(D) x<br>2,169(H)mm / 120kg(264lbs) | |
| Tube stand | | | |
| Configuration model | TS-FT6, TS-FT4 | TS-FC6, TS-FC4, TS-FC2<br>TS-FM6, TS-CSA, TS-CSE | Not applicable |
| Tube Rotation<br>Angle | ±135° | ±135° | |
| Configuration<br>model | DXR-8M, DXT-10M, DXT-<br>11M, DXR-12M, E7239X,<br>E7242X, E7884X, | E7239X, DXT-8M,<br>E7242X, DXT-11M<br>E7843X, DXT-10M<br>E7876X, DXT-12M<br>E7884X, E7252X<br>DXT-14U, RAD-14U,<br>DXT-15U, RAD-21<br>RAD-60, E7255FX,<br>E4254FX | No Impact of difference |
| Max. kV | 125kV, 150kV | 125kV, 150kV | |
| Collimator | | | |
| Configuration<br>model | DXC-RML | DXC-RML, DXC-RMH<br>MCR, R108, R302A,<br>R302MLP/A,<br>R302MFMLP/A | Yes, there is a difference.<br>Models have been tested against<br>International Safety and EMC<br>Standards. Any differences<br>between the subject device and<br>predicate device do not change or<br>add new potential safety risks. It is<br>our determination that there is "No<br>negative impact on safety or<br>effectiveness" and there are no new<br>potential<br>or increased safety risks concerning<br>this difference |
| Lamp Type | LED lamp<br>(Over 160LUX at 100cm<br>SID) | LED and Halogen lamp<br>(Over 160LUX at 100cm<br>SID) | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DRGEM. The letters "DRG" are in bold black font, while the letters "EM" are in a teal font. Below the logo is the text "Your Best Healthcare" in a smaller, gray font.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for DRGEM, a healthcare provider. The letters "DR" are in bold, black font, while "GEM" is in a teal color. Below the logo, the text "Your Best Healthcare" is written in a smaller, lighter font.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DRGEM. The letters "DR" are in black, and the letters "GEM" are in teal. Below the letters is the text "Your Best Healthcare" in a smaller font.
#### VIII. SUMMARY OF NON-CLINICAL Data [21 CFR 807.92(b) (1)]
Nonclinical Testing:
The GXR-ES/ ECS Series Diagnostic X-Ray System, has been assessed and tested and has passed predetermined testing criteria. The Validation Test Plan was designed to evaluate input functions, output functions, and actions performed by the subject device and followed the process documented in the System Validation Test Plan.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
| Std # | Safety/EMC Standards Description | FDA Rec.<br>Standard # |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| IEC 60601-1-3 | Medical electrical equipment<br>Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray<br>Equipment | 12-269 |
| IEC 60601-1-6 | IEC 60601-1-6 Edition 3.1 2013-10<br>Medical electrical equipment - Part 1-6: General requirements for basic safety and<br>essential performance - Collateral standard: Usability | 5-89 |
| IEC 60601-2-28 | IEC 60601-2-28<br>Medical electrical equipment Part 2: Particular requirements for the safety of X-ray<br>source assemblies and X-ray tube assemblies for medical diagnosis | 12-309 |
| IEC 60601-2-54 | IEC 60601-2-54<br>Medical electrical equipment<br>Part 2: Particular requirements for the basic safety and essential<br>performance of X-ray equipment for radiography and radioscopy | 12-317 |
| IEC 60601-1-2<br>(EMC) | IEC 60601-1-2 Edition 4.0 2014-02. Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances Requirements and tests. | 19-8 |
| IEC 62304:2006 | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION<br>Medical device software - Software life cycle processes | 13-79 |
| IEC 60601-1 | Medical electrical equipment, Part 1: General requirements for basic safety and<br>essential performance | 19-4 |
| ISO 14971:2019 | ISO 14971:2019 Third Edition, Medical devices - Applications of risk management to<br>medical devices. | 5-125 |
| ISO 15223-1 | ISO 15223-1 Fourth Edition 2021-07, Medical devices - Symbols to be used with<br>medical device labels, labelling, and information to be supplied - Part 1: General<br>requirements. | 5-134 |
| TR 60601-4-2 | TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and<br>interpretation - Electromagnetic immunity: performance of medical electrical equipment<br>and medical electrical systems | 19-19 |
| FDA Guidance | Pediatric Information for X-ray Imaging Device Premarket Notifications dated November 28, 2017 | |
| FDA Guidance | Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance<br>for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014 | |
| FDA Guidance | Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software<br>contained in Medical Devices, Document issued on: May 11, 2005 Medical Devices, Document issued<br>on: May 11 2005 | |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for DRGEM, a healthcare company. The letters "DR" are in bold, black font, while the letters "GEM" are in a teal color. Below the company name is the tagline "Your Best Healthcare" in a smaller, lighter font.
#### Summary:
Based on the performance as documented in the V&V Testing, the subject device was found to have a safe and effectiveness profile that is similar to the predicate device.
The above standards were used to develop and verify electrical safety, and EMC. GXR-ES/ ECS Series Diagnostic X-Ray System device has met all the requirements listed in the Standards except for inapplicable requirements.
The subject device conform to all applicable aspects of 21CFR 1020.30
# IX. CONCLUSIONS [21 CFR 807.92(b) (3)]
The 510(k) Pre-Market Notification for GXR-ES/ECS series Diagnostic X-ray System, contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device and the predicate device are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs comparably to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.