K232144 · Medartis AG · HRS · Aug 18, 2023 · Orthopedic
Device Facts
Record ID
K232144
Device Name
Sterile Products of the APTUS System
Applicant
Medartis AG
Product Code
HRS · Orthopedic
Decision Date
Aug 18, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The APTUS® Titanium Fixation System is indicated for use in hand and forearm fractures and arthrodeses. APTUS® Hand group: - · Management of the fractures of the distal, middle and proximal phalanges and metacarpals - · Management of all types of transversal fractures, spiral fractures near joints with or without joint involvement, shaft fractures, comminuted fractures, dislocation fractures, avulsion fractures - · DIP and PIP arthrodeses - APTUS® Radius 2.5 group: - · Management via radio volar approach of extra-aticular extension and flexion fractures, articular extension and flexion fractures, correction osteotomies for badly healed radius fractures - Management via dorsal approach of ractures that cannot be adequately reduced via volar approach, procedures for which the soft tissue conditions make a volar approach very difficult or impossible, correction osteotomies requiring stabilization from the dorsal side, carporadial fusions. The APTUS® 2.0/2.3 Four Corner Fusion Plate, an addition to the APTUS® Titanium Fixation System, is designed specifically for fusion of carpal bones including: hamate, capitate, lunate, triquetrum and is for use in patients suffering pain and/or loss of function due to osteoarthritis, post-traumatic arthritis, fractures, revision of failed partial wrist fusions, carpal instability, or rheumatoid arthritis. The fusion plate is used in conjunction with locking and non-locking screws that fix the plate to the carpal bones of the hand. The APTUS® K-Wire System is intended for use in fixation of bone reconstruction, and as guide pins for insertion of other implants. The APTUS® 1.5 TriLock is indicated for use in hand and forearm fractures, osteotomies and arthrodeses. APTUS® Hand group: - · Management of the fractures of the distal, middle and proximal phalanges and metacarpals - · Management of all types of transversal fractures, fractures near joints with or without joint involvement, shaft fractures, comminuted fractures, dislocation fractures, avulsion fractures - DIP and PIP arthrodeses - APTUS® Radius 2.5 group: - · Management via radio volar approach of extra-aticular extension and flexion fractures, articular extension and flexion fractures, correction osteotomies for badly healed radius fractures - Management via dorsal approach of ractures that cannot be adequately reduced via volar approach. procedures for which the soft tissue conditions make a volar approach very difficult or impossible, correction osteotomies requiring stabilization from the dorsal side, carporadial fusions. APTUS® Ulna Plates are indicated for fractures and osteotomies, in particular for the ulna. APTUS® Wrist 2.5 System is indicated for use in hand and forearm fractures and arthrodeses. APTUS® Wrist Arthrodesis Plates APTUS® Wrist Arthrodesis Plates are indicated for wrist arthrodesis. APTUS® Forearm Shaft Plates are indicated for management of fractures and osteotomies of the radius and ulna shaft.
Device Story
System of plates, screws, and K-wires for internal bone fixation in hand, wrist, and forearm. Plates feature anatomical designs; screws include locking and non-locking types; K-wires provided in various diameters/points. Used by surgeons in clinical settings to stabilize fractures, perform osteotomies, or facilitate arthrodeses. Subject devices are previously cleared components now provided in sterile condition. Benefits include stabilized bone healing and restored function. No change to clinical application or surgical technique compared to predicates.
Clinical Evidence
Bench testing only. No clinical data provided. Sterilization validation performed per ISO 11137-1, -2, -3. Sterile barrier shelf life validated per ISO 11607-2, ASTM F1980, ASTM D4332, ASTM D4169, ASTM F2096, DIN EN 868-5, and ASTM F88/F88M.
Indicated for patients requiring internal fixation for hand and forearm fractures, osteotomies, and arthrodeses; includes management of phalangeal, metacarpal, radius, and ulna fractures; carpal bone fusion; and bone reconstruction. Contraindications not specified.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
August 18, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medartis AG % Chelsea Kozior Regulatory Affairs Manager Medartis Inc. 1195 Polk Drive Warsaw, Indiana 46582
Re: K232144
Trade/Device Name: Sterile Products of the APTUS System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: July 17, 2023 Received: July 19, 2023
Dear Chelsea Kozior:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232144
#### Device Name
Sterile Products of the APTUS System
### Indications for Use (Describe)
APTUS® Titanium Fixation System
The APTUS® Titanium Fixation System is indicated for use in hand and forearm fractures and arthrodeses. APTUS® Hand group:
- · Management of the fractures of the distal, middle and proximal phalanges and metacarpals
- · Management of all types of transversal fractures, spiral fractures near joints with or without joint
- involvement, shaft fractures, comminuted fractures, dislocation fractures, avulsion fractures
- · DIP and PIP arthrodeses
- APTUS® Radius 2.5 group:
- · Management via radio volar approach of extra-aticular extension and flexion fractures, articular extension and flexion fractures, correction osteotomies for badly healed radius fractures
- Management via dorsal approach of ractures that cannot be adequately reduced via volar approach, procedures for which the soft tissue conditions make a volar approach very difficult or impossible, correction osteotomies requiring stabilization from the dorsal side, carporadial fusions.
### APTUS® 2.0/2.3 Four Corner Fusion Plate
The APTUS® 2.0/2.3 Four Corner Fusion Plate, an addition to the APTUS® Titanium Fixation System, is designed specifically for fusion of carpal bones including: hamate, capitate, lunate, triquetrum and is for use in patients suffering pain and/or loss of function due to osteoarthritis, post-traumatic arthritis, fractures, revision of failed partial wrist fusions, carpal instability, or rheumatoid arthritis. The fusion plate is used in conjunction with locking and non-locking screws that fix the plate to the carpal bones of the hand.
### APTUS® K-Wire System
The APTUS® K-Wire System is intended for use in fixation of bone reconstruction, and as guide pins for insertion of other implants.
### APTUS® 1.5 TriLock
The APTUS® 1.5 TriLock is indicated for use in hand and forearm fractures, osteotomies and arthrodeses. APTUS® Hand group:
- · Management of the fractures of the distal, middle and proximal phalanges and metacarpals
- · Management of all types of transversal fractures, fractures near joints with or without joint
- involvement, shaft fractures, comminuted fractures, dislocation fractures, avulsion fractures
- DIP and PIP arthrodeses
- APTUS® Radius 2.5 group:
- · Management via radio volar approach of extra-aticular extension and flexion fractures, articular extension and flexion fractures, correction osteotomies for badly healed radius fractures
- Management via dorsal approach of ractures that cannot be adequately reduced via volar approach.
procedures for which the soft tissue conditions make a volar approach very difficult or impossible, correction osteotomies requiring stabilization from the dorsal side, carporadial fusions.
### APTUS® Ulna Plates
APTUS® Ulna Plates are indicated for fractures and osteotomies, in particular for the ulna.
#### APTUS® Wrist 2.5 System
{3}------------------------------------------------
APTUS® Wrist 2.5 System is indicated for use in hand and forearm fractures and arthrodeses.
APTUS® Wrist Arthrodesis Plates APTUS® Wrist Arthrodesis Plates are indicated for wrist arthrodesis.
APTUS® Forearm Shaft Plates
APTUS® Forearm Shaft Plates are indicated for management of fractures and osteotomies of the radius and ulna shaft.
Type of Use (*Select one or both, as applicable*)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Prepared on: 2023-08-07
## Contact Details
| Applicant Name | Medartis AG | | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------|--|
| Applicant Address | Hochbergerstrasse 60E Basel 4057 Switzerland | | |
| Applicant Contact Telephone | +41 61 6333434 | | |
| Applicant Contact | Ms. Claudia De Santis | | |
| Applicant Contact Email | claudia.desantis@medartis.com | | |
| Correspondent Name | Medartis Inc. | | |
| Correspondent Address | 1195 Polk Drive Warsaw IN 46582 United States | | |
| Correspondent Contact Telephone | 610-731-8650 | | |
| Correspondent Contact | Ms. Chelsea Kozior | | |
| Correspondent Contact Email | chelsea.kozior@medartis.com | | |
| Device Name | 21 CFR 807.92(a)(2) | | |
| Device Trade Name | Sterile Products of the APTUS System | | |
| Common Name | Plate, Fixation, Bone (Primary) / Screw, Fixation, Bone / Pin, Fixation, Smooth | | |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories (Primary)<br>Smooth or threaded metallic bone fixation fastener | | |
| Regulation Number | 888.3030 (Primary) / 888.3040 | | |
| Product Code | HRS (Primary) / HWC / HTY | | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |
| K051567 | APTUS® Titanium Fixation System | HRS | |
| K090983 | APTUS® 2.0/2.3 Four Corner Fusion Plates | HRS | |
| K092038 | APTUS® K-Wire System | HTY | |
| K102537 | APTUS® 1.5 TriLock | HRS | |
| K103332 | APTUS® Ulna System | HRS | |
{5}------------------------------------------------
| K142906 | APTUS ® Wrist 2.5 System | HRS |
|---------|--------------------------|-----|
|---------|--------------------------|-----|
| K172170 | APTUS ® Wrist 2.5 System | HRS |
|---------|--------------------------|-----|
|---------|--------------------------|-----|
| K192297 | APTUS ® Wrist Arthrodesis System | HRS |
|---------|----------------------------------|-----|
|---------|----------------------------------|-----|
| K193554 | APTUS ® Forearm Shaft Plates | HRS |
|---------|------------------------------|-----|
|---------|------------------------------|-----|
| K191848 | APTUS ® Wrist Spanning Plates 2.5 (Reference Device) | HRS |
|---------|------------------------------------------------------|-----|
|---------|------------------------------------------------------|-----|
### Device Description Summary
21 CFR 807.92(a)(4)
The Sterile Products of the APTUS System is a plating, screw, and K-wire system for internal fixation of the hand, wrist, and/ or forearm. All plates have anatomical designs that are appropriate to their intended use. The subject device plates includes crew holes designed to accommodate appropriately sized bone screws.
Sterile Products of the APTUS System plate, screw, and K-wire system are the previously cleared predicate and additional predicate devices. The overal dimens of the subject device plates have the same lengths, widths, and thicknesses as the previously cleared predicate and additional predicate device screws have the same design as the previously cleared Medartis APTUS Screws under K051567, K103332, and K142906. These include both locking (cortical) screws. The subject device K-wires have design as the previously cleared APTUS K-Wires System under K092038. The K-wires are provided with a diameter of 1.2, 1.6, or 1.8 mm. All have a length of 150 mm. Single-ended trocar points are provided in 1.2 mm and 1.6 mm diameters, and single-ended bayonet (lancet) points are provided at all 3 diameters.
This submission includes all plates screws, and K-wires that have been previously cleared under the predicate devices by Medartis AG, however, they will be provided in a sterile condition.
All subject device plates are manufactured from unalloyed titanium conforming to ASTM F67.
All subject device screws are manufactured from titanium alloy conforming to ASTM F136.
All subject device K-wires are manufactured from stainless steel conforming to ASTM F138.
All subject devices are associated with the predicate device and/or an additional predicate device mentioned in the predicates and product list sections.
### Intended Use/Indications for Use
### 21 CFR 807.92(a)(5)
The APTUS® Titanium Fixation System
The APTUS® Titanium Fixation System is indicated for use in hand and forearm fractures, osteotomies and arthrodeses. APTUS® Hand group:
· Management of the fractures of the distal, middle and proximal phalanges and metacarpals
· Management of all types of transversal fractures, fractures near joints with or without joint involvement, shaft fractures, comminuted fractures, dislocation fractures, avulsion fractures
· DIP and PIP arthrodeses
APTUS® Radius 2.5 group:
· Management via radio volar approach of extra-articular extension and flexion fractures, articular extension and flexion fractures, correction osteotomies for badly healed radius fractures
· Management via dorsal approach of rare extension fractures that cannot be adequately reduced via volar approach, procedures for which the soft tissue conditions make a volar approach very difficult or impossible, correction osteotomies requiring stabilization from the dorsal side, carporadial fusions.
### APTUS® 2.0/2.3 Four Corner Fusion Plate
The APTUS® 2.0/2.3 Four Corner Fusion Plate, an addition to the APTUS® Titanium Fixation System, is designed specifically for fusion of carpal bones including: hamate, capitate, lunate, triquetrum and is for use in patients suffering pain and/or loss of function due to osteoarthritis, post-traumatic arthritis, fractures, revision of failed partial wrist fusions, carpal instability, or rheumatoid arthritis. The fusion plate is used in conjunction with locking and non-locking screws that fix the plate to the carpal bones of the hand.
### APTUS® K-Wire System
The APTUS® K-Wire System is intended for use in fixation of bone reconstruction, and as quide pins for insertion of other implants.
{6}------------------------------------------------
### APTUS® 1.5 TriLock
The APTUS® 1.5 TriLock is indicated for use in hand and forearm fractures, osteotomies and arthrodeses.
APTUS® Hand group:
- Management of the fractures of the distal, middle and proximal phalanges and metacarpals
· Management of all types of transversal fractures, fractures near joints with or without joint involvement, shaft fractures, comminuted fractures, dislocation fractures, avulsion fractures
· DIP and PIP arthrodeses
APTUS® Radius 2.5 group:
· Management via radio volar approach of extra-articular extension fractures, articular extension and flexion fractures, correction osteotomies for badly healed radius fractures
· Management via dorsal approach of rare extension fractures that cannot be adequately reduced via volar approach, procedures for which the soft tissue conditions make a volar approach very difficult or impossible, correction osteotomies requiring stabilization from the dorsal side, carporadial fusions.
APTUS® Ulna Plates
APTUS® Ulna Plates are indicated for fractures and osteotomies, in particular for the ulna.
APTUS® Wrist 2.5 System
APTUS® Wrist 2.5 System is indicated for use in hand and forearm fractures, osteotomies and arthrodeses.
APTUS® Wrist Arthrodesis Plates APTUS® Wrist Arthrodesis Plates are indicated for wrist arthrodesis.
APTUS® Forearm Shaft Plates
APTUS® Forearm Shaft Plates are indicated for management of fractures and osteotomies of the radius and ulna shaft.
### Indications for Use Comparison
### The subject device has the same intended use and the same or similar Indications for Use as the predicate cleared under K051567 or associated additional predicate device cleared under K090983, K102537, K103332, K142906, K172170, K193554,
### Technological Comparison
### The subject device has the same technological characteristics as the predicate and/or the associated additional predicate devices. The subject and the predicate devices are all manufactured from the same materials (Unalloyed titanium, ASTM F67, grade 4, Ti-6Al-4 V alloy, ASTM F136; and stainless steel, ASTMF138). The subject and predicate devices share the subject device has modified packaged to be provided sterile to the end user. The subject device also has modification
process to be provided sterile to the end user. The device of this submission, previously cleared under the reference device submission, do not raise any new questions of safety or effectiveness.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Performance Data
An analysis of the risk documentation and an engineering analysis was performed. These determined that the original testing of the sterilization process, previously cleared in the reference device submission under K191848, and the subsequent device performance is not adversely affected by the modifications to the results of the analysis demonstrated the modified sterility and packaging are substantially equivalent to the predicate devices.
Nor-clinical testing data referenced and relied won to demonstrate included: sterilization validation according to ISO 1137-1, ISO 11137-2, and ISO 11137-3. Sterile barrier shelf life testing according to ISO 11607-2, ASTM F1980, ASTM D4332, ASTM D4169, ASTM F2096, DIN EN 868-5, and ASTM F88/F88M.
The subject device is manufactured from the same matering processes as their associated predicate device. There is no change to the type and duration of tissue contact for the subject device from the associated predicate device. The device modification, changing the sterilization method, was previously cleared under the reference device submission. Therefore, the device modification, that is the subject of this submission, is not expected to impact the biological response.
Clinical data is not provided in this submission.
Conclusion
In summary, the Sterile Products of the APTUS System has the following similarities to the predicate device and additional predicate devices :
{7}------------------------------------------------
· has the same intended use and the same or similar indications for use ,
- · uses the same operating principle,
- · incorporates the same basic design,
- · incorporates the same materials, and
• has modified packaging so the device can be provided to the end user in a sterile condition.
The device modifications do not raise any new questions of the subject device. Therefore, the Sterile Products of the APTUS System is substantially equivalent to the predicate devices with respect to intended use, operating principles, and technical characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.