← Product Code [OZN](/productcode/OZN) · K232092

# Great Basin Toxigenic C. difficile Direct Test (CDF2) (K232092)

_Vela Operations USA · OZN · Nov 14, 2023 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232092

## Device Facts

- **Applicant:** Vela Operations USA
- **Product Code:** [OZN](/productcode/OZN.md)
- **Decision Date:** Nov 14, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3130
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K232092 · Nov 14, 2023 | Great Basin Toxigenic C. difficile Direct Test (CDF2) | Vela Operations USA | Excess remnants of clinical stool specimens from routine standard-of-care testing | The study used prospectively collected clinical remnant specimens to evaluate the performance (PPA/NPA) of the Toxigenic C. difficile Direct Test compared to an FDA-cleared molecular reference method. | Clinical remnant specimens; Standard of care; Method comparison; Prospective clinical study |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Prospective multi-center method comparison study; Prospective multi-center method comparison; Follow-up/Duration: June 2022 - February 2023; Study Period: June 2022 - February 2023 | Symptomatic individuals ≥ 2 years of age suspected of C. difficile infection (CDI); Sample Size: 829 specimens; Number of Sites: 3 | FDA-cleared molecular reference method | Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) |

## Indications for Use

The Great Basin Toxigenic C. difficile Direct Test, performed on the Great Basin PA500 Analyzer, is a qualitative in vitro diagnostic test for the detection of toxigenic Clostridioides difficile in unformed (liquid or soft) stool samples collected from patients suspected of having C. difficile infection (CDI). The automated assay utilizes polymerase chain reaction (PCR) to detect a conserved region of the toxin gene (tcdB) associated with toxin-producing C. difficile. The Toxigenic C. difficile Direct Test is intended for use as an aid in the diagnosis of CDI in conjunction with clinical and epidemiological risk factors.

## Device Story

Device is an automated in vitro diagnostic test for C. difficile infection (CDI). Input: unformed stool specimen. Process: specimen processed via Sample Preparation Device (SPD) to extract genomic DNA; automated hot-start PCR amplification of tcdB gene; biotin-labeled amplicons hybridized to probes on a silicon chip; HRP-conjugated antibody binding; TMB substrate addition produces colored precipitate. Output: optical signal detected by PA500 Analyzer; automated result interpretation. Used in clinical laboratories by trained personnel. Provides rapid (under 2 hours) qualitative detection of toxigenic C. difficile to aid clinical diagnosis.

## Clinical Evidence

Prospective multi-center method comparison study (n=829) compared CDF2 to an FDA-cleared molecular reference. Participants were symptomatic individuals ≥ 2 years old. Results: PPA 95.8% (91.2-98.5%), NPA 98.5% (97.3-99.3%). Discrepant analysis using a third molecular test supported the CDF2 results for 5/6 false negatives and 10/10 false positives. Analytical studies included LoD, inclusivity (22 strains), exclusivity (83 organisms), microbial interference, and chemical interference.

## Technological Characteristics

System uses hot-start PCR amplification and colorimetric detection via hybridization probes on a modified silicon chip. Energy source: electrical (PA500 Analyzer). Form factor: single-use, self-contained cartridge. Connectivity: integrated system. Software: automated result interpretation. Internal controls: Sample Processing Control (SPC), Hybridization Control (HC), and Detection Control (DC).

## Regulatory Identification

A Clostridium difficile toxin gene amplification assay is a device that consists of reagents for the amplification and detection of target sequences in Clostridium difficile toxin genes in fecal specimens from patients suspected of having Clostridium difficile infection (CDI). The detection of clostridial toxin genes, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of CDI caused by Clostridium difficile.

## Special Controls

In combination with the general controls of the FD&C Act, the Portrait Toxigenic C. difficile Assay is subject to the following special controls: 1. The special controls for the Clostridium difficile toxin gene amplification assay are contained in the guidance document: "Class II Special Controls Guidance Document: Toxin Gene Amplification Assays for the Detection of Clostridium difficile."

*Classification.* Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Toxin Gene Amplification Assays for the Detection of*Clostridium difficile;* Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.

## Predicate Devices

- GenePOC CDiff Assay ([K172569](/device/K172569.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K232092

B Applicant

Vela Operations USA

C Proprietary and Established Names

Great Basin Toxigenic C. difficile Direct Test (CDF2)

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  OZN | Class II | 21 CFR 866.3130 - Clostridium Difficile Toxin Gene Amplification Assay | MI - Microbiology  |

## II Submission/Device Overview:

A Purpose for Submission:

To obtain substantial equivalence for the Great Basin Toxigenic C. difficile Direct Test (CDF2).

B Measurand:

tcdB gene of toxigenic Clostridioides difficile

C Type of Test:

The Toxigenic C. difficile Direct Test (CDF2) is a polymerase chain reaction (PCR) which amplifies specific sequences that are hybridized onto probes immobilized on a silicon chip surface.

## III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K232092 - Page 2 of 16

## B Indication(s) for Use:

The Great Basin Toxigenic *C. difficile* Direct Test, performed on the Great Basin PA500 Analyzer, is a qualitative *in vitro* diagnostic test for the detection of toxigenic *Clostridioides difficile* in unformed (liquid or soft) stool samples collected from patients suspected of having *C. difficile* infection (CDI). The automated assay utilizes polymerase chain reaction (PCR) to detect a conserved region of the toxin gene (*tcdB*) associated with toxin-producing *C. difficile*.

The Toxigenic *C. difficile* Direct Test is intended for use as an aid in the diagnosis of CDI in conjunction with clinical and epidemiological risk factors

## C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

## D Special Instrument Requirements:

Great Basin PA500 Analyzer

## IV Device/System Characteristics:

### A Device Description:

The Great Basin Toxigenic *C. difficile* Direct Test (CDF2) performed on the PA500 Analyzer utilizes automated, hot-start PCR amplification technology to amplify specific nucleic acid sequences that are then detected using hybridization probes immobilized on a modified silicon chip surface, in a single-use, self-contained test cartridge.

A swab volume of the specimen (raw stool) is first processed using the Sample Preparation Device (SPD). During processing through the SPD, the specimen is infused with a synthetic bacterial sample processing control (SPC). An aliquot of the eluate obtained from the SPD, containing diluted specimen mixed with SPC, is loaded into the sample port of the CDF2 Test Cartridge.

Genomic DNA is extracted from microbial cells and diluted to reduce potential inhibitors of PCR. During the PCR process, biotin-labeled primers direct the amplification of specific nucleic acid sequences within a conserved region of the *C. difficile* Toxin B (*tcdB*) gene.

Following PCR, biotin-labeled, amplified target DNA sequences are hybridized to sequence specific probes immobilized on the silicon chip surface and incubated with anti-biotin antibody conjugated to the horseradish peroxidase enzyme (HRP). The unbound conjugate is washed away, then tetramethylbenzidine (TMB) is added to produce a colored precipitate at the location of the probe/target sequence complex.

The resulting signal is detected by the automated PA500 Optical Reader within the PA500 Analyzer System. The SPC undergoes the same extraction, amplification, and detection steps as the sample in order to monitor for inhibitory substances, as well as process inefficiency due to instrument or reagent failure. No operator intervention is required once the sample is loaded into

{2}

the sample port, and the Toxigenic C. difficile Direct Test cartridge is loaded into the PA500 Analyzer.

## B Principle of Operation:

The Great Basin System is a fully automated in vitro diagnostic system that includes: the PA500 Analyzer device, single-use SPDs, single-use Toxigenic C. difficile Direct Test Cartridges, and the PA500 Data Analysis Software Program.

The PA500 Analyzer is designed to perform automated sample preparation, target amplification via PCR, chip-based (optical) target detection, and integrated data analysis in less than two hours.

## V Substantial Equivalence Information:

A Predicate Device Name(s): GenePOC CDiff

B Predicate 510(k) Number(s): K172569

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | Toxigenic C. difficile Direct Test (CDF2) K232092 | GenePOC CDiff K172569  |
| --- | --- | --- |
|  Device Trade Name | Great Basin Toxigenic C. difficile Direct Test | GenePOC CDiff Assay  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | The Great Basin Toxigenic C. difficile Direct Test, performed on the Great Basin PA500 Analyzer, is a qualitative in vitro diagnostic test for the detection of toxigenic Clostridioides difficile in unformed (liquid or soft) stool samples collected from patients suspected of having C. difficile infection (CDI). The automated assay utilizes polymerase chain reaction (PCR) to | The GenePOC CDiff assay performed on the revogene instrument is a qualitative in vitro diagnostic test that utilizes automated sample processing and real-time polymerase chain reaction (PCR) to detect the toxin B (tcdB) gene of toxigenic Clostridium difficile (C. difficile) in unformed (liquid or soft) stool specimens obtained from patients suspected of having C. difficile  |

K232092 - Page 3 of 16

{3}

|   | detect a conserved region of the toxin gene (tcdB) associated with toxin-producing C. difficile.
The Toxigenic C. difficile Direct Test is intended for use as an aid in the diagnosis of CDI in conjunction with clinical and epidemiological risk factors. | infection (CDI). The GenePOC CDiff assay is intended to aid in the diagnosis of CDI.  |
| --- | --- | --- |
|  Specimen Type | Unformed (liquid or soft) stool specimens | Same  |
|  Assay Target | toxin B gene (tcdB) | Same  |
|  Sample Lysis and DNA Extraction | Automated (in cartridge) | Same  |
|  Amplification Technology | Multiplex PCR | Same  |
|  Result Interpretation | Automated | Same  |
|  General Device Characteristic Differences |  |   |
|  Detection Probes | Hybridization Probes | TaqMan Probes  |
|  Detection Technology | Colorimetric target specific hybridization to probe on a chip surface, optical reader, automated software with built-in result interpretation. | Real-time PCR (fluorescent signals) and embedded detection algorithm  |
|  Sample Volume | 1 swab | 5 ul  |
|  Sample per Run | One | Up to 8  |
|  Time to Result | 2 hours | 70 min  |

VI Standards/Guidance Documents Referenced:

1. Guidance for Clinical Investigators, Industry and FDA Staff: Financial Disclosure by Clinical Investigators. February 2013
2. Guidance on Informed Consent for In Vitro Diagnostics Device Studies Using Leftover Human Specimens that are not Individually Identifiable. April 2006
3. Class II Special Controls Guideline Document: Toxin Gene Amplification Assays for the Detection of Clostridium difficile – Guideline for Industry and Food and Drug Administration Staff. August 2015

K232092 - Page 4 of 16

{4}

4. CLSI MM13-A: Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods: Approved Guideline.
5. CLSI MM3-A2: Molecular Diagnostic Methods for Infectious Diseases: Approved Guideline

## VII Performance Characteristics (if/when applicable):

## A Analytical Performance:

### 1. Precision/Reproducibility:

A between-site reproducibility study was performed at three sites (two external and one internal) by two operators per site, over five non-consecutive days using two cartridge lots and 24 analyzers. A total of 90 negative sample replicates and 360 positive sample replicates were tested. Two toxigenic C. difficile strains were tested: ATCC 43255 (toxinotype 0, ribotype 087) and ATCC BAA-1805 (toxinotype IIIb, NAP1/ribotype 027).

A five-member panel, tested in triplicate, was contrived using a pool of C. difficile-negative clinical stool matrix at the following target concentrations:

- Low positive at 1.5X LoD
- Moderate positive at 3X LoD
- True negative (TN) samples without toxigenic C. difficile

The low positive (1.5X LoD) concentrations tested were 3.9 x 10⁴ CFU/mL and 1.1 x 10⁵ CFU/mL for ATCC 43255 and ATCC BAA-1805, respectively. The moderate positive (3X LoD) concentrations tested were 7.8 x 10⁴ CFU/mL and 2.2 x 10⁵ CFU/mL for ATCC 43255 and ATCC BAA-1805, respectively.

Results are shown by site and overall, in Table 1. For the between-site reproducibility, the overall percent agreement was 97.8% for the low positive (1.5X) samples and 100% for moderate positive (3X) samples for ATCC strain 43255. The overall percent agreement was 100% for the low positive (1.5X) and moderate positive (3X) samples of strain ATCC BAA-1805. The overall percent agreement was 98.9% for the negative samples. The two false negative results from Site 02 were attributed to assay variability and sampling error. The false positive result from Site 02 was attributed to unintentional sample contamination by the operator. There was no variability in results by operator or by lot.

Table 1. Reproducibility Results by Site and Overall

|  Panel | Strain | Assay Results/Expected |   |   |   | Agreement Total [95% CI]  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  Site 01 | Site 02 | Site 03 | All sites  |   |
|  Low Positive (1.5X LoD) | ATCC 43255 | 30/30 | 28/30 | 30/30 | 88/90 | 97.8% [92.2%-99.7%]  |
|   |  ATCC BAA-1805 | 30/30 | 30/30 | 30/30 | 90/90 | 100% [96%-100%]  |
|  Moderate Positive (3X LoD) | ATCC 43255 | 30/30 | 30/30 | 30/30 | 90/90 | 100% [96%-100%]  |
|   |  ATCC BAA-1805 | 30/30 | 30/30 | 30/30 | 90/90 | 100% [96%-100%]  |

K232092 - Page 5 of 16

{5}

K232092 - Page 6 of 16

|  Negative | N/A | 30/30 | 29/30 | 30/30 | 89/90 | 98.9%
[94%-100%]  |
| --- | --- | --- | --- | --- | --- | --- |

2. Linearity:

Not applicable

3. Analytical Specificity/Interference:

a. Analytical inclusivity:

The analytical inclusivity of the Toxigenic C. difficile Direct Test (CDF2) assay was evaluated by testing 22 strains of toxigenic C. difficile representing six different toxinotypes (Summarized in Table 2). Quantitated cultures of each C. difficile strain were diluted to $9.7 \times 10^{4}$ CFU/ml in clinical negative stool matrix, and three replicates per strain were tested using six cartridge lots. All strains were detected, which is acceptable.

Table 2. C. difficile Strain Evaluated for Inclusivity

|  Toxigenic C. difficile Strain | Toxinotype, Toxin  |
| --- | --- |
|  ATCC 9689 | Toxinotype 0, A+, B+  |
|  ATCC 700792  |   |
|  ATCC 17858  |   |
|  ATCC BAA-1382  |   |
|  ATCC BAA-1808  |   |
|  ATCC BAA-1871  |   |
|  ATCC BAA-1872  |   |
|  ATCC BAA-1874  |   |
|  ATCC BAA-2156  |   |
|  ATCC 51695  |   |
|  ATCC 43600  |   |
|  ATCC 43599  |   |
|  ATCC 43596  |   |
|  ATCC 43594  |   |
|  ATCC 17857  |   |
|  ATCC BAA-1804  |   |
|  ATCC 43598 | Toxinotype VIII, A-, B+  |
|  ATCC BAA-1812 | Toxinotype XII, A+, B+  |
|  ATCC BAA-1803 | Toxinotype IIIc, NAP1, A+, B+  |
|  ATCC BAA-1814 | Toxinotype XXII, A+, B+  |
|  ATCC BAA-1875 | Toxinotype V, A+, B+  |
|  ATCC BAA-2155 | Toxinotype XXII, A+, B+  |

b. Analytical specificity (exclusivity):

The potential cross-reactivity of the assay was assessed by testing concentrations of 82 non-targeted organisms (69 bacteria, one yeast, three parasites, nine viruses) and human DNA preparations that may be found in unformed stool specimens (see Table 3, Table 4, and Table 5). The test panel included commensal and pathogenic microorganisms from the intestinal tract, species phylogenetically related to $C$.

{6}

difficile, non-pathogenic strains of C. difficile, and human DNA. Bacteria and yeast were tested at a load of $\geq 10^{6}$ CFU/mL, except for a single strain of Clostridium butyricum ATCC 19398 which was tested at $1.0 \times 10^{5}$ CFU/mL. Nucleic acids from viruses and human DNA were tested at a load of $\geq 10^{6}$ copies/mL or $\geq 10^{6}$ TCID$_{50}$/mL. The test panel was contrived by spiking quantitated cell cultures or nucleic acid solutions into C. difficile-negative clinical stool matrix. Each organism was tested in triplicate.

Under the conditions of the study, Salmonella enterica subsp. Enterica serovar Newport gave a false positive result by the assay for one out of three replicates. However, no false positives were observed when the culture stock was regrown, and an additional three replicates were tested. All results were acceptable.

Table 3. Cross Reactivity Panel - Non-toxigenic C. difficile, other Clostridium sp.

|  Name | ID  |
| --- | --- |
|  Clostridium difficile (non-toxigenic) | ATCC 43593  |
|  Clostridium difficile (non-toxigenic) | ATCC 43601  |
|  Clostridium bifermentans | ATCC 638  |
|  Clostridium butyricum | ATCC 19398  |
|  Clostridium novyi | ATCC 19402  |
|  Clostridium perfringens | ATCC 13124  |
|  Clostridium scindens | ATCC 35704  |
|  Clostridium septicum | ATCC 12464  |
|  Clostridium sporogenes | ATCC 15579  |
|  Flavonifractor plautii
(formerly Clostridium orbiscindens) | ATCC 49531  |
|  Hathewaya histolytica (formerly
Clostridium histolyticum) | ATCC 19401  |
|  Paeniclostridium (Clostridium)
sordellii | ATCC 9714  |

Table 4. Cross Reactivity Panel - Other Bacteria

|  Name | ID | Name | ID  |
| --- | --- | --- | --- |
|  Abiotrophia defective | ATCC 49176 | Peptostreptococcus anaerobius | ATCC 27337  |
|  Acinetobacter baumannii | ATCC 19606 | Plesiomonas shigelloides | ATCC 51903  |
|  Aeromonas hydrophila | ATCC 35654 | Porphyromonas asaccharolytica | ATCC 27908  |
|  Alcaligenes faecalis | ATCC 15554 | Prevotella melaninogenica | ATCC 700524  |
|  Bacillus cereus | ATCC 14579 | Proteus mirabilis | ATCC 25933  |
|  Bacteroides fragilis | ATCC 23745 | Proteus vulgaris | ATCC 6896  |
|  Campylobacter jejuni | ATCC 49943 | Providencia alcalifaciens | ATCC 9886  |
|  Campylobacter coli | ATCC 49941 | Providencia rettgeri | ATCC 9250  |
|  Campylobacter fetus | ATCC 25936 | Pseudomonas aeruginosa | ATCC 10145  |
|  Citrobacter freundii | ATCC 8090 | Pseudomonas fluorescens | ATCC 13525  |

K232092 - Page 7 of 16

{7}

Table 5. Cross Reactivity Panel - Viruses and Other Organisms

|  Name | ID | Name | ID  |
| --- | --- | --- | --- |
|  Adenovirus type 40 | ATCC VR-931D | Human genomic DNA | Promega G147A  |
|  Adenovirus type 41 | ATCC VR-930D | Enterovirus 71 | ATCC VR-1775DQ  |
|  Cryptosporidium parvum | ATCC PRA-67D | Entamoeba histolytica | ATCC 30459DO  |
|  Rotavirus A | ATCC VR-2018DQ | Echovirus 4 | ATCC VR-1734D  |

K232092 - Page 8 of 16

{8}

K232092 - Page 9 of 16

|  Norovirus GI | ATCC VR-3234SD | Human Herpesvirus 5 (Cytomegalovirus) (DNA) | ATCC VR-538D  |
| --- | --- | --- | --- |
|  Norovirus GII | ATCC VR-3235SD | Giardia intestinalis | ATCC 50803D  |
|  Coxsackievirus B2 | ATCC VR-29 |  |   |

c. Microbial Interference

A microbial interference study was conducted to assess the potential inhibitory effect of 82 non-targeted microorganisms (See Table 6) that may be found in stool specimens. Pools of three to five organisms were prepared in C. difficile negative clinical stool matrix and tested in three replicates each in the presence of toxigenic C. difficile strain ATCC 43255. The bacteria/fungi, DNA/RNA, and viruses were spiked at $\geq 1 \times 10^{6}$ CFU/mL, $\geq 1 \times 10^{6}$ copies/mL, and $1 \times 10^{6}$ TCID50/mL, respectively. The toxigenic C. difficile was spiked at 2X LoD of this strain ($5.2 \times 10^{4}$ CFU/mL).

No interference was observed for the detection of C. difficile strain ATCC 43255.

Table 6. Microbial Interference Panel

|  Pool | Potentially Interfering Microorganisms  |
| --- | --- |
|  1 | Aeromonas hydrophila  |
|   |  Bacteroides fragilis  |
|   |  Campylobacter fetus  |
|   |  Campylobacter jejuni  |
|   |  Campylobacter coli  |
|  2 | Citrobacter freundii  |
|   |  Clostridium difficile (non-toxigenic)  |
|   |  Clostridium difficile (non-toxigenic)  |
|   |  Clostridium perfringens  |
|   |  Paeniclostridium (Clostridium) sordellii  |
|  3 | Enterobacter cloacae  |
|   |  Enterococcus faecium  |
|   |  Enterococcus faecalis  |
|   |  Escherichia coli O157:H7  |
|   |  Escherichia coli  |
|  4 | Escherichia fergusonii  |
|   |  Escherichia coli hermanii  |
|   |  Helicobacter pylori  |
|   |  Klebsiella pneumoniae  |
|   |  Peptostreptococcus anerobius  |
|  5 | Salmonella enterica serovar Choleraesuis  |
|   |  Salmonella enterica serovar Newington  |
|   |  Salmonella enterica serovar Newport  |

{9}

K232092 - Page 10 of 16
|  Pool | Potentially Interfering Microorganisms  |
| --- | --- |
|   | Salmonella enterica serovar Typhi  |
|   |  Salmonella enterica serovar Typhimurium  |
|  6 | Serratia liquefaciens  |
|   |  Serratia marcescens  |
|   |  Shigella boydii  |
|   |  Shigella flexneri  |
|   |  Shigella sonnei  |
|  7 | Candida albicans  |
|   |  Staphylococcus aureus  |
|   |  Staphylococcus aureus  |
|   |  Staphylococcus epidermidis  |
|   |  Yersinia enterocolitica  |
|  8 | Klebsiella oxytoca  |
|   |  Proteus mirabilis  |
|   |  Salmonella enterica subsp. Arizonae  |
|   |  Shigella dysenteriae  |
|   |  Streptococcus agalactiae  |
|  9 | Acinetobacter baumannii  |
|   |  Bacillus cereus  |
|   |  Edwardsiella tarda  |
|   |  Enterobacter aerogenes  |
|   |  Pseudomonas fluorescens  |
|  10 | Abiotrophia defective  |
|   |  Alcaligenes faecalis  |
|   |  Clostridium bifermentans  |
|   |  Clostridium butyricum  |
|   |  Clostridium novyi  |
|  11 | Hathewaya histolytica (Clostridium histolyticum)  |
|   |  Lactobacillus lactis  |
|   |  Listeria monocytogenes  |
|   |  Pseudomonas aeruginosa  |
|   |  Vibrio paraharmolyticus  |
|  12 | Clostridium sporogenes  |
|   |  Lactobacillus acidophilus  |
|   |  Plesiomonas shigelloides  |
|   |  Proteus vulgaris  |
|   |  Providencia rettgeri  |
|  13 | Flavonifractor plautii (formerly Clostridium orbiscindens)  |
|   |  Providencia alcalifaciens  |
|   |  Salmonella enterica serovar Newport  |
|   |  Escherichia coli  |
|   |  Escherichia coli  |
|  14 | Adenovirus type 40  |

{10}

K232092 - Page 11 of 16

|  Pool | Potentially Interfering Microorganisms  |
| --- | --- |
|   | Enterovirus 71  |
|   |  Norovirus GI  |
|   |  Norovirus GII  |
|   |  Rotavirus A  |
|  15 | Clostridium scindens  |
|   |  Clostridium septicum  |
|   |  Enterococcus faecalis (van B)  |
|   |  Porphyromonas asaccharolytica  |
|   |  Prevotella melaninogenica  |
|  16 | Adenovirus type 41  |
|   |  Coxsackievirus B2  |
|   |  Herpesvirus 5 (Cytomegalovirus)  |
|   |  Echovirus 4  |
|   |  Giardia intestinalis  |
|  17 | Cryptosporidium parvum  |
|   |  Entamoeba histolytica  |
|   |  Human gDNA  |

d. Interfering Substances

An interfering substances study was conducted to assess the potentially inhibitory effects of 30 substances (Table 7) that may be present in stool specimens. The substances were tested in triplicate in the presence of two strains of toxigenic *C. difficile* (ATCC 43255 and ATCC BAA-1805) in *C. difficile* negative clinical stool matrix. Substances were spiked at higher concentrations than expected in stool specimens, and *C. difficile* strains were spiked at 2X LoD (ATCC 43255 at 5.2 x 10⁴ CFU/mL, and ATCC BAA-1805 at 1.4 x 10⁵ CFU/mL).

No interference was observed for the 30 substances at the tested concentrations.

Table 7. Substance Interference Panel

|  Substance (commercial name) | Active Ingredient(s) | Concentration  |
| --- | --- | --- |
|  Vaginal antifungal / anti-itch (Nystatin) | Nystatin | 10% w/v  |
|  Creams / ointments (Cortizone-10) | Hydrocortisone | 30% w/v  |
|  Anti-hemorrhoidal creams / ointments (Preparation H) | Phenylephrine HCl | 20% w/v  |
|  Antacids (Tums) | Calcium Carbonate | 38% w/v  |
|  Mineral Oil (Fisher Bioreagents Cat# B2629-1) | Mineral Oil | 50% v/v  |
|  Enemas (Mesalazine Sigma Aldrich Cat# Y0000297) | Mesalazine | 2.5% w/v  |
|  Condom with spermicidal lubricant (Trojan) | Nonoxynol-9 | 40% w/v  |

{11}

K232092 - Page 12 of 16
|  Substance (commercial name) | Active Ingredient(s) | Concentration  |
| --- | --- | --- |
|  Anti-diarrheal medication
(Pepto Bismol) | Bismuth Subsalicylate | 50% v/v  |
|  Anti-diarrheal medication
(Imodium) | Loperamide Hydrochloride | 50% v/v  |
|  Laxatives
(Senna Glycoside Tablet) | Sennosides | 17% w/v  |
|  Antibiotics
(Vancomycin 100mg/mL in DMSO – Sigma Cat# SBR00001) | Vancomycin | 5% w/v  |
|  Antibiotics
(Metronidazole – MP Biomedicals Cat# 155710) | Metronidazole | 0.5% w/v  |
|  Non-steroidal anti-inflammatory
(Aleve) | Naproxen Sodium | 10% w/v  |
|  Moist towelettes (Wet Ones) | Benzalkonium Chloride | 50% v/v  |
|  Suppository
(Fleet) | Liquid Glycerin | 50% v/v  |
|  Personal lubricant
(KY) | Glycerin | 50% v/v  |
|  Personal lubricant
(Astroglide) | Silicone | 50% v/v  |
|  Aloe leaf juice flushable wipes
(Generic) | Butoxy PEG-4 PG-Amodimethicone | 50% v/v  |
|  Anti-fungal cream
(Miconazole-3) | Miconazole Nitrate | 30% w/v  |
|  Petroleum Jelly
(Vaseline) | Petroleum | 30% w/v  |
|  Biotin Supplement
(Nature Made, 2500 mcg/capsule) | Biotin | 51% w/v^{a}  |
|  Contrast agent | Barium Sulfate | 25% w/v  |
|  Stearic Acid
(ScienceLab Cat# SLS3742 | Stearic Acid | 1.1% w/v  |
|  Palmitic Acid
(Sigma Cat# P0500) | Palmitic acid | 1.5% w/v  |
|  Whole Blood
(Fisher Scientific Cat# NC0088224) | Blood components | 40% v/v^{b}  |
|  Biotin
(Sigma Cat# B4501) | Biotin | 0.5% w/v  |
|  Nonoxynol-9
(Sigma Aldrich Cat# PHR2731) | Nonoxynol-9 | 48% v/v  |
|  Mucin from Porcine Stomach
(Sigma Cat# M1778) | Mucin | 2.5% w/v  |
|  Topical antibiotic
(Neosporin) | Bacitracin, Neomycin, Polymyxin B | 50% w/v  |
|  Nystatin
(Sigma Cat# N6261) | Nystatin | 10% w/v  |

{12}

|  Substance (commercial name) | Active Ingredient(s) | Concentration  |
| --- | --- | --- |
|  Erythromycin
(Alfa Aesar Cat# J62279) | Erythromycin | 10% w/v  |

a Input listed is the total amount of the gel-capsule biotin supplement tested and equates to approximately 1.3 mg of biotin per swab.
b Raw stool was formulated to 40% whole blood by volume (v/v). One (1) swab volume of the mixture was evaluated in the CDF2 test.

## 4. Assay Reportable Range:

Not applicable

## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

### e. Quality Control:

The Great Basin Toxigenic *C. difficile* Direct Test (CDF2) assay uses three internal controls: a sample processing control (SPC), a hybridization control (HC), and a detection control (DC). The SPC is included in the Sample Preparation Device supplied with each kit. The SPC is extracted, amplified, and detected along with each specimen tested, and it verifies the efficacy of the sample preparation, PCR amplification, and detection processes. The HC is a control oligonucleotide attached to the silicon chip surface. A biotinylated oligonucleotide with complementary sequence to the HC is added to the hybridization buffer and the oligonucleotide pair provides verification that probe hybridization is functioning properly. The DC is a biotinylated oligonucleotide probe attached to the silicon chip surface that provides a control for the performance of the conjugate and detection reagents for proper signal development.

The use of two external controls is recommended: a positive external control (PEC), and negative external control (NEC). The PEC monitors for substantial reagent failure, and the NEC monitors for environmental contamination or carryover. Each laboratory must establish the number, type and frequency of testing external control materials per applicable regulations or accrediting agencies. External Control materials are recommended but not provided. External Controls are treated as if they are specimens.

### Optional PEC materials include:

- ZeptoMetrix *Clostridium difficile* Positive Control IVD CE (Catalog No. NATCDI-6MC-IVD).
- Remnants of a known *C. difficile* positive clinical stool sample or
- Culture of a known C. difficile strain (e.g., Microbiologics KWIK-STIK *C. difficile* ATCC 9689, catalog no. 0329)

### Optional NEC materials include:

- ZeptoMetrix *Clostridium difficile* Negative Control IVD CE (Catalog No. NATCSO-6MC-IVD) or
- Remnants of a known *C. difficile* negative clinical stool sample.

One PEC and one NEC were processed each day of analytical and clinical testing. Of the total 361 external controls that were run across all sites in the clinical study, 350 (97.0%)

K232092 - Page 13 of 16

{13}

were valid, eight (2.2%) were canceled by the end-user (test incomplete), and three (0.8%) were invalid due to a failed internal control. All failed results repeated successfully.

f. Sample Stability:

Specimen stability was evaluated analytically by testing pre-characterized positive and negative clinical samples under different storage time and temperature conditions. The results supported the stool specimen storage instructions described in the package insert.

Stool specimens can be stored at refrigerated (2°-8°C) for up to 72 hours (three days), or at room temperature (18°-22°C) for up to 24 hours.

g. Freeze/thaw Stability:

Studies were conducted to examine the stability of stool specimens when subjected to freeze/thaw cycles. Contrived positive stool samples contained 5X LoD and 2X LoD of C. difficile (ATCC 43255). The study results supported a recommendation that stool samples stored at -70°C can undergo no more than one freeze/thaw cycles.

6. Detection Limit:

The analytical sensitivity (Limit of Detection or LoD) of the Toxigenic C. difficile Direct Test (CDF2) assay was determined using negative clinical stool specimens spiked with various concentrations of toxigenic C. difficile bacterial suspensions. The testing to establish and confirm the LoD included four lots, multiple instruments, and operators. Two strains of toxigenic C. difficile (ATCC 43255, Toxinotype 0, tcdA+, tcdB+, and ATCC BAA-1805, NAP 1/027, Toxinotype IIIb, tcdA+, tcdB+) were tested in a minimum of 20 replicates per concentration in the confirmation study. The LoD was defined as the lowest concentration at which 95% or more of all replicates tested positive. The LoD was determined at 2.6 X 10⁴ CFU/mL for ATCC 43255 and 7.1 X 10⁴ CFU/ml for ATCC BAA-1805.

7. Assay Cut-Off:

Not applicable

B Comparison Studies:

1. Method Comparison with Predicate Device:

The performance of the Toxigenic C. difficile Direct Test (CDF2) was assessed with remnants of unpreserved stool specimens prospectively collected from symptomatic individuals suspected of having an C. difficile infection. Specimens were collected from three independent clinical sites. Stool specimens were tested with a high-performing FDA-cleared molecular C. difficile assay, which served as the basis for performance evaluation of this assay. Specimens with discrepant results were tested with an alternative FDA-cleared molecular C. difficile assay, performance of which is footnoted accordingly, but the original

K232092 - Page 14 of 16

{14}

performance was not changed. Prospective testing consisted of 829 stool specimens that met the inclusion criteria.

The ages of patients ranged from less than 2 years to over 89 years with 53.3% of the specimens coming from patients that were 19-64 years of age. Of the 829 patients tested, 52% were female, 47.4% were male, and 0.6% were undeclared gender.

Of the 829 eligible specimens, ten specimens generated false positive with the Toxigenic C. difficile Direct Test (CDF2) and six specimens generated false negative results.

Table 8 and 9 show a summary of the clinical performance of the Toxigenic C. difficile Direct Test (CDF2) for all sites individually and combined. The PPA and NPA values for the Toxigenic C. difficile Direct Test (CDF2) were 95.8% and 98.5%, respectively. There was no observable difference in performance of the Toxigenic C. difficile Direct Test (CDF2) with respect to study site, kit lot, or patient gender or age. These results are acceptable.

Table 8. Clinical Study Results
|  All Sites | Molecular Comparator  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  Toxigenic | Positive | 138 | 10* | 148  |
|  C. difficile | Negative | 6* | 675 | 681  |
|  Direct Test | Total | 144 | 685 | 829  |

* All 16 discrepant results were tested with the alternative FDA-cleared molecular C. difficile assay. None of the 10 false positive samples were positive and five of the six false negatives were negative with the alternative assay.

Table 9. Summary of Clinical Performance
|   | n | PPA
(95% CI) | NPA
(95% CI) | PPV
(95% CI) | NPV
(95% CI)  |
| --- | --- | --- | --- | --- | --- |
|  Toxigenic
C. difficile | 829 | 95.8%
(91.2 – 98.5) | 98.5%
(97.3 – 99.3) | 93.2%
(87.9 - 96.7) | 99.1%
(98.1 – 99.7)  |
|  Direct Test |  | 138/144 | 675/685 |  |   |

2. Matrix Comparison:
Not applicable

C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable

K232092 - Page 15 of 16

{15}

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not applicable

D Clinical Cut-Off:

Not applicable

E Expected Values/Reference Range:

In the clinical studies, a total of 829 specimens were enrolled at three geographically diverse clinical sites, from subjects ranging in age from less than two years old to more than 89 years old. Of these specimens, the overall percentage of positive results observed with the Toxigenic C. difficile Direct Test (CDF2) with these specimens was 17.4% [144/829; 95%CI: 14.9 –20.1%].

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K232092 - Page 16 of 16

---

**Source:** [https://fda-staging.innolitics.com/device/K232092](https://fda-staging.innolitics.com/device/K232092)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
