K232086 · Eos Imaging · LLZ · Oct 24, 2023 · Radiology
Device Facts
Record ID
K232086
Device Name
spineEOS
Applicant
Eos Imaging
Product Code
LLZ · Radiology
Decision Date
Oct 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
spineEOS software is indicated for assisting healthcare professionals with preoperative planning of spine surgeries. spineEOS provides access to EOS images with associated 3D datasets and measurements. spineEOS includes surgical planning tools that enable users to define a patient specific surgical strategy.
Device Story
Cloud-based software for preoperative spine surgery planning. Inputs: EOS imaging system X-rays and associated 3D datasets. Operation: User validates 3D models; utilizes planning tools for segmental alignment, interbody implants, osteotomy, spondylolisthesis, and rod curvature management. Output: Patient-specific surgical strategy and clinical measurements. Used by spine surgeons, EOS staff, and implant distributors in clinical settings. Healthcare providers use outputs to define and validate surgical plans; color-coded parameters display variance from normative values to assist decision-making. Benefits: Enables precise, patient-specific surgical strategy definition.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including design input review, unit testing, software/system integration, usability testing, and validation of clinical parameters.
Technological Characteristics
Cloud-based software; requires stable internet, 2.4 GHz processor, 4 GB RAM. Compatible with modern web browsers (Chrome, Firefox, Edge, Safari). Provides 2D/3D image display, distance/angle measurements, and surgical planning tools. Requires human intervention for interpretation. No specific hardware materials or sterilization applicable as it is software.
Indications for Use
Indicated for preoperative planning of spine surgeries in patients 7 years or older, including degenerative, deformative, and adolescent idiopathic scoliosis (AIS) cases. Requires trained healthcare professionals (spine surgeons, EOS staff, implant distributors).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
EOS imaging % Mathilde Masurel Design Quality and Regulatory Affairs Specialist 10 rue Mercoeur Paris, 75011 France
October 24, 2023
Re: K232086
Trade/Device Name: spineEOS Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: September 27, 2023 Received: September 27, 2023
Dear Mathilde Masurel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232086
Device Name spineEOS
Indications for Use (Describe)
spineEOS software is indicated for assisting healthcare professionals with preoperative planning of spine surgeries. spineEOS provides access to EOS images with associated 3D datasets and measurements. spineEOS includes surgical planning tools that enable users to define a patient specific surgical strategy.
Type of Use (Select one or both, as applicable)
| | <span style="padding-right: 20px;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------|
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for EOS imaging, an "AN cotec COMPANY". The logo is in orange and features the letters "EOS" stacked on top of the word "imaging". The "O" in "EOS" is shaped like a diamond. The words "AN cotec COMPANY" are in a smaller font size and are located below the word "imaging".
### 1 SUBMITTER
EOS imaging 10 rue Mercoeur Paris, France 75011 Phone: +33 1 55 25 60 60 Fax: +33 1 55 25 60 61
| Contact Person: | Mathilde MASUREL |
|-----------------|--------------------------------------------------|
| | Design Quality and Regulatory Affairs Specialist |
| | EOS imaging |
| Contact Phone: | +33 1 55 25 60 60 |
Date Summary Prepared: October 23, 2023
### 2 DEVICE
| Trade Name: | spineEOS |
|------------------------------|------------------------------------------------------------------|
| Common or Usual Name: | Cloud-based software |
| Classification Name: | System, image processing, radiological<br>(21 C.F.R. § 892.2050) |
| Regulatory Class: | Class II |
| Product Code: | LLZ |
| Premarket Submission Number: | K232086 |
### LEGALLY MARKETED PREDICATE DEVICES 3
| 510(k) | Product Name | Clearance Date |
|---------|--------------|----------------|
| K160407 | spineEOS | August 2016 |
### DEVICE DESCRIPTION 4
spineEOS is a software indicated for assisting healthcare professionals with preoperative planning of spine surgeries. EOS images (generated from EOS imaging's acquisition system) and associated 3D datasets are used as inputs of the software. The product manages clinical measurements and allows user to access surgical planning tools to define a patient specific surgical strategy. The product is indicated for adolescent and adult patients.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for EOS imaging, an orthopedic imaging company. The logo features the letters "EOS" in a stylized, sans-serif font, with the "O" shaped like a diamond. Below the letters "EOS" is the word "imaging" in a smaller, sans-serif font. Underneath "imaging" are the words "AN cctec COMPANY" in an even smaller font, indicating that EOS imaging is a subsidiary of cctec.
### INDICATIONS FOR USE 5
spineEOS software is indicated for assisting healthcare professionals with preoperative planning of spine surgeries. spineEOS provides access to EOS images with associated 3D datasets and measurements. spineEOS includes surgical planning tools that enable users to define a patient specific surgical strategy.
### TECHNOLOGICAL COMPARISON TO PREDICATES 6
The subject device was compared to the predicate device in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate device, does not impact substantial equivalence, or safety and effectiveness.
{5}------------------------------------------------
| AN alec company |
|-----------------|
| Characteristic | Cleared spineEOS (K160407) | Modified spineEOS | Substantially Equivalent? |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Using 3D data and models obtained from sterEOS workstation, spineEOS software is indicated for assisting healthcare professionals in viewing, measuring images as well as in preoperative planning of spine surgeries. The device includes tools for measuring spine anatomical components for placement of surgical implants. Clinical judgment and experience are required to properly use the software online. | spineEOS software is indicated for assisting healthcare professionals with preoperative planning of spine surgeries. spineEOS provides access to EOS images with associated 3D datasets and measurements. spineEOS includes surgical planning tools that enable users to define a patient specific surgical strategy. | Yes, the modified spineEOS has the same intended use as the cleared spineEOS. Specifically, both devices are indicated for assisting healthcare professionals with preoperative planning of spine surgeries.<br>The modified spineEOS has slightly modified indications for use. Specifically, the indications for use statement maintains the basic explanation of the device inputs and available tools.<br>In sum, the same information is provided to the end user; it is simply not organized in the same way. Thus, this change does not raise different questions of safety or effectiveness. |
| Regulatory Class/Code | Class II<br>LLZ<br>(21 CFR 892.2050) | Class II<br>LLZ<br>(21 CFR 892.2050) | Yes, same. |
| Device Classification Name | System, image processing, radiological | System, image processing, radiological | Yes, same. |
| User Population | spineEOS can only be used by a trained user including:<br>Spine surgeons to define and validate the surgical plan. | spineEOS can only be used by a trained user including:<br>Spine surgeons to define and validate the surgical plan.EOS staff and implant distributors to define and save the optional pre-planning. | Yes, similar. The modified spineEOS has EOS staff in addition to the users of cleared spineEOS. This minor change does not affect the safety and effectiveness of the product because EOS staff has the same account right as the implant distributors already considered in the predicate and needs to be properly trained in order to use the product. |
| Target Population | Patients 7 years or older who need spine surgery. spineEOS is | Patients 7 years or older who need spine surgery. spineEOS is | Yes, same. The modified spineEOS has the same target population as the cleared spineEOS. |
| Characteristic | Cleared spineEOS (K160407) | Modified spineEOS | Substantially Equivalent? |
| Hardware and<br>Software<br>Requirement | recommended in the preoperative<br>planning of spine surgeries for:<br>• Degenerative surgery<br>• Deformative adult spine<br>surgery<br>• Adolescent Idiopathic<br>Scoliosis (AIS) surgery. | recommended in the preoperative<br>planning for spine surgeries for:<br>• Degenerative Spine Surgery<br>• Deformative Spine Surgery<br>• Adolescent Idiopathic Scoliosis<br>(AIS) surgery. | Specifically, both devices are intended for patients 7<br>years or older who need spine surgery.<br>Modified spineEOS has slightly modified target<br>population statement as compared to the predicate.<br>Specifically, the target population statement<br>maintains the description of degenerative and<br>deformative spine surgery for patients 7 years or<br>older as well as the description of AIS surgery but in<br>a more consistent way. |
| | Web Browsers:<br>• Windows 7/8.1/10: Google<br>Chrome 47, Mozilla Firefox<br>43, Opera 34;<br>• For Macintosh OS X<br>Yosemite 10.10 et El Capitan<br>10.11: Google Chrome 47,<br>Mozilla Firefox 43, Opera 34<br>et Safari 9.<br><br>PC Configuration:<br>Stable internet connection - Dual<br>Core 2.4 GHz processor, 4 GB of<br>RAM and integrated graphics<br>board.<br><br>Screen Resolution:<br>The minimum screen resolution<br>ensuring full display of the | Web Browsers:<br>• Windows 10/11: Google<br>Chrome, Mozilla Firefox, Edge<br>• For macOS 12/13: Google<br>Chrome, Mozilla Firefox, Safari<br><br>PC Configuration:<br>Stable internet connection - Dual Core<br>2.4 GHz processor, 4 GB of RAM and<br>integrated graphics board.<br><br>Screen Resolution:<br>The minimum screen resolution<br>ensuring full display of the planning<br>page is 1920 x 1080 pixels.<br>At resolutions below 1600 x 900, the<br>application prompts the user to toggle | In sum, the same information is provided to the end<br>user; it is simply stated differently. Thus, this minor<br>change does not raise different questions of safety<br>or effectiveness.<br>Yes, similar. The differences between modified<br>spineEOS and the predicate take into account the<br>evolution of the web browsers. These minor<br>changes do not affect the safety and effectiveness<br>of the product. |
| Characteristic | Cleared spineEOS (K160407) | Modified spineEOS | Substantially Equivalent? |
| | planning page is 1920 x 1080 pixels.<br>At resolutions below 1600 x 900, the application prompts the user to toggle to full screen display to ensure correct visualization. | to full screen display to ensure correct visualization. | |
| Input Data | Patient's information X-rays images 3D model of the spine | Patient's information X-rays images 3D model of the spine | Yes, same. |
| Product<br>Workflow | Model Validation: allow users to validate or reject the 3D modeling.<br>Preoperative: allow user to consult the preoperative state of the patient<br>Planning: allow user to define patient specific surgical strategy<br>Rod: allow user to define the design of rods | Model Validation: allow users to validate or reject the 3D modeling.<br>Preoperative: allow user to consult the preoperative state of the patient<br>Planning: allow user to define patient specific surgical strategy<br>Rod: allow user to define the design of rods | Yes, same. |
| Tools Available<br>for Planning | Segmental Alignment Interbody Implant Osteotomy | Segmental Alignment Interbody Implant Osteotomy Spondylolisthesis Rod Curvature Management | Yes, similar. Modified spineEOS is an updated version of the cleared spineEOS. The fundamental tools of modified spineEOS are unchanged from the cleared spineEOS. Both devices have tools for segmental alignment, interbody implant, and osteotomy.<br>Modified spineEOS main difference from the cleared spineEOS is the introduction of the spondylolisthesis tool. This tool operates on the same principle than the existing segmental alignment tool, except that it allows movement of the vertebra on a different axis (translation for the spondylolisthesis tool vs. rotation for the segmental |
| Characteristic | Cleared spineEOS (K160407) | Modified spineEOS | Substantially Equivalent? |
| Software<br>Functionalities /<br>Modalities | Obtains an image and 3D model<br>transferred from other devices<br><br>Provides normative values used<br>to follow the impact of the<br>planning on the patient alignment<br><br>Provides color-coded clinical<br>parameters to display variance<br>from the defined normative values<br><br>Requires human intervention for<br>interpretation and manipulation of<br>images | Obtains an image and 3D model<br>transferred from other devices<br><br>Provides normative values used to<br>follow the impact of the planning on<br>the patient alignment<br><br>Provides color-coded clinical<br>parameters to display variance from<br>the defined normative values<br><br>Requires human intervention for<br>interpretation and manipulation of<br>images | alignment tool) and the possibility to manage the rod<br>curvature.<br>The introduction of this tool does not affect the<br>device safety and effectiveness.<br><br>Yes, same.<br><br>Yes, same.<br><br>Yes, same.<br><br>Yes, same. |
| Image<br>Manipulation<br>Functions | 2D images and 3D model display<br>and basic manipulation (zoom,<br>panning and angles<br>measurements) | 2D images and 3D model display and<br>basic manipulation (zoom, panning,<br>distance, and angles measurements) | Yes, same. In modified spineEOS, the possibility of<br>measuring distance directly in the image is based on<br>the same principle as the measurement of angles,<br>already available in the cleared spineEOS. It does<br>not affect the device safety and effectiveness.<br><br>Yes, same. |
| Measurement<br>Functions | Distances and Angles | Distances and Angles | Yes, same. |
| User Interface | Computer | Computer | Yes, same. |
| Software<br>Environment | Cloud-based software | Cloud-based software | Yes, same. |
# Table 5-1: Comparison for Substantial Equivalence
{6}------------------------------------------------
Traditional 510(k) Premarket Notification
ಲ್ಲಿ ಪ್ರದ
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for EOS imaging, an aztec company. The logo is vertically oriented, with the letters "EOS" stacked on top of each other in a sans-serif font. The word "imaging" is written to the right of the letters "EOS", and the phrase "an aztec company" is written below the word "imaging". The logo is orange in color.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for EOS imaging, an aztec company. The logo is vertically oriented, with the letters "EOS" stacked on top of each other in an orange, outlined font. To the right of the letters, the word "imaging" is written vertically in a smaller font. Below "imaging" is the phrase "AN aztec COMPANY" in an even smaller font.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for EOS imaging, an orthopedic imaging company. The logo is in orange and features the company name in a stylized font. Below the name, it says "AN cdetec COMPANY" in a smaller font. The logo is simple and modern, and the orange color is eye-catching.
### PERFORMANCE DATA 7
Nonclinical performance testing performed on the subject device, spineEOS, supports substantial equivalence to the predicate device. The following V&V testing was performed:
A. Verifications activities cover the following:
- o Design input review
- Unit testing ●
- Software integration 0
- o System integration
- B. Validation activities cover the following:
- Validation of the multifunctional requirements in terms of design. ●
- Usability testing was performed to demonstrate that spineEOS can be used safely . by assessing and mitigating usability risks.
- Validation of the clinical parameters o
Determination of substantial equivalence is not based on an assessment of clinical performance data.
### CONCLUSION 00
Based upon the information provided in this 510(k) submission, it has been determined that the subject device, spineEOS, is substantially equivalent to the legally marketed predicate device in regard to indications for use, intended use, design, technology, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.