Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green); Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl
K232079 · Wrp Asia Pacific Sdn. Bhd. · KGO · Mar 28, 2024 · General, Plastic Surgery
Device Facts
Record ID
K232079
Device Name
Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green); Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl
Applicant
Wrp Asia Pacific Sdn. Bhd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Mar 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn on the hands of healthcare personnel as a barrier for protection against cross-contamination between the healthcare personnel and patient. These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable natural rubber latex surgical gloves; sterile; powder-free. Used by healthcare personnel in clinical settings as a barrier against cross-contamination. Tested for permeation resistance against specific chemotherapy drugs and fentanyl citrate per ASTM D6978-05. Output is a physical barrier worn on hands. Provides protection during handling of hazardous drugs; however, specific drugs (Carmustine and Thiotepa) show low permeation times and are contraindicated for use with these gloves. Device performance verified through bench testing for physical properties, biocompatibility, and chemical permeation.
Clinical Evidence
No clinical data. Bench testing only. Testing included physical properties (ASTM D412), dimensions (ASTM D3767), watertight integrity (ASTM D5151), residual powder (ASTM D6124), protein content (ASTM D5712), biocompatibility (ISO 10993-5, -10, -11), and chemotherapy/fentanyl permeation (ASTM D6978-05).
Technological Characteristics
Natural rubber latex; microroughened surface; sterile (gamma radiation). Complies with ASTM D3577 (physical properties), ASTM D6124 (powder-free), and ASTM D6978-05 (chemotherapy permeation). Biocompatibility per ISO 10993. Sizes 5.5-9.0.
Indications for Use
Indicated for use by healthcare personnel as a protective barrier against cross-contamination between personnel and patients. Tested for use with specific chemotherapy drugs and fentanyl citrate.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder-Free Polymer Coated Latex Surgical Gloves, Sterile with Protein Labelling Claim (50 Micrograms or Less) (K020019)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2024
WRP Asia Pacific Sdn. Bhd. % Michael Scaglione U.S. Agent WRP USA Inc 3700 Massillon Road, Suite 340 Uniontown, Ohio 44685
## Re: K232079
Trade/Device Name: Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green); Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Regulation Number: 21 CFR 878.4460 Regulation Name: Non-powdered surgeon's glove Regulatory Class: Class I. reserved Product Code: KGO, LZC, OPJ, QDO Dated: March 5, 2024 Received: March 26, 2024
Dear Michael Scaglione:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Allan Guan -S
For Bifeng Qian, M.D., Ph.D. Assistant Director
{2}------------------------------------------------
DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232079
Device Name
Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy (Green); Natural Rubber Latex (Strip&Coat)/(Strip&Pack) Slove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl
### Indications for Use (Describe)
A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn on the hands of healthcare personnel as a barrier for protection against cross-contamination between the healthcare personnel and patient.
These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
NATURAL RUBBER LATEX SURGICAL GLOVE, POWDER FREE, STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS AND FENTANYL (GREEN):
| Chemotherapy Drug | Concentration Average Breakthrough Detection Time (Minutes) |
|-----------------------------------------|-------------------------------------------------------------|
| Cisplatin (1.0 mg/ml) | > 240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240 |
| Dacarbazine (10.0 mg/ml) | > 240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | > 240 |
| Etoposide (20.0 mg/ml) | > 240 |
| Fluorouracil (50.0 mg/ml) | > 240 |
| Ifosfamide (50.0 mg/ml) | > 240 |
| Methotrexate (25.0 mg/ml) | > 240 |
| Mitomycin C (0.5 mg/ml) | > 240 |
| Mitoxantrone (2.0 mg/ml) | > 240 |
| Paclitaxel (6.0 mg/ml) | > 240 |
| Vincristine Sulfate (1.0 mg/ml) | > 240 |
| *Thiotepa (10.0 mg/ml) | 14.7 |
| *Carmustine (BCNU) (3.3 mg/ml) | 11.9 |
| Opioid Drug | Fentanyl Citrate 100 mcg/2mL |
|----------------------------------------|-----------------------------------|
| Breakthrough Detection Time in Minutes | No breakthrough up to 240 minutes |
*WARNING: Testing showed an average breakthrough time of 11.9 minutes for Carmusine and an average breakthrough time of 14.7 minutes for Thiotepa. Do not use with Carmustine and Thiotepa.
### NATURAL RUBBER LATEX (STRIP&COAT)/(STRIP&PACK) SURGICAL GLOVE, POWDER FREE, STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS AND FENTANYL:
| Chemotherapy Drug | Average Breakthrough Detection Time (Minutes) |
|-----------------------------------------|-----------------------------------------------|
| Cisplatin (1.0 mg/ml) | > 240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240 |
| Dacarbazine (10.0 mg/ml) | > 240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | > 240 |
| Etoposide (20.0 mg/ml) | > 240 |
{4}------------------------------------------------
| Fluorouracil (50.0 mg/ml) | > 240 |
|---------------------------------|-------|
| Ifosfamide (50.0 mg/ml) | > 240 |
| Methotrexate (25.0 mg/ml) | > 240 |
| Mitomycin C (0.5 mg/ml) | > 240 |
| Mitoxantrone (2.0 mg/ml) | > 240 |
| Paclitaxel (6.0 mg/ml) | > 240 |
| Vincristine Sulfate (1.0 mg/ml) | > 240 |
| *Thiotepa (10.0 mg/ml) | 13.8 |
| *Carmustine (BCNU) (3.3 mg/ml) | 5.7 |
Opioid Drug Fentanyl Citrate 100 mcg/2mL Breakthrough Detection Time in Minutes No breakthrough up to 240 minutes
* WARNING: Testing showed an average breakthrough time of 5.7 minutes for Carmustine and an average breakthrough time of 13.8 minutes for Thiotepa. Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 1.0 Submitter:
| Name | : Muhammad Ameer Arief bin Mohd Mujab |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Address | WRP Asia Pacific Sdn. Bhd.<br>Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak<br>Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No. | : +60 3 8706 1486 |
| Fax No. | : +60 3 8706 1557 |
| Date of Summary<br>Prepared | : 28th March 2024 |
#### 2.0 Identification of the Subject Device:
| Trade Name | : | Natural Rubber Latex Surgical Glove, Powder Free,<br>Sterile, Tested for Use with Chemotherapy Drugs and<br>Fentanyl (Green); Natural Rubber Latex<br>(Strip&Coat)/(Strip&Pack) Surgical Glove, Powder<br>Free, Sterile, Tested for Use with Chemotherapy<br>Drugs and Fentanyl |
|-----------------------|---|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | : | Surgical Gloves |
| Classification Name | : | Non-powdered surgeon's glove |
| Device Classification | : | I |
| Regulation Number | : | 21 CFR 878.4460 |
| Product Code | : | KGO, LZC, OPJ, QDO |
#### 3.0 Predicate Device:
| | Predicate |
|------------------|----------------------------------------------------------------------------------------------------------------------|
| Manufacturer | WRP Asia Pacific Sdn. Bhd. |
| Device name | Powder-Free Polymer Coated Latex Surgical<br>Gloves, Sterile with Protein Labelling Claim (50<br>Micrograms or Less) |
| 510(k) Number | K020019 |
| Regulatory Class | I |
| Product Code | KGO |
{6}------------------------------------------------
#### 4.0 Description of The Device:
This is a disposable natural rubber surgical glove that is tested for use with chemotherapy drugs and fentanyl resistance. The glove is supplied in the following sizes: 5½, 6, 6½, 7, 7½, 8, 8½ and 9.
#### 5.0 Indication for use:
A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn on the hands of healthcare personnel as a barrier for protection against cross-contamination between the healthcare personnel and patient.
These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
| Chemotherapy Drug | Concentration | Minimum<br>Breakthrough<br>Detection Time<br>(Minutes) |
|----------------------------|---------------|--------------------------------------------------------|
| *Carmustine (BCNU) | 3.3 mg/ml | 11.9 |
| Cisplatin | 1.0 mg/ml | >240 |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml | >240 |
| Dacarbazine | 10.0 mg/ml | > 240 |
| Doxorubicin Hydrochloride | 2.0 mg/ml | > 240 |
| Etoposide | 20.0 mg/ml | > 240 |
| Fluorouracil | 50.0 mg/ml | > 240 |
| Ifosfamide | 50.0 mg/ml | > 240 |
| Methotrexate | 25.0 mg/ml | > 240 |
| Mitomycin C | 0.5 mg/ml | > 240 |
| Mitoxantrone | 2.0 mg/ml | > 240 |
| Paclitaxel | 6.0 mg/ml | > 240 |
| *Thiotepa | 10.0 mg/ml | 14.7 |
| Vincristine Sulfate | 1.0 mg/ml | > 240 |
### 5.1 Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl (Green)
{7}------------------------------------------------
*WARNING: Please note the following drugs have extremely low permeation times: Carmustine (BCNU): 11.9 minutes and Thiotepa: 14.7. Do not use with Carmustine and Thiotepa.
| Opioid Drug | Concentration | Breakthrough Detection<br>Time in Minutes |
|----------------------------|---------------|-------------------------------------------|
| Fentanyl Citrate Injection | 100mcg/2mL | No breakthrough up to 240<br>minutes |
5.2 Natural Rubber Latex Surgical Glove (Strip&Coat)/(Strip&Pack), Powder Free, Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl
| Chemotherapy Drug | Concentration | Minimum<br>Breakthrough<br>Detection Time<br>(Minutes) |
|----------------------------|---------------|--------------------------------------------------------|
| *Carmustine (BCNU) | 3.3 mg/ml | 5.7 |
| Cisplatin | 1.0 mg/ml | >240 |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml | >240 |
| Dacarbazine | 10.0 mg/ml | > 240 |
| Doxorubicin Hydrochloride | 2.0 mg/ml | > 240 |
| Etoposide | 20.0 mg/ml | > 240 |
| Fluorouracil | 50.0 mg/ml | > 240 |
| Ifosfamide | 50.0 mg/ml | > 240 |
| Methotrexate | 25.0 mg/ml | > 240 |
| Mitomycin C | 0.5 mg/ml | > 240 |
| Mitoxantrone | 2.0 mg/ml | > 240 |
| Paclitaxel | 6.0 mg/ml | > 240 |
| *Thiotepa | 10.0 mg/ml | 13.8 |
| Vincristine Sulfate | 1.0 mg/ml | > 240 |
*WARNING: Please note the following drugs have extremely low permeation times: Carmustine (BCNU): 5.7 minutes and Thiotepa: 13.8. Do not use with Carmustine and Thiotepa.
| Opiod Drug | Concentration | Breakthrough Detection<br>Time in Minutes |
|----------------------------|---------------|-------------------------------------------|
| Fentanyl Citrate Injection | 100mcg/2mL | No breakthrough up to 240<br>minutes |
{8}------------------------------------------------
#### Summary of the Technological Characteristics of the Device: See table 6.0
## Table 1
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| | | PREDICATE | CURRENT | |
| 510(k) Number<br>Manufacturer(s) | - | K020019<br>WRP Asia<br>Pacific Sdn.<br>Bhd. | K232079<br>WRP Asia<br>Pacific Sdn.<br>Bhd. | Same |
| Material | ASTM D3577 | Natural Rubber<br>Latex | Natural Rubber<br>Latex | Same |
| Color | - | Natural White | Green and<br>Natural | Different |
| Texture | - | Microroughened<br>surface | Microroughened<br>surface | Same |
| Physical Properties | ASTM D3577 | Meets | Meets | Same |
| Before Aging<br>Tensile Strength:<br>Ultimate Elongation:<br>Stress at 500%<br>Elongation: | | Min. 24MPa<br>Min. 750%<br>Max. 5.5MPa | | |
| After Aging<br>Tensile Strength:<br>Ultimate Elongation:<br>Stress at 500%<br>Elongation: | | Meets<br>Min. 18MPa<br>Min. 560%<br>N/A | Meets | Same |
| Thickness:<br>- Finger<br>- Palm<br>- Cuff | ASTM D3577 | Meets<br>Min. 0.10 mm<br>Min. 0.10 mm<br>Min. 0.10 mm | Meets<br>Min. 0.10 mm<br>Min. 0.10 mm<br>Min. 0.10 mm | Same |
| Dimension, Length | ASTM D3577 | Meets<br>5½: Min.<br>245mm<br>6.0: Min. 265<br>mm<br>6½: Min. 265<br>mm<br>7.0: Min. 265<br>mm<br>7½: Min. 265<br>mm<br>8.0: Min. 265<br>mm<br>8½: Min. 265<br>mm<br>9.0: Min. 265<br>mm | Meets<br>5½: Min.<br>245mm<br>6.0: Min. 265<br>mm<br>6½: Min. 265<br>mm<br>7.0: Min. 265<br>mm<br>7½: Min. 265<br>mm<br>8.0: Min. 265<br>mm<br>8½: Min. 265<br>mm<br>9.0: Min. 265<br>mm | Same |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
| | | PREDICATE | CURRENT | |
| 510(k) Number<br>Dimension, Width | ASTM D3577 | K020019<br>Meets<br>5½: 70 ± 6 mm<br>6.0: 76 ± 6 mm<br>6½: 83 ± 6 mm<br>7.0: 89 ± 6 mm<br>7½: 95 ± 6 mm<br>8.0: 102 ± 6 mm<br>8½: 108 ± 6 mm<br>9.0: 114 ± 6 mm | K232079<br>Meets<br>5½: 70 ± 6 mm<br>6.0: 76 ± 6 mm<br>6½: 83 ± 6 mm<br>7.0: 89 ± 6 mm<br>7½: 95 ± 6 mm<br>8.0: 102 ± 6 mm<br>8½: 108 ± 6 mm<br>9.0: 114 ± 6 mm | Same |
| Powder Free | ASTM D6124 | Meets requirements of ≤2.0 mg/glove for Powder-Free designation per ASTM D3577 | Meets requirements of ≤2.0 mg/glove for Powder-Free designation per ASTM D3577 | Same |
| Biocompatibility | Primary Skin Irritation - ISO 10993-10 | Passes (Not a primary skin irritant) | Passes (Not a primary skin irritant) | Same |
| | Dermal Sensitization- ISO 10993-10 | Passes (Not a contact sensitizer) | Passes (Not a contact sensitizer) | Same |
| | Acute Systemic Toxicity, ISO 10993-11 | N/A | It is concluded that the devices did not induce any acute systemic toxicity | Different |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
| | | PREDICATE | CURRENT | |
| 510(k) Number | - | K020019 | K232079 | - |
| | Material<br>Mediated<br>Pyrogenicity,<br>ISO 10993-11 | N/A | It is concluded<br>that the devices<br>are non-<br>pyrogenic | Different |
| | Cytotoxicity-<br>Direct Contact,<br>ISO 10993-5 | N/A | Non-cytotoxic for<br>S&C/S&P gloves | Different |
| | Cytotoxicity-<br>MEM Elution,<br>ISO 10993-5 | N/A | Under the<br>conditions of<br>this study, the<br>test article<br>extract showed<br>potential<br>toxicity to L929<br>cells for Green<br>gloves | Different |
| Watertight (1000ml) | ASTM D5151 | Meets 21 CFR<br>800.20 and<br>ASTM D3577<br>when tested in<br>accordance<br>with ASTM<br>D5151<br>Inspection<br>Level I, AQL<br>0.65 | Meets 21 CFR<br>800.20 and<br>ASTM D3577<br>when tested in<br>accordance with<br>ASTM D5151<br>Inspection Level<br>I, AQL 0.65 | Same |
| Intended use /<br>Indications for Use | - | A powder-free<br>sterile<br>surgeon's glove<br>is a disposable<br>device made of<br>natural rubber<br>intended to be<br>worn on the<br>hands of<br>healthcare<br>personnel as a<br>barrier for<br>protection<br>against cross-<br>contamination<br>between the<br>healthcare<br>personnel and<br>patient. | A powder-free<br>sterile<br>surgeon's glove<br>is a disposable<br>device made of<br>natural rubber<br>intended to be<br>worn on the<br>hands of<br>healthcare<br>personnel as a<br>barrier for<br>protection<br>against cross-<br>contamination<br>between the<br>healthcare<br>personnel and<br>patient.<br>These gloves<br>were tested for | Different |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
| | | PREDICATE | CURRENT | |
| 510(k) Number | - | K020019 | K232079 | - |
| | | | Chemotherapy<br>Drugs and<br>Fentanyl Citrate<br>as per ASTM<br>D6978<br>Standard<br>Practice for<br>Assessment of<br>Medical Gloves<br>to Permeation<br>by<br>Chemotherapy<br>Drugs. | |
| Size | Medical Glove<br>Guidance<br>Manual -<br>Labeling | 5½<br>6.0<br>6½<br>7.0<br>7½<br>8.0<br>8½<br>9.0 | 5½<br>6.0<br>6½<br>7.0<br>7½<br>8.0<br>8½<br>9.0 | Same |
| Single Use | Medical Glove<br>Guidance<br>Manual -<br>Labeling | Single use | Single use | Same |
| Sterility Status | Medical Glove<br>Guidance<br>Manual -<br>Labeling | Sterile | Sterile | Same |
| Sterility Method | ISO 11137-1<br>ISO 11137-2 | Gamma<br>Radiation | Gamma<br>Radiation | Same |
| Chemotherapy Drug<br>Permeation Test | ASTM D6978-<br>05 | | | |
| 1. Natural Rubber Latex Surgical Glove, Powder Free, Sterile, Tested for Use with Chemotherapy<br>Drugs and Fentanyl (Green) | | | | |
| * Carmustine (BCNU)<br>(3.3 mg/ml) | | N/A | 11.9 | Different |
| Cisplatin (1.0 mg/ml) | | N/A | >240 | Different |
| Cyclophosphamide<br>(Cytoxan) (20.0 mg/ml) | | N/A | >240 | Different |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
| | | PREDICATE | CURRENT | |
| 510(k) Number | - | K020019 | K232079 | - |
| Dacarbazine (10.0 mg/ml) | | N/A | > 240 | Different |
| Doxorubicin<br>Hydrochloride (2.0 mg/ml) | | N/A | > 240 | Different |
| Etoposide (20.0 mg/ml) | | N/A | > 240 | Different |
| Fluorouracil (50.0 mg/ml) | | N/A | > 240 | Different |
| Ifosfamide (50.0 mg/ml) | | N/A | > 240 | Different |
| Methotrexate (25.0 mg/ml) | | N/A | > 240 | Different |
| Mitomycin C (0.5 mg/ml) | | N/A | > 240 | Different |
| Mitoxantrone (2.0 mg/ml) | | N/A | > 240 | Different |
| Paclitaxel (6.0 mg/ml) | | N/A | > 240 | Different |
| * ThioTepa (10.0 mg/ml) | | N/A | 14.7 | Different |
| Vincristine Sulfate (1.0 mg/ml) | | N/A | > 240 | Different |
| Warning Statement | | N/A | *WARNING:<br>Please note the<br>following drugs<br>have extremely<br>low permeation<br>times:<br>Carmustine<br>(BCNU): 11.9<br>minutes and<br>Thiotepa: 14.7<br>minutes. Do not<br>use with<br>Carmustine and<br>Thiotepa. | Different |
| Fentanyl Citrate (100 mcg/2mL) | ASTM D6978-<br>05 | N/A | > 240 | Different |
| 2. Natural Rubber Latex Surgical Glove (Strip&Coat)/(Strip&Pack), Powder Free, Sterile, Tested<br>for Use with Chemotherapy Drugs and Fentanyl | | | | |
| * Carmustine (BCNU)<br>(3.3 mg/ml) | | N/A | 5.7 | Different |
| Cisplatin (1.0 mg/ml) | | N/A | >240 | Different |
| Characteristics | | PREDICATE | CURRENT | Comparison |
| 510(k) Number | | K020019 | K232079 | |
| Cyclophosphamide<br>(Cytoxan) (20.0<br>mg/ml) | | N/A | >240 | Different |
| Dacarbazine (10.0<br>mg/ml) | | N/A | > 240 | Different |
| Doxorubicin<br>Hydrochloride (2.0<br>mg/ml) | | N/A | > 240 | Different |
| Etoposide (20.0<br>mg/ml) | | N/A | > 240 | Different |
| Fluorouracil (50.0<br>mg/ml) | | N/A | > 240 | Different |
| Ifosfamide (50.0<br>mg/ml) | | N/A | > 240 | Different |
| Methotrexate (25.0<br>mg/ml) | | N/A | > 240…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.