N1 is a dental portable X-ray imaging system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use N1 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
Device Story
Portable dental X-ray imaging system; generates X-rays for dental examinations. Components: control unit, user interface, X-ray aperture (collimator), back-scatter shielding glass, remote irradiation switch. Powered by rechargeable 22.2V Li-ion polymer battery. Operated by dentist or radiologist in dental clinics. Device captures radiographic images via intraoral imaging sensors (not included). User interface and remote switch facilitate irradiation. Output used by clinicians to diagnose dental conditions. Benefits: portable diagnostic imaging for teeth/jaws/oral structures.
Clinical Evidence
No clinical studies submitted. Substantial equivalence supported by bench testing, software verification/validation, and electrical/EMC safety testing.
Technological Characteristics
Handheld portable X-ray system; 70kV tube voltage; 3mA tube current; 0.3mm focal spot; 1.5mm Al total filtration. Powered by 22.2V DC Li-ion polymer battery. Includes back-scatter shielding. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-65 (X-ray performance), IEC 62133-2 (battery safety).
Indications for Use
Indicated for dental diagnosis and treatment of diseases of teeth, jaws, and oral structures in adults and children using intraoral imaging sensors; for use by trained dental practitioners or radiologists.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the FDA logo, which features an eagle. To the right of the eagle is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is stacked, with "FDA" on the top line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
November 28, 2023
Osstem Implant Co., Ltd. % Peter Lee QA/RA Manager Hiossen Inc. 85 Ben Fairless Dr. FAIRLESS HILLS PA 19030
Re: K232012
Trade/Device Name: N1 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: EHD Dated: October 25, 2023 Received: October 25, 2023
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232012
Device Name N1
Indications for Use (Describe)
N is a dental portable X-ray imaging system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use N1 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dsstem Implant Co., Ltd 2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea Tel: +82 70 4871 0191 Fax: +82 70 4015 0508 www.osstem.com
### 510(k) Summary - K232012
### Date: November 26, 2023
### 1. Company and Correspondent making the submission
| - Submitter's Name | : Osstem Implant Co., Ltd. |
|------------------------|-------------------------------------------------------------------------------------------------|
| - Address | : 2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si<br>Gyeonggi-do, 15079, Republic of Korea |
| - Contact | : Miss. Jimin hyun |
| - Phone | : +82-70-4871-0191 |
| - Correspondent's Name | : Hiossen Inc. |
| - Address | : 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact | : Mr. Peter Lee |
- +1-267-759-7031 Phone -
### 2. Proposed Device
| - Trade or (Proprietary) Name | : N1 |
|-------------------------------|---------------------------------|
| - Classification Name | : Extraoral source x-ray system |
| - Regulation Number | : 21 CFR872.1800 |
| - Devce Classification | : Class II |
| - Classification Product Code | : EHD |
| - 510(K) number | : K232012 |
### 3. Predicate Device
| - | Trade or (Proprietary) Name | : NOMAD DENTAL X-RAY SYSTEM |
|---|-----------------------------|---------------------------------|
| - | Classification Name | : Extraoral source x-ray system |
| - | Regulation Number | : 21 CFR872.1800 |
| - | Device Classification | : Class II |
| - | Classification Product Code | : EHD |
| - | 510(k) number | : K051795 |
### 4. Description
- N1 is a portable dental X-ray imaging device powered by a rechargeable lithium polymer battery pack. It generates X-rays optimized for dental examinations (on teeth, etc.). The N1's X-ray generator equipped with an X-ray tube consists of control unit, user interface, X-ray aperture (collimator), a back-scatter shielding glass, and remote irradiation switch. The x-ray detectors (digital detectors or analog film) are not accessories for use with the N1. N1 is a product designed to diagnose the conditions of teeth through X-irradiation using its intraoral imaging sensors. Software is included for operation and configuration of the N1. It is of Moderate level of concern and it's not based on the predicate software.
### 5. Substantial Equivalence Matrix
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, gray, sans-serif font. There is a small circle with the number 6 in the upper right corner of the logo.
Ossterm Implant Co., Ltd..
2 Floor, B-dong, S1, Mayu-ro 238beon-gil, Gibeung-si, Gyeonggi-do, 15079, Republic of Korea
Tel: +82 70 4871 0191 Fax: +82 70 4015 0508 www.osstem
| | | Proposed Devices | Predicate Devices | Remark |
|-------------------------|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Device Name | | N1 | NOMAD DENTAL X-RAY<br>SYSTEM | Different |
| 510(k) No. | | Proposed | K051795 | Different |
| Manufacturer | | Osstem Implant Co., Ltd. | ARIBEX,INC. | Different |
| Indications for Use | | N1 is a dental portable X-ray<br>imaging system that captures<br>radiographic images for dental<br>diagnosis using intraoral<br>imaging sensors. Only trained<br>and qualified dental<br>practitioner or radiologist shall<br>use N1 to diagnose and treat<br>diseases related to the teeth,<br>jaws, and/or other oral<br>structures in adults and<br>children. | The NOMADTM Dental X-ray<br>System is intended to be used<br>by trained dentists and dental<br>technicians as an extraoral x-ray<br>source for producing diagnostic<br>x-ray images using intraoral<br>image receptors. Its use is<br>intended for both adult and<br>pediatric subjects. | Same |
| Fixed structure | | Handheld device | Handheld device | Same |
| Remote switch | | USB Type-C Cable | N/A | Different |
| Power | | 22.2V | 14.4V | Similar<br>(included<br>within<br>range of<br>predicate<br>product) |
| Battery | | Rechargeable 22.2V DC<br>Li-ion Polymer Battery Pack | Rechargeable 14.4V DC<br>NiCd battery pack | Different |
| Tube voltage[kV] | | 70kV | 70kV | Same |
| Tube current[mA] | | 3mA | 2.3mA | Similar<br>(included<br>within<br>range of<br>predicate<br>product) |
| | Focus | 0.3mm | 0.4mm | Similar<br>(included<br>within<br>range of<br>predicate<br>product) |
| X-ray tube | Angle | 13' | 12' | Similar<br>(included<br>within<br>predicate<br>product) |
| | Inherent<br>Filtration | 0.5mm AI | 0.8mm AI | Similar<br>(included<br>within<br>range of<br>predicate<br>product) |
| | X-ray<br>Coverage<br>(FOV) | 20 mm x 30 mm (600 mm²) | Φ 60 mm (2826 mm²) | Similar<br>(included<br>within |
| | | 30 mm x 40 mm (1200 mm²) | | range of<br>predicate<br>product) |
| Total filtration | | 1.5mm Al | 1.5mm Al | Same |
| User | | Dentist or Radiologist | Dentist or Radiologist | Same |
| Operation environment | | Dental Clinic | Dental Clinic | Same |
| Site of application | | Tooth | Tooth | Same |
| Contact area of patient | | Buy a sterilized plastic bag at<br>the dental clinic and cover the<br>product with a plastic bag when<br>using | Buy a sterilized plastic bag at<br>the dental clinic and cover the<br>product with a plastic bag when<br>using | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo of OSSTEM IMPLANT. The logo is composed of the word "OSSTEM" in a bold, sans-serif font, with the letters in a bright orange color. Below "OSSTEM", the word "IMPLANT" is written in a smaller, gray, sans-serif font. The overall design is clean and modern, suggesting a professional and technologically advanced company.
# Dsstem Implant Co., Ltd
Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea el: +82 70 4871 0191 Fax: +82 70 4015 0508 www.osstem.com
### 6. Indications for Use
- -N1 is a dental portable X-ray imaging system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use N1 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
### 7. Summary of Non-clinical Performance Testing
Software validation and verification test, EMC/Electrical Safety tests, and performance tests were conducted and the test results support that the subject device is substantially equivalent to the predicate device
### Non-Clinical Performance Data
We performed the applicable non-clinical tests in the FDA guidance document for solid state x-ray detectors. The digital detectors used for testing are not part of the subject device.
### Software Verification and Validation Testing
Software was designed and developed according to a software development process. Also, Software verification and validation test were conducted and documented in accordance with FDA guidance: "The content of premarket submissions for software contained in medical devices, on May 11, 2005". The software for this device contains "Moderate" level of concern software.
### Safety, EMC and Performance Data
N1 complies with the electrical safety and electromagnetic compatibility requirements established by the standards IEC60601-1, IEC60601-1-2, IEC 60601-2-65 The N1 battery has been tested following the standards IEC 62133-2. This successful test results indicate that the lithium battery that operates the N1 system is effective and safe.
### 8. Summary of Clnical Testing
No clinical studies are submitted.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, with a small circle containing the number 6 in the upper right corner. Below "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern.
### 9. Conclusion
There were some differences in some specifications, but there was no significant difference in the safety / performance of the product, and it was confirmed that the Portable dental X-ray imaging device had better performance than the equivalent device in some specifications.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this submission, we conclude that the subject device is safe and effective and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.