← Product Code [MBI](/productcode/MBI) · K232005

# HEALICOIL PK Suture Anchor with Needles, ULTRATAPE (Blue); HEALICOIL PK Suture Anchor with Needles, ULTRATAPE (Blue Cobraid) (K232005)

_Smith & Nephew, Inc. · MBI · Aug 4, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K232005

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Aug 4, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is intended for use for the reattachment of soft tissue to bone.

## Device Story

Device is a soft tissue fixation system consisting of a suture anchor preloaded with ULTRATAPE suture and needles, mounted on an insertion device. Used by surgeons in orthopedic procedures to reattach soft tissue to bone. Operates as a mechanical fastener; anchor is inserted into bone to secure suture, which is then used to reattach tendon or ligament tissue. Benefits include stable fixation for soft tissue repairs. No electronic or software components.

## Clinical Evidence

Bench testing only. Mechanical testing performed on the device and all components, including fixation, insertion & fixation, cyclic loading, and needle attachment, met all required specifications.

## Technological Characteristics

Soft tissue fixation device consisting of a suture anchor, preloaded ULTRATAPE suture, and needles on an insertion device. Materials and design are consistent with the predicate (K152566). Mechanical fixation principle. No energy source, connectivity, or software.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- HEALICOIL PK Suture Anchor ([K152566](/device/K152566.md))

## Reference Devices

- TWINFIX Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 and Needles, SL ([K123425](/device/K123425.md))
- FAST-FIX FLEX ([K203393](/device/K203393.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

August 4, 2023

Smith & Nephew, Inc. Camille Fleischer Senior Regulatory Affairs Specialist 150 Minuteman Road Andover, Massachusetts 01810

Re: K232005

Trade/Device Name: HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE' (Blue); HEALICOIL PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: July 6, 2023 Received: July 6, 2023

Dear Camille Fleischer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Jesse Muir -S

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K232005

#### Device Name

HEALICOIL◊PK Suture Anchor with Needles, ULTRATAPE◊ (Blue); HEALICOILS PK Suture Anchor with Needles, ULTRATAPES (Blue Cobraid)

#### Indications for Use (Describe)

The HEALICOIL◊ PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL◊ PK Suture Anchor with Needles, ULTRATAPE (Blue Cobraid) are intended for use only for the reattachment of soft tissue to bone for the following indications:

Foot & Ankle

- Medial or lateral instability repairs/reconstructions
- Achilles tendon repairs/reconstructions

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

T:+ 1 978 749 1000 T:+ 1 800 343 8386 (USA toll free) www.smith-nephew.com

Image /page/3/Picture/2 description: The image shows the logo for Smith+Nephew. The logo is in a bright orange color. The words are written in a sans-serif font and are slightly bolded.

K232005

# 510(k) Summary

Prepared: 03 August 2023

| Submitter Information | Contact Information                  |
|-----------------------|--------------------------------------|
| Smith & Nephew, Inc.  | Mrs. Camille Fleischer               |
| Endoscopy Division    | Senior Regulatory Affairs Specialist |
| 150 Minuteman Road    | Phone: (978) 749-1057                |
| Andover, MA 01810     |                                      |

| Device Name & Classification |                                                                                                                                                             |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name             | HEALICOIL             ® PK Suture Anchor with Needles, ULTRATAPE® (Blue)<br>HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue             Cobraid) |
| Common Name                  | Soft Tissue Fixation Device                                                                                                                                 |
| Classification Name          | Fastener, fixation, biodegradable, soft tissue; fastener, fixation,<br>nondegradable, soft tissue                                                           |
| Classification Regulation    | 21 CFR 888.3040                                                                                                                                             |
| Class                        | II                                                                                                                                                          |
| Product Code(s)              | MBI                                                                                                                                                         |
| Panel                        | Orthopedic                                                                                                                                                  |

## Legally Marketed Predicate Devices

The Smith & Nephew HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is substantially equivalent in intended use and fundamental scientific technology to the following legally marketed devices in commercial distribution:

| Description                | Submission Number | Clearance Date |
|----------------------------|-------------------|----------------|
| HEALICOIL PK Suture Anchor | K152566           | 02DEC2015      |

## Legally Marketed Reference Devices

| Description                                                                    | Submission Number | Clearance Date |
|--------------------------------------------------------------------------------|-------------------|----------------|
| TWINFIX Ti 3.5 Suture Anchor with two<br>ULTRABRAID Sutures #2 and Needles, SL | K123425           | 20FEB2013      |
| FAST-FIX FLEX                                                                  | K203393           | 01FEB2021      |

{4}------------------------------------------------

#### Device Description

The Smith & Nephew HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is a fixation device intended to provide reattachment of soft tissue to bone. The device consists of a suture anchor with a preloaded ULTRATAPE suture with needles assembled on an insertion device.

#### Intended Use

The HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is intended for use for the reattachment of soft tissue to bone.

## Indications for Use

The HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) are intended for use only for the reattachment of soft tissue to bone for the following indications:

#### Foot & Ankle

- Medial or lateral instability repairs/reconstructions
- Achilles tendon repairs/reconstructions ●

## Technological Characteristics

The change is modifying the design of the Smith & Nephew HEALICOIL PK Suture Anchor (K152566) to include a new inserter and the addition needles to the preloaded ULTRATAPE suture. The Smith & Nephew HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE' (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is similar in intended use, indications for use, materials, design, and manufacturing process as the predicate; some of the same components as the reference device TWINFIX Ti 3.5 Suture Anchor with two ULTRABRAID Sutures #2 and Needles, SL (K123425) and the same packaging configuration as the reference device FAST-FIX FLEX (K203393) will be used. HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE° (Blue Cobraid) are substantially equivalent the predicate. The differences between the proposed device and predicate device are considered minor and do not raise new questions of safety or effectiveness.

#### Performance Data

Mechanical testing performed on the device and all components met the required specifications for the completed tests. A summary of the test results has been provided for fixation, insertion & fixation, cyclic loading and needle attachment. The results for all testing were passing.

{5}------------------------------------------------

## Conclusion

The substantial equivalence of the HEALICOIL° PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is based on similarities in indications for use, design features, operational principles, material biocompatibility and composition. Modifications in design are minor and do not raise additional questions of safety or effectiveness compared to the predicate device. Based on the similarities the HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue) and HEALICOIL® PK Suture Anchor with Needles, ULTRATAPE® (Blue Cobraid) is substantially equivalent to the predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K232005](https://fda-staging.innolitics.com/device/K232005)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).