K231985 · Shenzhen Mecun Medical Supply Co., Ltd. · GEI · Feb 16, 2024 · General, Plastic Surgery
Device Facts
Record ID
K231985
Device Name
Disposable Grounding Pad
Applicant
Shenzhen Mecun Medical Supply Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Indications for Use
The disposable grounding pad is an accessory for high-frequency surgery in monopolar applications. It is self-adhesive, ready-to-use, disposable, and single-patient-use.
Device Story
Disposable grounding pad functions as a return electrode for monopolar electrosurgery; completes electrical circuit between patient and high-frequency (HF) generator. Device consists of conductive adhesive hydrogel layer, aluminum foil/PET conductive laminate, PE-foam backing, and release liner. Applied to patient's muscular or well-vascularized convex skin site near surgical field. Compatible with HF generators equipped with Contact Quality Monitoring (CQM) systems; triggers alarm if safe contact is lost. Used in clinical settings by healthcare professionals. Benefits include safe return of electrical current to generator, preventing patient burns. Single-patient use; disposable.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing per IEC 60601-2-2. Accelerated aging validated a 2-year shelf-life.
Indicated for adult patients (>15kg) undergoing high-frequency monopolar electrosurgery. Used as a return electrode to complete the electrical circuit between the patient and the electrosurgical generator.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2024
Shenzhen Mecun Medical Supply Co., Ltd. Zheng Bo General Manager 2nd Level. 2nd Building, Fuqiang S&T Park, 6 Ailian Industrial Park, Zhugushi, Wulian Community Shenzhen, Guangdong province 518000, China
#### Re: K231985
Trade/Device Name: Disposable Grounding Pad Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: July 5, 2023 Received: July 5, 2023
#### Dear Zheng Bo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.02.16
13:44:10 -05'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Disposable Grounding Pad
Indications for Use (Describe)
The disposable grounding pad is an accessory for high-frequency surgery in monopolar applications. It is self-adhesive, ready-to-use, disposable, and single-patient-use.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is submitted as required by requirements of SMDA and 21 CFR §807.92.
#### 1. Submitter
| Submitter's Name | Shenzhen Mecun Medical Supply Co., Ltd. |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Zheng Bo |
| Address | 2nd Level, 2nd Building, Fuqiang S&T Park, 6 Ailian<br>Industrial Park, Zhugushi, Wulian Community, Longgang<br>Street, Longgang District, Shenzhen City, China |
| Telephone | +86-755-86062204 |
| Fax number | +86-755-86062204 |
| E-mail | james@szcsqy.com; manager@mecun.com |
| Data preparation | February 10, 2024 |
#### 2. Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|----------------------------------------------------------------|
| Trade Name: | Disposable grounding pad |
| Model | MA2000-WAB |
| Classification name: | Electrosurgical cutting and coagulation device and accessories |
| Classification: | II |
| Review Panel: | General & Plastic Surgery |
| Product Code: | GEI |
| Regulation Number: | 878.4400 |
## 3. Predicate Device Information
| Trade Name | BOWA Neutral Electrodes |
|---------------------|-------------------------------------------------------------------|
| 510(k) Number | K173877 |
| Classification name | Electrosurgical cutting and coagulation device and<br>accessories |
| Classification: | II |
| Review Panel: | General & Plastic Surgery |
| Product code | GEI |
| Regulation No. | 878.4400 |
{4}------------------------------------------------
#### 4. Device Descriptions
Mecun disposable grounding pad has a cable and a neutral electrode, the electrode consists of a conductive adhesive area surrounded by a non-conductive border.
The neutral electrode is intended to be equipped with a compatible generator which has a CQM, loss of safe contact between the neutral electrode and the patient will result in an alarm.
The disposable grounding pad is a non-sterile, non-active conductor intended to be fastened to a patient and connected to a high frequency surgical equipment, to create a circuit for the return of electrical current to the generator after its emission to perform electrosurgery on the patient. It is fastened to the patient's body typically where the greatest surface area can be covered in closest proximity to the surgical site. It may or may not include a return cable(s). This medical device is intended for disposable and single-patient use.
#### 5. Indications for Use
The disposable grounding pad is an accessory for high-frequency surgery in monopolar applications. It is self-adhesive, ready-to-use, disposable, and single-patient-use.
#### 6. Comparisons of technological characteristics with the predicate device
The subject disposable grounding pad and BOWA neutral electrodes (K173877) are similar in intended use, compositions, functions, scientific technology, materials and method of operation, the following table provides a comparison summary:
| Comparison<br>item | Subject Device | Predicate Device (K173877) | |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Manufacturer | Shenzhen Mecun Medical<br>Supply Co., Ltd. | BOWA-electronics GmbH & Co.<br>KG | Remark |
| Product name | Disposable Grounding Pad | BOWA Neutral Electrodes | |
| Product Code | GEI | GEI | Same |
| Regulation<br>Number | 878.4400 | 878.4400 | Same |
| Classification | II | II | Same |
| Intended<br>use&<br>Indication<br>s for Use | The disposable grounding pad<br>is an accessory for high-<br>frequency surgery in<br>monopolar applications. It is<br>self-adhesive, ready-to-use,<br>disposable, and single-patient-<br>use. | Disposable neutral electrodes<br>are self-adhesive, ready-to-use<br>and single-use products and are<br>an accessory for HF surgery in<br>monopolar applications. The<br>electrodes complete the<br>electrical circuit between the<br>patient and the HF generator on<br>the passive side. | Same |
| Prescription or<br>OTC | Prescription | Prescription | Same |
| Mechanism of<br>Action | Neutral electrodes serve to<br>return the current from the<br>patient to the electrosurgical<br>unit (ESU) during HF-surgery<br>in Monopolar application. | Neutral electrodes serve to<br>return the current from the<br>patient to the electrosurgical unit<br>(ESU) during HF-surgery in<br>Monopolar application. | Same |
| Technology<br>overview | Multi-layer device consists of:<br>Backing material;<br>Conductive layer; | Multi-layer device consists of:<br>Backing material;<br>Conductive layer; | Same |
| | Conductive adhesive<br>hydrogel;<br>Cover material | Conductive adhesive hydrogel;<br>Cover material | |
| Intended<br>population | Adult | Neonates, Children, Adults | Similar<br>(included) |
| Anatomical<br>location | Muscular or well vascularized<br>convex skin site, as close as<br>possible to the operating field | Muscular or well vascularized<br>convex skin site, as close as<br>possible to the operating field | Same |
| Weight range<br>according to<br>IEC 60601-2-2 | >15kg (33lbs)<br>Adults | >15kg (33lbs) Adults<br>>5kg (11lbs) Children and Adults<br>Between 5 and 15kg (11 to<br>33lbs) Children<br><5kg (11lbs) Neonates | Similar<br>(included) |
| Conductive<br>area | 124 cm² Adults | 140 cm² Adults<br>110 cm² Children and Adults<br>70 cm² Children<br>40 cm² Neonates | Similar<br>(included) |
| Material | Conductive laminate: Al-<br>foil/PET and medical grade<br>hydrogel<br>Backing: PE-foam<br>Cover: release liner | Conductive laminate: Al-foil/PET<br>and medical grade hydrogel<br>Backing: PE-foam<br>Cover: release liner | Same |
| Maximum<br>power rating | 124 cm²: 400W | 140 cm² not limited<br>110 cm² not limited<br>70 cm² limited to 200W<br>40 cm² limited to 100W | Similar<br>(included) |
| Maximum<br>current density<br>(power setting:<br>400W, load:<br>500Ω) | 6.29 mA/cm²<br>(Electrode pad: 124cm²) | 5.79 mA/cm² (Electrode pad:<br>140cm²)<br>6.45 mA/cm² (Electrode pad:<br>110cm²) | Similar<br>(included) |
| Maximum<br>power density | Adult type: 3.23W/cm² | Adult type: not limited | Similar<br>(included) |
| Maximum<br>permitted<br>voltage for the<br>cable | 500 Vp | 500 Vp | Same |
| Self-adhesive | Yes | Yes | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Single-<br>use/disposable | Yes | Yes | Same |
| Shelf-life | 2 years | 3 years | Note 1 |
| Biocompatibility | ISO10993-5, ISO10993-10 | ISO10993-5, ISO10993-10 | Same |
| Electrical<br>Performance<br>and Safety | IEC60601-1, IEC60601-1-2,<br>ISO60601-2-2 | IEC60601-1, IEC60601-1-2,<br>ISO60601-2-2 | Same |
| Compatibility<br>with HF<br>generators<br>(ESU) | Yes, if ESU is equipped with a<br>CQM system which fulfils IEC<br>60601-1 | Yes, if ESU is equipped with a<br>CQM system which fulfils IEC<br>60601-1 | Same |
| Accessory | Provide with cable | Provide with or without cable | Similar<br>(included) |
{5}------------------------------------------------
{6}------------------------------------------------
Note 1:
The shelf-life of subject device has been validated by accelerated aging test, the difference does not raise any new safety and effectiveness issue.
# 7. Performance data
The subject device conforms to the following standards:
- 7.1 Biocompatibility testing
The biocompatibility evaluation for Neutral electrode was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. The subject device is contacted with patient's intact skin during the electrosurgery operation for duration of less than 24 hours. The biocompatibility testing includes the following:
- . Cytotoxicity
- Intracutaneous Reactivity .
- Skin Sensitization .
7.2 Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the following standards:
- . IEC 60601-1: 2005 + A1:2012 + A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: 2014 + A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC
- . IEC 60601-2-2:2017 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
## 7.3 Performance of appearance and size
The appearance and size of the disposable grounding pad, 3 lots of pads and 5 pieces for each were inspected, the testing results showed the appearance and size of the proposed device met the design requirements.
## 8. Conclusions
Based on device comparison information and performance data, the subject device is as safety and effectiveness as predicate device, and the differences do not raise any new issue of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.