← Product Code [DTZ](/productcode/DTZ) · K231982

# Quantum Perfusion Blood Oxygenator VT75-E1, Quantum Perfusion Blood Oxygenator VT75-E2 (K231982)

_Qura S.R.L · DTZ · Sep 12, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K231982

## Device Facts

- **Applicant:** Qura S.R.L
- **Product Code:** [DTZ](/productcode/DTZ.md)
- **Decision Date:** Sep 12, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4350
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

VT75-E1 Quantum Perfusion Blood Oxygenator is a diffusion membrane oxygenator, designed to oxygenate and remove carbon dioxide from venous blood during cardiac surgery requiring cardiopulmonary bypass and to measure blood pressure and temperature during the procedure. The device is limited to 6 hours of use. Device is intended for Paediatric patients with a Body Surface Area (BSA) ranging from 0.6 m² to 1 m². VT75-E2 Quantum Perfusion Blood Oxygenator is a diffusion membrane oxygenator, designed to oxygenate and remove carbon dioxide from venous blood during cardiac surgery requiring cardiopulmonary bypass and to measure blood pressure during the procedure. The device is limited to 6 hours of use. Device is intended for Paediatric patients with a Body Surface Area (BSA) ranging from 0.6 m² to 1 m².

## Device Story

Diffusion membrane oxygenator for cardiopulmonary bypass; oxygenates venous blood and removes CO2. Input: venous blood via inlet connector; sweep gas via hollow fiber membrane. Principle: blood flows over outer surface of Polymethylpentene (PMP) hollow fibers while sweep gas flows inside; gas exchange occurs across membrane. Output: oxygenated blood with controlled CO2 levels. Used in cardiac surgery by perfusionists. Integrated sensors measure blood pressure (VT75-E1/E2) and temperature (VT75-E1 only). Surfaces treated with phosphorylcholine-based biocompatible coating. Single-use; sterile; 6-hour limit. Benefits: provides respiratory support during bypass; real-time monitoring of pressure/temperature aids clinical decision-making.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by non-clinical bench testing: gas transfer performance, temperature/pressure sensor verification, mechanical blood cell damage, device pressure drop, mechanical integrity, connector resistance, coating durability, and air handling. Biocompatibility testing performed per ISO 10993-1.

## Technological Characteristics

Hollow fiber membrane oxygenator; fibers made of Polymethylpentene (PMP). Surfaces treated with phosphorylcholine-based biocompatible coating. Integrated pressure sensor (0-800 mmHg) and temperature probe (VT75-E1). Sterilized by ethylene oxide (EtO). Single-use. Compliant with ISO 10993 series for biocompatibility.

## Regulatory Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

## Predicate Devices

- Medos Hilite Pediatric Oxygenator 2400 LT ([K090450](/device/K090450.md))

## Reference Devices

- Quantum Perfusion Blood Oxygenator ECC ([K203424](/device/K203424.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 12, 2024

Qura S.r.1. % Raffaella Tommasini QA&RA Director Spectrum Medical s.r.l. Via di Mezzo, 23 Mirandola, Modena 41037 Italy

Re: K231982

Trade/Device Name: Quantum Perfusion Blood Oxygenator VT75-E1, Quantum Perfusion Blood Oxygenator VT75-E2

Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary bypass oxygenator Regulatory Class: Class II Product Code: DTZ Dated: August 7, 2024 Received: August 7, 2024

Dear Raffaella Tommasini:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a white document with a red Adobe logo in the center. The document has a red spine on the left side with two silver rivets. The top right corner of the document has a red triangle.

For the best experience, open this PDF portfolio in Acrobat X or Adobe Reader X, or later.

Get Adobe Reader Now!

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Spectrum Medical. The text "Spectrum" is in blue, followed by a red line that resembles a heart rate monitor. The word "Medical" is in blue and follows the red line. 

www.spectrummedical.com

## 510(K) SUMMARY

| I.                     | <b>SUBMITTER</b>                                              |
|------------------------|---------------------------------------------------------------|
| Submitter Name:        | Spectrum Medical S.r.l.                                       |
| Submitter Address:     | Via di Mezzo, 23 41037 Mirandola (MO) Italy                   |
| Contact Person:        | Raffaella Tommasini, QA&RA Director – Spectrum Medical S.r.l. |
| Phone:                 | +39 0535 1803050                                              |
| e-mail:                | raffaella.tommasini@spectrummedical.com                       |
| Fax:                   | +39 0535 1803051                                              |
| Date Summary Prepared: | December 22th, 2023                                           |

### ll. DEVICE

| Proprietary Name:    | Quantum Perfusion Blood Oxygenator VT75-E1                                                                                                                    |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                      | Quantum Perfusion Blood Oxygenator VT75-E2                                                                                                                    |
| Common Name:         | Oxygenator                                                                                                                                                    |
| Classification Name: | Oxygenator, Cardiopulmonary Bypass                                                                                                                            |
| Regulatory Class:    | II                                                                                                                                                            |
| Product Code:        | DTZ                                                                                                                                                           |
| Panel:               | Cardiovascular Devices, Office of Health Technology 2<br>(OHT2) / Division of Health Technology 2 B (Circulatory<br>Support, Structural and Vascular Devices) |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Spectrum Medical. The word "Spectrum" is written in blue, followed by a red line that resembles a heartbeat. The word "Medical" is written in blue after the red line.

www.spectrummedical.com

### III. PREDICATE AND REFERENCE DEVICES

| Predicate Device:              |                                           |
|--------------------------------|-------------------------------------------|
| Proprietary Name:              | Medos Hilite Pediatric Oxygenator 2400 LT |
| Registered Establishment Name: | GISH BIOMEDICAL INC                       |
| Regulation Number:             | 870.4350                                  |
| Common Name:                   | Oxygenator                                |
| Classification Name:           | Cardiopulmonary Bypass Oxygenator         |
| Regulatory Class:              | II                                        |
| Product Code:                  | DTZ                                       |
| 510(k) Number:                 | K090450                                   |

## Reference Device:

| Proprietary Name:              | Quantum Perfusion Blood Oxygenators ECC |
|--------------------------------|-----------------------------------------|
| Registered Establishment Name: | Spectrum Medical S.r.l.                 |
| Regulation Number:             | 870.4350                                |
| Common Name:                   | Oxygenator                              |
| Classification Name:           | Cardiopulmonary Bypass Oxygenator       |
| Regulatory Class:              | II                                      |
| Product Code:                  | DTZ                                     |
| 510(k) Number:                 | K203424                                 |

### IV. DEVICE DESCRIPTION

The Quantum Perfusion Blood Oxygenator (acronym VT-E) diffusion membrane device is designed to oxygenate blood and remove carbon dioxide from venous blood during cardiac surgical procedures requiring cardiopulmonary bypass for a maximum duration of 6 hours.

Blood enters the oxygenator through the blood inlet connector, flows through a blood chamber, touching the outer surface of hollow fibers membrane; while the sweep gas flows into the hollow fiber membrane. The hollow fibers are made of Polymethylpentene (PMP). In this chamber, carbon dioxide moves from the blood to

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Spectrum Medical. The text "Spectrum Medical" is written in blue font. A red line in the shape of an EKG reading is placed between the two words.

www.spectrummedical.com

the gas compartment, while oxygen enters into the red blood cells. Then, blood exits the oxygenator with the desired level of oxygen content and saturation, and carbon dioxide content. Sweep gas composition and flow rate are used to control saturation, and oxygen and carbon dioxide content of blood at the outlet of the oxygenator. The device is non-toxic, non-pyrogenic, sterilized by ethylene oxide and packaged in a

single box.

All the device surfaces in contact with blood are treated with a phosphorylcholinebased biocompatible coating.

### INTENDED USE / INDICATIONS FOR USE V.

## VT75-E1

Quantum Perfusion Blood Oxygenator is a diffusion membrane oxygenator, designed to oxygenate and remove carbon dioxide from venous blood during cardiac surgery requiring cardiopulmonary bypass and to measure blood pressure and temperature during the procedure. The device is limited to 6 hours of use.

Device is intended for Paediatric patients with a Body Surface Area (BSA) ranging from 0.6 m² to 1 m².

## VT75-E2

Quantum Perfusion Blood Oxygenator is a diffusion membrane oxygenator, designed to oxygenate and remove carbon dioxide from venous blood during cardiac surgery requiring cardiopulmonary bypass and to measure blood pressure during the procedure. The device is limited to 6 hours of use.

Device is intended for Paediatric patients with a Body Surface Area (BSA) ranging from 0.6 m² to 1 m².

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Spectrum Medical. The word "Spectrum" is in blue, followed by a red heartbeat symbol. The word "Medical" is in blue and follows the heartbeat symbol. 

## www.spectrummedical.com

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES VI.

| Device                | Proposed Device - Quantum Perfusion Blood<br>Oxygenator<br>Spectrum Medical S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Device - Medos Hilite Pediatric<br>Oxygenator 2400 LT<br>GISH BIOMEDICAL INC                                                                                                                                                                                                                                                                                                                                                                            | Reference Device -<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>Spectrum Medical<br>S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Device                | Proposed Device - Quantum Perfusion Blood<br>Oxygenator<br>Spectrum Medical S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device - Medos Hilite Pediatric<br>Oxygenator 2400 LT<br>GISH BIOMEDICAL INC                                                                                                                                                                                                                                                                    | Reference Device -<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>Spectrum Medical<br>S.r.l.                                                                                                                                                                    |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                  | Quantum Perfusion Blood Oxygenator VT75-E1<br>Quantum Perfusion Blood Oxygenator VT75-E2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Medos Hilite Pediatric Oxygenator 2400<br>LT                                                                                                                                                                                                                                                                                                                                                                                                                      | Quantum Perfusion<br>Blood Oxygenator ECC<br>VT160-E1<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>VT160-E2<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>VT200-E1<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>VT200-E2                                                                                                                                                                                                                                                                                                      |                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                           | contact with blood are<br>treated with a<br>phosphorylcholine-<br>based biocompatible<br>coating.                                                                                                                                                                |
| 510(k) Number         | K231982                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K090450                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K203424                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Regulation #          | 870.4350                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 870.4350                                                                                                                                                                                                                                                                                                                                                  | 870.4350                                                                                                                                                                                                                                                         |
| Device<br>description | The Quantum Perfusion Blood Oxygenator<br>(acronym VT-E) diffusion membrane device is<br>designed to oxygenate blood and remove<br>carbon dioxide from venous blood during<br>cardiac surgical procedures requiring<br>cardiopulmonary bypass for a maximum<br>duration of 6 hours.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The MEDOS HILITE 2400 LT Hollow Fiber<br>Oxygenators consist of a hollow fiber<br>membrane oxygenator and<br>extracorporeal heat exchanger. The<br>MEDOS HILITE 2400 LT hollow fibre<br>membrane consists of a<br>Polymethylpentene gas plasma tight mat.<br>The unique mat design increases the<br>interaction between blood and gas.                                                                                                                            | The<br>Quantum<br>Perfusion<br>Blood<br>Oxygenator ECC<br>(acronym VT-E)<br>diffusion membrane<br>device is designed to<br>oxygenate blood and<br>remove carbon dioxide                                                                                                                                                                                                                                                                                                                                                               | Regulation Name       | Cardiopulmonary Bypass Oxygenator                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Cardiopulmonary Bypass Oxygenator                                                                                                                                                                                                                                                                                                                         | Cardiopulmonary<br>Bypass Oxygenator                                                                                                                                                                                                                             |
| Device                | Proposed Device - Quantum Perfusion Blood<br>Oxygenator<br>Spectrum Medical S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Device – Medos Hilite Pediatric<br>Oxygenator 2400 LT<br>GISH BIOMEDICAL INC                                                                                                                                                                                                                                                                                                                                                                            | Reference Device -<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>Spectrum Medical<br>S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Product Code          | DTZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | DTZ                                                                                                                                                                                                                                                                                                                                                       | DTZ                                                                                                                                                                                                                                                              |
|                       | Blood enters the oxygenator through the blood<br>inlet connector, flows through a blood<br>chamber, touching the outer surface of hollow<br>fibres membrane; while the sweep gas flows<br>into the hollow fibre membrane. The hollow<br>fibres are made of Polymethylpentene (PMP).<br>In this chamber, carbon dioxide moves from the<br>blood to the gas compartment, while oxygen<br>enters into the red blood cells. Then, blood<br>exits the oxygenator with the desired level of<br>oxygen content and saturation, and carbon<br>dioxide content. Sweep gas composition and<br>flow rate are used to control saturation, and<br>oxygen and carbon dioxide content of blood at<br>the outlet of the oxygenator. The device is non-<br>toxic, non-pyrogenic, sterilized by ethylene<br>oxide and packaged in a single box. All the<br>device surfaces in contact with blood are<br>treated with a phosphorylcholine-based<br>biocompatible coating. | creating a highly efficient blood<br>oxygenator. The heat exchanger consists<br>of a polyester non-porous hollow fiber<br>configured heat exchanger as the primary<br>element to affect heat exchange. This<br>element is encased by a polycarbonate<br>housing, which directs the blood around<br>the outside of the fibres while water flow<br>through the inner lumen of the fibres and<br>therefore effects heat exchange while<br>minimizing priming volume. | from venous blood<br>during cardiac surgical<br>procedures requiring<br>cardiopulmonary<br>bypass for a maximum<br>duration of 6 hours.<br>Blood enters the<br>oxygenator through<br>the blood inlet<br>connector, flows<br>through a blood<br>chamber, touching the<br>outer surface of hollow<br>fibres membrane;<br>while the sweep gas<br>flows into the hollow<br>fibre membrane. The<br>hollow fibres are made<br>of Polymethylpentene<br>(PMP). In this chamber,<br>carbon dioxide moves from the blood to the                 | Classification        | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | II                                                                                                                                                                                                                                                                                                                                                        | II                                                                                                                                                                                                                                                               |
| Device                | Proposed Device - Quantum Perfusion Blood Oxygenator<br>Spectrum Medical S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device – Medos Hilite Pediatric<br>Oxygenator 2400 LT<br>GISH BIOMEDICAL INC                                                                                                                                                                                                                                                                                                                                                                            | Reference Device -<br>Quantum Perfusion Blood Oxygenator ECC<br>Spectrum Medical S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                                                               | Indication for<br>Use | VT75-E1<br>Quantum Perfusion Blood Oxygenator VT75-E1<br>is a diffusion membrane oxygenator, designed<br>to oxygenate and remove carbon dioxide from<br>venous blood during cardiac surgery requiring<br>cardiopulmonary bypass and to measure blood<br>pressure and temperature during the<br>procedure. The device is limited to 6 hours to<br>use.<br>VT75-E2<br>Quantum Perfusion Blood Oxygenator VT75-E2<br>is a diffusion membrane oxygenator, designed<br>to oxygenate and remove carbon dioxide from | The hilite 2400 LT is a hollow fibre<br>membrane oxygenator is indicated for use<br>in procedures requiring the<br>extracorporeal oxygenation of and carbon<br>dioxide removal from human blood. The<br>2400 LT is a paediatric oxygenators<br>intended for use at blood flow rates of 0.5<br>to 2.4 litres per minute for periods of up to<br>six hours. | VT200-E1 and VT160-<br>E1<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>(VT-E1) is a diffusion<br>membrane<br>oxygenator, designed<br>to oxygenate and<br>remove carbon dioxide<br>from venous blood<br>during cardiac surgery<br>requiring<br>cardiopulmonary |
|                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | gas compartment,<br>while oxygen enters<br>into the red blood cells.<br>Then, blood exits the<br>oxygenator with the<br>desired level of oxygen<br>content and<br>saturation, and carbon<br>dioxide content.<br>Sweep gas composition<br>and flow rate are used<br>to control saturation,<br>and oxygen and carbon<br>dioxide content of<br>blood at the outlet of<br>the oxygenator. The<br>device is non-toxic,<br>non-pyrogenic,<br>sterilized by ethylene<br>oxide and packaged in<br>a single box. All the<br>device surfaces in |                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                  |

{8}------------------------------------------------

## www.spectrummedical.com

{9}------------------------------------------------

SpectrumMedical

{10}------------------------------------------------

SpectrumMedical

## www.spectrummedical.com

{11}------------------------------------------------

SpectrumMedical

| Device                                                                      | Proposed Device - Quantum Perfusion Blood<br>Oxygenator<br>Spectrum Medical S.r.l.                                        | Predicate Device - Medos Hilite Pediatric<br>Oxygenator 2400 LT<br>GISH BIOMEDICAL INC   | Reference Device -<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>Spectrum Medical<br>S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                    |
|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                             | cardiopulmonary bypass and to measure blood<br>pressure during the procedure. The device is<br>limited to 6 hours of use. |                                                                                          | blood pressure and<br>temperature during<br>the procedure. The<br>device is limited to 6<br>hours to use.<br>VT200-E2 and VT160-<br>E2<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>(VT-E2) is a diffusion<br>membrane<br>oxygenator, designed<br>to oxygenate and<br>remove carbon dioxide<br>from venous blood<br>during cardiac surgery<br>requiring<br>cardiopulmonary<br>bypass and to measure<br>blood pressure during<br>the procedure. The<br>device is limited to 6<br>hours of use. |
| Target<br>population                                                        | Paediatric<br>(0.6 m² < BSA < 1 m²)                                                                                       | Paediatric                                                                               | Adult                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device                                                                      | Proposed Device - Quantum Perfusion Blood<br>Oxygenator<br>Spectrum Medical S.r.l.                                        | Predicate Device - Medos  Hilite Pediatric<br>Oxygenator  2400 LT<br>GISH BIOMEDICAL INC | Reference Device -<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>Spectrum Medical<br>S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                    |
| Target User                                                                 | Perfusionist                                                                                                              | Perfusionist                                                                             | Perfusionist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Biocompatibility<br>requirements                                            | In compliance with ISO 10993 series                                                                                       | In compliance with ISO 10993 series                                                      | In compliance with ISO<br>10993 series                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Main Contacting<br>Materials                                                | Fiber: Polymethylpentene                                                                                                  | Fiber: Polymethylpentene                                                                 | Fiber:<br>Polymethylpentene                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sensor Operating<br>Pressure Range                                          | 0 to 800 mmHg                                                                                                             | N/A                                                                                      | 0 to 800 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sensor<br>Sensitivity                                                       | 5μV/V/mmHg ± 1%                                                                                                           | N/A                                                                                      | 5μV/V/mmHg ± 1%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Sensor Accuracy                                                             | 0 to 50 mmHg: ±1 mmHg + 1% of reading                                                                                     | N/A                                                                                      | 0 to 50 mmHg: ±1<br>mmHg + 1% of reading                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                             | 50 to 800 mmHg: ±1 mmHg + 3% of reading                                                                                   |                                                                                          | 50 to 800 mmHg: ±1<br>mmHg + 3% of reading                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Sensor<br>Overpressure<br>Protection                                        | -400 to + 6464 mmHg                                                                                                       | N/A                                                                                      | -400 to + 6464 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Blood side<br>Connector Type                                                | 1/4" (9.525mm)                                                                                                            | 1/4" (9.525mm)                                                                           | 3/8" (9.525mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Blood flow range<br>[I/min]                                                 | 0.5 - 2.4                                                                                                                 | 0.5 - 2.4                                                                                | 0.5 - 5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Max pressure<br>drop @ max flow<br>rate [mmHg]                              | 80                                                                                                                        | 150                                                                                      | 80                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device                                                                      | Proposed Device - Quantum Perfusion Blood<br>Oxygenator<br>Spectrum Medical S.r.l.                                        | Predicate Device - Medos Hilite Pediatric<br>Oxygenator 2400 LT<br>GISH BIOMEDICAL INC   | Reference Device -<br>Quantum Perfusion<br>Blood Oxygenator ECC<br>Spectrum Medical<br>S.r.l.                                                                                                                                                                                                                                                                                                                                                                                                    |
| Exchange surface<br>[m²]                                                    | $0.8 \pm 0.02 \text{ m}^2$                                                                                                | $0.65 \text{ m}^2$                                                                       | $1.5\pm0.05 \text{ m}^2$                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Priming Volume<br>[ml]                                                      | $75 \pm 5$                                                                                                                | 95                                                                                       | $145 \pm 10 \text{ ml}$                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| O2 Gas Transfer<br>[mlO2/min]<br>@ max flow rate<br>(QBlood:Qgas=1:<br>1)   | ≥120                                                                                                                      | ≥120                                                                                     | ≥250                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| CO2 Gas Transfer<br>[mlCO2/min]<br>@ max flow rate<br>(QBlood:Qgas=1:<br>1) | ≥96                                                                                                                       | ≥96                                                                                      | ≥200                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Single-use                                                                  | Yes                                                                                                                       | Yes                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Sterile Condition                                                           | Sterile                                                                                                                   | Sterile                                                                                  | Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Sterilization<br>Method                                                     | EtO sterilization process                                                                                                 | EtO sterilization process                                                                | EtO sterilization<br>process                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{12}------------------------------------------------

SpectrumMedical

## www.spectrummedical.com

{13}------------------------------------------------

SpectrumMedical

## www.spectrummedical.com

Table 1 – Comparative Data with Predicate and Reference Devices

{14}------------------------------------------------

SpectrumMedical

Spectrum Medical S.r.l. Via di Mezzo. 23 41037 Mirandola (MO) FAX: +39 0535 1803051

www.spectrummedical.com

### VII. PERFORMANCE DATA

## NON-CLINICAL TESTING

The following activities were performed to demonstrate product safety and effectiveness, considering the proposed change and related impact:

- -Gas transfer performance;
- -Temperature probe and Pressure sensor verification;
- -Mechanical Blood Cell Damage;
- -Device pressure Drop;
- -Mechanical Integrity;
- -Mechanical resistance of connectors;
- -Coating coverage and durability;
- Air handling. -

## ANIMAL TESTING

No animal studies have been performed except for mandatory biocompatibility tests according to International Standard ISO 10993-1:2018 [Recognition Nr. 2-258] and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

## CLINICAL TESTING

No clinical data have been included in the current Traditional 510(k) submission to support substantial equivalence to legally marketed predicate devices.

### VIII. CONCLUSIONS

Based on the indications for use, main technological characteristics and results of nonclinical testing, the Quantum Perfusion Blood Oxygenator devices, VT75-E1 and VT75-E2, have been demonstrated to be appropriate for the intended use and are considered substantially equivalent to predicate device, Medos Hilite Pediatric Oxygenator 2400 LT (K090450) and reference Spectrum Medical S.r.l own marketed Quantum Perfusion Blood Oxygenator ECC devices (K203424).

---

**Source:** [https://fda-staging.innolitics.com/device/K231982](https://fda-staging.innolitics.com/device/K231982)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).