OPT96X Model: The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway. This device is designed to be used in a hospital, sub-acute facility, or long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional. MYDUETXXX Model: The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway. This device is designed to be used in a long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional.
Device Story
Nasal cannula patient interface for delivery of high-flow warmed and humidified respiratory gases; used with compatible respiratory gas humidifiers (Airvo 2, myAirvo 2, Airvo 3, MR850, F&P 950). Device consists of tubing sub-assembly, headgear sub-assembly, and nasal prongs. Used in hospitals, sub-acute facilities, long-term care facilities, or home environments. Operated by healthcare professionals or lay users (home). Provides non-invasive respiratory support to spontaneously breathing patients. Benefits include improved patient comfort and airway hydration through warmed/humidified gas delivery. Single-patient use; disposable (up to 14 days hospital/sub-acute, 30 days home/long-term care).
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1, ISO 18562-1), mechanical performance (ISO 5356-1), usability engineering (IEC 62366-1), and transit testing (ISTA 2A). No clinical data presented.
Technological Characteristics
Nasal cannula interface with tubing and headgear sub-assemblies. Features color-coded sizing (Small: Orange, Medium: Blue, Large: Green). Headgear uses loop-and-buckle adjustment. Compatible with various F&P humidifiers. Non-sterile. Shelf life: 3 years. Storage/transport: -10 °C to +50 °C. Flow ranges vary by humidifier model (5-70 L/min).
Indications for Use
Indicated for spontaneously breathing pediatric (10 years and up) and adult patients requiring high flow warmed and humidified respiratory gases delivered to the upper airway via nasal cannula.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
February 7, 2024
Fisher & Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand
Re: K231956
Trade/Device Name: F&P Optiflow+ Duet Nasal Cannula Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Class II Product Code: BTT Dated: February 6, 2024 Received: January 8, 2024
Dear Reena Daken:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director, Respiratory Devices Team DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia,
{2}------------------------------------------------
Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231956
### Device Name
F&P Optiflow+ Duet Nasal Cannula
# Indications for Use (Describe)
OPT96X Model:
The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway.
This device is designed to be used in a hospital, sub-acute facility, or long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional.
## MYDUETXXX Model:
The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway.
This device is designed to be used in a long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
As Required by 21 CFR 807.92
| I.<br>SUBMITTER | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Company Name and Address | Fisher & Paykel Healthcare Limited<br>15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand<br>Telephone: +64 9 574 0100 |
| Prepared and Submitted by | Jung Yun Lee<br>Regulatory Affairs Associate |
| Contact Person | Reena Daken<br>Regulatory Affairs Manager<br>Telephone: +64 9 574 0100<br>Email: reena.daken@fphcare.co.nz |
| Date prepared | 30 June 2023 |
| II.<br>DEVICE | |
| Name of Device | F&P Optiflow+ Duet Nasal Cannula |
| Common/Usual Name | Nasal Cannula |
| Classification Name | Respiratory Gas Humidifier |
| Regulatory Class | Class II (21 CFR §868.5450) |
| Product Code | BTT |
#### III. PREDICATE DEVICE
- Predicate device: . o F&P Optiflow+ Nasal Cannula OPT94X, K162553
#### IV. DEVICE DESCRIPTION
The F&P Optiflow+ Duet Nasal Cannula is a nasal cannula interface for use with a respiratory gas humidifier and flow generator to deliver Nasal High Flow (NHF) therapy to spontaneously breathing patients.
The F&P Optiflow+ Duet Nasal Cannula is a prescription-only device, provided in a non-sterile state and intended to be used in a hospital, sub-acute facility, or long-term (managed) care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional. The device is single patient use only for up to 14 days in the hospital and up to 30 days in the home/long-term care facilities.
{5}------------------------------------------------
A list of the subject device product codes can be found below:
| Model | Product Code | Product Description |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|------------------------------------------|
| OPT96X<br>This device is designed to be used in a hospital, sub-acute facility or long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional. | OPT962 | Nasal Cannula - Small |
| | OPT964 | Nasal Cannula - Medium |
| | OPT966 | Nasal Cannula - Large |
| MYDUETXXX<br>This device is designed to be used in a long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional. | MYDUETSMALL | myAirvo Small Duet Nasal Cannula 2-pack |
| | MYDUETMEDIUM | myAirvo Medium Duet Nasal Cannula 2-pack |
| | MYDUETLARGE | myAirvo Large Duet Nasal Cannula 2-pack |
#### V. INDICATIONS FOR USE
The indications for use for the OPT96X Model and the MYDUETXXX Model are below:
# OPT96X Model:
The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway.
This device is designed to be used in a hospital, sub-acute facility, or long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional.
# MYDUETXXX Model:
The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway.
This device is designed to be used in a long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed by a healthcare professional.
{6}------------------------------------------------
| Design/Technological<br>Characteristic | Subject Device<br>(F&P Optiflow+ Duet Nasal Cannula) | Predicate Device (K162553)<br>(F&P Optiflow+ Nasal Cannula,<br>OPT94X) | Comments |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | | | |
| Product Code | BTT | BTT | |
| Device Classification | 21 CFR §868.5450 | 21 CFR §868.5450 | Identical |
| Classification Panel | Anesthesiology | Anesthesiology | |
| Indications for use | | | |
| Indications for use | OPT96X Model<br>The F&P Optiflow+ Duet is a nasal cannula patient interface for use with specified respiratory gas humidifiers to treat spontaneously breathing pediatric (10 years and up) and adult patients who would benefit from receiving high flow warmed and humidified respiratory gases to the upper airway.<br><br>This device is designed to be used in a hospital, sub-acute facility or long-term care facility by appropriately qualified healthcare professionals, or in the home by lay users operating the device as prescribed | Nasal cannula patient interface for delivery of humidified respiratory gases. | The Indications for Use statement for the subject device is more specific than the predicate device.<br><br>Target population and operating environment have been defined for the subject devices, for clarity. |
| Design/Technological<br>Characteristic | Subject Device<br>(F&P Optiflow+ Duet Nasal Cannula) | Predicate Device (K162553)<br>(F&P Optiflow+ Nasal Cannula,<br>OPT94X) | Comments |
| | MYDUETXXX Model<br><br>The F&P Optiflow+ Duet is a nasal cannula<br>patient interface for use with specified<br>respiratory gas humidifiers to treat<br>spontaneously breathing pediatric (10 years<br>and up) and adult patients who would<br>benefit from receiving high flow warmed<br>and humidified respiratory gases to the<br>upper airway. | | |
| | This device is designed to be used in a long-<br>term care facility by appropriately qualified<br>healthcare professionals, or in the home by<br>lay users operating the device as prescribed<br>by a healthcare professional. | | |
| Availability | Prescription use.<br>(Part 21 CFR 801 Subpart D) | Prescription use.<br>(Part 21 CFR 801 Subpart D) | Identical |
| Patient Population | Pediatric (10 years and up) to adult. | Mainly Adult Patients. | Patient population for the subject device<br>is more specific than the predicate device. |
| Patient Acuity | Spontaneously breathing patients. | Spontaneously breathing patients. | Identical |
| Patient Monitoring | Appropriate patient monitoring must be<br>used at all times. | Appropriate patient monitoring must be<br>used at all times. | Identical |
| Use Environment | OPT96X Series:<br>Home and Hospital environment, sub-acute | Home and Hospital environment, long-term<br>(managed) care facilities | The addition of sub-acute facilities does<br>not affect / change the user group and<br>Identical |
| Design/Technological<br>Characteristic | Subject Device<br>(F&P Optiflow+ Duet Nasal Cannula) | Predicate Device (K162553)<br>(F&P Optiflow+ Nasal Cannula,<br>OPT94X) | Comments |
| | facilities, long-term (managed) care<br>facilities | | patient population compared to the<br>predicate. |
| | MYDUET Series:<br>Home and long-term (managed) care<br>facilities | | |
| Reusability | Single patient-use only, < 14 days in<br>hospital, including sub-acute facilities | Single patient-use only, < 14 days in<br>hospital | Identical |
| | Single patient-use only, < 30 days in home,<br>and long-term care facilities | Single patient-use only, < 30 days in home<br>and long-term care facilities | |
| Number of Cannula<br>Sizes | Both Hospital and Home models are<br>available in:<br>Small<br>Medium<br>Large | Both Hospital and Home models are<br>available in:<br>Small<br>Medium<br>Large | Identical |
| Compatibility with other devices | | | |
| Humidifier<br>Compatibility | Airvo 2 and myAirvo 2 (K131895)<br>Airvo 3 (K221338)<br>MR850 humidifier (K110019)<br>F&P 950 Respiratory Humidifier<br>(K220703) | Airvo 2 and myAirvo 2 (K131895)<br>MR850 humidifier (K110019) | The predicate was not cleared for use<br>with the Airvo 3 or the F&P 950 System. |
| Design/Technological<br>Characteristic | Subject Device<br>(F&P Optiflow+ Duet Nasal Cannula) | Predicate Device (K162553)<br>(F&P Optiflow+ Nasal Cannula,<br>OPT94X) | Comments |
| Nasal Cannula Design | | | |
| Overall Cannula<br>Construction | The F&P Optiflow+ Duet Nasal Cannula is<br>made up of two sub-assemblies that connect<br>to the Nasal Prong component. | The F&P Optiflow+ Nasal Cannula is made<br>up of two sub-assemblies that connect to the<br>Nasal Prong component. | Identical |
| | The two sub-assemblies are: | The two sub-assemblies are: | |
| | Tubing sub-assembly Headgear sub-assembly | Tubing sub-assembly Headgear sub-assembly | |
| Headgear Adjustment | The headgear size adjustment (i.e. headstrap<br>length and tightness adjustment) method<br>comprises of the Headstrap component<br>looping through (and being retained by) two<br>slots on the Headstrap Buckle component. | The headgear size adjustment (i.e. headstrap<br>length and tightness adjustment) method<br>comprises of the Headstrap component<br>looping through (and being retained by) two<br>slots on the Headstrap Buckle component. | Identical |
| Size Identification Color<br>Coding | Size Small: Orange<br>Size Medium: Blue<br>Size Large: Green | Size Small: Orange<br>Size Medium: Blue<br>Size Large: Green | Identical |
| Performance Specifications | | | |
| F&P Airvo 2 System<br>Flow Range | OPT962 (Small) 10 – 50 L/min<br>OPT964 (Medium) 10 – 60 L/min<br>OPT966 (Large) 10 – 60 L/min | OPT942 (Small) 10 – 50 L/min<br>OPT944 (Medium) 10 – 60 L/min<br>OPT946 (Large) 10 – 60 L/min | Identical |
| F&P Airvo 3 System<br>Flow Range | OPT962 (Small) 10 – 60 L/min<br>OPT964 (Medium) 10 – 70 L/min<br>OPT966 (Large) 10 – 70 L/min | OPT942 (Small) 10 – 60 L/min<br>OPT944 (Medium) 10 – 70 L/min<br>OPT946 (Large) 10 – 70 L/min | Identical |
| Design/Technological<br>Characteristic | Subject Device<br>(F&P Optiflow+ Duet Nasal Cannula) | Predicate Device (K162553)<br>(F&P Optiflow+ Nasal Cannula,<br>OPT94X) | Comments |
| MR850 System Flow<br>Range | OPT962 (Small) 5 - 60 L/min<br>OPT964 (Medium) 5-60 L/min<br>OPT966 (Large) 5 - 60 L/min | OPT942 (Small) 5 - 60 L/min<br>OPT944 (Medium) 5-60 L/min<br>OPT946 (Large) 5 - 60 L/min | Identical |
| F&P 950 System Flow<br>Range | OPT962 (Small) 5 - 70 L/min<br>OPT964 (Medium) 5 - 70 L/min<br>OPT966 (Large) 5 - 70 L/min | N/A | The predicate was not cleared for use<br>with the F&P 950 System. |
| F&P myAirvo System<br>Flow Range | MYDUETSMALL 10-50 L/min<br>MYDUETMEDIUM 10-60 L/min<br>MYDUETLARGE 10-60 L/min | OPT942E (Small) 10-50 L/min<br>OPT944E (Medium) 10-60 L/min<br>OPT946E (Large) 10-60 L/min | Identical |
| Shelf life | 3 years | Shelf life not claimed on labeling | The subject device claims a 3-year shelf<br>life. |
| Transport and Storage<br>Temperature range | -10 °C to +50 °C | -10 °C to +50 °C | Identical |
| Sterility | Device not provided sterile | Device not provided sterile | Identical |
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVIC VI.
ison of features, performance data, and intended use demonstrate that the F&P Solo Mask Range is substantially equivals The com
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
#### VII. PERFORMANCE DATA
# Summary of Non-Clinical Tests
The F&P Optiflow+ Duet Nasal Cannula has been tested to applicable requirements to the following standards:
- ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process
- ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare . applications - Part 1: Evaluation and testing within a risk management process
- ISO 5356-1:2015 Anesthetic and respiratory equipment Conical connectors: Part 1: ● Cones and sockets
- IEC 62366-1:2015+A1:2020 Medical devices Application of usability engineering to . medical devices
- ISTA 2A:2011 International Safety Transit Association Guidelines Procedure 2A: . Packaged-Products weighing 150 lb (68 kg) or less
# VIII. CONCLUSIONS
The F&P Optiflow+ Duet Nasal Cannula is substantially equivalent to the predicate based on patient population, intended uses, comparison of the technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.