Biatain Fiber Ag (33570 - 5 cm x 5 cm / 2 in x 2 in); Biatain Fiber Ag (33572 - 10 cm x 12 cm / 4 in x 5 in); Biatain Fiber Ag (33574 - 15 cm x 15 cm / 6 in x 6 in); Biatain Fiber Ag (33576 - 2.5 cm x 46 cm / 1 in x 18 in); Biatain Fiber Ag (33578 - 20 cm x 30 cm / 8 in x 12 in)

K231953 · Coloplast Corp. · FRO · Mar 22, 2024 · SU

Device Facts

Record IDK231953
Device NameBiatain Fiber Ag (33570 - 5 cm x 5 cm / 2 in x 2 in); Biatain Fiber Ag (33572 - 10 cm x 12 cm / 4 in x 5 in); Biatain Fiber Ag (33574 - 15 cm x 15 cm / 6 in x 6 in); Biatain Fiber Ag (33576 - 2.5 cm x 46 cm / 1 in x 18 in); Biatain Fiber Ag (33578 - 20 cm x 30 cm / 8 in x 12 in)
ApplicantColoplast Corp.
Product CodeFRO · SU
Decision DateMar 22, 2024
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The product is indicated for use on moderate to highly exuding acute and chronic wounds such as: - Diabetic ulcers - Leg ulcers (arterial ulcers, venous ulcers and leg ulcers of mixed etiology) - Pressure injuries (stage 2-4) - Traumatic wounds - Partial thickness (second degree) burns - Donor sites - Post-operative surgical wounds - Exudate absorption in oncology wounds, such as fungoides-cutaneous tumors, fungating carcinoma, cutaneous metastasis, Kaposi's sarcoma, and angiosarcoma

Device Story

Biatain Fiber Ag is a sterile, primary wound dressing composed of non-woven gelling carboxymethyl cellulose (CMC) fibers and non-gelling bicomponent (BiCo) fibers. It incorporates ionic silver (Ag+) for antimicrobial efficacy and ethyl lauroyl arginate (LAE) as a surfactant and chelator. The dressing is embossed with a hexagon pattern to provide structural integrity. Used in clinical settings by healthcare professionals, the dressing is applied directly to the wound bed. Upon contact with wound exudate, the dressing absorbs fluid and transforms into a conformable gel, maintaining a moist environment that supports autolytic debridement. The ionic silver within the gelled dressing kills microorganisms held within the dressing, reducing bacterial growth. The device is intended to manage exudate and support healing in various wound types. It may require a secondary dressing depending on the clinical condition of the wound.

Clinical Evidence

No human clinical trials were conducted. Evidence includes biocompatibility testing per ISO 10993-1:2018 (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity, implantation) and bench testing of physical/mechanical properties (absorption, fluid retention, wet strength, shrinkage, expansion). Antimicrobial effectiveness was demonstrated via in vitro AATCC 100 testing, showing ≥4 log reduction against 9 organisms (Gram-positive, Gram-negative, yeast, mold) for up to 7 days. An animal study (porcine full-thickness wound model) confirmed the dressing does not hinder the wound healing process compared to the predicate.

Technological Characteristics

Primary dressing composed of sodium carboxymethyl cellulose (CMC) fibers and bicomponent (BiCo) fibers. Contains 0.63% (w/w) ionic silver and ethyl lauroyl arginate (LAE). Features a hexagon-embossed pattern for structural integrity. Sterile, non-active, single-use device. Functions via passive absorption and gel formation; antimicrobial action via silver release. No software or electronic components.

Indications for Use

Indicated for patients with moderate to highly exuding acute and chronic wounds, including diabetic ulcers, leg ulcers, pressure injuries (stage 2-4), traumatic wounds, partial thickness burns, donor sites, post-operative surgical wounds, and oncology wounds (e.g., fungating carcinoma, Kaposi's sarcoma).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 22, 2024 Coloplast Corp. Laura Kelly Principal Regulatory Affairs Specialist 1601 West River Road North Minneapolis, Minnesota 55411 Re: K231953 Trade/Device Name: Biatain Fiber Ag (33570 - 5 cm x 5 cm / 2 in x 2 in); Biatain Fiber Ag (33572 - 10 cm x 12 cm / 4 in x 5 in); Biatain Fiber Ag (33574 - 15 cm x 15 cm / 6 in x 6 in); Biatain Fiber Ag (33576 - 2.5 cm x 46 cm / 1 in x 18 in); Biatain Fiber Ag (33578 - 20 cm x 30 cm / 8 in x 12 in) Regulatory Class: Unclassified Product Code: FRO Dated: February 22, 2024 Received: February 22, 2024 Dear Laura Kelly: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory- assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Yu-chieh Chiu -S Yu-Chieh Chiu, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231953 Device Name Biatain Fiber Ag Indications for Use (Describe) The product is indicated for use on moderate to highly exuding acute and chronic wounds such as: - · Diabetic ulcers - · Leg ulcers (arterial ulcers, venous ulcers and leg ulcers of mixed etiology) - · Pressure injuries (stage 2-4) - Traumatic wounds - · Partial thickness (second degree) burns - · Donor sites - · Post-operative surgical wounds · Exudate absorption in oncology wounds, such as fungoides-cutaneous tumors, fungating carcinoma, cutaneous metastasis, Kaposi's sarcoma, and angiosarcoma | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## TRADITIONAL 510(k) SUMMARY | Submitted by: | Coloplast A/S<br>Holtedam 1<br>3050<br>Humlebaek<br>Denmark | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Laura Kelly<br>Principal Regulatory Affairs Specialist<br>Coloplast Corp.<br>1601 West River Road North<br>Minneapolis MN 55411<br>Phone: +1 (612) 940-9983<br>Email: uslke@coloplast.com | | Date of Summary: | March 22, 2024 | | SUBJECT DEVICE | | | Trade or Proprietary<br>Name: | Biatain Fiber Ag | | Classification Name: | Dressing, Wound, Drug | | Device Classification: | Unclassified | | Product Code: | FRO | | Review Panel: | General & Plastic Surgery | | Predicate Device: | K173675, AQUACELTM Ag+ EXTRA Enhanced<br>HydrofiberTM Dressing with Silver and Strengthening Fiber | | Device Description: | Biatain Fiber Ag is a soft, sterile, primary wound dressing<br>made mainly from non-woven gelling carboxymethyl<br>cellulose (CMC) fibers. In addition, non-gelling<br>bicomponent (BiCo) fibers are incorporated to obtain<br>strength and integrity of the products both in dry and<br>gelled condition. Ionic silver (Ag+) is added for<br>antimicrobial efficacy within the dressing, and ethyl<br>lauroyl arginate (LAE) is added as combined surfactant<br>and chelator within the dressing. | {5}------------------------------------------------ | | Subject device<br>Biatain Fiber Ag (K231953) | Predicate device<br>AQUACEL Ag+ EXTRA (K173675) | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The product is indicated for use on moderate<br>to highly exuding acute and chronic wounds<br>such as:<br>Diabetic ulcers Leg ulcers (arterial ulcers, venous ulcers<br>and leg ulcers of mixed etiology) Pressure injuries (stage 2-4) Traumatic wounds Partial thickness (second degree) burns Donor sites Post-operative surgical wounds Exudate absorption in oncology wounds,<br>such as fungoides-cutaneous tumors,<br>fungating carcinoma, cutaneous metastasis,<br>Kaposi's sarcoma, and angiosarcoma Prescription use only. | Over-the-Counter Use:<br>AQUACELTM Ag+ EXTRA Enhanced HydrofiberTM<br>Dressing with Silver and Strengthening Fiber may be<br>used for the management of:<br>Abrasions Lacerations Minor cuts Minor scalds and burns Prescription Use:<br>Under the supervision of a healthcare professional:<br>AQUACELTM Ag+ EXTRA Enhanced HydrofiberTM<br>Dressing with Silver and Strengthening Fiber may be<br>used for the management of:<br>Wounds as an effective barrier to bacterial<br>penetration of the dressing as this may help<br>reduce infection; Partial thickness (second degree) burns; Diabetic<br>foot ulcers, leg ulcers, (venous stasis ulcers,<br>arterial ulcers and leg ulcers of mixed etiology)<br>and pressure ulcers/sores (partial & full<br>thickness); Surgical wounds left to heal by secondary<br>intention such as dehisced surgical incisions; Surgical wounds that heal by primary intent such<br>as dermatological and surgical incisions (e.g.,<br>orthopedic and vascular); Traumatic wounds; Wounds that are prone to bleeding, such as<br>wounds that have been mechanically or<br>surgically debrided and donor sites; Oncology wounds with exudate, such as<br>fungoides-cutaneous tumors, fungating<br>carcinoma, cutaneous metastasis, Kaposi's<br>sarcoma, and angiosarcoma; Management of painful wounds; Infected wounds | | Components | Composed of sodium carboxymethyl<br>cellulose (CMC) fibers with bicomponent<br>(BiCo) fibers and 0.63% (w/w) ionic silver<br>with ethyl lauroyl arginate (LAE).<br>The device is reinforced with BiCo fibers<br>embossed to create a hexagon pattern. | Composed of sodium carboxymethylcellulose<br>(HydrofiberTM) and 1.2% ionic silver with<br>EDTA and benzethonium chloride. | | Mode of<br>action | The dressing absorbs exudate and forms a<br>gel. The gelled dressing retains exudate,<br>thereby maintaining a moist wound<br>environment to support the healing process.<br>The dressing aids in the removal of<br>unnecessary material from the wound<br>(autolytic debridement). | The dressing absorbs wound fluid and creates a<br>soft, conformable gel, which maintains a moist<br>wound environment to support the healing<br>process.<br>The dressing aids in the removal of unnecessary<br>material from the wound (autolytic debridement)<br>without damaging newly formed tissue. | | | When the wound exudate is absorbed by the<br>Biatain Fiber Ag dressing, the free ionic<br>silver exerts its activity by killing the<br>microorganisms held within the dressing. | The silver preservative in the dressing reduces<br>the growth of microorganisms within the<br>dressing. The silver preservative in the dressing kills<br>wound bacteria held in the dressing and provides<br>an effective barrier to bacterial penetration of the<br>dressing. | | Characteristics | Sterile. Absorbs exudate. Forms a conformable gel The preservative action of ionic silver<br>reduces bacterial growth within the<br>dressing. Should be used with a secondary dressing<br>depending on the clinical condition of the<br>wound. | Sterile. Absorbs exudate (including bacteria) or blood. Forms a soft conformable gel. Ionic silver within the dressing functions as a<br>preservative to reduce bacterial growth within<br>the dressing. May require a secondary dressing. | ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS {6}------------------------------------------------ The differences in technological characteristics of the subject device do not raise additional questions of safety and effectiveness. ## PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. #### Biocompatibility: A biocompatibility evaluation, in accordance with ISO 10993-1:2018, has been conducted and concludes that the dressings have a safe toxicological profile for their intended use. The battery of testing included the following tests: - . Cytotoxicity - . Irritation - . Sensitization - Acute systemic toxicity ● - Sub-acute systemic toxicity . - . Pyrogenicity - . Implantation Testing of physical/mechanical properties has been conducted on key performance parameters: - Absorption capacity . - . Fluid retention capacity - . Wet strength (gelled condition) - . Lateral shrinkage upon absorption - Volume expansion upon absorption ● - Lateral spreading of fluid . ## Bench Testing: {7}------------------------------------------------ | Antimicrobial<br>Effectiveness within the<br>Dressing: | The preservative action of silver and LAE reduces bacterial<br>growth within the dressing. This was confirmed by <i>in vitro</i><br>tests. | |--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | ≥4 log reduction (AATCC 100) antimicrobial effectiveness<br>against 9 organisms, 3 Gram positive bacteria, 4 Gram<br>negative bacteria, 1 yeast, and 1 mold for up to 7 days<br>within the dressing. | | Animal Study: | A full thickness wound healing model in porcine study was<br>conducted to evaluate local tissue effects and wound healing<br>performance of a wound dressing containing silver (Biatain<br>Fiber Ag). The study demonstrated that Biatain Fiber Ag<br>and the predicate device did not hinder the wound healing<br>process. | | CONCLUSION | Based on the identical indications for use, principles of<br>operations, technological characteristics, and performance<br>testing, Biatain Fiber Ag is as safe and as effective as the<br>predicate device, ConvaTec's AQUACELTM Ag+ EXTRA<br>Enhanced HydrofiberTM Dressing with Silver and<br>Strengthening Fiber (K173675). |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...