Rapid Fentanyl (FYL) Test Strip, Rapid Fentanyl (FYL) Test Dipcard
Applicant
Co-Innovation Biotech Co., Ltd.
Product Code
DJG · Clinical Toxicology
Decision Date
Mar 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Indications for Use
The Rapid Fentanyl (FYL) Test Strip is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The Rapid Fentanyl (FYL) Test Dipcard is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
Device Story
Rapid Fentanyl (FYL) Test Strip and Dipcard are lateral flow immunochromatographic assays for qualitative fentanyl detection in human urine. Input: human urine sample. Principle: competitive binding antigen-antibody immunochemistry. Output: visual qualitative result (positive/negative) on test strip/dipcard. Used at point-of-care sites by healthcare professionals. No instrument required. Results are preliminary; clinical judgment and confirmatory testing (GC/MS or LC/MS) required for final diagnosis. Benefits: rapid screening for fentanyl presence to assist clinical decision-making.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated across 3 lots, 6 operators, and 3 point-of-care sites (1620 observations). Analytical specificity/interference tested against various opioids and common medications; no interference observed. Method comparison performed using 480 clinical urine samples confirmed by LC/MS.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Form factors: test strip and test dipcard. Storage: 4-30°C. No instrument/platform required. Materials and manufacturing procedures are identical between strip and dipcard formats.
Indications for Use
Indicated for qualitative screening of Fentanyl in human urine at 1 ng/mL cutoff. Intended for prescription use in point-of-care settings. Provides preliminary results; requires confirmation via GC/MS or LC/MS.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K231904
B Applicant
Co-Innovation Biotech Co., Ltd.
C Proprietary and Established Names
Rapid Fentanyl (FYL) Test Strip, Rapid Fentanyl (FYL) Test Dipcard
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DJG | Class II | 21 CFR 862.3650 - Opiate Test System | TX - Clinical Toxicology |
## II Submission/Device Overview:
A Purpose for Submission: New Device
B Measurand: Fentanyl in human urine.
C Type of Test: Qualitative competitive binding, lateral flow immunochromatographic assay.
## III Intended Use/Indications for Use:
A Intended Use(s): See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
The Rapid Fentanyl (FYL) Test Strip is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
The Rapid Fentanyl (FYL) Test Dipcard is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in-vitro diagnostic use only.
D Special Instrument Requirements:
N/A
IV Device/System Characteristics:
A Device Description:
Rapid Fentanyl (FYL) Test Strip and Rapid Fentanyl (FYL) Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of fentanyl at or above the cut-off concentration of 1 ng/mL. Test Strip and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.
Content of the kit:
Test Strip, one test in one pouch or Test Dipcard, one test in one pouch.
Desiccant. The desiccant is only for storage purposes only and is not used in the test procedures.
Leaflet with instructions for use.
B Principle of Operation:
The Rapid Fentanyl (FYL) Test Strip is a competitive immunoassay that is used to screen for the presence of Fentanyl and metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.
When the test is activated, the urine is absorbed into test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal
K231904 - Page 2 of 10
{2}
antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.
When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C), of each strip, if the test has been performed properly.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Superbio Fentanyl Urine Detection Kit, Superbio Immunofluorescence Analyzer EASY-11
B Predicate 510(k) Number(s):
K220046
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K231904 | K220046 |
| --- | --- | --- |
| Device Trade Name | Rapid Fentanyl (FYL) Test Strip, Rapid Fentanyl (FYL) Test Dipcard | Superbio Fentanyl Urine Detection Kit |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | Qualitative detection of fentanyl in urine, | Same |
| Intended Users | Prescription use only. | Same |
| Calibrator and Cutoff Values | 1 ng/ml | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Methodology | Competitive binding, lateral flow immunochromatographi c assay based on the principle of antigen antibody immunochemistry | Same |
K231904 - Page 3 of 10
{3}
| General Device Characteristic Differences | | |
| --- | --- | --- |
| Configurations | Test Strip and Dipcard | Test Card |
| Platform required | None | Immunofluorescence Analyzer |
| Storage/Operating Temp | 4-30°C | 4-35°C |
VI Standards/Guidance Documents Referenced:
None
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Precision studies were performed using 3 lots of test Strip and test Dipcard. Drug free specimens were spiked with target drug fentanyl at 0, ±75% cutoff, ±50% cutoff, ±25% cutoff and +100% cutoff of drug. The concentrations of the target drugs were confirmed with LC/MS. Each concentration of the urine specimen was divided into aliquots. Each aliquot was blindly labeled by a nonparticipant. Separate sets of blinded coded samples were assigned and randomized prior to testing. The study was conducted by 6 operators/nurses at 3 Point-of-Care sites. Two operators per location tested 3 aliquots at each concentration for each lot per day (3 runs/day) for 10 non-consecutive days using one device lot per location. One operator tested the test strip format, and the second operator tested the test dipcard format. There were 1620 observations by 3 sites at 9 concentrations.
Rapid Fentanyl (FYL) Test Strip:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Strip Lot 1 | | Test Strip Lot 2 | | Test Strip Lot 3 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Positive | Negative | Positive | Negative | Positive | Negative |
| 0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 0.25ng/ml | -75% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 0.5ng/ml | -50% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 0.75ng/ml | -25% cutoff | 60 | 4 | 56 | 2 | 58 | 6 | 54 |
| 1ng/ml | cutoff | 60 | 36 | 24 | 34 | 26 | 32 | 28 |
| 1.25ng/ml | +25% cutoff | 60 | 54 | 6 | 56 | 4 | 56 | 4 |
| 1.5ng/ml | +50% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 1.75ng/ml | +75% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 2ng/ml | +100% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
K231904 - Page 4 of 10
{4}
Rapid Fentanyl (FYL) Test Dipcard:
| Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Dipcard Lot 1 | | Test Dipcard Lot 2 | | Test Dipcard Lot 3 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Positive | Negative | Positive | Negative | Positive | Negative |
| 0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 0.25ng/ml | -75% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 0.5ng/ml | -50% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60 |
| 0.75ng/ml | -25% cutoff | 60 | 4 | 56 | 4 | 56 | 2 | 58 |
| 1ng/ml | cutoff | 60 | 34 | 26 | 34 | 26 | 36 | 24 |
| 1.25ng/ml | +25% cutoff | 60 | 56 | 4 | 58 | 2 | 56 | 4 |
| 1.5ng/ml | +50% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 1.75ng/ml | +75% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
| 2ng/ml | +100% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0 |
2. Linearity:
Not applicable, this device is intended for qualitative use only
3. Analytical Specificity/Interference:
Cross Reactivity:
The sponsor performed a study to verify potential cross-reactants using the target drug, drug metabolites, and other related compounds. The target drugs, drug metabolites and other components that are likely to interfere in urine samples were added to drug-free urine specimens at a concentration of 100μg/ml. The study was conducted by 8 laboratory professionals. Two Laboratory professionals spiked the compound controls and divided them into individual containers, the other 6 laboratory professionals conducted the testing using the instructions for use package insert. Each sample was tested in 3 replicates using 3 lots of Rapid Fentanyl (FYL) Test Strip and the same protocol for Rapid Fentanyl (FYL) Test Dipcard. One operator conducted each individual lot (and product type; strip and dipcard) separately.
The formula for calculating the percent of cross-reactivity was calculated by dividing the cutoff concentration by the minimum concentration required to obtain a positive result and then multiplying by 100%. The lowest concentration that caused a positive result for each compound are listed below for both the Rapid Fentanyl (FYL) Test Strip and the same protocol for Rapid Fentanyl (FYL) Dipcard:
Rapid Fentanyl (FYL) Test-Strip:
| Compound | Quantity equivalent to cutoff(ng/mL) conducted on 3 lots | % Cross Reactivity |
| --- | --- | --- |
| Norfentanyl | 10,000 | 0.01 |
K231904 - Page 5 of 10
{5}
Rapid Fentanyl (FYL) Dipcard:
| Compound | Quantity equivalent to cutoff(ng/mL) conducted on 3 lots | % Cross Reactivity |
| --- | --- | --- |
| Norfentanyl | 10,000 | 0.01 |
| Acetyl fentanyl | 1.2 | 83.3 |
| Acetyl norfentanyl | 10,000 | 0.01 |
| Acrylfentanyl | 1.5 | 66.7 |
| Butyryl fentanyl | 1.6 | 62.5 |
| Carfentanil | 500 | 0.2 |
| (±)-3-cis-methylfentanyl | 5 | 20 |
| 4-Fluoro-isobutyrylfentanyl | 3 | 33.3 |
| Furanyl fentanyl | 1.8 | 55.6 |
| ω-1-Hydroxyfentanyl | 20,000 | 0.01 |
| (±) β-hydroxythiofentanyl | 2.8 | 35.7 |
| Isobutyryl fentanyl | 1.5 | 66.7 |
| Ocfentanil | 1.8 | 55.6 |
| Para-fluorobutyrylfentanyl (p-FBF) | 3 | 33.3 |
| Para-fluoro fentanyl | 3 | 33.3 |
| Sufentanil | 625 | 0.16 |
| Valeryl fentanyl | 2.5 | 40 |
| Compound | Quantity equivalent to cutoff(ng/mL) conducted on 3 lots | % Cross Reactivity |
| --- | --- | --- |
| Norfentanyl | 10,000 | 0.01 |
| Acetyl fentanyl | 1.2 | 83.3 |
| Acetyl norfentanyl | 10,000 | 0.01 |
| Acrylfentanyl | 1.5 | 66.7 |
| Butyryl fentanyl | 1.6 | 62.5 |
| Carfentanil | 500 | 0.2 |
| (±)-3-cis-methylfentanyl | 5 | 20 |
| 4-Fluoro-isobutyrylfentanyl | 3 | 33.3 |
| Furanyl fentanyl | 1.8 | 55.6 |
| ω-1-Hydroxyfentanyl | 20,000 | 0.01 |
| (±) β-hydroxythiofentanyl | 2.8 | 35.7 |
| Isobutyryl fentanyl | 1.5 | 66.7 |
K231904 - Page 6 of 10
{6}
| Ocfentanil | 1.8 | 55.6 |
| --- | --- | --- |
| Para-fluorobutyryl fentanyl (p-FBF) | 3 | 33.3 |
| Para-fluoro fentanyl | 3 | 33.3 |
| Sufentanil | 625 | 0.16 |
| Valeryl fentanyl | 2.5 | 40 |
## Interference Study on the Effect of Specific Gravity and pH:
The specific Gravity study was conducted using 5 increments of specific gravity levels: 1.000, 1.010, 1.020, 1.030, and 1.040 on 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard with drug free urine spiked with Fentanyl at $0.5\mathrm{ng / ml}$ (50% below cutoff) and $1.5\mathrm{ng / ml}$ (50% above cutoff). All concentrations were confirmed with LC/MS. No discrepant results were observed.
The pH study was conducted on 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard using seven increments of pH level ranging from 3 to 9 in aliquot negative urine spiked with Fentanyl at $0.5\mathrm{ng / ml}$ (50% below cutoff) and $1.5\mathrm{ng / ml}$ (50% above cutoff). All concentrations were confirmed with LC/MS. No discrepant results were observed.
## Interference
The sponsor conducted two different studies for interference, using their Test Strip and Test Dipcard for each. The first interference study was conducted using opioid compounds (a separate study from their Cross-Reactivity study), and the second interference study was conducted using commonly ingested medications/substances. Both are described below:
### A. Interfering Substances (Opioid compounds):
To test the interference of opioid compounds. This study was conducted by 8 laboratory professionals. 2 laboratory professionals spiked the compound controls and divided them into individual containers. The other 6 laboratory professionals conducted the following testing using the English package insert as guide and read the results at 5 minutes.
Potential interfering opioid compounds, each at a concentration of $100~\mu \mathrm{g / mL}$ (except Despropionly fentanyl (4-ANPP) was tested at $50\mu \mathrm{g / ml}$, norcarfentanil was tested at $5\mu \mathrm{g / mL}$ and remifentanil was tested at $10~\mu \mathrm{g / mL}$), were added to drug-free urine specimens spiked with fentanyl to concentrations at $50\%$ below and $50\%$ above cutoff levels at $0.5\mathrm{ng / ml}$ and $1.5\mathrm{ng / ml}$ level and were tested in replicates of 3 on 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard. The compounds showed no interference with the Rapid Fentanyl (FYL) Test Strip or the Rapid Fentanyl (FYL) Test Dipcard.
The following opioid compounds showed no interference with the Rapid Fentanyl (FYL) Test Strip or the Rapid Fentanyl (FYL) Test Dipcard:
6-Acetyl morphine, Alfentanil, Amphetamine, Buprenorphine, Buprenorphineglucuronide, Codeine, Despropionyl fentanyl (4-ANPP),
K231904 - Page 7 of 10
{7}
Dextromethorphan, Dihydrocodeine, EDDP, EMDP, Fluoxetine, Heroin, Hydrocodone, Hydromorphone, Ketamine, Levorphanol, Meperidine, Methadone, Morphine, Morphine-3-glucuronide, Naloxone, Naltrexone, Norbuprenorphine, Norcarfentanil, Norcodeine, Norketamine, Normeperidine, Normorphine, Noroxycodone, Oxycodone, Oxymorphone, Pentazocine (Talwin), Pipamperone, Remifentanil, Risperidone, Tapentadol, Thioridazine, Tilidine, Tramadol, Tramadol-O-Desmethyl, Tramadol-NDesmethyl, Trazodone, Dextromethorphan, Meperidine, and Pentazocine
B. Interfering Substances (Commonly ingested medications/substances):
This study was conducted by 8 laboratory professionals. 2 laboratory professionals spiked the compound controls and divided them into individual containers. The other 6 laboratory professionals conducted the following testing using the English package insert as guide and read the results at 5 minutes.
Potential interfering substances, each at a concentration of 100 µg/mL (or specified concentrations noted in the tables below), were added to drug-free urine specimens spiked with fentanyl to concentrations at 50% below and 50% above cutoff levels at 0.5 ng/ml and 1.5 ng/ml level and were used to test 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard. The following substances showed no interference with the Rapid Fentanyl (FYL) Test Strip or the Rapid Fentanyl (FYL) Test Dipcard:
Acetaminophen, Acetone (1000 mg/dL), Acetophenetidin, Acetylsalicylic acid, Albumin (100 mg/dL), Albuterol, Aminopyrine, Amitriptyline (35 µg/ml), Amobarbital, Amoxicillin, Ampicillin, Apomorphine, Ascorbic acid, Aspartame, Atropine, Benzilic acid, Benzoic acid, Benzoylecgonine, Bilirubin, Boric Acid (1%), Bupropion (50 µg/ml), Caffeine, Carbamazepine, Chloral hydrate, Chloramphenicol, Chlorothiazide, Chlorpromazine, Cholesterol, Clomipramine (50 µg/ml), Clonidine, Cortisone, Cotinine, Creatinine, Cyclobenzaprine (10 µg/ml), Deoxycorticosterone, Desipramine (50 µg/ml), Dextromethorphan, Diclofenac, Diflunisal, Digoxin, Diphenhydramine, DL-Tryptophan, DL-Tyrosine, Doxepin (50 µg/ml), Ecgonine methyl ester, Ephedrine, Erythromycin, Ethanol (1%), Fenoprofen, Fluphenazine, Furosemide, Galactose (10 mg/dL), Gamma Globulin (500 mg/dL), Gentisic acid, Glucose (3000 mg/dL), Hemoglobin, Hydralazine, Hydrochlorothiazide, Hydrocortisone, Hydroxytyramine, Ibuprofen, Imipramine (30 µg/ml), Isoproterenol, Isoxsuprine, Ketoprofen, Labetalol, Lidocaine (50 µg/ml), Loperamide, Maprotiline (50 µg/ml), Meperidine, Meprobamate, Methapyrilene (10 µg/ml), Methaqualone (50 µg/ml), Methoxyphenamine, Metronidazole (300 µg/ml), N-Acetylprocainamide, NaCl (4000 mg/dL), Nalidixic acid, Naloxone, Naltrexone, Naproxen, Niacinamide, Nicotine (10 µg/ml), Nifedipine, Norethindrone, Nortriptyline (25 µg/ml), Noscapine, OHydroxyhippuric acid, Octopamine, Oxalic acid (100 mg/dL), Oxazepam, Oxolinic acid, Oxymetazoline, Papaverine, Penicillin G, Perphenazine, Phencyclidine, Phenelzine, Phenobarbital, Prednisone, Propoxyphene (50 µg/ml), Propranolol, Pseudoephedrine, Quinine, Ranitidine, Riboflavin (7.5 mg/dL), Ketamine, Salicylic acid, Secobarbital, Serotonin (5-Hydroxytyramine), Sulfamethazine, Sulindac, Tetrahydrocortisone 3-(β-D-glucuronide), Tetrahydrocortisone 3-acetate, Tetrahydrozoline, Thiamine, Thioridazine, Triamterene, Trifluoperazine, Trimethoprim, Tyramine, Urea (2000 mg/dL), Uric acid, Valproic acid (250 µg/ml), Venlafaxine, Verapamil, Zomepirac, β-Estradiol.
K231904 - Page 8 of 10
{8}
4. Assay Reportable Range:
Not applicable, this device is intended for qualitative use only.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The device is traceable to a commercially available standard that was verified by LC/MS-MS
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Analytical performance of the device around the claimed cutoff is described in the precision section VII.A1. above (1ng/ml).
B Comparison Studies:
1. Method Comparison with Predicate Device:
Four hundred eighty clinical urine samples collected from several hospitals and drug relief reformatories. All samples were unaltered clinical specimens, and the concentrations were confirmed with LC/MS. The samples were divided among the 6 nurses. All aliquots were blindly labeled by a nonparticipant and were randomized prior to testing.
The study was conducted by 6 nurses at three Point-of-Care sites. Each of the three sites were responsible for conducting testing on one lot of 80 samples for both the Rapid Fentanyl (FYL) Test Strip, and the Rapid Fentanyl (FYL) Test Dipcard (160 samples at each of the 3 sites=480). Each nurse only tested one format of Rapid Fentanyl Test, one tested with strip format another tested with Dipcard.
Comparison data of Rapid Fentanyl (FYL) Test (Strip)
| Format | LC/MS | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. (drug free) | Neg. (<-50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) | | Pos. (> +50% cutoff) |
| Lot 1 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Lot 2 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Lot 3 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
K231904 - Page 9 of 10
{9}
Comparison data of Rapid Fentanyl (FYL) Test (Dipcard)
| Format | LC/MS | | | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. (drug free) | Neg. (<-50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) | | Pos. (> +50% cutoff) |
| Test Dipcard Lot 1 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Test Dipcard Lot 2 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
| Test Dipcard Lot 3 | Positive | 0 | 0 | 1 | 8 | 31 | 80 |
| | Negative | 27 | 5 | 7 | 1 | 0 | |
2. Matrix Comparison:
Not applicable. This device is intended to be used with urine samples only.
C Clinical Studies:
1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable.
D Clinical Cut-Off: Not applicable.
E Expected Values/Reference Range: Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K231904 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.